K Number
K152648

Validate with FDA (Live)

Date Cleared
2016-02-18

(155 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ennova Self-adhesive Electrode is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use.

Device Description

Ennova Self-adhesive Electrode is used as an accessory to the TENS or EMS device unit, which transmits electrical current to patient skin. The electrical current is first transmitted via the lead wire or snap button then transmitted to the conductive gel which is adhered to patient skin. Ennova Self-adhesive Electrode is composed of a cover, connector lead wire or snap button, conductive carbon film, conductive hydrogel, and an electrode carrier liner. It is non-sterile and intended for single adult patient(age=18) multiple application use. Ennova Self-adhesive Electrode has various shapes and sizes. To connect with a nerve or muscle stimulator, this Electrode has lead wire type and snap button type. For lead wire type electrode, the wire is at least 40mm long, with 2.0~3.5mm diameter female socket. For snap button type electrode, the connector is male snap button. The individual electrodes are electrically connected in pairs. When not in use, the hydrogel face is covered by a PET (polyethylene terephthalate) release liner.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Ennova Self-adhesive Electrode. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study.

Therefore, many of the requested details about a study proving device acceptance criteria cannot be extracted from this document, as a clinical study was explicitly not included in this submission.

Here's an analysis based on the information available:

1. A table of acceptance criteria and the reported device performance:

The document doesn't define "acceptance criteria" in the traditional sense of a clinical trial's primary or secondary endpoints. Instead, it outlines design specifications and performance standards for the device and compares them to predicate devices to establish substantial equivalence.

Criterion TypeAcceptance Criteria (from predicate/standards)Reported Device Performance (Ennova Self-adhesive Electrode)
Intended UseTransmit electrical current to patient skin for transcutaneous electrical stimulation devices (TENS, EMS). For OTC or Prescription use.Transmit electrical current to patient skin for transcutaneous electrical stimulation devices (TENS, EMS). For OTC or Prescription use.
Target PopulationSingle patient use, multiple application, adult patients (age ≥ 18).Single adult patient (age ≥ 18), multiple application use.
BiocompatibilityISO 10993-5:2009/(R) 2014 (Cytotoxicity), ISO 10993-10:2010 (Irritation and Skin Sensitization)Passed Biocompatibility test per ISO10993-5:2009; ISO10993-10:2010
Shelf LifeNot explicitly detailed as "acceptance criteria," but "Shelf life test" was performed.Passed Shelf life test.
ImpedanceRange as seen in predicate devices (e.g., Round: 300~1000 Ω)Reported ranges for specific shapes. E.g., Round: 397612 Ω; Rectangle: 432624 Ω; Oval: 412642 Ω; Butterfly: 407616 Ω. (Stated as "within the range specified by the predicate device")
AdhesionNot explicitly detailed as "acceptance criteria," but "Adhesive test" was performed.Passed Adhesive test.
Electrical SafetyAAMI / ANSI ES60601-1:2005/(R) 2012, IEC 60601-2-2 Edition 5.0 2009-02 (specifically lead wire characteristics).Lead wire meets Clause 8.5.2.3 of AAMI/ANSI ES60601-1.
Current DensityRanges derived from predicate devices, ensuring below 0.25 W/cm² to reduce thermal burns.Values provided for various shapes (e.g., Round: Min.0.36mA/cm² Max.1.70mA/cm²). All reported to be within safety limits as compared to predicate devices.
Maximum Power DensityLess than 0.25 W/cm² (FDA guidance)Values provided for various shapes (e.g., Round: 0.0117W/cm²). All reported to be within safety limits.
Other PerformanceGood electrical conductivity, good adhesive property.Good electrical conductivity, good adhesive property.
SterilityASTM F1980 (Reapproved 2011) - for sterile barrier systems (though the device itself is non-sterile).Complies with ASTM F1980 (for sterile barrier systems).

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not applicable for a clinical study in this submission. The tests performed are bench tests on the physical device components. The document does not specify the number of individual electrodes or batches tested for these non-clinical evaluations.
  • Data Provenance: The tests were conducted by Suzhou Ennova Electronics Technology Co., Ltd. in China ("Bench tests were conducted on the proposed Self-adhesive Electrode to verify that the proposed device met all design specifications"). The data is non-clinical, originating from laboratory testing. It is neither retrospective nor prospective in the clinical study sense.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This submission relies on non-clinical bench testing against recognized industry standards (e.g., ASTM, ISO, AAMI/ANSI, IEC) and comparison to legally marketed predicate devices. Ground truth in the context of expert consensus (e.g., for image interpretation) is not relevant here. The "ground truth" for these tests comes from objective measurements and adherence to specified performance metrics in the standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods are typically used in clinical studies, particularly for subjective assessments or when reconciling differing expert opinions. The tests performed here are objective bench tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a self-adhesive electrode, not an AI-powered diagnostic tool. No MRMC study was conducted or is relevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This is a hardware component (electrode), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this submission is based on established industry standards (e.g., ISO, ASTM, AAMI/ANSI, IEC) and the performance characteristics of legally marketed predicate devices. For example, biocompatibility is assessed against ISO standards, and electrical performance against AAMI/ANSI and IEC standards. The efficacy is inferred from the substantial equivalence to predicate devices that have already demonstrated safety and effectiveness for their intended use.

