K Number
K153150

Validate with FDA (Live)

Date Cleared
2015-11-24

(22 days)

Product Code
Regulation Number
866.4750
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NOVA View® Automated Fluorescence Microscope is an automated system consisting of a fluorescence microscope and software that acquires, analyzes, stores and displays digital images of stained indirect immunofluorescent slides. It is intended as an aid in the detection and classification of certain antibodies by indirect immunofluorescence technology. The device can only be used with cleared or approved in vitro diagnostic assays that are indicated for use with the device. A trained operator must confirm results generated with the device.

Device Description

NOVA View is an automated fluorescence microscope that acquires, analyses, stores and displays digital images of stained indirect immunofluorescent slides.

The NOVA View AUTOLoader is an optional hardware accessory that performs the automated transfer of slide carriers to and from NOVA View, thereby providing a continuous load capability without human interaction.

AUTOLoader hardware components consist of a NOVA View alignment base, 3-position stack base, 3 slide carrier stacks (labelled as "Pending", "Completed", "and "Error"), telescoping arm with rotary gripper, and a 2D barcode scanner station. The AUTOLoader can be connected to up to two NOVA View devices.

AI/ML Overview

Below is a summary of the acceptance criteria and study information for the NOVA View® Automated Fluorescence Microscope with AUTOLoader, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes the NOVA View® Automated Fluorescence Microscope with AUTOLoader as an aid in the detection and classification of certain antibodies. It explicitly states that "A trained operator must confirm results generated with the device." This implies that the device is not intended as a standalone diagnostic tool, but rather as one that assists a human expert.

Given this context and the fact that this is a Special 510(k) submission for the addition of an AUTOLoader, the primary focus of the performance evaluation appears to be on demonstrating that the addition of the AUTOLoader does not negatively impact the existing functionality of the NOVA View, and that the AUTOLoader itself performs its mechanical tasks reliably and accurately.

The document does not provide specific numerical acceptance criteria or performance metrics (e.g., sensitivity, specificity, accuracy) for the diagnostic performance of the device (i.e., its ability to correctly detect and classify antibodies). Instead, it focuses on demonstrating that the system functionality is maintained with the addition of the AUTOLoader.

Acceptance Criteria for AUTOLoader Functionality (Inferred):

Acceptance Criteria CategoryReported Device Performance
Software FunctionalityNew AUTOLoader module interfaces correctly with NOVA View software (version 2.1.4). Regression testing performed to confirm no change in NOVA View functionality.
Automated Slide Handling (Loading)AUTOLoader successfully picks up slide carriers from "Pending" stack and places them on NOVA View stage.
Automated Barcode ScanningAUTOLoader captures images of barcodes on slides.
Automated Slide Handling (Unloading)AUTOLoader successfully picks up slide carriers after scanning and transports them to "Completed" or "Error" stack.
Continuous Load CapabilityAUTOLoader provides continuous load capability without human interaction for multiple carriers.
Integration with NOVA ViewAUTOLoader connects to NOVA View and operates as an integrated system.

Note on Diagnostic Performance: The document explicitly states "Analytical performance characteristics n/A" and "Clinical performance n/a." This indicates that this specific submission for the AUTOLoader addition did not involve new analytical or clinical performance studies related to the diagnostic accuracy of antibody detection. The regulatory submission leverages the established performance of the predicate device (NOVA View without AUTOLoader) for its core diagnostic function.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in the traditional sense for evaluating diagnostic performance. The studies described are focused on the functionality of the AUTOLoader and the integration of new software.

  • Software Testing: "All new functions were tested during software verification, and regression testing has been performed to demonstrate that NOVA View functionality has not changed." The sample size (number of test cases, scenarios, etc.) for this software testing is not provided.
  • AUTOLoader Mechanical Testing: "This procedure [automated handling of slide carriers] is automatically repeated with the rest of the carriers that are in the Pending stack." While the number of carriers per stack (up to 12) is mentioned, the total number of carriers or repeated cycles used for testing the AUTOLoader's mechanical reliability is not specified.
  • Data Provenance: Not explicitly stated as the primary focus is on system functionality and software verification rather than a clinical dataset.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable in the context of the studies described. The "ground truth" here pertains to the correct functioning of the AUTOLoader and software, which is evaluated through verification and validation processes rather than expert clinical consensus on diagnostic outcomes. The device's diagnostic "results" still require confirmation by a "trained operator."

