K Number
K152959

Validate with FDA (Live)

Date Cleared
2016-04-14

(190 days)

Product Code
Regulation Number
872.3820
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sealapex/Sealapex Xpress is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

Device Description

Sealapex/Sealapex Xpress is a non-eugenol, radiopaque, calcium hydroxide polymeric resin root canal filling material. It is indicated for use as a root canal sealing and filling material, and is used during an endodontic procedure to seal off the prepared root canal apical foramen and tubules from blood, exudates, and infection. The proposed is a two-part, base/catalyst - paste/paste system that is mixed in equal portions. Then, the mixture is carried to the root canal with endodontic obturation points or directly dispensed in the root canal depending on delivery method, e.g. tube or dual-barrel syringe.

The proposed is available packaged in tubes (Sealapex) and dual-barrel syringes (Sealapex Xpress) used with single-use automix tips. The tubes are dispensed onto a mixing pad, hand mixed and then applied to the canal with an endodontic point. Sealapex Xpress has the same formulation and indications for use as Sealapex, but is branded as "Sealapex Xpress" since it offers the convenience of single-use automix tips that remove the need for hand mixing. The dual-barrel and single-use automix tips allow the dental professional to deliver the desired volume of Sealapex Xpress directly into the prepped root canal.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for Sealapex/Sealapex Xpress:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied to be adherence to the ISO 6876:2012 standard for "Dentistry - Root Canal Sealing Materials" and substantial equivalence to the predicate device, Sealapex 4 (K010940).

CharacteristicAcceptance Criteria (Implied)Predicate Device (Sealapex 4) PerformanceProposed Device (Sealapex/Sealapex Xpress) Performance
Physical and Mechanical Properties
Work Time @ 37°C, 100% RH> 60 minutes> 60 minutes> 60 minutes
Setting time @ 37°C, 100% RH< 24 hours< 24 hours< 24 hours
Consistency (Flow) (Internal Test Method)No substantial difference from predicate48.5 mm51.0 mm
Consistency (Flow) (ISO 6876:2012 Method)No substantial difference from predicate22.8 mm23.1 mm
Film ThicknessNot more than 50 µm (ISO 6876:2012)20.3 µm33.85 µm
Solubility and DisintegrationN/A (Not reported, likely considered satisfactory if passed other ISO tests)N/AN/A
Radiopacity> 3.3mm Al (ISO 6876:2012 Minimum)> 3.3mm Al6.0 mm Al

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for the performance tests. It mentions "Performance testing on stability, work time, set time and slump were performed on the modified formula." and "The sealer was tested to meet ISO 6876:2012 standard." and "The predicate and proposed devices produced Flow results..." This implies that samples were tested, but the number of samples tested for each characteristic is not provided.

The data provenance is from non-clinical bench testing. It doesn't specify the country of origin for the data generation, but the manufacturing location for the proposed device is noted as Scafati Salerno, Italy. The predicate device's tests would have been performed by Kerr Corporation in Romulus, MI, USA. The current testing was performed on the "modified formula" which has been on the EU market for several years.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes a material science/bench testing study, not a clinical study involving experts establishing ground truth for a diagnostic device. Therefore, this information is not applicable to this submission. The "ground truth" for this study is defined by the objective, measurable standards set by ISO 6876:2012.

4. Adjudication Method for the Test Set

As this is a bench testing study evaluating physical and mechanical properties against an international standard (ISO 6876:2012) and comparison to a predicate device, an adjudication method for a "test set" in the human-reader sense is not applicable. The results are quantitative measurements against predefined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This submission is for a root canal filling material, and the evaluation focuses on its physical and mechanical properties, not on diagnostic accuracy involving human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

No, a standalone performance study in the context of an algorithm or AI was not done. This device is a material, not a software algorithm.

