K Number
K153233

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-01-06

(58 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Galaxy UNYCO System is intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.

Temporary stabilization of the tibia and foot in conditions and procedures, such as:

  • · comminuted open or closed tibial fractures
  • · polytrauma patient
  • · damage control orthopedics for fractures with severe soft tissue injuries
  • peri-prosthetic or peri-implant fractures
  • · joint dislocations, intra- and extra-articular injuries where spanning fixation is needed
  • · intra-operative fracture reduction
  • · intermediate stabilization in staged surgery
  • · infected non-union pending second stage treatment bone-loss or other reconstructive procedures

The Galaxy UNYCO System is compatible with Galaxy Fixation System and bicortical screws. Galaxy Fixation System and bicortical screws must be used when Galaxy UNYCO is not indicated or available. The product is indicated for non-weight-bearing use.

Device Description

The Orthofix Galaxy UNYCO System consists of a series of sterile kits that include UNYCO Cancellous Screws, Large Multiscrew Clamps for UNYCO Screws, Rods Ø 12mm, a radiolucent Foot Unit and specific application tools. External fixation systems are modular, therefore different frame configurations are possible. The Orthofix components in the Galaxy UNYCO System are not intended to replace normal healthy bone or to withstand the stresses of weight bearing.

AI/ML Overview

Acceptance Criteria and Device Performance for Galaxy UNYCO System (K153233)

The provided documentation describes the acceptance criteria and study findings for the Orthofix Galaxy UNYCO System, a device intended for temporary bone stabilization.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Criteria/StandardReported Device PerformanceComments
Mechanical PerformanceMechanical strength as per ASTM F 1541-02Found to be equivalent or better than the predicate devices.Testing met or exceeded requirements.
Material CompatibilityNot explicitly stated as a numerical criterion, but implied by equivalence to predicate.Utilizes similar materials to predicate devices.Part of establishing substantial equivalence.
Intended UseTemporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.Device is intended for these uses.Confirmed through indications for use.
SafetyRisk Management Plan, adherence to applicable standards.Potential hazards evaluated and controlled; testing met or exceeded requirements.No clinical data was deemed necessary for safety and effectiveness.
MRI CompatibilityAdherence to ASTM F2182 "Standard test method for measurement of radio frequency induced heating near passive implants during magnetic resonance imaging"MRI compatibility testing was conducted per ASTM F2182.Specific results not detailed, but testing was performed.

2. Sample Size Used for the Test Set and Data Provenance:

The document primarily relies on mechanical testing and material comparison for demonstrating substantial equivalence. It does not mention a "test set" in the context of clinical data or patient data.

  • Test Set (for mechanical and MRI testing): The specific number of devices/components tested for mechanical performance (ASTM F 1541) and MRI compatibility (ASTM F2182) is not explicitly stated in the provided summary.
  • Data Provenance: The studies are described as non-clinical technical assessments (mechanical and MRI compatibility testing) rather than clinical studies with patient data. Therefore, there is no mention of country of origin or retrospective/prospective nature regarding patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not applicable as the studies were non-clinical mechanical and MRI compatibility tests. Ground truth in this context would be defined by the technical specifications of the standards (ASTM F 1541, ASTM F2182).

4. Adjudication Method for the Test Set:

This information is not applicable as the studies were non-clinical mechanical and MRI compatibility tests. Adjudication methods are typically associated with human reviewer assessments in clinical or diagnostic studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study was mentioned. The submission explicitly states: "Clinical data was not needed to support the safety and effectiveness of the Subject Device."

6. Standalone Performance (Algorithm Only):

No standalone performance study for an algorithm was mentioned. The device is a physical external fixation system, not a software algorithm.

7. Type of Ground Truth Used:

The "ground truth" for the evaluation of the Galaxy UNYCO System was based on:

  • Engineering Standards: Specifically, the requirements and methodologies outlined in ASTM F 1541 for mechanical performance and ASTM F2182 for MRI compatibility.
  • Predicate Device Characteristics: Comparison to the established design, materials, and mechanical performance of legally marketed predicate devices (Synthes Large External Fixation, MR Conditional (K082650) and Orthofix Galaxy Unyco Diaphyseal Tibia Kit (K142052)).

8. Sample Size for the Training Set:

This information is not applicable. The device is a physical medical device, not an algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting community and connection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 6, 2016

Orthofix SRL % Ms. Cheryl Wagoner Consultant Wagoner Consulting LLC P.O. Box 15729 Wilmington, North Carolina 28408

Re: K153233

Trade/Device Name: Galaxy UNYCO System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: November 6, 2015 Received: November 9, 2015

Dear Ms. Wagoner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K153233

Device Name Galaxy UNYCO System

Indications for Use (Describe)

The Galaxy UNYCO System is intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.

