K Number
K152701

Validate with FDA (Live)

Date Cleared
2016-04-22

(214 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Matrix Mini ECG Monitor is intended for continuous measurement of heart rate, respiration rate and detection of cardiac standstill (asystole), ventricular tachycardia and ventricular fibrillation in general medical and surgical floors, general hospital and alternate professional healthcare environments. The system is indicated for use in pediatric and adult patients.

Device Description

The Matrix Mini ECG Monitor system provides a cardiac monitor which monitors ECG derived heart rate, ECG electrode impedance respiration rate, and specific ECG arrhythmias (including asystole, ventricular tachycardia, and ventricular fibrillation). The Matrix Mini ECG Monitor system consists of a display module, a data acquisition module and an ECG patient cable.

The display module and the data acquisition module are connected together with a USB interface. The data acquisition module is connected to the patient with the ECG patient cable, which is attached to standard ECG electrodes applied to the patient's skin surface.

The display module consists of a tablet computer which provides the graphical display, user interface, data processing, data reporting/review, data communication, alarm, and data storage functions. The data acquisition module provides the patient electrical isolation, interface to ECG skin surface electrodes attached to patient, conversion of analog ECG electrode data to digital data, data processing, and data communication functions. An accessory ECG patient cable is provided to connect the data acquisitions module to standard skin surface ECG electrodes. The accessory ECG patient cables are available in 3 or 5 ECG lead configurations.

The Matrix Mini ECG Monitor provides an ECG heart rate monitor that includes software capable of limited ECG analysis and ECG electrode impedance respiration rate measurement. The Matrix Mini ECG monitor includes the basic functions of the 3 or 5 lead ECG monitoring system hardware standard to a Cardiac Monitor including a Cardiotachometer and rate alarm (product code DRT). The primary predicate device is a cardiac monitor and has a product code of DRT and also provides functions for ECG electrode impedance respiration rate measurement (product code BZQ).

The Matrix Mini ECG Monitor device software can report asystole, ventricular tachycardia and ventricular fibrillation, which may be thought of as arrhythmias, we selected a supplemental predicate device (K120774) with the product code DSI. Both the software device component and hardware components are presented below compared to the currently marketed predicate devices.

The Matrix Mini ECG Monitor uses software to process and display the ECG electrode derived cardiac information to the user. The analog ECG electrode data is acquired and converted to digital data using conventional, well proven, and widely used, electronic digital data acquisition technology. The technology has operational and performance characteristics that comply with IEC 60601-2-27:2011, "Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment". The asystole, ventricular tachycardia and ventricular fibrillation algorithms use conventional software ECG analysis techniques and have been evaluated using ANSI/AAMI EC57:2012 "Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms"

AI/ML Overview

The Matrix Mini ECG Monitor is intended for continuous measurement of heart rate, respiration rate and detection of cardiac standstill (asystole), ventricular tachycardia and ventricular fibrillation in general medical and surgical floors, general hospital and professional healthcare environments, in pediatric and adult patients.

1. A table of acceptance criteria and the reported device performance

Feature/MetricAcceptance CriteriaReported Device Performance
Heart Rate
Range21 to 249 beats/minute (Predicate Standard)20 to 300 beats/minute
Accuracy± 3 beats/minute or ± 3% whichever is greater± 3 beats/minute or ± 3% whichever is greater
Bandwidth, Normal Mode0.5 to 40 Hz (Predicate Standard)0.05 to 70 Hz
Alarm Ranges (Low)Heart Rate Low Limits: 21 to 245 beats/minuteHeart Rate Low Limits: 20 to 300 beats/minute
Alarm Ranges (High)Heart Rate High Limits: 25 to 249 beats/minuteHeart Rate High Limits: 20 to 300 beats/minute
Samples per second180 samples/sec (Predicate Standard)250 samples/sec
Respiration Rate
Range5 to 100 breaths/minute5 to 100 breaths/minute
Accuracy± 3 breaths/minute± 2 breaths/minute
Alarm Ranges (Low)Respiration Low limits: 5 to 100 breaths/minuteRespiration Low limits: 5 to 100 breaths/minute
Alarm Ranges (High)Respiration High limits: 5 to 100 breaths/minuteRespiration High limits: 5 to 100 breaths/minute
Arrhythmia Detection
Types DetectedAsystole, Ventricular tachycardia, Ventricular Fibrillation (Predicate Standard)Asystole, Ventricular tachycardia, Ventricular Fibrillation