4. Adjudication Method for the Test Set

Not applicable for the described functionality and software verification testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No MRMC comparative effectiveness study is mentioned. The device is intended as an "aid" for a "trained operator" who "must confirm results." This inherently suggests a human-in-the-loop system, but a formal study comparing human performance with and without the device's assistance is not part of this specific submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No standalone performance study is explicitly described. The "Indications for Use" clearly state that "A trained operator must confirm results generated with the device," indicating it's not a standalone diagnostic device.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

For the specific performance studies described in this submission (AUTOLoader addition):

  • Functional 'Ground Truth': The expected operational behavior of the AUTOLoader (e.g., correctly picking up, barcode scanning, placing carriers) and the established, unchanged functionality of the NOVA View software. This is assessed through engineering and software verification standards rather than clinical ground truth types.

8. The Sample Size for the Training Set

Not applicable. The document does not describe any machine learning or AI components that would require a "training set" in the conventional sense for diagnostic algorithm development. The software update is for controlling the AUTOLoader and preserving existing NOVA View functionality.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set for AI/ML was mentioned.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure, possibly representing the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 12, 2016

INOVA DIAGNOSTICS, INC. GABRIELLA LAKOS DIRECTOR OF ASSAY DEVELOPMENT 9900 OLD GROVE ROAD SAN DIEGO CA 92131

Re: K153150

Trade/Device Name: NOVA View automated fluorescence microscope with AUTOLoader Regulation Number: 21 CFR 866.4750 Regulation Name: Automated indirect immunofluorescence microscope and software-assisted system

Regulatory Class: II Product Code: PIV Dated October 29, 2015 Received: November 2, 2015

Dear Dr. Lakos:

This letter corrects our substantially equivalent letter of November 24, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809]), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kelly Oliner -S

FOR

Leonthena R. Carrington, MS, MBA, MT (ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health (OIR) Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

NOVA View® Automated Fluorescence Microscope with AUTOLoader

Indications for Use (Describe)

NOVA View® Automated Fluorescence Microscope is an automated system consisting of a fluorescence microscope and software that acquires, analyzes, stores and displays digital images of stained indirect immunofluorescent slides. It is intended as an aid in the detection and classification of certain antibodies by indirect immunofluorescence technology. The device can only be used with cleared or approved in vitro diagnostic assays that are indicated for use with the device. A trained operator must confirm results generated with the device.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Inova Diagnostics

Special 510(k) Summary

NOVA View® Automated Fluorescence Microscope with AUTOLoader

510(k) Summary

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Table of Contents

1.Submitter
2.Purpose of submission
3.Devices in the submission
4.Primary contact
5.Secondary contact
6.Product Information and Classification
6.1Proprietary and established names
6.2Regulatory information
7.Intended Use
8.Indications for Use
9.Predicate device
10. Substantial equivalence information
11. Regulatory history
12. Type of the Product
13. System Description:
13.1.Modes of Operation
13.2.Software
14. Device description
15. Principle of the Method and Summary of the Procedure
16. Analytical performance characteristics
17. Clinical performance

{4}------------------------------------------------

1. Submitter

Inova Diagnostics, Inc 9900 Old Grove Road, San Diego, CA, 92131

2. Purpose of submission

Special (510) k submission.

The reason for submission is the addition of an optional hardware accessory, AUTOLoader to the NOVA View device.

3. Devices in the submission

NOVA View Automated Fluorescence Microscope with AUTOLoader

4. Primary contact

Gabriella Lakos, Director of Assay development Inova Diagnostics, Inc 9900 Old Grove Road, San Diego, CA, 92131 Phone: 858-586-9900/1393 email: glakos@inovadx.com

5. Secondary contact

Ronda Elliott, VP of Quality Systems and Regulatory Affairs Inova Diagnostics, Inc. 9900 Old Grove Road, San Diego, CA, 92131 Phone: 858-586-9900/1381 Fax: 858-863-0025/1351 email: relliott@inovadx.com

