K Number
K152894

Validate with FDA (Live)

Device Name
VIVIX-S 1717N
Manufacturer
Date Cleared
2016-02-26

(149 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VIVIX-S 1717N (FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW) is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography.

Device Description

Models FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW intercept X-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create an electrical signals. After the electrical signals are generated, these are converted to a digital value, and an image will be displayed on the monitor.

These devices should be integrated with an operating PC and an X-Ray generator to digitalize Xray images and transfer the digitalized images for radiography diagnostic.

Advanced digital image processing allows considerably efficient diagnosis, all kinds of information management, and image information sharing on the network.

Models XRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW are digital X-ray flat panel detectors, and each model has a 10 x 12 inch imaging area.

FXRD-1717NA and FXRD-1717NB communicate by using a wired communication feature (Gigabit Ethernet communication method by connecting to a tether cable), while FXRD-1717NAW and FXRD-1717NBW communicate by using a wireless communication feature (IEEE 602.lla/b/g/n).

The scintillator used in FXRD-1717NA and FXRD-1717NAW is Csl. Gadox was used for FXRD-1717NB and FXRD-1717NBW.

The FXRD-1717N series is designed to be used with any certified X-ray generators that features DR Trigger mode and is marketed legally. When the DR Trigger mode is not desired, then the connection with the generator can be maintained with AED mode. FXRD-1717N is not designed to function as an X-ray control. The AED mode does not require integration procedure since there is no connection requirement between the X-ray System and the detector. The subject device can receive any types of x-ray signals without SW.

For the DR Trigger mode, the generator interface cable connects the SCU and the X-ray generator. The head of the cable is connected with one of the port (EXT-INF) of the SCU, and the other end of the cable (which is stripped) is connected to the generator's socket.

AI/ML Overview

This document (K152894) describes a 510(k) premarket notification for the VIVIX-S 1717N digital flat panel X-ray detector. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a specific clinical acceptance criterion with a novel AI device. Therefore, much of the requested information about AI performance, multi-reader multi-case studies, and detailed ground truth for training data is not applicable or available in this document.

However, I can extract and infer information about the acceptance criteria for a medical imaging device's performance through comparison to a predicate device, as well as the study conducted to demonstrate this equivalence.

Here's the breakdown based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" for quantitative clinical performance metrics of the device in a table format as one might expect for a new AI diagnostic device. Instead, it demonstrates substantial equivalence to predicate devices based on technological characteristics and image quality assessments. The "acceptance criteria" are implicitly met if the device's performance metrics are comparable to or better than the predicate's.

ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (Subject Device)
DQE (Gadox)0.5 lp/mm: 37, 1 lp/mm: 31, 2 lp/mm: 20, 3 lp/mm: 11"Similar performance characteristics" to predicate. Implicitly, meets or exceeds these values. Values shown for predicate: 0.5 lp/mm: 37, 1 lp/mm: 31, 2 lp/mm: 20, 3 lp/mm: 9 (Reference), 3 lp/mm: 11 (Primary Predicate). The table for the subject device is blank under DQE, implying it is substantially equivalent to these.
DQE (CsI)0.5 lp/mm: 60, 1 lp/mm: 54, 2 lp/mm: 45, 3 lp/mm: 31"Similar performance characteristics" to predicate. Implicitly, meets or exceeds these values. Values shown for predicate: 0.5 lp/mm: 59, 1 lp/mm: 53, 2 lp/mm: 45, 3 lp/mm: 27 (Reference), 3 lp/mm: 31 (Primary Predicate). The table for the subject device is blank under DQE, implying it is substantially equivalent to these.
MTF (Gadox)0.5 lp/mm: 80, 1 lp/mm: 58, 2 lp/mm: 25, 3 lp/mm: 11"Similar performance characteristics" to predicate. Implicitly, meets or exceeds these values. Values shown for predicate: 0.5 lp/mm: 80, 1 lp/mm: 56, 2 lp/mm: 24, 3 lp/mm: 10 (Reference), 3 lp/mm: 11 (Primary Predicate). The table for the subject device is blank under MTF, implying it is substantially equivalent to these.
MTF (CsI)0.5 lp/mm: 88, 1 lp/mm: 72, 2 lp/mm: 44, 3 lp/mm: 25"Similar performance characteristics" to predicate. Implicitly, meets or exceeds these values. Values shown for predicate: 0.5 lp/mm: 81, 1 lp/mm: 58, 2 lp/mm: 28, 3 lp/mm: 15 (Reference) / 0.5 lp/mm: 88, 1 lp/mm: 72, 2 lp/mm: 44, 3 lp/mm: 25 (Primary Predicate). The table for the subject device is blank under MTF, implying it is substantially equivalent to these.
Spatial Resolution3.5 lp/mm (Predicate)3.5 lp/mm (implicitly assumed to be achieved as it's not listed as a difference)
Diagnostic CapabilityEquivalent to K122865 (ViVIX-S wireless)"The new x-ray detectors FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW provide images of equivalent diagnostic capability to the predicate device (K122865)."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated in terms of number of images or patients. The document only mentions "a single-blinded concurrence study."
  • Data Provenance: Not specified in the document (e.g., country of origin). The study is described as a "clinical test, which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices." This suggests a prospective or retrospective collection of images for the comparison, but details are not provided. The study confirms "equivalent diagnostic capability" of the new detectors to the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not specified. The study is referred to as a "single-blinded concurrence study," which implies some form of assessment by experts, but the exact method of combining opinions or establishing ground truth isn't detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? The document describes a "single-blinded concurrence study" to assess "equivalent diagnostic capability." While this involves readers and cases, the text does not indicate it was designed as a comparative effectiveness study comparing human readers with AI vs. without AI assistance. The device itself is an X-ray detector, not an AI interpretation algorithm.
  • Effect Size of Human Readers with AI vs. without AI assistance: Not applicable, as this is a device for image acquisition, not AI for image interpretation.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

  • Was a standalone study done? Not applicable in the context of an AI algorithm. This device is an X-ray detector. Its performance was evaluated based on physical detector characteristics (DQE, MTF, spatial resolution) and a "concurrence study" of its images against a predicate, implying human interpretation of the images produced by the detector.

7. Type of Ground Truth Used for the Test Set

  • Type of Ground Truth: The "single-blinded concurrence study" evaluated "equivalent diagnostic capability." This implies that the ground truth for image quality was established through expert assessment of the diagnostic content of the images produced by the subject device compared to the predicate device. It is a form of expert consensus on image diagnosticity, rather than pathology or long-term outcomes data primarily.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This document describes a medical imaging device (X-ray detector), not an AI algorithm that requires a training set. The performance data relates to the physical characteristics of the detector and the quality of the images it produces.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable, as this is a medical imaging device, not an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, connected by flowing lines. The profiles are stacked vertically, creating a sense of unity and forward movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 26, 2016

Vieworks Co., Ltd. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 2651 E. Chapman Avenue, Suite 110 FULLERTON CA 92831

Re: K152894

Trade/Device Name: VIVIX-S 1717N Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: January 28, 2016 Received: February 2, 2016

Dear Ms. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara
For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K152894

Device Name VIVIX-S 1717N

Indications for Use (Describe)

VIVIX-S 1717N (FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW) is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92] (K152894)

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR807.92

1. Date Prepared [21 CFR 807.92(a) (1)]

01/28/2016

2. Submitter's Information [21 CFR 807.92(a) (1)]

Name of Sponsor: Vieworks Co., Ltd. Address: (Gwanyang-dong) 41-3, Burim-ro 170beon-gil,Dongan-gu, Anyang-si, Gyeonggi-do, 431-060 Republic of Korea Contact Name: Im, Yoonjae / RAQA team Manager 3006013411 Registration Number: Name of Manufacturer: Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

Trade Name:VIVIX-S 1717N
Model Name:FXRD-1717NAFXRD-1717NBFXRD-1717NAWFXRD-1717NBW
Common Name:Digital Flat Panel X-ray Detector
Classification Name:Stationary X-ray System
Classification Panel:Radiology
Classification Regulation:21 CFR 892.1680
Product Code:MQB
Device Class:2

