K Number
K152960

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-07-22

(289 days)

Product Code
Regulation Number
882.4360
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ELAN 4 Electro Motor System is intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

Device Description

The ELAN 4 Electro motor system is an electrical motor system consisting of a control unit with different sizes and types of handpieces, each containing its own integrated motor, and attachments such as burrs, saw blades, drills, etc. The system allows for high-speed dissection at up to 80,000 RPM while also allowing low speed cutting between 1,000 and 20,000 RPM. System settings are controlled with the touch screen panel of the control unit, while all functions of the handpieces can be controlled from the sterile field by the foot control.

AI/ML Overview

The provided document details the acceptance criteria and study results for the ELAN 4 Electro Motor System.

1. Table of Acceptance Criteria and Reported Device Performance:

The document presents several tables detailing acceptance criteria and results for different handpieces of the ELAN 4 Electro Motor System. Below is a consolidated representation of these, focusing on the general themes rather than repeating every specific test if the acceptance criteria and results are similar across different components.

Requirement / Test CategoryAcceptance CriteriaReported Device Performance (Results)
Handpieces (GA)
Material change (Titanium alloy to Stainless steel)No new risk to patient. (Verified by testing and biocompatibility testing per ISO 10993-1)Pass. Does not raise new questions of safety and effectiveness.
Material change (PEEK (blue) to PEEK (black))No new risk to patient. (Verified by testing)Pass. Does not raise new questions of safety and effectiveness.
Motor and handpiece has no coupling (one piece)No new risk to patient. (Verified by testing)Pass. Does not raise new questions of safety and effectiveness.
Tool adaptation at distal end of handpieceNo new risk to patient. (Verified by testing)Pass. Does not raise new questions of safety and effectiveness.
Burr release only in off position mechanismNo new risk to patient. (Verified by testing)Pass. Does not raise new questions of safety and effectiveness.
Bench Testing of ELAN 4 Attachments (General)
Contour milling performance (Drill hand pieces GA861 and GA864)As representative of the standard tool portfolio "XS", contour milling is per application and processing cycle with the test pattern to be milled in bone replacement material (ash hardwood). Definition of the tools used.The 120 application cycles, i.e., a contour milling, could be performed without affecting by the test pattern itself. Pass.
Coupling ToolfunctionsTool is manually decoupled (with actuator).120 times "OK", i.e., at 120 cycles of use, the tool was automatically coupled and uncoupled manually. There were no problems with tool coupling. PASS.
Function/Safety after Cleaning, Disinfection, SterilizationNo impairment of function and safety by cleaning, disinfection and steam sterilization.During the preparation and application of 120 cycles, there were no function and safety impairments caused by cleaning, disinfection and steam sterilization. PASS. (Weight remained constant before/after processing cycle).
Overall Functional and Safety Impairments during/after cyclesNo impairment of function and safety during and after at least 100 applications and processing cycles.Since all other 20 acceptance criteria were "satisfied" rated, this also applies to requirement (5). After and during 120 application and processing cycles, there was no functional and safety impairments. PASS.
Label ReadabilityLabels are still readable after min. 100 application and processing cycles.After 120 cycles all relevant information are still identifiable/readable. PASS.
Motor Cable Coupling/Uncoupling EaseMotor cable can be coupled and uncoupled.120 times "OK", i.e. at 120 cycles of use, the motor cables could be easily coupled and uncoupled. PASS.
No Operation Interruption from Cable ReleaseNo interruption of operation by releasing the cable.120 times "no", i.e. at 120 cycles of use, there was no interruption of operation through the cable. PASS.
No Operation Interruption from Electrical Component FailureNo interruption of operation due to failure of electrical components.120 times "no", i.e. at 120 cycles of use, there was no operation interruption due to failure of electrical components. PASS.