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size used for the test set or the data provenance. It refers to the use of ANSI/AAMI EC57:2012 "Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms" for evaluating the arrhythmia algorithms. This standard typically involves testing against standardized arrhythmia databases, but the specific details of the test set used (e.g., its size, origin, and whether it was retrospective or prospective) are not provided in this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

The document does not specify the number of experts or their qualifications for establishing the ground truth of any test sets. The mention of ANSI/AAMI EC57:2012 suggests that ground truth would likely be established based on accepted labeled arrhythmia databases, but no details are given regarding expert involvement in this specific context.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

The document does not describe any specific adjudication method for the test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. The submission states, "No clinical studies were utilized for the purpose of obtaining safety or effectiveness data." This indicates that no studies involving human readers, with or without AI assistance, were performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, standalone performance testing was conducted. The document states that the asystole, ventricular tachycardia, and ventricular fibrillation algorithms use conventional software ECG analysis techniques and "have been evaluated using ANSI/AAMI EC57:2012 'Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms'". This standard is for evaluating the performance of the algorithms themselves, without human readers.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Given the reference to ANSI/AAMI EC57:2012, the ground truth for the arrhythmia detection algorithms would likely be derived from a standardized arrhythmia database, which typically uses expert-annotated ECG recordings, forming a type of expert consensus ground truth. However, the document does not explicitly state the type of ground truth used for this specific evaluation.

8. The sample size for the training set

The document does not specify the sample size for the training set of the algorithms.

9. How the ground truth for the training set was established

The document does not specify how the ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 22, 2016

Global Instrumentation, LLC Craig Sellers Regulatory Affairs Manager 8104 Cazenovia Road Manlius, New York 13104

Re: K152701

Trade/Device Name: Matrix Mini ECG Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI, DRT, BZO Dated: March 14, 2016 Received: March 17, 2016

Dear Craig Sellers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152701

Device Name Matrix Mini ECG Monitor

Indications for Use (Describe)

The Matrix Mini ECG Monitor is intended for continuous measurement of heart rate, respiration rate and detection of cardiac standstill (asystole), ventricular tachycardia and ventricular fibrillation in general medical and surgical floors, general hospital and alternate professional healthcare environments. The system is indicated for use in pediatric and adult patients.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

[As described in 21 CFR 807.92]

Submitted by:Global Instrumentation LLC.8104 Cazenovia RoadManlius, NY 13104 USA
Contact Person:Craig SellersRegulatory Affairs ManagerPhone: (315) 682-0272 Ext. 120Fax: (315) 685-0278E-mail: craigs@globalinstrumentation.com
Date Prepared:March 14, 2016
Trade Name:Matrix Mini ECG Monitor
Common Name:Cardiac monitor(including Cardiotachometer and rate alarm)
Classification Reference:Class II, 870.3247 Cttj { j o kc F gvgevqt"cpf "Creto (includingST segment Measurement and alarm)Classification Product Code – DUKSubsequent Product Codes – DTV, BZQ
Predicate Devices:Atlas Monitor(Primary Predicate Device)Cardiac monitor (including Cardiotachometer and rate alarm)Product Code – DRT (Atlas also includes BZQ function)Welch Allyn, Inc.510(k) Number K984033Acuity Central Monitoring Station(Supplemental Predicate Device)Arrhythmia Detector and Alarm (including ST-segmentmeasurement and alarm)Product Code - DSIWelch Allyn, Inc.510(k) Number K120774

{4}------------------------------------------------

Description of the Device:

The Matrix Mini ECG Monitor system provides a cardiac monitor which monitors ECG derived heart rate, ECG electrode impedance respiration rate, and specific ECG arrhythmias (including asystole, ventricular tachycardia, and ventricular fibrillation). The Matrix Mini ECG Monitor system consists of a display module, a data acquisition module and an ECG patient cable.

The display module and the data acquisition module are connected together with a USB interface. The data acquisition module is connected to the patient with the ECG patient cable, which is attached to standard ECG electrodes applied to the patient's skin surface.