8. The sample size for the training set:

  • Not applicable. This device's approval is based on substantial equivalence demonstrated through bench testing and comparison to predicates, not through machine learning model training.

9. How the ground truth for the training set was established:

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures facing right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 18, 2016

Suzhou Ennova Electronics Technology Co.,Ltd % Doris Dong Manager Shanghai CV Technology Co., Ltd Room 1706, No.128 Songle Rd., Songjiang Area Shanghai, 201600 CHINA

Re: K152648

Trade/Device Name: Ennova Self-adhesive Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: January 12, 2016 Received: January 19, 2016

Dear Doris Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael J. Hoffmann -A

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152648

Device Name Ennova Self-adhesive Electrode

Indications for Use (Describe)

Ennova Self-adhesive Electrode is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information:

510(k) Number:K152648
Date:January 12, 2016
Type of 510(k) Submission:Traditional
Basis for 510(k) Submission:New device
Submitter/Manufacturer:Suzhou Ennova Electronics Technology Co., Ltd.Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, JiangsuProvince, 215124 China
Contactor:Doris DongShanghai CV Technology Co., Ltd.Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, 201600 ChinaE-mail: doris_d@126.comTel: 86 21-31261348 / Fax: 86 21-37824346

2. Device Description:

Proprietary Name:Ennova Self-adhesive Electrode
Common Name:Cutaneous electrode
Classification Name:Cutaneous electrode
Product Code:GXY
Device Class:II
Regulation Number:882.1320
Review Panel:Neurology
Indications for use:Ennova Self-adhesive Electrode is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use.
Device Description:Ennova Self-adhesive Electrode is used as an accessory to the TENS or EMS device unit, which transmits electrical current to patient skin. The electrical current is first transmitted via the lead wire or snap button then transmitted to the conductive gel which is adhered to patient skin. Ennova Self-adhesive Electrode is composed of a cover, connector lead wire or snap button, conductive carbon film, conductive hydrogel, and an electrode carrier liner. It is non-sterile and intended for single adult patient(age=18) multiple application use. Ennova Self-adhesive Electrode has various shapes and sizes. To connect with a nerve or muscle stimulator, this Electrode has lead wire type and snap button type. For lead wire type electrode, the wire is at least 40mm long, with 2.0~3.5mm diameter female socket. For snap button type electrode, the connector is male snap button. The individual electrodes are electrically connected in pairs. When not in use, the hydrogel face is covered by a PET (polyethylene terephthalate) release liner.

{4}------------------------------------------------

The conductive hydrogel is imported from Axelgaard, USA, which was 510(k) cleared with number of K130987. And the whole Ennova Self-adhesive Electrode together with the conductive hydrogel has again passed the Biocompatibility test, Shelf life test, Impedance test, Adhesive test, and so on.

3. Predicate Device Identification

510k numberDevice nameManufacturerDate cleared
K090198Jiajian Self-adhesive ElectrodeWuxi Jiajian Medical Instrument Co., Ltd2009
K150971Zekang Self-adhesive ElectrodeJiangsu Zekang Medical Technology Co., Ltd.2015

4. Non-Clinical Test Conclusion

Bench tests were conducted on the proposed Self-adhesive Electrode to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ASTM F1980 (Reapproved 2011), Standard guide for accelerated aging of sterile barrier systems for medical devices. (Sterility)

  • ISO 10993-5:2009/(R) 2014, Biological evaluation of medical devices -- Part 5: Tests for In Vitro

Cytotoxicity. (Biocompatibility)

  • ISO 10993-10 Third Edition 2010-08-01, Biological evaluation of medical devices - Part 10: Tests for Irritation and Skin Sensitization. (Biocompatibility)

  • AAMI / ANSI ES60601-1:2005/(R) 2012, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance

  • IEC 60601-2-2 Edition 5.0 2009-02, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

5. Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data was not including in this submission.

6. Substantially Equivalent Comparison Conclusion

Detailed comparison data is included in Section 9 of "Substantial Equivalence Discussion" of this 510(k) submission.