No Operation Interruption from Mechanical Component FailureNo interruption of operation due to failure of mechanical components.120 times "no", i.e. at 120 cycles of use, there was no interruption due to failure of mechanical components. PASS.
Tool Securely Locked During UseTool is securely locked during the function test and the use.120 times "no", i.e. at 120 cycles of use the tool during application was securely locked and there was thus no interruption. PASS.
Elan 4 Bench Testing Craniotome Handpiece Summary Table
Tool coupling/uncouplingTool can be consistently coupled and uncoupled.After 123 cycles the tool was automatically coupled and manually decoupled with no problems. PASS.
Reprocessing Procedure (Cleaning, Disinfecting, Steam Sterilization)Reprocessing procedure does not cause any functional or safety impairments.No performance or safety impairments caused by cleaning, disinfecting or sterilization. Weight remained constant. PASS.
Overall Functional and Safety ImpairmentsAll other testing requirements must be passed.Since all other acceptance criteria were satisfied this requirement can be determined to have met acceptance criteria. PASS.
Label ReadabilityLabels and other identifying information remains readable after set amount of reprocessing cycles.After 123 cycles all relevant information are still identifiable/readable. PASS.
Motor Cable ConsistencyMotor cable consistently couples and uncouples.After 123 cycles motor cables are easily coupled/uncoupled. PASS.
No Cable Release Safety/Function ImpairmentNo safety or function impairments due to cable release.After 123 cycles there was no interruption of operation due to cable release. PASS.
No Electrical Component FailureNo safety or function impairments due to failure of electrical components.After 123 cycles there was no interruption of operation due to failure of electrical components. PASS.
No Mechanical Component FailureNo safety or function impairments due to failure of mechanical components.After 123 cycles there was no interruption of operation due to failure of mechanical components. PASS.
Tool Lock Mechanism IntegrityNo safety or function impairments due to failure or tool lock mechanism.After 123 cycles the tool was securely locked and there was no interruption. PASS.
Temperature Measurement (Hand-piece)Temperature remains in acceptable range throughout testing.10th, 51st, and 101st cycle are all PASS.
Elan 4 Bench Testing Perforator Summary Table
Sterilization TestReprocessing does not result in any functional or safety issues.PASS.
Bracket System TestMotors held securely during processing.PASS.
Function TestTool securely locked and could be safely removed.PASS.
Plug TestingMotor cable was safe, easy and error-free coupling and uncoupling.PASS.
Lubrication TestingLack of lubrication did not cause impairment of function or safety.PASS.
Perforation TestPerforator broke through material and turned off automatically after breaking through.PASS.
Weight and Temperature TestPerforator hand piece maintained a consistent weight and temperature range throughout testing.PASS.
Elan 4 Bench Testing Reciprocating Saw Handpiece Summary Table
Cleaning and Disinfection TestNo functional or safety issues caused by cleaning and disinfection procedures.PASS.
Sterilization TestNo functional or safety issues caused by sterilization procedure.PASS.
Bracket Test (for saw)Saw held securely during reprocessing, transport and storage.PASS.
Lubrication Test (for saw)No impairment of safety or function due to lack of lubrication.PASS.
Function Test (for saw)Tool is static and locked securely. Tool can be safely removed. Oscillation of tool functions. No functional or safety issues as a result of material changes.PASS.
Motor Cable Test (for saw)Motor cable can be coupled and uncoupled safely, easily, and without error.PASS.
Performance Test (Sawing motion)Sawing motion (oscillation) is consistent and cut is at least 6cm.PASS.
Temperature Measurement Test (for saw)Temperature remains in acceptable range throughout testing.PASS.
Elan 4 Bench Testing Sagittal Saw Handpiece Summary Table
Cleaning TestNo function or safety issues as a result of the cleaning and disinfection procedure.PASS.
Sterilization TestNo function or safety issues as a result of the sterilization procedure.PASS.
Bracket Test (for saw)Saw remained secure during reprocessing, transport and storage.PASS.
Lubrication Test (for saw)No safety or function issues due to lack of lubrication.PASS.
Motor Cable Test (for saw)No impairment of the connection and disconnection process.PASS.
Performance Test (Sawing motion)Sawing motion and cut is constant and consistent.PASS.
Temperature Test (for saw)Temperature remains in acceptable range throughout testing.PASS.

