K Number
K153302

Validate with FDA (Live)

Device Name
coflex-IF
Date Cleared
2016-09-08

(300 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The coflex-IF is a posterior, non-pedicle supplemental fixation device intended for use with an interbody cage as an adjunct to fusion at a single level in the lumbar spine (L1 - S1). It is intended for attachment to the spinous processes for the purpose of achieving stabilization to promote fusion in patients with disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies - with up to Grade 1 spondylolisthesis.

Device Description

The coflex-IF system is an implant system for interspinous fixation within lumbar interbody fusion surgery. It consists of a single, U-shaped component, fabricated from medical grade titanium alloy (Ti6Al4V). A set of two wings extends vertically from the superior long arm of the device, with a second set of wings extending below the inferior long arm. A screw and sleeve are inserted through a prepared hole and fixes the crimped wings to the superior and inferior spinous processes.

The purpose of this Special 510(k) is to add a notch in the posterior portion of the U-shaped body. The modifications do not affect the intended use of the device or alter the fundamental scientific technology of the device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the coflex-IF device, based on the provided FDA 510(k) summary:

  • Note: This document is a 510(k) summary for a spinal implant (coflex-IF), which is a physical medical device, not a software or AI-based diagnostic device. Therefore, many of the requested categories (like sample size for test/training sets, experts, adjudication methods, MRMC studies, standalone algorithm performance, and ground truth types for AI models) are not applicable to this type of submission. The information provided focuses on engineering and mechanical testing for substantial equivalence to a predicate device.

Acceptance Criteria and Reported Device Performance

Acceptance Criteria (What was measured)Reported Device Performance (Result compared to criteria)
Mechanical Stability and Properties (per ASTM F 1717-13):The "worst case device" was subjected to these tests. The results demonstrated that the acceptance criteria (defined in an unprovided "Design Control Activities Summary") were met, and the performance was equivalent to the predicate device.
Static Compression BendingAchieved equivalence to predicate device.
Static TorsionAchieved equivalence to predicate device.
Compression Bending FatigueAchieved equivalence to predicate device.

Study Information (Where applicable, based on a physical medical device)

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not explicitly stated as "sample size" in the context of human subjects or data, but "The worst case device" was tested. This refers to a manufactured physical device chosen as representative of the most challenging conditions or configurations.
    • Data Provenance: Not applicable in the context of human data. The testing is described as mechanical performance testing of the device itself.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth for a physical device's mechanical performance is established through standardized engineering tests (e.g., ASTM F 1717-13) and comparison to the predicate device, not expert consensus on human data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Mechanical testing results are objective measurements of physical properties, not subject to human adjudication in the way clinical diagnostic interpretations would be.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical spinal implant, not an AI software/diagnostic device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical spinal implant, not an AI algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For mechanical performance, the "ground truth" is defined by the standards set by ASTM F 1717-13 and the established performance of the predicate device. The goal is substantial equivalence, meaning the new device performs acceptably within the same safety and effectiveness parameters as the legally marketed predicate.
  7. The sample size for the training set:

    • Not applicable. This is a physical device, not an AI model requiring a training set of data. Device design and manufacturing parameters are based on engineering principles and material science, not machine learning.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, there is no AI training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus symbol is composed of three abstract human profiles facing to the right. The text is written in a simple, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 8, 2016

Paradigm Spine, LLC % Mr. Justin Eggleton Senior Director, Spine Regulatory Affairs Musculoskeletal Clinical Regulatory Advisers, LLC 1331 H Street NW, 12th Floor Washington, District of Columbia 20005

Re: K153302

Trade/Device Name: coflex-IF Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: July 28, 2016 Received: July 29, 2016

Dear Mr. Eggleton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153302

Device Name coflex-IF

Indications for Use (Describe)

The coflex-IF is a posterior, non-pedicle supplemental fixation device intended for use with an interbody cage as an adjunct to fusion at a single level in the lumbar spine (L1 - S1). It is intended for attachment to the spinous processes for the purpose of achieving stabilization to promote fusion in patients with disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies - with up to Grade 1 spondylolisthesis.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Device Trade Name:coflex-IF
Manufacturer:Paradigm Spine, LLC505 Park Avenue, 14th FloorNew York, NY 10022Phone: (212) 583-9700Fax:(212) 826-9509
Contact:Mr. Justin EggletonMusculoskeletal Clinical Regulatory Advisers, LLC1331 H Street NW, 12th FloorWashington, DC 20005Phone: (202) 552-5804Email: jeggleton@mcra.com
Date Prepared:September 6, 2016
Classification:21 CFR §888.3050, Spinal Interlaminal Fixation Orthosis
Class:II
Product Code:PEK
Primary Predicate:coflex-F system (K112595, K093438)
Additional Predicates:Lanx ASPEN (K071877)NuVasive Affix (K073278)

Indications For Use:

The coflex-IF is a posterior, non-pedicle supplemental fixation device intended for use with an interbody cage as an adjunct to fusion at a single level in the lumbar spine (L1-S1). It is intended for attachment to the spinous processes for the purpose of achieving stabilization to promote fusion in patients with degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies - with up to Grade 1 spondylolisthesis.

Device Description:

The coflex-IF system is an implant system for interspinous fixation within lumbar interbody fusion surgery. It consists of a single, U-shaped component, fabricated from medical grade titanium alloy (Ti6Al4V). A set of two wings extends vertically from the superior long arm of the device, with a second set of wings extending below the inferior long arm. A screw and sleeve are inserted through a prepared hole and fixes the crimped wings to the superior and inferior spinous processes.

{4}------------------------------------------------

The purpose of this Special 510(k) is to add a notch in the posterior portion of the U-shaped body. The modifications do not affect the intended use of the device or alter the fundamental scientific technology of the device.

Predicate Device(s):

The modified coflex-IF is substantially equivalent to the predicate coflex-F (Paradigm Interspinous Fusion Plate) previously cleared in K112595 and K093438 with respect to indications, design, function, and materials.

Performance Testing

Mechanical testing was performed on coflex-IF, providing sufficient proof for functional safety and performance regarding the mechanical stability and properties. The following three tests were performed (according to ASTM F 1717-13): static compression bending, static torsion, and compression bending fatigue. Results demonstrated equivalent performance to the predicate device.

Substantial Equivalence:

The worst case device was subjected to mechanical testing. The results demonstrate that the acceptance criteria defined in the Design Control Activities Summary were met and the coflex-IF is substantially equivalent to the predicate device(s).

Conclusion:

The coflex-IF is substantially equivalent to previously cleared devices with respect to its indications for use, design, function, materials, and performance.

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.