K Number
K153016

Validate with FDA (Live)

Date Cleared
2016-10-20

(371 days)

Product Code
Regulation Number
884.5300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

aoni Natural Rubber Latex Condom [Smooth/Dotted/Combined]

a) Smooth 52mm/54mm/55mm natural

b) Dotted 52mm natural

c) Combined [smooth+girdling] 52mm natural

d) Combined [dotted+girdling] 52mm natural

The Condom is used for contraception and for prophylactic purpose ( to help prevent pregnancy and the transmission of sexually transmitted infections.)

Device Description

aoni Natural Rubber Latex Condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. These condoms conform to ISO 4074:2002 Natural Rubber Latex Condoms – Requirements and Test Methods, ASTM D3492-14 Standard Specification for Rubber Contraceptives (Male Condom), and ISO 10993-1 Biological Evaluation of Medical Devices - Part 1(2009). These condoms have repeated contact potential for a duration of <24 hours and contact the mucosal membrane. They are for home use.

aoni Natural Rubber Latex Condom include four types as following:

[Smooth]: aoni Natural rubber latex condom [Smooth] are straight walled bulbous ended with smooth surface, having a nominal length of 180mm, a nominal width of 52mm/54mm/55mm, a nominal thickness of 0.036-0.080mm, Natural color

[Dotted]: aoni Natural rubber latex condom [Dotted] are straight walled bulbous ended with Dotted surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

[Combined: (smooth+girdling)]: aoni Natural rubber latex condom [Combined: (smooth+girdling)] are girdling walled bulbous ended with smooth surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

[Combined: (dotted+girdling)]: aoni Natural rubber latex condom [Combined: (dotted+girdling)] are girdling walled bulbous ended with dotted surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

AI/ML Overview

Acceptance Criteria and Device Performance for aoni Natural Rubber Latex Condom

This document summarizes the acceptance criteria and reported device performance for the "aoni Natural Rubber Latex Condom," as described in the provided 510(k) summary (K153016).


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the aoni Natural Rubber Latex Condom are based on recognized international and national standards for rubber contraceptives and biocompatibility. The reported device performance indicates that the product met these specifications.

Table 1: Acceptance Criteria and Reported Device Performance

Parameter / TestAcceptance CriteriaReported Device PerformanceStandard(s) Applied
Physical Performance Tests
Water Leakage (Freedom from Holes)Met specifications of ISO 4074:2002 and ASTM D3492-14"met the specifications"ISO 4074:2002, ASTM D3492-14
Burst Volume and PressureMet specifications of ISO 4074:2002 and ASTM D3492-14"met the specifications"ISO 4074:2002, ASTM D3492-14
Dimensions (Length, Width, Thickness)Within specified ranges (e.g., Nominal Length 180mm, Width 52/54/55mm, Thickness 0.036-0.080mm)Conformed to nominal dimensions and thickness rangesISO 4074:2002, ASTM D3492-14
Biocompatibility Tests
CytotoxicityMet acceptance criteria of ISO 10993-5:2009"passed each biocompatibility test"ISO 10993-5:2009
Skin IrritationMet acceptance criteria of ISO 10993-10:2010"passed each biocompatibility test"ISO 10993-10:2010
Vaginal IrritationMet acceptance criteria of ISO 10993-10:2010"passed each biocompatibility test"ISO 10993-10:2010
SensitizationMet acceptance criteria of ISO 10993-10:2010"passed each biocompatibility test"ISO 10993-10:2010
Acute Systemic ToxicityMet acceptance criteria of ISO 10993-11:2006"passed each biocompatibility test"ISO 10993-11:2006
Shelf Life5 years (60 months)"has a five years (60 month) shelf life based on the results of real time storage and testing"21 CFR 801.435, ISO 4074:2012 (Note: Document specifies 2012, while performance tests mention 2002 version for ISO 4074)

Note on Standards: The document refers to ISO 4074:2002 for performance testing and ISO 4074:2012 for shelf life. Both are relevant versions of the standard for natural rubber latex condoms.