2. Sample Size and Data Provenance for the Test Set:

  • Sample Size:
    • For the "ELAN 4 Bench Testing Handpieces Summary Table" and "ELAN 4 Bench Testing Craniotome Handpiece Summary Table," the sample size for most tests was 120-123 cycles.
    • For the "ELAN 4 Bench Testing Perforator Summary Table," "ELAN 4 Bench Testing Reciprocating Saw Handpiece Summary Table," and "Elan 4 Bench Testing Sagittal Saw Handpiece Summary Table," the sample size for most tests was 100 cycles.
  • Data Provenance: The studies described are bench tests, meaning they were conducted in a controlled laboratory environment rather than on human or animal subjects. The document does not specify the country of origin of the data, but the context implies it was conducted by Aesculap, Inc., whose address is in Center Valley, PA, USA. These are prospective tests designed to demonstrate device performance against predefined criteria.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • This information is not provided in the document. The studies are bench tests, and the "ground truth" implicitly refers to engineering specifications and performance criteria that the device is expected to meet. It's not a clinical study requiring human experts to establish diagnostic ground truth.

4. Adjudication Method for the Test Set:

  • This information is not applicable as the studies are bench tests with clearly defined pass/fail criteria based on engineering specifications. There is no mention of human adjudication for these technical performance tests.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not done. The document describes bench testing comparing the ELAN 4 Electro Motor System to its predicate device based on technological characteristics and performance in a controlled environment. It does not involve human readers or AI assistance.

6. Standalone Performance Study (Algorithm only without human-in-the loop performance):

  • Yes, the studies described are analogous to standalone performance studies for mechanical devices. The ELAN 4 Electro Motor System was tested in a standalone capacity (i.e., the device itself, without human intervention in the evaluation of the test results beyond operating the equipment and recording data) against predefined performance criteria. There is no "algorithm" in the sense of AI; it's the mechanical/electrical device's performance that is being evaluated.

7. Type of Ground Truth Used:

  • The ground truth used is primarily engineering specifications and predefined performance criteria. For example, a burr should not release during use, labels should remain readable after a certain number of cycles, and temperature should remain within an acceptable range. These are objective, measurable parameters.

8. Sample Size for the Training Set:

  • This concept is not applicable to the described studies. The document describes verification and validation bench testing for a physical medical device, not a machine learning or AI model that requires a training set.

9. How the Ground Truth for the Training Set was Established:

  • This information is not applicable as it is not a study involving a training set for an AI/machine learning model. The ground truth for the verification and validation (bench testing) was established through engineering specifications and safety requirements relevant to the device's function.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 22, 2016

Aesculap, Inc. Mr. Peter Stoll Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, PA 18034

Re: K152960

Trade/Device Name: ELAN 4 Electro Motor System Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBE Dated: June 3, 2016 Received: June 23, 2016

Dear Mr. Stoll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152960

Device Name ELAN 4 Electro Motor System

Indications for Use (Describe)

The ELAN 4 Electro Motor System is intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over The Counter Use (21 CFR 801 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY (as required by 21 CFR 807.92) ELAN 4 Greet q'Motor System

July 21, 2016

COMPANY:Aesculap®, Inc.3773 Corporate Parkway Center Valley,PA 18034Establishment Registration Number: 2916714
CONTACT:Peter Stoll610-984-9076 (phone)610-791-6882 (fax)
TRADENAME:ELAN 4 Electro Motor System
COMMON NAME:Drills, Burrs, Trephines & Accessories (Simple, Powered)
CLASSIFICATION NAME:Powered simple cranial drills, burrs, trephines and their accessories
CLASSIFICATION PANEL:Neurology
REGULATION NUMBER:21 CFR 882.4310

PRODUCT CODE: HBE

SUBSTANTIAL EQUIVALENCE

The ELAN 4 Electro Motor System is substantially equivalent to Aesculap's Microspeed Uni Motor System cleared via K053526.

DEVICE DESCRIPTION

The ELAN 4 Electro motor system is an electrical motor system consisting of a control unit with different sizes and types of handpieces, each containing its own integrated motor, and attachments such as burrs, saw blades, drills, etc. The system allows for high-speed dissection at up to 80,000 RPM while also allowing low speed cutting between 1,000 and 20,000 RPM. System settings are controlled with the touch screen panel of the control unit, while all functions of the handpieces can be controlled from the sterile field by the foot control.

