K Number
K153264

Validate with FDA (Live)

Date Cleared
2016-05-19

(189 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EasyPass Guidewire is used for selective cannulation of the biliary ducts including, but not limited to the common bile duct, cystic duct, and right and left hepatic ducts. The EasyPass guidewire is designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

Device Description

The EasyPass Guidewire consists of a Nitinol core wire with a proximal 2-color spiral patterned PTFE jacket and a copolymer distal section. The distal stainless steel spring section of the guidewire provides flexibility and incorporates a hydrophilic coating. The EasyPass Guidewire is a sterile packed, disposable instrument intended for single-patient use only.

AI/ML Overview

This document is a 510(k) premarket notification for the EasyPass Guidewire. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study where specific acceptance criteria are defined and statistically proven.

Therefore, the requested information cannot be fully provided in the typical format for a study proving device meets acceptance criteria. However, I can extract the relevant information regarding the performance data provided.


1. Table of Acceptance Criteria and Reported Device Performance

As this is a 510(k) submission, "acceptance criteria" are not explicitly stated in a quantitative manner with pass/fail thresholds for specific performance metrics in the way one might expect for a de novo or PMA submission. Instead, the performance is demonstrated by comparing the device to a predicate and showing it meets "the same performance requirements." The "acceptance criteria" are therefore implicitly that the device performs equivalently to the predicate device in the specified tests.

Acceptance Criteria (Implicit: Performance equivalent to predicate)Reported Device Performance
Cytotoxicity (per ISO 10993)Performed; determined to be safe and effective as predicate
Irritation (per ISO 10993)Performed; determined to be safe and effective as predicate
Sensitization (per ISO 10993)Performed; determined to be safe and effective as predicate
USP Class VI Testing (for color additives)Performed; determined to be safe and effective as predicate
Fracture TestPerformed; determined to be safe and effective as predicate
Bend TestPerformed; determined to be safe and effective as predicate
Friction TestPerformed; determined to be safe and effective as predicate
Scratch TestPerformed; determined to be safe and effective as predicate
Tensile TestPerformed; determined to be safe and effective as predicate
Radiographic Image TestPerformed; determined to be safe and effective as predicate

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (N) used for each non-clinical performance test. It only states that these tests were conducted. The data provenance is "non-clinical performance data" generated from bench testing and biocompatibility assessments of the proposed EasyPass Guidewire. There is no information regarding the country of origin or whether it's retrospective/prospective as these are not clinical tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This document describes non-clinical testing (bench and biocompatibility), not studies involving expert evaluation of clinical data or images.

4. Adjudication method for the test set

Not applicable. No expert adjudication was described for these non-clinical tests.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical guidewire, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was mentioned.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a medical guidewire, not an algorithm.

7. The type of ground truth used

For the non-clinical performance data, the "ground truth" would be established by the standardized methods and specifications relevant to each test (e.g., ISO 10993 for biocompatibility, material specifications for fracture/bend/tensile tests, etc.). The goal was to demonstrate performance equivalent to the predicate device.

8. The sample size for the training set

Not applicable. This is a medical device, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. (See #8)

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 18, 2016

Medi-Globe Corporation Scott Karler Regulatory Affairs Coordinator 7850 South Hardy Drive, Suite. 112 Tempe, AZ 85284

Re: K153264

Trade/Device Name: EasyPass Guidewire Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCY Dated: April 7, 2016 Received: April 13, 2016

Dear Scott Karler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

510(k) Number (if known)K153264
Device NameEasyPass Guidewire
Indications for Use (Describe)The EasyPass Guidewire is used for selective cannulation of the biliary ducts including, but not limited to the common bile duct, cystic duct, and right and left hepatic ducts. The EasyPass guidewire is designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

(AN/8) 1886 AQ3 MAO3

ခုေပြေ

ປີ ປະຈຸບັນ ( ໄປນີ້ ) ແລວ ເຄລາ ຄວາ ສາມາດ ສາມາດສານປະການ ລົງສຸ

{3}------------------------------------------------

ົວຂອງ ໄດ້ ເວີຍ ແລ່ງວັນນ້ອຍການປະກວດນາງງາມ ປີ ປຸກາ ຂອງໄປປຸ່ນ ແລະວ່າໄປປຸ່ນ ແວ່ງວ່າ ຂໍ້ມັ

iseqs redio your and exemites nebud sint pribremori in noilemorii in noileemore ent əri bulbuları ,əsrlədən vəd suvorların əsasən və bateminə sinə aid ini və əmir nəbud əri ətələr və bərən əsasən əsasən və qalını qarşının ətəb çərir. Bu və bir və illər və qalında və bir mill of the same and the mails are and of the common and and