{5}------------------------------------------------

6. Product Information and Classification

6.1 Proprietary and established names

Proprietary name:NOVA View® with AUTOLoader
Common name:Automated Fluorescence Microscope with AUTOLoader
Classification name:Automated indirect immunofluorescence microscope and softwareassisted system for clinical use

6.2 Regulatory information

Regulation DescriptionAutomated indirect immunofluorescence microscope and software-assisted system for clinical use
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodePIV
Regulation Number866.4750
Device ClassII

7. Intended Use

NOVA View Automated Fluorescence Microscope is an automated system consisting of a fluorescence microscope and software that acquires, analyzes, stores and displays digital images of stained indirect immunofluorescent slides. It is intended as an aid in the detection of certain antibodies by indirect immunofluorescence technology. The device can only be used with cleared or approved in vitro diagnostic assays that are indicated for use with the device. A trained operator must confirm results generated with the device.

8. Indications for Use

NOVA View Automated Fluorescence Microscope is an automated system consisting of a fluorescence microscope and software that acquires, analyzes, stores and displays digital images of stained indirect immunofluorescent slides. It is intended as an aid in the detection of certain antibodies by indirect immunofluorescence technology. The device can only be used with cleared or approved in vitro diagnostic assays that are indicated for use with the device. A trained operator must confirm results generated with the device.

{6}------------------------------------------------

9. Predicate device

NOVA View Automated Fluorescence Microscope

10. Substantial equivalence information

Comparison to the predicate device
ItemNOVA View with AUTOLoaderPredicate Device: NOVA View
Intended useNOVA View is an automated systemconsisting of a fluorescence microscope andsoftware that acquires, analyses, stores anddisplays digital images of stained indirectimmunofluorescent slides. It is intended as anaid for the trained operator in the detectionand classification of certain antibodies withthe indirect immunofluorescent technology.NOVA View can only be used with reagentsthat have been approved for use on thesystem. Results generated by the NOVA Viewmust be confirmed by trained operator.Same
TechnologyAutomated indirect immunofluorescencemicroscope and software-assisted system forclinical useSame

11. Regulatory history

De novo submission: DEN140039

12. Type of the Product

Device technology: indirect immunofluorescence and computer controlled digital imaging and image analysis.

13. System Description

Modes of Operation 13.1.

The instrument does not process samples. The instrument acquires digital images of representative areas of stained indirect immunofluorescent slides can be loaded onto the device either manually, or by the AUTOLoader. The use of AUTOLoader is optional.

{7}------------------------------------------------

13.2. Software

The NOVA View instrument is operated by the NOVA View software. Software version is 2.1.4. The AUTOLoader is controlled by the NOVA View software. A module ("AUTOLoader" module) was added to the NOVA View software. This module interfaces between the NOVA View software and the AUTOLoader firmware. In the AUTOLoader tab the user can view the AUTOLoader log files, error messages (if applicable) and can interact with the AUTOLoader. All new functions were tested during software verification, and regression testing has been performed to demonstrate that NOVA View functionality has not changed.

Level of Concern

Based on FDA Guidance, "Guidance for the Contents of Premarket Submissions for Software Contained in Medical Devices" the NOVA View software 2.1.4 has "Moderate" level of concern.

14. Device description

NOVA View is an automated fluorescence microscope that acquires, analyses, stores and displays digital images of stained indirect immunofluorescent slides.

The NOVA View AUTOLoader is an optional hardware accessory that performs the automated transfer of slide carriers to and from NOVA View, thereby providing a continuous load capability without human interaction.

AUTOLoader hardware components consist of a NOVA View alignment base, 3-position stack base, 3 slide carrier stacks (labelled as "Pending", "Completed", "and "Error"), telescoping arm with rotary gripper, and a 2D barcode scanner station. The AUTOLoader can be connected to up to two NOVA View devices.

15. Principle of the Method and Summary of the Procedure

Slides (from NOVA Lite DAPI ANA Kit) are placed in 5-slide carriers fit on the microscope stage of NOVA View. Carriers to be scanned on the NOVA View are placed in the Pending stack of the AUTOLoader. One stack can accommodate up to 12 carriers. The AUTOLoader is started by the initiation of the NOVA View software. The AUTOLoader picks up the first slide carrier from the Pending stack, transports it to the barcode station, takes off the slide carrier lid, captures an image of the barcodes on the slides, re-places the slide carrier lid, and then proceeds to place the carrier on the NOVA View stage for scanning. After scanning is complete, the AUTOLoader picks up the slide carrier and transports it to the Completed or Error slide carrier stack (depending on the scanning). This procedure is automatically repeated with the rest of the carriers that are in the Pending stack.