INDICATIONS FOR USE

The ELAN 4 Electro motor system is intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

TECHNOLOGICAL CHARACTERISTICS (compared to predicates)

The modifications made to the ELAN 4 Electro motor system do not affect the fundamental scientific technology. The design, materials, and principal of operation have not changed for these devices. The ELAN 4 Electro Motor System offers similar operating speeds, power sources, and attachments as the predicate device. A table of the modifications between the predicate and the ELAN 4 Electro is below. The performance of the device was determined to be equivalent or improved for the device, in comparison to the predicate, due to each of the changes. Material changes such as changing to stainless steel or another type of PEEK were evaluated and determined to

{4}------------------------------------------------

present no new risk to patients. Additionally, new features such as the distal tool adaptation and burr release mechanism were determined to present no new risk to patient and improved the performance of the device by reducing temperature and ensuring burrs do not release during use. This comparison supports the claim of substantial equivalence to the predicate device for the proposed intended use.

Device ModificationRiskVerificationActivityAcceptanceCriteriaResults ofVerification
handpieces (GA)
material change fromtitinium alloy to stainlesssteelNo new risk topatient.Verificationtesting andbiocompatibilitytesting per ISO10993-1EngineeringspecificationPass. Does notraise newquestions ofsafety andeffectiveness.
material change fromPEEK (blue) to PEEK(black)No new risk topatient.Verificationtesting (also seeAppendix Abelow)EngineeringspecificationPass. Does notraise newquestions ofsafety andeffectiveness.
motor and handpiece hasno coupling (one piece)no new risk topatientVerificationtesting (also seeAppendix Abelow)EngineeringspecificationPass. Does notraise newquestions ofsafety andeffectiveness.
tool adaption at distal endof handpieceno new risk topatientVerificationtesting (also seeAppendix Abelow)EngineeringspecificationPass. Does notraise newquestions ofsafety andeffectiveness.
burr release only in offposition mechanismno new risk topatientVerificationtesting (also seeAppendix Abelow)EngineeringspecificationPass. Does notraise newquestions ofsafety andeffectiveness.

Comparison To Predicate

SystemElan 4 Electro (subject of this submission)Microspeed Uni
K#K152960K053526
Indications for UseThe Elan 4 Electro motor system is intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.Aesculap's Microspeed Uni is intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Control Unit2 of 11
Maximum80,000 rpm80,000 rpm

{5}------------------------------------------------

speed
Voltage100-240 V100 - 240V
Frequency50-60Hz50 - 60Hz
IrrigationPumpYesYes
Motor
High speedmotorMin/Max10,000/ 80,000 rpm5,000 rpm / 80,000 rpm
Low speedmotorMin/Max1,000/ 20,000 rpm3,000 rpm / 40,000 rpm
MotorrotationHigh speedLeft and right hand rotationLeft and right hand rotation
Low speedLeft and right hand rotationLeft and right hand rotation
MaterialsStainless SteelTitanium Alloy
Features
ControlmechanismFootHand and foot
Ability toconnectvariousHandpiecesYesYes
Motorconnections22
Footswitchconnections11
Control UnitTouch screenTouch screen
AppliedstandardsUL2601-1IEC60601-1IEC60601-1-2IEC62304UL2601-1IEC60601-1IEC60601-1-2

PERFORMANCE DATA

ELAN 4 Electro Motor System has been tested in accordance with FDA recognized standards IEC 60601-1, IEC 60601-1-2, IEC 62304, and UL 2601-1. Bench testing was performed with each handpiece to demonstrate equivalent or improved performance in comparison to the predicate device. Bench testing summaries are below. Each handpiece demonstrated equivalent or improved performance compared to the acceptance criteria.

