K Number
K153191

Validate with FDA (Live)

Device Name
Calcium Bridge
Manufacturer
Date Cleared
2016-02-26

(115 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Calcium Bridge is a resin-based composite recommended for use as a liner or initial layer beneath all composite resins, or as flowable composite in preventive resin restorations. For Use only by a Licensed Dentist. Rx Use Only.

Device Description

Calcium Bridge is a resin-based, flowable polymerizable composite paste recommended as a lining material beneath composite resins or as a preventive resin restoration (PRR) material (a minimally invasive procedure to protect a tooth). A PRR restoration is not intended to prevent caries but intended to be used in the treatment of caries. Because Calcium Bridge is highly radio-opaque this allows for easy radio-graphic observation compared to enamel, dentin, and possible secondary caries.

AI/ML Overview

This document is a 510(k) summary for a dental device called "Calcium Bridge." It compares the device to a predicate device, "Tetric EvoCeram," to demonstrate substantial equivalence. It does not contain information about a study proving the device meets acceptance criteria related to AI or digital image analysis. Therefore, it's not possible to provide the requested information.

The document discusses the following properties, which could be considered acceptance criteria for a dental material:

Table of Acceptance Criteria and Reported Device Performance (as inferred from the provided test values):

PropertyAcceptance Criteria (Predicate)Reported Device Performance (Calcium Bridge)
Light Curing CapabilityCures Less Than 20 secondsCures Less Than 20 seconds
Radiopacity (Aluminum Equivalent)Equal to 2.0 mm Aluminum3.5 mm Aluminum
FluorescenceVisible Blue Glow DetectedVisible Blue Glow Detected
Compressive Strength250 MPaGreater Than 250 MPa
Depth of CureGreater Than 1.5 mmGreater Than 1.5 mm
Biocompatibility (ISO 10993-5, Cytotoxicity)(N/A - Not Available for Predicate)Zone of Inhibition 0.25 cm

Missing Information (based on your request, but not present in the document):

The document does not describe a study involving human readers, AI, or ground truth establishment in the context of diagnostic performance. Therefore, I cannot answer the following parts of your request:

  • Sample size used for the test set and the data provenance.
  • Number of experts used to establish the ground truth for the test set and their qualifications.
  • Adjudication method for the test set.
  • Multi-reader multi-case (MRMC) comparative effectiveness study, including effect size.
  • Standalone (algorithm only) performance.
  • Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  • Sample size for the training set.
  • How the ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with a flowing, wave-like design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 26, 2016

Denali Corporation Ms. Jan G. Stannard President 134 Old Washington Street Hanover, Massachusetts 02339

Re: K153191

Trade/Device Name: Calcium Bridge Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF Dated: January 15, 2016 Received: January 19, 2016

Dear Ms. Stannard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for Denali Corporation. The logo consists of a blue triangle with a blue sphere and a white arc around it on the left, and the words "denali corporation" in gray on the right. The text is in a bold, sans-serif font.

INDICATIONS FOR USE STATEMENT

510 (k) Number K153191 (if known)

Device Name

Calcium Bridge

Indications for Use:

Calcium Bridge is a resin-based composite recommended for use as a liner or initial layer beneath all composite resins or as flowable composite in preventive resin restorations. For Use only by a Licensed Dentist. Rx Use Only.

Please do not write below this line. Continue on another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

or

Over-The-Counter Use

{3}------------------------------------------------

510 (k) Summary

Image /page/3/Picture/1 description: The image shows the logo for Denali Corporation. The logo includes a blue and black geometric shape with a blue sphere and a curved line around it. The text "denali corporation" is written in gray next to the logo. Below the company name is the address "134 Old Washington Street, Hanover, MA 02339 USA" also written in gray.