2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical testing for performance and biocompatibility. It does not refer to a "test set" in the context of clinical studies with human subjects.

  • Sample Size for performance and biocompatibility testing: The specific sample sizes for each non-clinical test (e.g., number of condoms for burst testing, number of animals for toxicity tests) are not explicitly stated in the provided 510(k) summary. The summary only broadly states that "Each variant of aoni Natural Rubber Latex Condom was tested".
  • Data Provenance: The testing was conducted by Guangzhou Daming United Rubber Products Ltd. or its contracted laboratories. The information does not specify the country of origin for the data for individual tests beyond the manufacturer's location in Guangzhou, China. All testing appears to be retrospective (i.e., conducted on finished products to demonstrate compliance with standards) rather than prospective clinical trials.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This section is not applicable as the evaluation described is for ex-vivo device performance and biocompatibility testing against established standards, not a clinical study requiring expert assessment of "ground truth" for a test set (e.g., diagnosis of disease). The "ground truth" in this context is defined by the technical specifications and requirements outlined in the referenced ISO and ASTM standards.


4. Adjudication Method for the Test Set

This section is not applicable. The evaluation described involves objective physical and chemical testing against pre-defined performance criteria in established standards, not interpretation by human adjudicators.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC comparative effectiveness study was not performed as this device (a condom) is not a diagnostic device that relies on human interpretation for an outcome. The evaluation is based on objective physical and chemical tests.


6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

This section is not applicable. The device is a physical medical device (condom), not an algorithm or AI-driven system. Therefore, standalone algorithm performance is not relevant.


7. Type of Ground Truth Used

The "ground truth" for demonstrating device compliance is defined by:

  • Performance Specifications: Quantitative and qualitative requirements outlined in ISO 4074: Natural Rubber Latex Condoms – Requirements and Test Methods and ASTM D3492-14 Standard Specification for Rubber Contraceptives (Male Condom). These standards define acceptable ranges for parameters like burst volume and pressure, freedom from holes, and dimensions.
  • Biocompatibility Standards: Criteria for various biological responses (cytotoxicity, irritation, sensitization, systemic toxicity) outlined in ISO 10993-1: Biological Evaluation of Medical Devices - Part 1, and its sub-parts (e.g., ISO 10993-5, ISO 10993-10, ISO 10993-11).

In essence, the ground truth is derived from established scientific and engineering standards and accepted laboratory test methods.


8. Sample Size for the Training Set

This section is not applicable. As a physical device, there is no "training set" in the context of machine learning or AI models. The manufacturing process is validated (and products routinely tested) to ensure consistent quality and adherence to specifications.


9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the reasons stated in point 8. The "ground truth" for manufacturing a condom to meet specifications is established through the development and validation of manufacturing processes to consistently produce a product that passes the non-clinical tests detailed above.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 20, 2016

Guangzhou Daming United Rubber Products Ltd. Yongbing Xie Chairman Secretary No. 1 Daming Street, Port Industrial Zone Huada District Guangzhou, Guangdong 510802 CHINA

Re: K153016 Trade/Device Name: aoni Natural Rubber Latex Condom Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: HIS Dated: August 17, 2016 Received: September 21, 2016

Dear Yongbing Xie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K153016

Device Name

aoni Natural Rubber Latex Condom [Smooth/Dotted/Combined]

Indications for Use (Describe)

aoni Natural Rubber Latex Condom [Smooth/Dotted/Combined]

a) Smooth 52mm/54mm/55mm natural

b) Dotted 52mm natural

c) Combined [smooth+girdling] 52mm natural

d) Combined [dotted+girdling] 52mm natural

The Condom is used for contraception and for prophylactic purpose ( to help prevent pregnancy and the transmission of sexually transmitted infections.)