New DevicePredicate Device 1Predicate Device 2
501(k) numberK152648K150971K090198
Trade Name:Ennova Self-adhesive ElectrodeZekang Self-adhesive ElectrodeJiajian Self-adhesive Electrode
Common Name:Cutaneous electrodeCutaneous electrodeCutaneous electrode
ClassificationCutaneous electrodeCutaneous electrodeCutaneous electrode
Name:
Product Code:GXYGXYGXY
Regulation882.1320882.1320882.1320
Number:
MedicalNeurologyNeurologyNeurology
Specialty:
Device Class:IIIIII
Indications foruse:Ennova Self-adhesive ElectrodeZekang Self-adhesive Electrode isThe self-adhesive electrode is
is intended to transmit electricalintended to transmit electrical currentintended to be used to apply
current to patient skin for useto patient skin for use withelectrical stimulation current to
with transcutaneous electricaltranscutaneous electrical stimulationthe patient's skin.
stimulation devices. Somedevices. Some common types of theExample electrical stimulations
common types of the stimulationstimulation devices include, but arefor current applications of the
devices include, but are not limitnot limit to TENS (Transcutaneouselectrodes are: TENS and EMS.
to TENS (TranscutaneousElectrical Nerve Stimulation) and
Electrical Nerve Stimulation)EMS (Electrical Muscular
and EMS (Electrical MuscularStimulation). The electrode is for
Stimulation). The electrode is forOTC (Over-The -Counter) or
OTC (Over-The -Counter) orPrescription use.
Prescription use.
Targetpopulation:Single patient use and multipleapplicationSingle patient use and multipleapplicationSingle patient use and multipleapplication
Design (shape):Round, rectangle, oval andbutterfly, according toRound, rectangle, oval, gourd,butterfly and saddle, according toRound, rectangle, oval, gourd,butterfly and saddle, according to
customized specification.customized specification.customized specification.
Size of electrodesRoundMin.Ø32mmMax.Ø70mmRoundMin.Ø20mmMax.Ø70mmRoundMin.Ø20mmMax.Ø80mm
RectangleMin.40×40mmMax.100×130mmRectangleMin.20×20mmMax.150×150mmRectangleMin.30×10mmMax.80×130mm
OvalMin.50×120mmMax.100×240mmOvalMin.50×120mmMax.100×240mmOvalMin.50×120mmMax.100×240mm
ButterflyMin.55×75mmMax.95×165mmButterflyMin.55×75mmMax.95×165mmButterflyMin.55×75mmMax.95×165mm
GourdMin.85×50mmMax.170×100mmGourdMin.85×50mmMax.170×100mm
SaddleMin.75×110mmMax.150×220mmSaddleMin.75×110mmMax.150×220mm
Patient contactareaRoundMin.804mm²Max.3846.5mm²RoundMin.314mm²Max.3846.5mm²RoundMin.314mm²Max.5024mm²
RectangleMin.1600 mm²Max.13000mm²RectangleMin.400 mm²Max.22500mm²RectangleMin.300 mm²Max.10400mm²
OvalMin.4710 mm²Max.18840mm²OvalMin.4710 mm²Max.18840mm²OvalMin.4710 mm²Max.18840mm²
ButterflyMin.4125 mm²Max.15675mm²ButterflyMin.4125 mm²Max.15675mm²ButterflyMin.4125 mm²Max.15675mm²
GourdMin.4250 mm²Max.17000mm²GourdMin. 4250 mm²Max.17000mm²
SaddleMin.8250 mm²Max.33000mm²SaddleMin.8250 mm²Max.33000mm²
ImpedanceparametersRound$397\text{~}612 \Omega$Round$300\text{~}500 \Omega$Round$300\text{~}1000 \Omega$
Rectangle$432\text{~}624 \Omega$Rectangle$300\text{~}700 \Omega$Rectangle
Oval$412\text{~}642 \Omega$Oval$500\text{~}700 \Omega$Oval
Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, Jiangsu Province, 215124 China
Butterfly407~616 ΩButterfly400~600 ΩButterfly400~600 Ω
Gourd400~600 ΩGourd
Saddle400~800 ΩSaddle
Currentdensity(UseI RMS= 13.7mA forcalculation )RoundMin.0.36mA/cm2Max.1.70mA/cm2RoundMin.0.36mA/cm2Max.4.36mA/cm2RoundMin.0.27mA/cm2Max.4.36mA/cm2
RectangleMin.0.11mA/cm2Max.0.86mA/cm2RectangleMin.0.06mA/cm2Max.3.43mA/cm2RectangleMin.0.13mA/cm2Max.4.57mA/cm2
OvalMin.0.07mA/cm2Max.0.29mA/cm2OvalMin.0.07mA/cm2Max.0.29mA/cm2OvalMin.0.07mA/cm2Max.0.29mA/cm2
ButterflyMin.0.09mA/cm2Max.0.33mA/cm2ButterflyMin.0.09mA/cm2Max.0.33mA/cm2ButterflyMin.0.09mA/cm2Max.0.33mA/cm2
GourdMin.0.08mA/cm2Max.0.32mA/cm2GourdMin.0.08mA/cm2Max.0.32mA/cm2
SaddleMin.0.04mA/cm2Max.0.17mA/cm2SaddleMin.0.04mA/cm2Max.0.17mA/cm2
Maximum PowerDensity (W/cm2)Round0.0117W/cm2Round0.0299W/cm2Round0.0299W/cm2
Rectangle0.0059W/cm2Rectangle0.0235W/cm2Rectangle0.0313W/cm2
Oval0.002W/cm2Oval0.002W/cm2Oval0.002W/cm2
Butterfly0.0023W/cm2Butterfly0.0023W/cm2Butterfly0.0023W/cm2
Gourd0.0022W/cm2Gourd0.0022W/cm2
Saddle0.0011W/cm2Saddle0.0011W/cm2
(FDA guidance states that a maximum average power density should be less than 0.25 watts per squarecentimeter of electrode conductive surface area to reduce the risk of thermal burns.)
MaximumCurrent forPower Density @500 OhmsUse Jiajian® TENS(K112288)Use Jiajian® TENS(K112288) forcalculation, the Maximum R.M.S current is 13.7mAr.m.s.Use Jiajian® TENS(K112288) forcalculation, the Maximum R.M.S current is 13.7mAr.m.s.Calculation formulas as follow:Maximum Output Voltage:36V@500ΩPulse Duration: 60300µSWaveform:Pulsed MonophasicFrequency:0.5120Hz$Ir.m.s= *72mA=13.7 mAr.m.s$
MaterialComposition- Top cover material- Conductive carbon film:PU, Conductive carbon- Biocompatible conductivehydrogel- Electrode carrier liner- Top cover material- Conductive carbon film:PU, Conductive carbon- Biocompatible conductive hydrogel- Electrode carrier liner- Top cover material- Electrically conductive carboncloth- Biocompatible conductivehydrogel- Electrode carrier liner
Lead wire typeandcharacteristicsLead wire with female socket, orsnap button with male snapconnectorLead wire with female socket, or snapbutton with male snap connectorLead wire with female socket