The display module consists of a tablet computer which provides the graphical display, user interface, data processing, data reporting/review, data communication, alarm, and data storage functions. The data acquisition module provides the patient electrical isolation, interface to ECG skin surface electrodes attached to patient, conversion of analog ECG electrode data to digital data, data processing, and data communication functions. An accessory ECG patient cable is provided to connect the data acquisitions module to standard skin surface ECG electrodes. The accessory ECG patient cables are available in 3 or 5 ECG lead configurations.

The Matrix Mini ECG Monitor provides an ECG heart rate monitor that includes software capable of limited ECG analysis and ECG electrode impedance respiration rate measurement. The Matrix Mini ECG monitor includes the basic functions of the 3 or 5 lead ECG monitoring system hardware standard to a Cardiac Monitor including a Cardiotachometer and rate alarm (product code DRT). The primary predicate device is a cardiac monitor and has a product code of DRT and also provides functions for ECG electrode impedance respiration rate measurement (product code BZQ).

The Matrix Mini ECG Monitor device software can report asystole, ventricular tachycardia and ventricular fibrillation, which may be thought of as arrhythmias, we selected a supplemental predicate device (K120774) with the product code DSI. Both the software device component and hardware components are presented below compared to the currently marketed predicate devices.

The Matrix Mini ECG Monitor uses software to process and display the ECG electrode derived cardiac information to the user. The analog ECG electrode data is acquired and converted to digital data using conventional, well proven, and widely used, electronic digital data acquisition technology. The technology has operational and performance characteristics that comply with IEC 60601-2-27:2011, "Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment". The asystole, ventricular tachycardia and ventricular fibrillation algorithms use conventional software ECG analysis techniques and have been evaluated using ANSI/AAMI EC57:2012 "Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms"

{5}------------------------------------------------

The Matrix Mini monitoring device is a prescription device to be used by authorized health care professionals using standard institutional procedures and standards of care for patient monitoring. Staff training in the operation of the Matrix Mini ECG Monitor device is essential for optimal use. Users should be skilled at the level of a technician, nurse, physician, health care provider or medical specialist, with the knowledge and experience to acquire and interpret patients' ECG data. Individuals using the Matrix Mini ECG Monitor should be familiar with its operation as described in the manual, and they should understand all warnings and cautions in the manual.

Shelf Life

The product does not have a shelf life. A shelf-life is not applicable because of low likelihood of time-dependent product degradation. The device uses technologies, components and materials that do not significantly change in characteristics with time when used/stored in the specified operating and storage conditions. The materials and technologies utilized ensure that the specified storage, and usage, conditions are not expected to affect device safety or effectiveness

Indications for Use:

The Matrix Mini ECG Monitor is intended for continuous measurement of heart rate, respiration rate and detection of cardiac standstill (asystole), ventricular tachycardia and ventricular fibrillation in general medical and surgical floors, general hospital and professional healthcare facilities. The system is indicated for use in pediatric and adult patients.

Predicate Device Selection

The primary predicate device selected is the Atlas Monitor (Welch Allyn, K984033, Product Code DRT). The Atlas device also provides the function of ECG electrode impedance respiration rate measurement (product code BZQ). A supplemental predicate device selected is the Acuity Central Monitoring Station Monitor (Welch Allyn, K120774, Product Code DSI). The Acuity device includes detection and alarming for asystole, ventricular tachycardia, and ventricular fibrillation arrhythmias. These devices are presently in commercial distribution globally including the United States of America. The Matrix Mini ECG Monitor has similar technological characteristics, indications for use, and performance characteristics as the predicate devices. The Matrix Mini ECG Monitor is similar in design, function and application to the predicate device. All materials and technologies are currently utilized in in other medical devices therefore, no new issues of safety or effectiveness are introduced by using this device.

Technological Characteristics:

The Matrix Mini ECG Monitor (System) uses conventional electronic digital data acquisition techniques to digitize and process ECG signals from skin surface ECG electrodes attached to a patient. The System consists of the Matrix Mini Display Module and the Matrix Mini Acquisition Module which are connected by an electrically isolated USB communication interface. The display module provides the data/waveform display and the user interface functions of the system. The acquisition module acquires the ECG signal from ECG electrodes on the surface of the patient's skin and provides the electrical isolation required to ensure patient safety. Refer to Table 1 at the end of this document for a detailed comparison

{6}------------------------------------------------

of the subject device to the predicate device.