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) SUMMARY K153016

Submitted by:

Contact Person: Date Prepared: Proprietary Name: Common Name: Classification Name: Product Code: Classification: Predicate Device:

Description of the Device:

Guangzhou Daming United Rubber Products Ltd. NO.1 Daming Street. Port Industrial Zone Huadu District Guangzhou, Guangdong Province 510802 China Phone: 086-020-22680888-8296 E-mail: dm500@ausny.com Yongbing Xie 20 October, 2016 aoni Natural Rubber Latex Condom Condom Condom (21 CFR §884.5300) HIS - Condom Class II LifeStyles Zero Lubricated Latex Condom 510(k) # K120394 Predicate has not been subject to a design-related recall aoni Natural Rubber Latex Condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. These condoms conform to ISO 4074:2002 Natural Rubber Latex Condoms – Requirements and Test Methods, ASTM D3492-14 Standard Specification for Rubber Contraceptives (Male Condom), and ISO 10993-1 Biological Evaluation of Medical Devices - Part 1(2009). These condoms have repeated contact potential for a duration of <24 hours and contact the mucosal membrane. They are for home use.

aoni Natural Rubber Latex Condom include four types as following:

[Smooth]: aoni Natural rubber latex condom [Smooth] are straight walled bulbous ended with smooth surface, having a nominal length of 180mm, a nominal width of 52mm/54mm/55mm, a nominal thickness of 0.036-0.080mm, Natural color[Dotted]: aoni Natural rubber latex condom [Dotted] are straight walled bulbous ended with Dotted surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

[Combined: (smooth+girdling)]: aoni Natural rubber latex condom [Combined: (smooth+girdling)] are girdling walled bulbous ended with smooth surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

[Combined: (dotted+girdling)]: aoni Natural rubber latex condom

{4}------------------------------------------------

[Combined: (dotted+girdling)] are girdling walled bulbous ended with dotted surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.This condom has the same intended use as the predicate condom.The condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections.)
Intended Use of the Device:

Technological Characteristics

The condom is substantially equivalent to the predicate device, the Latex lubricated condom (K120394).

The indication for use statement for aoni Natural Rubber Latex Condom is consistent with the Latex lubricated condom (K120394), additionally, the design is in conformance with ASTM Latex Condom Standard D3492, and that the condom is made of natural rubber latex. The only differences between our condom and predicate condom (K120394) is condom dimensions, type of lubricant, and fragrance.

Predicate Product Comparison

ParametersGuangzhou Daming United RubberProducts Ltd.Ansell Healthcare Product, LLC
510(k) Number-K120394
Intended Useaoni Natural Rubber Latex Condom isused for contraception and forprophylactic purposes (to help preventpregnancy and the transmission ofsexually transmittedinfections.)LifeStyles Zero Condoms is used forcontraception and for prophylacticpurposes (to help prevent pregnancyand the transmission of sexuallytransmitted diseases.)
MaterialNatural rubber latexNatural rubber latex
DesignStraight-walled, smooth, cylindrical,reservoirStraight-walled, cylindrical, reservoir
Nominal Length180mm185mm
Nominal width52mm , 54mm , 55mm51mm
Nominalthickness0.036-0.080mm0.04mm
ColorNaturalNatural
ScentedYesNo
LubricantWater-based lubricantSilicone (silicone oil)
BiocompatibilityISO 10993Acceptance criteria metISO 10993Acceptance criteria met
Table 2 Predicate Product Comparisons - aoni Natural rubber latex condom [Dotted]
ParametersGuangzhou Daming United RubberProducts Ltd.Ansell Healthcare Product, LLC

Table 1 Predicate Product Comparisons - aoni Natural rubber latex condom [Smooth]