7. The Type of Ground Truth Used

The "ground truth" used for this study is based on:

  • International Standard Specifications: Specifically, ISO 6876:2012 for "Dentistry - Root Canal Sealing Materials."
  • Predicate Device Performance: The performance characteristics of the legally marketed Sealapex 4 (K010940) served as a benchmark for comparison to demonstrate substantial equivalence.

8. The Sample Size for the Training Set

This type of study for a material science device does not involve a "training set" as would be used for an AI/ML algorithm. Therefore, this question is not applicable.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" in the context of this material science submission, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 14, 2016

Sybron Dental Specialties Jennifer Dzidrums 1332 S. Lone Hill Ave. Glendora, CA 91740

Re: K152959-S001

Trade/Device Name: SEALAPEX/ SEALAPEX XPRESS Regulation Number: 21 CFR 872.3820 Regulation Name: Resin, Root Canal Filling Regulatory Class: Class II Product Code: KIF Dated: March 07, 2016 Received: March 08, 2016

Dear Jennifer Dzidrums:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erin I. Keith -S

Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name Sealapex/Sealapex Xpress

Indications for Use (Describe)

Sealapex / Sealapex Xpress is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart G)

| X | Prescription Use (Part 21 CFR 801 Subpart D)

|_ | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 6. 510(k) SUMMARY FOR SEALAPEX/ SEALAPEX XPRESS

Image /page/3/Picture/2 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters "sds" in a bold, sans-serif font, with a curved line above the letters. Below the letters, the words "Sybron Dental Specialties" are written in a smaller, sans-serif font. The logo is blue and white.

Sealapex/ Sealapex Xpress

    1. Submitter Information:
      Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687

Contact Person: Jennifer Dzidrums Telephone Number: 909-962-5650 909-962-5694 Fax Number:

Date Prepared: April 8, 2016

    1. Device Name:
· Proprietary Name:Sealapex and Sealapex Xpress
· Classification Name:Resin, Root Canal Filling
  • 872.3820 • CFR Number:
  • · Device Class: II
  • Product Code: KIF

3. Predicate Device:

The proposed Sealapex Sealapex Xpress is substantially equivalent to the legally marketed device, Sealapex 4 (K010940), cleared on June 20, 2001, product code KIF.

4. Description of Device:

Sealapex/Sealapex Xpress is a non-eugenol, radiopaque, calcium hydroxide polymeric resin root canal filling material. It is indicated for use as a root canal sealing and filling material, and is used during an endodontic procedure to seal off the prepared root canal apical foramen and tubules from blood, exudates, and infection. The proposed is a twopart, base/catalyst - paste/paste system that is mixed in equal portions. Then, the mixture is carried to the root canal with endodontic obturation points or directly dispensed in the root canal depending on delivery method, e.g. tube or dual-barrel syringe.

The proposed is available packaged in tubes (Sealapex) and dual-barrel syringes (Sealapex Xpress) used with single-use automix tips. The tubes are dispensed onto a mixing pad, hand mixed and then applied to the canal with an endodontic point. Sealapex Xpress has the same formulation and indications for use as Sealapex, but is branded as "Sealapex Xpress" since it offers the convenience of single-use automix tips that remove

{4}------------------------------------------------

the need for hand mixing. The dual-barrel and single-use automix tips allow the dental professional to deliver the desired volume of Sealapex Xpress directly into the prepped root canal.

Accessories Used with SealapexXpressManufacturer of Accessory
Dual-barrel syringeSulzer Mixpac AGRuetistrasse 7Haag Sankt gallen, Switzerland9469
Auto-mix tipsSulzer Mixpac AGRuetistrasse 7Haag Sankt gallen, Switzerland9469

Table 6.1 Description of Accessories used with the Proposed Sealapex Xpress

5. Statement of Indications for Use:

Sealapex/Sealapex Xpress is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

    1. Technological Characteristics:
      Sealapex/Sealapex Xpress is intended to fill the microscopic gaps between the canal filler/obturator and canal walls. Reactive components that let Sealapex Xpress sealer set in root-canal are Calcium Oxide in base paste and Salicylate moieties in accelerator (catalyst) paste. Material hardens upon mixing of the base and accelerator pastes in the presence of moisture in the root canal.