{8}------------------------------------------------

16. Analytical performance characteristics

n/A

17. Clinical performance

n/a

§ 866.4750 Automated indirect immunofluorescence microscope and software-assisted system.

(a)
Identification. An automated indirect immunofluorescence microscope and software assisted system is a device that acquires, analyzes, stores, and displays digital images of indirect immunofluorescent slides. It is intended to be used as an aid in the determination of antibody status in clinical samples. The device may include a fluorescence microscope with light source, a motorized microscope stage, dedicated instrument controls, a camera, a computer, a sample processor, or other hardware components. The device may use fluorescent signal acquisition and processing software, data storage and transferring mechanisms, or assay specific algorithms to suggest results. A trained operator must confirm results generated with the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The labeling for the device must reference legally marketed assays intended for use with the device.
(2) Premarket notification submissions must include the following information:
(i) A detailed description of the device that includes:
(A) A detailed description of instrumentation and equipment, and illustrations or photographs of non-standard equipment or methods, if applicable;
(B) Detailed documentation of the software, including, but not limited to, stand-alone software applications and hardware-based devices that incorporate software, if applicable;
(C) A detailed description of appropriate internal and external quality controls that are recommended or provided. The description must identify those control elements that are incorporated into the recommended testing procedures;
(D) Detailed description and specifications for sample preparation, processing, and storage, if applicable;
(E) Methodology and protocols for detecting fluorescence and visualizing results; and
(F) Detailed specification of the criteria for test results interpretation and reporting.
(ii) Data demonstrating the performance characteristics of the device, which must include:
(A) A comparison study of the results obtained with the conventional manual method (
i.e., reference standard), the device, and the reading of the digital image without aid of the software, using the same set of patient samples for each. The study must use a legally marketed assay intended for use with the device. Patient samples must be from the assay-specific intended use population and differential diagnosis population. Samples must also cover the assay measuring range, if applicable;(B) Device clinical performance established by comparing device results at multiple U.S. sites to the clinical diagnostic standard used in the United States, using patient samples from the assay-specific intended use population and the differential diagnosis population. For all samples, the diagnostic clinical criteria and the demographic information must be collected and provided. Clinical validation must be based on the determination of clinical sensitivity and clinical specificity using the test results (
e.g., antibody status based on fluorescence to include pattern and titer, if applicable) compared to the clinical diagnosis of the subject from whom the clinical sample was obtained. The data must be summarized in tabular format comparing the result generated by automated, manual, and digital only interpretation to the disease status;(C) Device precision/reproducibility data generated from within-run, between-run, between-day, between-lot, between-operator, between-instruments, between-site, and total precision for multiple nonconsecutive days (as applicable) using multiple operators, multiple instruments and at multiple sites. A well-characterized panel of patient samples or pools from the associated assay specific intended use population must be used;
(D) Device linearity data generated from patient samples covering the assay measuring range, if applicable;
(E) Device analytical sensitivity data, including limit of blank, limit of detection, and limit of quantitation, if applicable;
(F) Device assay specific cutoff, if applicable;
(G) Device analytical specificity data, including interference by endogenous and exogenous substances, if applicable;
(H) Device instrument carryover data, if applicable;
(I) Device stability data including real-time stability under various storage times and temperatures, if applicable; and
(J) Information on traceability to a reference material and description of value assignment of calibrators and controls, if applicable.
(iii) Identification of risk mitigation elements used by the device, including description of all additional procedures, methods, and practices, incorporated into the directions for use that mitigate risks associated with testing.
(3) Your 21 CFR 809.10 compliant labeling must include:
(i) A warning statement that reads “The device is for use by a trained operator in a clinical laboratory setting”;
(ii) A warning statement that reads “All software-aided results must be confirmed by the trained operator”;
(iii) A warning statement that reads “This device is only for use with reagents that are indicated for use with the device”; and
(iv) A description of the protocol and performance studies performed in accordance with paragraph (b)(2)(ii) of this section and a summary of the results, if applicable.