The Matrix Mini ECG monitor uses software to process the ECG electrical signals and display the ECG and respiration information to the user. The technology has operational and performance characteristics that comply with IEC 60601-2-27:2011, "Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment" The asystole, ventricular tachycardia and ventricular fibrillation algorithms use conventional software ECG analysis techniques and have been evaluated using ANSI/AAMI EC57:2012 "Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms"

The Matrix Mini ECG Monitor uses the same technologies as the predicate devices. The Matrix Mini ECG Monitor is technically equivalent to the Atlas Monitor (K984033) and the Acuity Central Monitoring Station (K120774) with respect to technical and performance characteristic including, data storage, patient management, data review, measured physiological parameters and alarm management.

Non-Clinical Tests:

Verification and validation testing was conducted to ensure expected performance of the Matrix Mini FCG Monitor

The Matrix Mini ECG Monitor was also tested to evaluate its safety and effectiveness based on the following standards and guidance documents:

  • IEC 60601-2-27 Edition 3.0 2011-03 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment [Including: Corrigendum 1 (2012)]
  • ISO 14971: 2007 Medical devices Application of risk management to medical ● devices
  • IEC 62304: 2015 - Medical device software - Software life cycle processes
  • . ANSI/AAMI EC57:2012 - Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms
  • ANSI AAMI EC53:2013 - ECG trunk cables and patient leadwires
  • . IEC 60601-1-2:2014 - Medical Electrical Equipment - Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Test
  • . IEC 60601-1-8 Medical electrical equipment – Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • . FDA guidance Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm (October 28, 2003), was utilized in the planning and conduct of all performance testing.
  • FDA Guidance for the Content of Premarket Submissions for ● Software Contained in Medical Devices (May 11, 2005). The device's software has been validated in accordance with this guidance.

{7}------------------------------------------------

The Matrix Mini ECG Monitor system has been tested and found to be in compliance with the applicable recognized safety, performance and electromagnetic compatibility standards. The standards used are consistent with the standards and testing used for the predicate devices. The results of the testing demonstrate that the device is as safe, as effective, and performs as well as, or better than, the predicates.

Clinical Performance Data:

No clinical studies were utilized for the purpose of obtaining safety or effectiveness data.

Conclusion:

Based on the information presented in this 510(k) premarket notification, the Global Instrumentation LLC Matrix Mini ECG Monitor is considered substantially equivalent (as safe, as effective and performs as well as) the currently marketed devices cited in this submission.