The paste base formulation of the proposed Sealapex Xpress and the predicate device, Sealapex 4 (K010940), is very similar, the only difference being an increase in the amount of Resin and Titanium Dioxide, along with a slight addition of a Silica in the proposed Sealapex/Sealapex Xpress base formulation.

The paste catalyst formulation of the proposed Sealapex Xpress and the predicate device, Sealapex 4 (K010940), is very similar, the only difference being a decrease in amount of filler, a decrease in Titanium Dioxide, along with the slight addition of a Silica in the proposed Sealapex/Sealapex Xpress catalyst formulation. Titanium Dioxide (USP) is a common non-toxic white pigment approved by FDA as a food additive, and the minor change does not affect performance or biocompatibility of the paste.

Fumed silica filler serves as a rheological modifier. It does not participate in the setting reaction. Both fillers in the proposed base and catalyst formulations are similar in physical characteristics and surface treated to make the filler particle more hydrophobic and to improve the compatibility with the formulation.

The Resin is an inert, water-insoluble material that helps thicken the paste and does not participate in setting reaction. The slight increase in the amount of Resin should stabilize the consistency of the paste and does not adversely affect the equivalence and performance of the product.

{5}------------------------------------------------

The slightly modified formulation for the proposed Sealapex Xpress has successfully been on the EU market for several years. Performance testing on stability, work time, set time and slump were performed on the modified formula to validate that the proposed products meet the current product specifications and are substantially equivalent to the predicate Sealapex 4 (K010940).

    1. Non-Clinical Performance Data:
      Verification activities were performed in accordance with design control requirements as specified in 21 CFR 820.30 and ISO 13485 Medical Devices- Quality Management Systems and the results demonstrated substantial equivalence to the predicate.
    1. Summary of Performance Testing Performance Testing Bench
      Performance testing for the proposed Sealapex/ Sealapex Xpress included testing for Consistency (Flow), Film Thickness, and Radiopacity per ISO 6876:2012 - Dentistry -Root Canal Sealing Materials.

In the Physical and Mechanical Properties of Table 6.2, the film thickness has a 13.5 micron (um) difference between the predicate and proposed devices. This difference does not affect the substantial equivalence of the subject and predicate devices. The sealer was tested to meet ISO 6876:2012 standard, which states that a sealer shall have a film thickness of not more than 50 um. The proposed Sealapex and Sealapex Xpress are below the maximum 50 um allowance. Therefore, the predicate and proposed sealers will flow sufficiently into the root canal and adapt to canal shape to fill voids.

Table 6.2 displays both an internal and the ISO 6876:2012 test method for determining Flow. The internal test method for Flow differs from the current ISO 6876:2012 test procedure due to the amount of sample used for testing.

The predicate and proposed devices produced Flow results of 48.5 mm and 51.0 mm, respectively, based on internal test methods. This data shows no substantial difference between the predicate and proposed devices. Per ISO 6876:2012, the predicate and proposed devices show no substantial difference based on Flow test results of 22.8 mm and 23.1 mm, respectively.

The data analyzed from the various tests substantiate that the proposed Sealapex/ Sealapex Xpress is substantially equivalent to the predicate Sealapex 4 (K010940).

    1. Substantial Equivalence:
      Sealapex 4 (K010940) is an existing device which was granted market clearance by FDA in 2001. Sybron Dental Specialties, on behalf of SybronEndo, seeks only to slightly modify the formula of the existing device cleared under K010940. Additionally, the device would be produced at a sister facility in Scafati, Italy; and the manufacturer name will change to SybronEndo.

Table 6.2 depicts the comparison of performance and mechanical properties and modification associated with the proposed Sealapex Xpress.