T 781.826.9190 F 781.826.4465 denalicorporation.com

February 26, 2016denalicorporation.com
ADDRESSDENALI CORPORATION134 Old Washington Street / Hanover, MA 02339-1629
OWNER/CONTACT PERSONDr. Jan G. Stannard
TEL: 781-826-9190 / FAX: 781-826-4465j.stannard@denalicorporation.com
TRADE NAMECalcium Bridge
COMMON NAMECavity Liner, Flowable Composite
CLASSIFICATION NAMETooth Shade Resin Material, Class II (21 CFR 872.3690, Product Code EBF)
REGISTRATION3006367836
PREDICATE DEVICESTetric EvoCeram (K042819)
EQUIVALENCEThe predicate product has been found to be substantially equivalentunder the 510(k) premarket notification process as Class II DentalDevices under CFR EBF 872.3690, Tooth Shade Resin Material.
DEVICE DESCRIPTIONCalcium Bridge is a resin-based, flowable polymerizable composite paste recommended as a lining material beneath composite resins or as a preventive resinrestoration (PRR) material (a minimally invasive procedure to protect a tooth). A PRRrestoration is not intended to prevent caries but intended to be used in the treatment ofcaries. Because Calcium Bridge is highly radio-opaque this allows for easy radio-graphic observation compared to enamel, dentin, and possible secondary caries.
INDICATIONS FOR USECalcium Bridge is a resin-based composite recommended for use as a liner or initiallayer beneath all composite resins, or as flowable composite in preventive resinrestorations.For Use only by a Licensed Dentist. Rx Use Only.
TECHNOLOGICALCHARACTERISTICSSUMMARYCalcium Bridge has the same technological characteristics as the predicate deviceTetric Evoceram including: design, composition, biocompatibility, performance,ageing, intended use, physical characteristics, setting time, depth of cure, radio-opacity fluorecensing ability and mechanical properties (tensile and compressivestrength), as summarized in the Table below.
Summary of Characteristics of Calcium Bridge and the Predicate Product
PropertyCalcium BridgeTetric Evoceram
Light Curing Capability, Cures Less Than 20 secondsCures Less Than 20 secondsCures Less Than 20 seconds
Radiopaque, Aluminum Equivalent to Equal 1 mm3.5 mm Aluminum2.0 mm Aluminum
Fluorescence, Visible Blue Glow Under UV LightingVisible Blue Glow DetectedVisible Blue Glow Detected
Compressive Strength (MPa)Greater Than 250 MPa250 MPa
Depth of Cure Greater Than 1.5 mmGreater Than 1.5 mmGreater Than 1.5 mm
Biocompatiblity, ISO 10993-5 (Cytoxicity)Zone of Inhibition 0.25 cmN/A - Not Available

{4}------------------------------------------------

510 (k) Summary

Image /page/4/Picture/1 description: The image shows the logo for Denali Corporation. The logo includes a stylized graphic of a blue sphere orbiting a geometric shape. The text "denali corporation" is next to the graphic, and below that is the address "134 Old Washington Street, Hanover, MA 02339 USA".

T 781.826.9190 F 781.826.4465 denalicorporation.com

The differences between Calcium Bridge and the predicate product in the above Table are that of radiopacity and compressive strength. In this case, Calcium Bridge is more radiopague than the predicate device. Calcium Bridge was also found to have a higher compressive strength than the predicate product. These differences do not change the performance or usage between the subject and the predicate device.

BIOCOMPATIBILITYCalcium Bridge was found to be biocompatible according to ISO 10993-5"Evaluation of Medical Devices-Cytotoxicity", and is consistent in formulationto Tetric EvoCeram, in its intended use and biocompatibility.
SUBSTANTIALEQUIVALENCEDETERMINATION ANDSUMMARYCalcium Bridge is substantially equivalent in design, composition, biocompatibility(ISO 10993-5, Evaluation of Medical Devices - Cytotoxicty), performance, ageing,and intended use to Tetric EvoCeram. This assessment is based upon a comparisonof the physical and mechanical characteristics (tensile strength, compressivestrength), depth of cure, setting time, radio-opacity, and fluorecesing ability.The results of these tests are equivalent to the values of Tetric EvoCeram.
CONCLUSIONSDifferences in the Indications For Use between the predicate device and the subjectdevice reflect the current usage of the term "flowable composite". A flowablecomposite is used today for the predicate device's Indications For Use. The usage ofthe term flowable composite does not effect any usage or performance differencebetween the subject and predicate device. Calcium Bridge has been found to besubstantially equivalent in design, composition, ageing, biocompatibility,performance, and intended use to Tetric EvoCeram.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.