{8}------------------------------------------------

TABLE 1

CharacteristicPredicate DevicesSubject DeviceDifferences
Device NameAtlas Monitor (primary predicate device)Matrix Mini ECG MonitorN/A
ManufacturerWelch Allyn, Inc.,Global Instrumentation, LLC.N/A
510(k) NumberK984033N/AN/A
Product CodeDRTDRT, DSI, BZQSame
Regulation Name870.2300 Cardiac monitor (includingCardiotachometer and rate alarm)870.2300 Cardiac monitor (includingCardiotachometer and rate alarm)Same
Indications for UseThe indications for use for the AtlasMonitor, include the monitoring of thefollowing human physiological vital signs:Blood Oxygenation (SpO2) measurement,ECG waveform derived from 3 or 5 leadmeasurement, Respiration rate/waveformderived from ECG or CO2, Temperaturemeasurement via YSI 400 series probes,Noninvasive Blood Pressure (NIBP)measurement, CO2, End tidal sidestream/waveform (ETCO2), Heart Ratederived from selected source (ECG,SpO2).The target populations are adult andpediatric populations. The monitor isintended for use within the healthcarefacility setting.The Matrix Mini ECG Monitor isintended for continuous measurement ofheart rate, respiration rate, and detectionof cardiac standstill (asystole),ventricular tachycardia and ventricularfibrillation in general medical andsurgical floors, general hospital andprofessional healthcare environments.The system is indicated for use inpediatric and adult patients.SimilarThe Matrix Mini ECG Monitor isequivalent to the Atlas Monitor(K984033) in ECG, Respiration andheart rate parameter performance.The target populations are the same.Some functions are removed sincethe subject device is only for heartrate, respiration rate, and detectionof cardiac standstill (asystole),ventricular tachycardia andventricular fibrillationThe Matrix Mini ECG arrhythmiaanalysis is limited to cardiacstandstill (asystole), ventriculartachycardia and ventricular
CharacteristicPredicate DevicesSubject DeviceDifferences
fibrillation. This is equivalent toprimary predicate in conjunctionwith Welch Allyn Acuity CentralMonitoring Station (K120774).
Target PopulationAdult and pediatric populationsPediatric and adult patients of anygender.Same
Where UsedThe monitor is intended for use within thehealthcare facility setting.The product is intended for use in aclinical environment general medical andsurgical floors, general hospital, andprofessional healthcare environments.Similar - more detail added todescription.
DesignMultiparameter device used to monitorhuman physiological vital signs.Multiparameter device used to monitorECG and respiration humanphysiological vital signs.SimilarMatrix Mini ECG monitor providesthe same functions for ECG, heartrate, and respiration parametersderived from ECG electrodes.Matrix Mini does not providemonitoring for temperature, SpO2,CO2, NIBP or ETCO2.The Matrix Mini provides the sameECG, heart rate, and respirationparameters that is provided by thepredicate devices. For ECG andrespiration parameters, the MatrixMini is equivalent to the predicatedevices.
TechnologiesECG and ECG respiration sensing withconventional patient skin surface appliedECG/Respirationelectrodes.dataECG and ECG respiration sensing withconventional patient skin surface appliedelectrodes. ECG/Respiration dataSame
CharacteristicPredicate DevicesSubject DeviceDifferences
acquisition performed with solid stateamplifiers, analog to digital converter andprocessing of data controlled bymicroprocessor/microcontroller. Data isconverted from analog to digital formatand digital data is then processed andanalyzed. Microcontroller based systemprovides electronic user interface withalarm functions, electronic waveformdisplays and numeric data display. Internalbattery power supply. External AC powersupply/battery charging.acquisition performed with solid stateamplifiers, analog to digital converter andprocessing of data controlled bymicroprocessor/microcontroller. Data isconverted from analog to digital formatand digital data is then processed andanalyzed. Microcontroller based systemprovides electronic user interface withalarm functions, electronic waveformdisplays and numeric data display.Internal battery power supply. ExternalAC power supply/battery charging.Similar
Heart Rate Range21 to 249 beats/minute20 to 300 beats/minuteSimilarMatrix Mini detects wider range ofheart rates. This will provideperformance that is equal to, orbetter than, that of the predicatedevices.
Heart Rate Accuracy$\pm$ 3 beats/minute or $\pm$ 3% whichever isgreater$\pm$ 3 beats/minute or $\pm$ 3% whichever isgreaterSame
Bandwidth, NormalMode0.5 to 40 Hz0.05 to 70 HzSimilarMatrix Mini has wider range offrequency response bandwidth. Thiswill provide performance that isequal to, or better than, that of thepredicate devices.
Leads3 wire or 5 wire, available in AHA or IECcolors3 wire or 5 wire, available in AHA or IECcolorsSame