Table 6.2: Predicate Sealanex 4 (K010940) and Proposed Sealapex Xpress Comparison Table

{6}------------------------------------------------

ElementPredicate Device- Sealapex 4Proposed Sealapex/Sealapex Xpress
510(k)K010940K152959
Trade NameSealapex 4Sealapex and Sealapex Xpress
Target UsersLicensed dental professionalsLicensed dental professionals
Indications for UseSealapex 4 is a calciumhydroxide, polymeric resin, rootcanal filling material that isused in conjunction with guttapercha or silver endodonticpointsSealapex/Sealapex Xpress is acalcium hydroxide, polymericresin, root canal filling materialthat is used in conjunction withgutta percha or silver endodonticpoints
DeviceDescriptionSealapex 4 is a non-eugenol,radiopaque, calcium hydroxidepolymeric resin root canalfilling material. Sealapex 4 is atwo-part, base/catalyst –paste/paste system. The two-part system is packaged in tubesthat is hand mixed, then placedinto the prepped root canal.Sealapex/Sealapex Xpress is anon-eugenol, radiopaque,calcium hydroxide polymericresin root canal filling material.Sealapex/Sealapex Xpress is atwo-part, base/catalyst –paste/paste system. The two partsystem is packaged in tubes thatare hand mixed, then placed intothe prepped root canal, or in adual-barrel syringe (SealapexXpress) that does not requirehand mixing.
CommonNameRoot Canal SealerRoot Canal Sealer
ClassificationNameResin, Root Canal Filling perCFR § 872.3820Resin, Root Canal Filling perCFR § 872.3820
ClassIIII
Product CodeKIFKIF
Base FormulaContains:ResinTitanium DioxideContains:Increased range of ResinIncreased range of TitaniumDioxideAdded: Silica
CatalystFormulaContains:FillerTitanium DioxideContains:Decreased range of FillerIncreased range of TitaniumDioxideAdded: Silica
LegalManufacturer/SpecificationDeveloperKerr Corporation28200 Wick RoadRomulus, MI 48174USASybronEndo1332 S. Lone Hill Ave.Glendora, CA 91740USA
ManufacturingLocationKerr Corporation28200 Wick RoadRomulus, MI 48174USAKerr Italia, S.R.L.Via Passanti, 332Scafati Salerno, I-84018ITALY
PackagingLaminate TubesLaminate Tubes and
Polypropylene dual-barrelsyringes used with single-useautomix tips.
Physical and Mechanical Properties
FormTwo Pastes(Catalyst & Base)Two Pastes(Catalyst & Base)
Mix RatioEqual VolumesEqual Volumes
Work Time@ 37°C(98.6° F),100% RH> 60 minutes> 60 minutes
Setting time@ 37°C(98.6° F),100% RH< 24 hours< 24 hours
Consistency(Flow) perInternal TestMethod andSpecification48.5 mm51.0 mm
Consistency(Flow) perISO6876:2012Method22.8 mm23.1 mm
FilmThickness20.3 µm33.85 µm
Solubility andDisintegrationN/AN/A
Radiopacity> 3.3mm Al6.0 mm Al

{7}------------------------------------------------

10. Conclusion as to Substantial Equivalence:

The slight modifications of Sealapex Xpress base and catalyst formulations and the move of the manufacturing location to Scafati, Italy does not affect the intended use of the device nor does it alter the fundamental scientific technology of the device. The nonclinical testing demonstrates that the proposed Sealapex Xpress performs as well as the predicate device based on the specifications set by ISO 6876:2012 testing requirements. The proposed Sealapex/Sealapex Xpress are substantially equivalent to the predicate Sealapex 4 (K010940).

§ 872.3820 Root canal filling resin.

(a)
Identification. A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.(b)
Classification. (1) Class II if chloroform is not used as an ingredient in the device.(2) Class III if chloroform is used as an ingredient in the device.
(c)
Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976. Any other root canal filling resin shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.