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

Primary Predicate Device:

510(k) Number:K122866
Product Code:MQB
Applicant:Vieworks Co., Ltd.
Trade Name:ViVIX-S with VXvue
Model Name:FXRD-1417SA
FXRD-1417SB

{4}------------------------------------------------

FXRD-1717SA
FXRD-1717SB
Decision Date:01/11/2013

Reference Predicate Device

510(k) Number:K122865
Product Code:MQB
Applicant:Vieworks Co., Ltd.
Trade Name:ViVIX-S Wireless
Model Name:FXRD-1417WAFXRD-1417WB
Decision Date:02/01/2013

5. Description of the Device [21 CFR 807.92(a) (4)]

  • o General Description
    Models FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW intercept X-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create an electrical signals. After the electrical signals are generated, these are converted to a digital value, and an image will be displayed on the monitor.

These devices should be integrated with an operating PC and an X-Ray generator to digitalize Xray images and transfer the digitalized images for radiography diagnostic.

Advanced digital image processing allows considerably efficient diagnosis, all kinds of information management, and image information sharing on the network.

Differences between models O

Models XRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW are digital X-ray flat panel detectors, and each model has a 10 x 12 inch imaging area.

FXRD-1717NA and FXRD-1717NB communicate by using a wired communication feature (Gigabit Ethernet communication method by connecting to a tether cable), while FXRD-1717NAW and FXRD-1717NBW communicate by using a wireless communication feature (IEEE 602.lla/b/g/n).

The scintillator used in FXRD-1717NA and FXRD-1717NAW is Csl. Gadox was used for FXRD-1717NB and FXRD-1717NBW.

The FXRD-1717N series is designed to be used with any certified X-ray generators that features DR Trigger mode and is marketed legally. When the DR Trigger mode is not desired, then the connection with the generator can be maintained with AED mode. FXRD-1717N is not designed to function as an X-ray control. The AED mode does not require integration procedure since there is no connection requirement between the X-ray System and the detector. The subject device can receive any types of x-ray signals without SW.

For the DR Trigger mode, the generator interface cable connects the SCU and the X-ray generator. The head of the cable is connected with one of the port (EXT-INF) of the SCU, and the other end of the cable (which is stripped) is connected to the generator's socket.

{5}------------------------------------------------

6. Intended Use [21 CFR 807.92(a)(5)]

VIVIX-S 1717N (FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW) is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography.

7. Technological Characteristics [21 CFR 807.92(a) (6)]

Comparisons with the predicate, devices show the technological characteristics of the proposed FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW devices to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices.

8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

When compared to the predicate devices (K122865 andK122866), the FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW presented in this submission have the same:

  • Intended Use
  • Technological characteristics
  • Operating principle
  • · Design features
  • Communication Method
  • Scintillator Materials
  • Resolution

There are similar performance characteristics as follow.

  • Performance (MTF)
  • Performance (DQE)

There are no significant difference between the FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.

ParameterReference PredicateDevicesPrimary PredicateDevicesSubject Device
510(k) NumberK122865K122866K152894
ManufacturerVieworks Co., Ltd.
Model NameFXRD-1417WA,FXRD-1417WBFXRD-1417SA,FXRD-1417SB,FXRD-1717SA,FXRD-1717NA,FXRD-1717NB,FXRD-1717NAW,

