K Number
K152795

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-10-28

(30 days)

Product Code
Regulation Number
878.5020
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RiverLon Suture is intended to be used in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological procedures.

Device Description

The Riverpoint Medical RiverLon Nylon Sutures are inert, non-absorbable, sterile surgical sutures composed of the long-chain aliphatic polymers Nylon 6 and Nylon 6/6. RiverLon sutures are available in common sizes and length with or without pre-attached needles.

AI/ML Overview

The provided text does not contain a study that proves the device meets specific acceptance criteria in the format requested. The document is a 510(k) summary for Riverpoint Medical's RiverLon Nylon Suture, focusing on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and results.

However, it does mention some general performance requirements and tests performed. Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document broadly states that the RiverLon Sutures meet requirements established by the United States Pharmacopeia (USP). Specific acceptance criteria (e.g., minimum tensile strength values, diameter ranges) are not provided, nor are the precise reported device performance metrics beyond the general statement of compliance.

Acceptance Criteria (Not explicitly stated with values, implied by USP)Reported Device Performance
USP performance requirements for needle attachmentMeets USP requirements
USP performance requirements for diameterMeets USP requirements
USP performance requirements for tensile strengthMeets USP requirements
Biological Evaluation (per ISO 10993-1:2009)Materials evaluated per ISO 10993-1:2009

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified.
  • Data Provenance: Not specified. The document only mentions that the sutures are "tested" per USP requirements. No information on country of origin or whether the data is retrospective or prospective is present.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable and not provided. The testing relates to physical and chemical properties of the suture, primarily against pharmacopeial standards (USP), not clinical or expert-reviewed interpretative data.

4. Adjudication method for the test set

Not applicable. The testing is against defined physical/chemical standards, not interpretive data requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a surgical suture, not an AI-powered diagnostic or assistive technology.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a surgical suture, not an algorithm.

7. The type of ground truth used

The "ground truth" for the device's technical characteristics is compliance with the United States Pharmacopeia (USP) performance requirements for surgical sutures, as well as biological evaluation standards like ISO 10993-1:2009. These are often established through standardized testing methodologies and specifications.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable.


Summary of the Study Mentioned:

The document describes a "Performance Data" section stating that:

  • "The Riverpoint Medical RiverLon Sutures meet requirements established by the United States Pharmacopeia."
  • "The RiverLon Sutures are tested per USP performance requirements for needle attachment, diameter, and tensile strength."
  • "FDA Guidance 'Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA' was followed during the preparation of this submission."
  • "Materials used were evaluated per ISO 10993-1:2009 -Biological Evaluation of Medical Devices."

This indicates that standard pharmacopeial and international standards testing was performed to demonstrate the device's compliance. However, the specific details of these tests (e.g., number of samples, exact values of acceptance criteria, detailed results, specific test methods used beyond "per USP") are not provided in this 510(k) summary. The purpose of this document is primarily to establish substantial equivalence to a predicate device (K100006 – Riverpoint Medical RiverLon (Nylon)) based on similar materials, manufacturing processes, intended use, and meeting established performance standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a stylized eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2015

Riverpoint Medical Mr. Edwin Anderson Director of Quality and Regulatory 825 Northeast 25th Avenue Portland, Oregon 97232

Re: K152795

Trade/Device Name: RiverLon (Nylon) Suture Regulation Number: 21 CFR 878.5020 Regulation Name: Nonabsorbable polyamide surgical suture Regulatory Class: Class II Product Code: GAR Dated: September 24, 2015 Received: September 28, 2015

Dear Mr. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

  • for
    Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for Riverpoint Medical. The logo consists of the letters RP inside of a rounded square with a green border. To the right of the square is the text "RIVERPOINT" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font below it.

Indications for Use Statement

K152795 510(k) Number (if known):

Device Name: RiverLon (Nylon) Suture

Indications for Use:

RiverLon Suture is intended to be used in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological procedures.

X Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page _________________________________________________________________________________________________________________________________________________________________________

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Riverpoint Medical. The logo consists of a green square with rounded corners, inside of which are the letters "RP" in white. To the right of the square is the text "RIVERPOINT" in bold, black letters, with the word "MEDICAL" in smaller letters underneath.

510(k) SUMMARY

Riverpoint Medical RiverLon Nylon Suture Line Extension

Submitter Information

Submitter's Name:Riverpoint Medical
Address:825 NE 25th Ave.Portland, OR 97232
Phone Number:(503) 517-8001 or 866 445-4923
Fax Number:(503) 517-8002
Registration Number:3006981798
Contact Person:Edwin Anderson(503) 517-8001
Date of Preparation:September 24th, 2015

Device Name

Trade Name:RiverLon (Nylon) Suture
Common or Usual Names:Nylon Suture, Non-absorbable Surgical Sutures
Classification Name:Nonabsorbable Polyamide Surgical Suture

Device Classification

FDA Class:II
Product Classification:878.5020: Suture, Nonabsorbable, Synthetic, Polyamide
Classification Code:GAR
Review PanelGeneral & Plastic Surgery
Premarket ReviewOffice of Device Evaluation
Division of Surgical Devices, Plastic and Reconstructive
Surgery Devices Branch Two - Skin/Wound
Dressing/Aesthetic Injectables

Predicate Device

K100006 – Riverpoint Medical RiverLon (Nylon)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Riverpoint Medical. The logo consists of a green rounded square with the letters "RP" in white inside. To the right of the square is the text "RIVERPOINT" in bold, black letters, with the word "MEDICAL" in smaller letters underneath.

Device Description

The Riverpoint Medical RiverLon Nylon Sutures are inert, non-absorbable, sterile surgical sutures composed of the long-chain aliphatic polymers Nylon 6 and Nylon 6/6. RiverLon sutures are available in common sizes and length with or without pre-attached needles.

Intended Use / Indications for Use

The Riverpoint Medical RiverLon Suture is intended to be used in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological procedures. The Riverpoint Medical RiverLon Nylon Sutures are provided sterile. RiverLon Suture are intended for single use and are not to be resterilized.

Performance Data

The Riverpoint Medical RiverLon Sutures meet requirements established by the United States Pharmacopeia. The RiverLon Sutures are tested per USP performance requirements for needle attachment, diameter, and tensile strength. FDA Guidance "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA" was followed during the preparation of this submission. Materials used were evaluated per ISO 10993-1:2009 -Biological Evaluation of Medical Devices.

Substantial Equivalence and Comparison of Technical Characteristics

The RiverLon Nylon Suture line extension is as safe and effective as the previously cleared RiverLon Nylon Sutures. The RiverLon Suture has the same intended use and indications for use, the same principles of operation, and similar technical characteristics as the predicate device. Both the RiverLon Nylon Suture and the predicate device are sterilized using the same processes, are composed of the same material, and are tested per USP performance requirements for needle attachment, diameter, and tensile strength. The minor difference in technical characteristics is limited to USP size, and the line extension introduces a size outside of the originally cleared range of sizes. These differences do not raise new questions of safety or effectiveness, therefore the RiverLon Nylon Suture line extension is substantially equivalent to the currently marketed predicate device.

Conclusion

The information provided in this Special 510(k) demonstrates that the Riverpoint Medical RiverLon Nylon Suture line extension is substantially equivalent to the predicate device.

§ 878.5020 Nonabsorbable polyamide surgical suture.

(a)
Identification. Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.