Temporary stabilization of the tibia and foot in conditions and procedures, such as:

  • · comminuted open or closed tibial fractures
  • · polytrauma patient
  • · damage control orthopedics for fractures with severe soft tissue injuries
  • peri-prosthetic or peri-implant fractures
  • · joint dislocations, intra- and extra-articular injuries where spanning fixation is needed
  • · intra-operative fracture reduction
  • · intermediate stabilization in staged surgery
  • · infected non-union pending second stage treatment bone-loss or other reconstructive procedures

The Galaxy UNYCO System is compatible with Galaxy Fixation System and bicortical screws. Galaxy Fixation System and bicortical screws must be used when Galaxy UNYCO is not indicated or available. The product is indicated for non-weight-bearing use.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Orthofix, followed by the text "510(k) Summary". The Orthofix logo is a blue, abstract shape. The text is in a sans-serif font and is black.

(as required by 21 CFR 807.92)

SubmitterOrthofix Srl
Via delle Nazioni, 9 37012 Bussolengo (VR) - Italy
Telephone+ 39 045 6719.000
Fax+ 39 045 6719.380
Contact PersonGianluca RicadonaQuality & Regulatory Affairs Manager
AddressVia delle Nazioni, 937012 Bussolengo (VR) - Italy
Telephone+ 39 045 6719.000
Fax+ 39 045 6719.380
emailGianlucaRicadona@orthofix.it
Date PreparedNovember 6, 2015
Trade NameGalaxy UNYCO System
Common NameExternal Fixation Device and Accessories
Panel CodeOrthopedic
ClassificationNameSingle/multiple component metallic bone fixation appliances andaccessories.
ClassClass II
RegulationNumber21 CFR 888.3030
Product CodeKTT
Predicate Device Name510(k) NumberManufacturer
Synthes Large External Fixation, MR ConditionalK082650Synthes
Orthofix Galaxy Unyco Diaphyseal Tibia KitK142052Orthofix Srl
DescriptionThe Orthofix Galaxy UNYCO System consists of a series of sterile kits thatinclude UNYCO Cancellous Screws, Large Multiscrew Clamps forUNYCO Screws, Rods Ø 12mm, a radiolucent Foot Unit and specificapplication tools. External fixation systems are modular, therefore differentframe configurations are possible. The Orthofix components in the GalaxyUNYCO System are not intended to replace normal healthy bone or towithstand the stresses of weight bearing.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Indications andIntended UseThe Galaxy UNYCO System is intended to be used for temporary bonestabilization in trauma and orthopedic procedures of the lower limb priorto definitive treatment.Temporary stabilization of the tibia and foot in conditions and procedures,such as:• comminuted open or closed tibial fractures• polytrauma patient• damage control orthopedics for fractures with severe soft tissue injuries• peri-prosthetic or peri-implant fractures• joint dislocations, intra- and extra-articular injuries where spanningfixation is needed
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{4}------------------------------------------------

Traditional 510(k) Premarket Notification

Galaxy Unyco System

• intra-operative fracture reduction
• intermediate stabilization in staged surgery
• infected non-union pending second stage treatment bone-loss or other
reconstructive procedures
The Galaxy UNYCO System is compatible with Galaxy Fixation System
and bicortical screws. Galaxy Fixation System and bicortical screws must
be used when Galaxy UNYCO is not indicated or available.
The product is indicated for non-weight-bearing use.
TechnologicalCharacteristicsand SubstantialEquivalenceDocumentation was provided to demonstrate that the Galaxy UnycoSystem is substantially equivalent to the legally marketed predicates. Thedevices and accessories included in the Galaxy Unyco System and thepredicate devices are all external fracture fixation systems as defined in21 CFR 888.3030. The Galaxy Unyco System is substantially equivalentto the predicate devices in intended use, site of application, patientpopulation, conditions-of-use, mechanical performances, basic design,operating principles, and materials. The Galaxy Unyco System iscomparable to its predicate in size and materials. Testing in accordancewith ASTM F 1541-02 shows the mechanical strength of the subject deviceto be equivalent or better than the predicate devices.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Performance DataThe potential hazards have been evaluated and controlled through a RiskManagement Plan.All testing met or exceeded the requirements as established by the testprotocols and applicable standards. A review of the mechanical dataindicates that the components of the Subject device are capable ofwithstanding expected loads without failure. The Subject device wastherefore found to be substantially equivalent to the predicate devices.Clinical data was not needed to support the safety and effectiveness ofthe Subject Device.The following mechanical testing was performed:ASTM F 1541 "Standard Specification and Test Methods for ExternalSkeletal Fixation Devices" MRI compatibility testing was also conducted per: ASTM F2182 "Standard test method for measurement of radiofrequency induced heating near passive implants during magneticresonance imaging"
ConclusionBased on design, materials, intended use, technological characteristics,and comparison to predicate devices, the Galaxy UNYCO System hasbeen shown to be substantially equivalent to legally marketed predicatedevices..

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.