{5}------------------------------------------------

Suzhou Ennova Electronics Technology Co., Ltd.

Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, Jiangsu Province, 215124 China

{6}------------------------------------------------

Suzhou Ennova Electronics Technology Co., Ltd.

Suzhou Ennova Electronics Technology Co., Ltd.
Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, Jiangsu Province, 215124 China

{7}------------------------------------------------

Suzhou Ennova Electronics Technology Co., Ltd.

Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, Jiangsu Province, 215124 China

thickness
BiocompatibilityISO10993-5:2009;ISO10993-5:2009;ISO 10993-5: 1999;
ISO10993-10:2010ISO10993-10:2010ISO 10993-10: 2002
Electrical safetyLead wire meets Clause 8.5.2.3Lead wire meets Clause 8.5.2.3 ofLead wire meets 56.3C of IEC
of AAMI/ANSI ES60601-1AAMI/ANSI ES60601-160601-1
OtherPerformanceGood electrical conductivity,good adhesive propertyGood electrical conductivity, goodadhesive propertyGood electrical conductivity,good adhesive property
Discussion:
Similarities between New deviceand Predicate Device:Similar intended use, composition, safety performance,complied standards
Differences between New deviceand Predicate Device:Note 1:The Size and Patient contact area are a little different between proposed device andpredicate device, based on the impedance dispersion test, these differences will not raiseany questions of safety and effectiveness of the subject device.Note 2:Ennova devices used AG700 Series Hydrogel from Axelgaard which was 510(k) clearedwith number of K130987. Jiajian devices used ValuTrode@GEL from Axelgaard which was510(k) cleared with number of K970426. The conductive hydrogel of Zekang device isimported from Covidien, USA, which was 510(k) cleared with number of K100418. Sincethey are all from cleared devices, these differences will not raise any questions of safety andeffectiveness of the subject device.Note 3:There are a little difference of the impedance between the proposed device and predicatedevice, but the impedance values of proposed device are within the range specified by thepredicate device, so these small difference will not raise any questions of safety andeffectiveness.
Conclusion:

Based on successful biocompatibility testing of the skin contacting conductive hydrogel, the electrical performance of the insulated lead wire components and electrode current distribution test results, Ennova Self-adhesive Electrode is safe and effective when used as an interface between a user's skin and an approved nerve and muscle stimulation device. The conclusions drawn from the non-clinical tests demonstrate that the device is as effective, and performs as well as the legally marketed devices identified in the submission. Thus the subject device is substantially equivalent to the predicate devices.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).