{5}------------------------------------------------

510(k) Number-K120394
Intended Useaoni Natural Rubber Latex Condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmittedinfections.)LifeStyles Zero Condoms is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases.)
MaterialNatural rubber latexNatural rubber latex
DesignStraight-walled, dotted, cylindrical, reservoirStraight-walled, cylindrical, reservoir
Nominal Length180mm185mm
Nominal width52mm51mm
Nominal thickness0.036-0.080mm0.04mm
ColorNaturalNatural
ScentedYesNo
LubricantWater-based lubricantSilicone (silicone oil)
BiocompatibilityISO 10993Acceptance criteria metISO 10993Acceptance criteria met

Table 3 Predicate Product Comparisons - aoni Natural rubber latex condom [Combined:

ParametersGuangzhou Daming United Rubber Products Ltd.Ansell Healthcare Product, LLC
510(k) Number-K120394
Intended Useaoni Natural Rubber Latex Condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmittedinfections.)LifeStyles Zero Condoms is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases.)
MaterialNatural rubber latexNatural rubber latex
DesignStraight-walled with girdling, smooth, reservoirStraight-walled, cylindrical, reservoir
Nominal Length180mm185mm
Nominal width52mm51mm
Nominal thickness0.036-0.080mm0.04mm
ColorNaturalNatural
ScentedYesNo

{6}------------------------------------------------

LubricantWater-based lubricantSilicone (silicone oil)
BiocompatibilityISO 10993Acceptance criteria metISO 10993Acceptance criteria met

Table 4 Predicate Product Comparisons - aoni Natural rubber latex condom [Combined: dotted+airdling]

ParametersGuangzhou Daming United RubberProducts Ltd.Ansell Healthcare Product, LLC
510(k) Number-K120394
Intended Useaoni Natural Rubber Latex Condom isused for contraception and forprophylactic purposes (to help preventpregnancy and the transmission ofsexually transmittedinfections.)LifeStyles Zero Condoms is used forcontraception and for prophylacticpurposes (to help prevent pregnancyand the transmission of sexuallytransmitted diseases.)
MaterialNatural rubber latexNatural rubber latex
DesignStraight-walled with girdling, dotted,reservoirStraight-walled, cylindrical, reservoir
Nominal Length180mm185mm
Nominal width52mm51mm
Nominalthickness0.036-0.080mm0.04mm
ColorNaturalNatural
ScentedYesNo
LubricantWater-based lubricantSilicone (silicone oil)
BiocompatibilityISO 10993Acceptance criteria metISO 10993Acceptance criteria met

Non-clinical testing:

Biocompatibility : Biocompatibility testing was performed to evaluate the biocompatibility of the aoni Natural Rubber Latex Condom in accordance with ISO 10993-1:2009, the device passed each biocompatibility test indentified below:

  • Cytotoxicity testing in according to ISO 10993-5:2009 -
  • Skin irritation testing in according to ISO 10993-10:2010 -
  • -Vaginal irritation testing in according to ISO 10993-10:2010
  • -Sensitization testing in according to ISO 10993-10:2010
  • -Acute systemic toxicity testing in according to ISO 10993-11:2006

Performance testing: Each variant of aoni Natural Rubber Latex Condom was tested and met the specifications of ISO 4074: Natural Rubber Latex Condoms - Requirements and Test Methods, ASTM D3492-14 Standard Specification for Rubber Contraceptives (Male Condom),

{7}------------------------------------------------

Shelf life:

aoni Natural Rubber Latex Condom has a five years (60 month) shelf life based on the results of real time storage and testing of 21 CFR 801.435 and ISO 4074:2012.

Conclusion:

The aoni Natural Rubber Latex Condom is substantially equivalent to the predicate device in indications for use, materials. Safety and performance testing to ASTM D3492-14 Standard Specifications for Rubber Contraceptives (Male Condoms) and ISO 10993 Biological Evaluation of Medical Devices has concluded that the aoni Natural Rubber Latex Condom are substantially equivalent to the predicate devices (K120394) in term of safety and effectiveness.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.