ElectrodesDisposable snap electrodes applied topatient skin surfaceDisposable snap electrodes applied toSame
CharacteristicPredicate DevicesSubject DeviceDifferences
patient skin surface.
Gain10 mm/mV, adjustable, visual scaleindicator10 mm/mV, adjustable, visual scaleindicatorSame
Lead DisplaySingle, user selectable: I, II, III, VSingle, user selectable: I, II, III, VSame
Heart Rate DisplayNumericNumericSame
Waveform DisplayOne row minimum of ECG waveformdisplayOne row minimum of ECG waveformdisplaySame
Leads Off ConditionDetected and displayed (selected leadonly)Detected and displayed (selected leadonly)Same
Heart Rate AlarmsHigh and Low Heart RateHigh and Low Heart RateSame
Alarm RangesHeart Rate Low Limits: 21 to 245beats/minute Heart Rate High Limits: 25to 249 beats/minuteHeart Rate Low Limits: 20 to 300beats/minute Heart Rate High Limits: 20to 300 beats/minuteSimilarMatrix Mini supports a wider rangeof alarm limits. This ensuresperformance in this area is equal to,or better than, that of the predicatedevices.
Data AcquisitionTechniqueAnalog to Digital ConvertorAnalog to Digital ConvertorSame
CharacteristicPredicate DevicesSubject DeviceDifferences
Samples per second(sample rate)180 samples/sec250 samples/secSimilarMatrix Mini has higher samplingrate. This will provide data qualitythat is equal to, or better than, that ofthe predicate devices.
Input Impedance>2.5 Megohms at 10Hz>2.5 Megohms at 10HzSame
Input ProtectionDefibrillator and Electrosurgery protectedDefibrillator and Electrosurgery protectedSame
Defibrillator Recoveryper EC13:1992 Clause3.1.2.1a)YesYesSame
Pacemaker DisplayPacemaker signals displayed as captured.Pacemaker signals displayed as captured.Same
Pacemaker DetectionYesYesSame
Pacemaker RejectYesYesSame
Lead Off DetectionYesYesSame
ECG Respiration/LeadOff Detection Current50nA max for RA, LA, LL, V; 200nAmax for RL50nA max for RA, LA, LL, V; 200nAmax for RLSame
ECG RespirationMeasurementTechniqueTrans-thoracic impedanceTrans-thoracic impedanceSame
ECG Respiration #LeadsSingleSingleSame
ECG Respiration Range5 to 100 breaths/minute5 to 100 breaths/minuteSame
CharacteristicPredicate DevicesSubject DeviceDifferences
ECG RespirationAccuracy$\pm$ 3 breaths/minute$\pm$ 2 breaths/minuteSimilarMatrix Mini has better accuracy.This ensures performance in thisarea is equal to, or better than, thatof the predicate devices.
Respiration AlarmsHigh and Low Respiration RateHigh and Low Respiration RateSame
Respiration AlarmRangesRespiration Low limits: 5 to 100breaths/minute Respiration High limits: 5to 100 breaths/minuteRespiration Low limits: 5 to 100breaths/minute Respiration High limits: 5to 100 breaths/minuteSame
Respiration RateDisplayNumericNumericSame
ECG analysis and detection of limited arrhythmias similar to Acuity (K120774 supplemental predicate)in conjunction with the Atlas Monitor (K984033 primary predicate). The below provides the comparisonto K120774.
Device NameAcuity Central Monitoring Station(supplemental predicate device)Matrix Mini ECG MonitorN/A
CharacteristicPredicate DevicesSubject DeviceDifferences
ManufacturerWelch Allyn, Inc.,Global Instrumentation, LLC.N/A
510(k) NumberK120774N/AN/A
Product CodeDSIDRT, DSI, BZQSimilar, Acuity and Matrix Miniboth include DSI product code.
Regulation Name870.1025, Arrhythmia Detector and Alarm(including ST-segment measurement andalarm)870.2300 Cardiac monitor (includingCardiotachometer and rate alarm)Acuity and Matrix Mini ECGMonitor both provide ECG analysisand detection of waveforms to detectthe following arrhythmia: asystole,ventricular tachycardia, andventricular fibrillation.
Indications for UseThe Acuity Central Monitoring Station isintended to be used by clinicians for thecentral monitoring of neonatal, pediatric,and adult patients in health care facilities.In addition to the central monitoring ofpatient data, waveforms, alarms and alerts,The Acuity software can includeoperational modules to provide extendedrecording of patient data (Full Disclosure),arrhythmia monitoring and ST analysis.* Full disclosure stores patient data for upto 96 hours* Arrhythmia monitoring module providesreal-time monitoring and alarms forspecific changes in cardiac rhythms. TheThe Matrix Mini ECG Monitor isintended for continuous measurement ofheart rate, respiration rate, and detectionof cardiac standstill (asystole),ventricular tachycardia and ventricularfibrillation in general medical andsurgical floors, general hospital andSimilarThe Matrix Mini ECG Monitor isequivalent to the Atlas Monitor inECG, Respiration and heart rateparameter performance.The target populations are the same.Some functions removed since thesubject device is only for ECG.The Matrix Mini ECG Monitorincludes arrhythmia analysis that islimited to cardiac standstill(asystole), ventricular tachycardiaand ventricular fibrillation. This isequivalent to the Atlas primarypredicate (K984033) in conjunctionwith the arrhythmia monitoring