Bench Testing of ELAN 4 Attachments

Elan 4 Bench Testing Handpieces Summary Table

RequirementAcceptanceSampleResults
contour milling couldbe performed withoutaffecting the testpattern itself.as representative of thestandard tool portfolio "XS"is per application andprocessing cycle with thetest pattern to be milled{Drill hand pieces GA861and GA864} in bonereplacement material {ashhardwood}. Definition ofthe tools used,120 cyclesThe 120 applicationcycles, i.e. "a"contour millingcould be performedwithout affecting bythe test pattern itselfPass
CouplingToolfunctionsTool ismanuallydecoupled {withactuator)120 times120 times "OK", ieat 120 cycles of use,the tool wasautomaticallycoupled anduncoupledmanually. Therewere no problemswith tool coupling.PASS
Determining theweight before theRDG / after theRDG / after theoiling / after theSterilization hasshown that theweight remainedconstant before andjust after the entireprocessing cycle.No impairment of function andsafety by cleaning,Disinfection and steamsterilization120 cyclesDuring the preparationand application of 120cycles, there were nofunction and safetyimpairments caused bycleaning, disinfectionand steam sterilization.PASS
To evaluate theacceptance criteria, allanalysis of all of theother 20 acceptancecriteria can be used.Since all other 20acceptance criteriawere "satisfied" rated,this also applies torequirement (5). Afterand during 120application andprocessing cycles,there was nofunctional and safetyimpairments.No impairment offunction and safetyduring and after at least100 applications andprocessing cycles120 cyclesSince all otheracceptance criteriawere satisfied thisrequirement can bedetermined to havemet acceptancecriteria. PASS
After 120thprocessing andapplication cycles allrelevant information(manufacturer, partnumber, serialnumber) the labelsare still identifiable /readable.Labels are still readableafter min.100 applicationand processing cycles120 cyclesAfter 120 cycles allrelevant informationare stillidentifiable/readable.PASS
120 times "OK", at120 cycles of use,the motor cablescould be easilycoupled anduncoupled.Motor cable can becoupled and uncoupled120 cyclesAfter 120 cycles themotor cables could beeasilycoupled/uncoupled.PASS
120 times "no", ie at120 cycles of use,there was nointerruption ofoperation throughthe cableNo interruption ofoperation by releasingthe cable120 cyclesAfter 120 cycles therewas no interruption ofoperation through thecable. PASS
120 times "no", at120 cycles of use,there was nooperationinterruption due tofailure of electricalcomponentsinterruption of operationdue to failure ofelectrical components120 cyclesAfter 120 cycles nooperation interruptiondue to failure ofelectricalcomponents. PASS
120 times "no", ie at120 cycles of use,there was nointerruption due tofailure of mechanicalcomponents.No interruption ofoperation due to failureof mechanicalcomponents120 cyclesAfter 120 cycles nooperation interruptiondue to failure ofmechanicalcomponents. PASS
120 times "no", at120 cycles of use thetool duringapplication wassecurely locked andthere was thus nointerruptionTool is securely lockedduring the function testand the use120 cyclesAfter 120 cycles thetool was securelylocked and there wasno interruption. PASS

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Elan 4 Bench Testing Craniotome Handpiece Summary Table

RequirementAcceptance CriteriaSample SizeResults
Tools can beattachedautomaticallyand releasedmanuallyTool can beconsistently coupledand uncoupled123 cyclesAfter 123 cycles the toolwas automaticallycoupled and manuallydecoupled with noproblems. PASS
No performance orsafety impairmentswere caused bycleaning,disinfecting or steamsterilizationReprocessing proceduredoes not cause anyfunctional or safetyimpairments123 cyclesNo performance orsafety impairmentscaused by cleaning,disinfecting orsterilization. Weightremained constant
To evaluate theacceptance criteria,all analysis of all ofthe other acceptancecriteria can be used.Since all otheracceptance criteriawere "satisfied"rated, this alsoapplies torequirement (5).After and during 123application andprocessing cycles,there was nofunctional and safetyimpairments.All other testingrequirements must bepassed123 cyclesSince all otheracceptance criteria weresatisfied thisrequirement can bedetermined to have metacceptance criteria.PASS
After 123rdprocessing andapplication cycles allrelevant information(manufacturer, partnumber, serialnumber) the labelsare still identifiable /readable.Labels and otheridentifyinginformation remainsreadable after setamount ofreprocessing cycles123 cyclesAfter 123 cycles allrelevant information arestillidentifiable/readable.PASS
123 times'OK' motorcable couldbe easilycoupled anduncoupledMotor cableconsistently couplesand uncouples123 cyclesAfter 123 cycles motorcables are easilycoupled/uncoupled.PASS
123 times'no' at 123 cycles therewas nointerruptionof operationdue to releaseof cableNo safety or functionimpairments due tocable release123 cyclesAfter 123 cycles therewas no interruption ofoperation due to cablerelease. PASS
123 times'no' at 123 cycles therewas nointerruptionof operationdue to failureof electricalcomponentsNo safety or functionimpairments due tofailure of electricalcomponents123 cyclesAfter 123 cycles therewas no interruption ofoperation due to failureof electricalcomponents. PASS
123 times'no' at 123 cycles therewas nointerruptionof operationdue to failureofmechanicalcomponentsNo safety or functionimpairments due tofailure of mechanicalcomponents123 cyclesAfter 123 cycles therewas no interruption ofoperation due to failureof mechanicalcomponents. PASS
123 times'no' at 123 cycles thetool wassecurelylocked andthere wasthus nointerruptionNo safety or functionimpairments due tofailure or tool lockmechanism123 cyclesAfter 123 cycles the toolwas securely locked andthere was nointerruption. PASS
Temperaturemeasurement:Checktemperatureof hand-pieceat 3 definedtimes duringthe course ofthe 123 cycletestTemperature remainsin acceptable rangethroughout testing123 cycles10th, 51st, and 101stcycle are all PASS