:of ,nebrud sint prisuber 101 suoitzaggus pribuloni ,unitsellos noilsmonni siril 10 vog.zdy.spy@jtfillelsAAP 1912 (AR9) ISA noilloubes വാഗ്യാപക Office of Chief Information Officer noilerisinimbA pung bung boo รอว่างาอร์ ทรการ Home HillseH To เมือกการตุลอ

้า ติดใจอีกดว ธ , ot bnoqser of beiluper ใดต ร่านตราคาร เวทราคาราชอาณา เอก เยก บุคคล พ.ศ. 2019 บุคคลจากคำ

{4}------------------------------------------------

510(k) SUMMARY

Date of Preparation: 11/09/2015

A. Submitter Information:

Medi-Globe Corporation 7850 S. Hardy Drive, #112 Tempe, Arizona 85284

Contact Person: Scott Karler Telephone: (480) 897-2772 ext. 208 Fax: (480) 897-2878 Email: skarler@mediglobe.com

B. Proposed Device Information:

  • Trade Name: EasyPass Guidewire
  • 2 Common Name: Endoscopic Guidewire, Gastroenterology-Urology
  • ಷ Classification Name: Endoscope and Accessories 21 CFR876.1500
  • Σ Product Code: ΟCY

C. Predicate Device:

  • Trade Name: Taxi Endoscopic Guidewire
  • ಷ Manufacturer: Lake Region Manufacturing, Inc.
  • と Clearance Number: K081708
  • Σ Common Name: Endoscopic Guidewire, Gastroenterology-Urology
  • Classification Name: Endoscope and Accessories 21 CFR876.1500
  • Product Code: OCY

D. Device Description:

The EasyPass Guidewire consists of a Nitinol core wire with a proximal 2-color spiral patterned PTFE jacket and a copolymer distal section. The distal stainless steel spring section of the guidewire provides flexibility and incorporates a hydrophilic coating. The EasyPass Guidewire is a sterile packed, disposable instrument intended for single-patient use only.

E. Indications for Use:

The EasyPass Guidewire is used for selective cannulation of the biliary ducts including, but not limited to the common bile duct, cystic duct, and right and left hepatic ducts. The EasyPass guidewire is designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

{5}------------------------------------------------

F. Technological Characteristics Summary:

The EasyPass Guidewire is a 2-color spiral patterned guidewire used in endoscopic procedures. It is comprised of a Nitinol core wire with a 2-color PTFE jacket as a visualization aid during use of the guidewire. The distal section of the EasyPass Guidewire consists of a stainless steel spring encased in a polymer jacket and is hydrophilically coated. The predicate device, Taxi Endoscopic Guidewire. (K081708) incorporates the same design. intended use. diagnostic/therapeutic effect, method of introduction/use, technical characteristics and general range of descriptive features as the proposed. EasyPass Guidewire. There are no technological differences when comparing the proposed EasyPass Guidewire to the predicate device.

G. Non-Clinical Performance Data:

Design verification data and biocompatibility testing has demonstrated that the proposed EasyPass Guidewire meets the same performance requirements and is as safe and effective as predicate device, Taxi Endoscopic Guidewire, (K081708).

The following non-clinical performance tests were conducted on the proposed EasyPass Guidewire:

Biocompatibility testing per ISO 10993: Cytotoxicity Irritation Sensitization

USP Class VI Testing performed on color additives

  • Bench/performance testing consisted of: Fracture Test Bend Test Friction Test Scratch Test Tensile Test Radiographic Image Test

Conclusions of Non-Clinical Performance Data: H.

Medi-Globe Corporation has demonstrated that the proposed EasyPass Guidewire is as safe and effective as the predicate device, (Taxi Endoscopic Guidewire, K081708). The EasyPass Guidewire is considered to have the same intended use, diagnostic/therapeutic effect. method of introduction/use. technical characteristics and general range of descriptive features as the predicate device. (Taxi Endoscopic Guidewire. K081708). No clinical testing was relied on for evaluation of the proposed device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.