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

CharacteristicPredicate DevicesSubject DeviceDifferences
clinician is responsible for determining theclinical significance of each detectedarrhythmia event or alarm. The arrhythmiamodule is not intended for use withneonatal patients.module functions of the Welch AllynAcuity Central Monitoring Station.(K120774)
* ST analysis module provides real-timemonitoring and alarms for ST segmentdeviations, from a reference beat, forpatients with suspected heart disease andanomalies. The clinician is responsible fordetermining the clinical significance ofeach selected ST segment deviation oralarm.The ST analysis module is not intendedfor use with neonatal patients.
Target PopulationAdult and pediatric populations (for thearrhythmia module) which provides theasystole, ventricular tachycardia andventricular fibrillation arrhythmia analysis.Pediatric and adult patients of anygender.SameThe arrhythmia module of the AllynAcuity Central Monitoring Station isintended for pediatric and adultpatients. The arrhythmia module ofis not intended for use with neonatalpatients.
Where UsedThe product is intended for use withinhealthcare facilities.The product is intended for use in aclinical environment general medical andsurgical floors, general hospital, andprofessional healthcare environments.Similar – more detail added todescription for Matrix Mini. Nosignificant difference betweenMatrix Mini and predicate devices.All devices are intended for use

{16}------------------------------------------------

CharacteristicPredicate DevicesSubject DeviceDifferences
within healthcare facilities.
DesignMultiparameter device used to monitorhuman physiological vital signs includingdetection of ECG Arrhythmias includingasystole, ventricular tachycardia, andventricular fibrillation.Multiparameter device used to monitorECG and respiration humanphysiological vital signs.SimilarMatrix Mini does not providemonitoring for temperature, SpO2,CO2, NIBP or ETCO2.The Matrix Mini ECG monitor doesnot perform ST analysis.Matrix Mini ECG monitor has thesame design functions as thepredicate devices for ECG, heartrate, and respiration parametersderived from ECG electrodes,including detection of asystole,ventricular tachycardia, andventricular fibrillation.
TechnologiesUses software running on computerhardware to monitor and analyze digitalpatient physiological data. Uses personalcomputer technologies to provide userinterface to system. Supports optionalnetwork and printer connections.Uses software running on computerhardware to monitor and analyze digitalpatient physiological data. Uses personalcomputer technologies to provide userinterface to system. Supports optionalnetwork and printer connections.Same
Heart Rate ArrhythmiadetectionAsystole, Ventricular tachycardia andVentricular FibrillationAsystole, Ventricular tachycardia andVentricular FibrillationSame

{17}------------------------------------------------

CharacteristicPredicate DevicesSubject DeviceDifferences
ST analysisYesNoST analysis is not required by MatrixMini ECG monitor indications foruse.Matrix Mini ECG Monitorarrhythmia detection limited toasystole, ventricular tachycardia andventricular fibrillation.
AlarmsCan configure, observe, and respond to analarm condition.Can configure, observe, and respond to analarm condition.Same
Heart Rate Arrhythmiaalarm settingsYesYesSame
Alarm reviewYesYesSame
Waveform displayYes - ECG waveformsYes - ECG waveformsSame
Patient Data SourceCentral station interfaces with wirelessand hardwired patient monitors for patientdataMatrix Mini ECG Monitor is a hardwiredpatient monitor for patient data. Connectsdirectly to patient.SimilarThe Matrix Mini ECG monitorprovides for monitoring of thepatient at the patient location using ahardwired interface. The MatrixMini ECG monitor does not providefor remote monitoring of the patient.The Matrix Mini ECG monitorprovides all required information atthe patient location.
LogsWaveform and Error logsWaveform and Error logsSame
Logging capabilityYes-logs a variety of informationincluding alarms, alerts, connection andYes-logs a variety of informationincluding alarms, alerts, connection andSame

{18}------------------------------------------------

CharacteristicPredicate DevicesSubject DeviceDifferences
disconnection of monitors, significantactions by clinicians, etc. Logs provide anaudit trail for system failures and customercomplaints. System can be set to capturemore information for troubleshooting.disconnection of monitors, significantactions by clinicians, etc. Logs providean audit trail for system failures andcustomer complaints. System can be setto capture more information fortroubleshooting.
Full disclosure datastorageUp to 96 hours available to customersUp to 96 hours available to customersSame
Data interfaces topatient data acquisitiondeviceRS232 and optional network connectionsUSBSimilarThe Matrix Mini ECG monitor doesnot support use of wireless or remotemonitoring of the patient. TheMatrix Mini uses a hard wired serialdata interface (USB) which is asreliable, or more reliable, than thedata interfaces used by the predicatedevices. The subject deviceperformance in this area is as goodas, or better than, that of thepredicate device.
Optional data interfacesOptional network connections to externaldevices such as printers.Optional network connections to externaldevices such as printers.Same

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.