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Elan 4 Bench Testing Perforator Summary Table

RequirementAcceptance CriteriaSample SizeResults
Processing test (See belowfor sub-tests)
Sterilization Test(Processing sub-test 1)Reprocessing does notresult in any functional orsafety issues100 cyclesPASS
Bracket System Test(Processing sub-test 2)Motors held securelyduring processing100 cyclesPASS
Function Test (Processing sub-test 3)Tool securely lockedand could be safelyremoved100 cyclesPASS
Plug Testing(Processing sub-test 4)Motor cable was safe,easy and error-freecoupling anduncoupling100 cyclesPASS
Lubrication Testing(Processing sub-test 5)Lack of lubrication didnot cause impairmentof function or safety100 cyclesPASS
Perforation TestPerforator broke throughmaterial and turned offautomatically afterbreaking through100 cyclesPASS

{12}------------------------------------------------

Weight andTemperature TestPerforator hand piecemaintained a consistentweight and temperaturerange throughout testing100 cyclesPASS
----------------------------------------------------------------------------------------------------------------------------------------------------------

Elan 4 Bench Testing Reciprocating Saw Handpiece Summary Table

RequirementAcceptance CriteriaSample SizeResults
Cleaning andDisinfection TestNo functional or safety issuescaused by cleaning anddisinfection procedures100 cyclesPASS
SterilizationTestNo functional or safety issuescaused by sterilizationprocedure.100 cyclesPASS
Bracket TestSaw held securely duringreprocessing, transport andstorage100 cyclesPASS
Lubrication TestNo impairment of safety orfunction due to lack oflubrication100 cyclesPASS
Function TestTool is static and lockedsecurely. Tool can be safelyremoved. Oscillation of toolfunctions. No functional orsafety issues as a result ofmaterial changes100 cyclesPASS
Motor CableTestMotor cable can be coupledand uncoupled safely, easily,and without error.100 cyclesPASS
PerformanceTestSawing motion (oscillation)is consistent and cut is atleast 6cm.100 cyclesPASS
TemperatureMeasurementTestTemperature remains inacceptable range throughouttesting100 cycles(temperaturemeasured every 20cycles)PASS

{13}------------------------------------------------

RequirementAcceptance CriteriaSample SizeResults
Cleaning TestNo function or safety issues asa result of the cleaning anddisinfection procedure100 cyclesPASS
Sterilization TestNo function or safety issues as aresult of the sterilizationprocedure100 cyclesPASS
Bracket TestSaw remained secure duringreprocessing, transport and100 cyclesPASS
Lubrication TestNo safety or function issuesdue to lack of lubrication100 cyclesPASS
Motor Cable TestNo impairment of theconnection and disconnectionprocess100 cyclesPASS
Performance TestSawing motion and cut isconstant and consistent100 cyclesPASS
Temperature TestTemperature remains inacceptable range throughouttesting100 cyclesPASS

Elan 4 Bench Testing Sagittal Saw Handpiece Summary Table

SUBSTANTIAL EQUIVALENCE CONCLUSION

Bench testing results along with a comparison between the technology and materials used in the ELAN 4 Electro Motor System and the predicate device, demonstrates that the ELAN 4 Electro Motor System meets or exceeds the performance of the predicate device. Bench testing for each handpiece was conducted with pass/fail criteria that would demonstrate equivalent or improved performance in comparison to the predicate as intended in the specified use conditions. The bench testing results and technological comparison demonstrate that the ELAN 4 Electro Motor System performs comparably to the predicate device currently marketed.

§ 882.4360 Electric cranial drill motor.

(a)
Identification. An electric cranial drill motor is an electrically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.(b)
Classification. Class II (performance standards).