{6}------------------------------------------------

Common NameFXRD-1717SBFXRD-1717NBW
Classification NameSolid State X-Ray Imager (Flat Panel/Digital Imager)
Classification PanelRadiology
Classification Regulation21 CFR 892.1680
Product CodeMQB
Device Class2
Intended UseThe VIVIX-S Wireless isindicated for digitalimaging solutiondesigned as a generalradiographic systemfor human anatomy. Itis intended to replacefilm or screen basedradiographic systemsin all general purposesof diagnosticprocedures. It is not tobe used formammography.VIVIX-S with VXvue isindicated for digitalimaging solutiondesigned for generalradiographic systemfor human anatomy. Itis intended to replacefilm or screen basedradiographic systemsin all general purposediagnostic procedures.Not to be used formammography.
FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW areindicated for digitalimaging solutiondesigned as a generalradiographic systemfor human anatomy. Itis intended to replacefilm or screen basedradiographic systemsin all general purposesof diagnosticprocedures. It is not tobe used formammography.
DesignPanel ShapeSquare PanelSquare Panel
Field of View14 x 17inch14 x 17inch17 x 17inch
17 x 17inch
Dimensions(H x W x D)444 x 460 x15mm470 x 470x 35mm444 x 460 x15 mm
Pixel Pitch0.14mm0.14mm
Materials ScintillatorCsl: TI,Gd2O2S:TbCsl: TI,Gd2O2S:Tb
Csl: TI,Gd2O2S:Tb
Performance DQEGadoxGadox
0.5 lp/mm3737
1 lp/mm3131
2 lp/mm2020
3 lp/mm911
DQECslCsl
0.5 lp/mm5960
1 lp/mm5354
2 lp/mm4545
3 lp/mm2731
MTFGadoxGadox
0.5 lp/mm8080
1 lp/mm5658
2 lp/mm2425
3 lp/mm1011
MTFCslCsl
0.5 lp/mm8188
1 lp/mm5872
2 lp/mm2844
3 lp/mm1525
Resolution3.5 lp/mm3.5 lp/mm
Communication MethodWirelessWiredWired,Wireless

{7}------------------------------------------------

9. Summary of Non-Clinical Data

A comparison test was conducted between the subject devices (FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW) and the predicate device (K122866) on the items such as DQE, MTF and spatial resolution. Test results support that the subject device is substantially equivalent to the predicate device.

These detectors comply with the following international and FDA-recognized consensus standards:

[IEC 60601-1 2005, Edition 3.0]Medical Electrical Equipment -- Part 1: GeneralRequirements For Basic Safety And Essential Performance
[IEC 60601-1-2, Edition 3]Medical Electrical Equipment - Part 1-2: GeneralRequirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility -Requirements And Tests
[IEC 62220-1, Edition 1.0]Medical electrical equipment - Characteristics of digital X-ray imaging devices - Part 1: Determination of the detectivequantum efficiency
[NEMA PS 3.1 - 3.20 2011]Digital Imaging and Communications in Medicine (DICOM)Set
[ETSI] [EN 300 328 1.8.1]Electromagnetic compatibility and Radio spectrum Matters(ERM); Wideband transmission systems; Data transmissionequipment operating in the 2,4 GHz ISM band and usingwide band modulation techniques; Harmonized ENcovering the essential requirements of article 3.2 of theR&TTE Directive
[ETSI EN 301 893 1.7.1]Broadband Radio Access Networks (BRAN); 5 GHz highperformance RLAN; Harmonized EN covering the essentialrequirements of article 3.2 of the R&TTE Directive
[IEC 62133, Edition 2.0]IEC 62133 Edition 2.0 2012-12 Secondary Cells AndBatteries Containing Alkaline Or Other Non-AcidElectrolytes•Safety Requirements For Portable SealedSecondary Cells, And For Batteries Made From Them, ForUse In Portable Applications [Including: Corrigendum 1(2013)

{8}------------------------------------------------

10. Summary of Clinical Data

A single-blinded concurrence study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted, and the study confirmed that the new x-ray detectors FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW provide images of equivalent diagnostic capability to the predicate device (K122865), the ViVIX-S wireless (FXRD-1417WA and FXRD-1417WB), and its results demonstrate substantial equivalence.

11. Conclusion [21 CFR 807.92(b) (3)]

The FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW Digital X-ray detectors are substantially equivalent to the currently marketed and predicate devices (K122865, and K122866) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1-2, IEC62220-1 and NEMA PS 3.1 - 3.20, IEC 62133 and the clinical test, which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.

The results of these tests demonstrate that FXRD-1717NA, FXRD-1717NB, FXRD-1717NAW and FXRD-1717NBW Digital X-ray detectors meet the acceptance criteria and are adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well the predicate devices.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.