K Number
K152374

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-01-13

(145 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LL TENS 160A and LL TENS 160B are used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

Device Description

LL TENS 160A and LL TENS 160B are handheld battery powered TENS devices for pain relief intended for over-the-counter use. The device would generate electrical pulses and transmit it to the electrodes, which are attached to the patient's (user's) skin. Consequently, the electrical pulses would then pass through the skin to the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain.

LL TENS 160A and LL TENS 160B has two output channels and four preset programs. The program mode is displayed on LCD. The user can adjust the output intensity by 14 steps.

LL TENS 160A and LL TENS 160B have same hardware, software and mechanical structure. The difference is that LL TENS 160A's LCD display is placed on the bottom of keys while LL TENS 160B's LCD display is on the top of keys.

Since the device is battery powered, there is no connection to AC mains supply.

AI/ML Overview

This document is a 510(k) Summary for the LL TENS 160A and LL TENS 160B devices, which are Transcutaneous Electrical Nerve Stimulators (T.E.N.S.) intended for over-the-counter use. The purpose of the document is to demonstrate substantial equivalence to a legally marketed predicate device, "Pain Buddy" (K102051).

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly define "acceptance criteria" as pass/fail thresholds for clinical performance metrics in the way a diagnostic AI device might. Instead, it focuses on demonstrating substantial equivalence to a predicate device by comparing technical characteristics and safety parameters against established standards.

The "acceptance criteria" implicitly revolve around ensuring the new device's electrical stimulation parameters, safety features, and usability are within acceptable limits as defined by relevant IEC standards and are comparable to or safer than the predicate device.

CharacteristicAcceptance Criteria (Implicit from Predicate & Standards)Reported Device Performance (LL TENS 160A/160B)Met Criteria?
Indications for UseSame as predicate: Symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.Same as predicate.Yes
Patient PopulationAdultAdultYes
Prescription/OTCOTCOTCYes
Environment of UseClinics, hospital, and home environmentClinics, hospital, and home environmentYes
Maximum Phase Charge (500 Ω)Less than safety limit (not explicitly stated, but inferred from conclusion statement)18.1 µCYes (stated as "less than the limit of the safety standard")
Maximum Average Current (500 Ω)Below 10 mA0.9 mAYes (stated as "below the 10mA limit")
Maximum Average Power Density (500 Ω)Less than 0.25 W/cm²0.0025 W/cm²Yes (stated as "less than 0.25W/cm²")
Electrical SafetyCompliance with IEC 60601-1:2005 + A1:2012Not explicitly detailed, but implied by adherence to standards.Yes (stated as "The relevant standards including..." followed by a list of IEC standards)
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2007Not explicitly detailed, but implied by adherence to standards.Yes
Home Healthcare Environment SafetyCompliance with IEC 60601-1-11:2010Not explicitly detailed, but implied by adherence to standards.Yes
UsabilityDevice can be used correctly and safely by users.Usability study performed and showed users were able to use the device correctly and safely.Yes
Overall Safety & EffectivenessDifferences from predicate device do not raise new questions of safety and effectiveness.Concluded that differences do not raise questions on safety and effectiveness, and the device is as safe as the predicate.Yes

Note: The document focuses on demonstrating equivalency rather than specific performance metrics against pre-defined, numerical acceptance criteria for clinical efficacy, as this is a 510(k) for a TENS device, not an AI diagnostic. The electrical parameters are compared to the predicate, with the overarching "acceptance" being that any differences do not impact safety or effectiveness beyond what is accepted for similar devices.

2. Sample Size Used for the Test Set and Data Provenance

For the usability study, the document states "A usability study has been performed," but it does not specify the sample size of participants in the test set.

  • Data Provenance: Not explicitly stated, meaning no information on country of origin or if it was retrospective or prospective. Given the nature of a usability study, it would typically be prospective, involving human participants performing tasks with the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

For the usability study, the document does not mention the number of experts used to establish ground truth (e.g., assessing correct usage or safety issues), nor does it specify their qualifications. Typically, usability studies would involve human factors engineers or independent evaluators, but this is not detailed here.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method for the usability study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This document pertains to Transcutaneous Electrical Nerve Stimulators (TENS devices), which are physical medical devices for pain relief, not AI software or diagnostic imaging devices that would typically involve human readers (e.g., radiologists) or AI assistance. The concept of "AI vs. without AI assistance" is not applicable here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, a standalone (algorithm only) performance study was not done. This is not an AI device. The "device" in question is a hardware TENS unit. The document mentions "software uses same mechanism as Pain Buddy," but this refers to the embedded firmware controlling the TENS unit, not a standalone algorithm in the AI sense.

7. The type of ground truth used

For the usability study, the "ground truth" would be the observed ability of users to operate the device correctly and safely, likely assessed against predefined operational requirements and safety criteria. This would be qualitative and quantitative data gathered during the usability testing sessions.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device that requires a "training set." The device is a TENS unit.

9. How the Ground Truth for the Training Set was Established

Not applicable. As above, there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other, representing the human element of the department's work.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 13, 2016

Fuji Dynamics Ltd Man Man Chung Unit 1-3, 23/F., Laws Commercial Plaza 788 Cheung Sha Wan Road, Kowloon Hong Kong

Re: K152374

Trade/Device Name: LL TENS 160A, LL TENS 160B Regulation Number: 21 CFR 882.5890 Regulation Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Regulatory Class: Class II Product Code: NUH Dated: December 8, 2015 Received: December 11, 2015

Dear Man Man Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

William J.Heetderks -S

for Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152374

Device Name

LL TENS 160A LL TENS 160B

Indications for Use (Describe)

The LL TENS 160A and LL TENS 160B are used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

Type of Use (Select one or both, as applicable):

Experimental Use (Applicable if an IDE with a Pediatric Extrapolation Plan)
Compassionate Use (Applicable if an IDE with a Pediatric Extrapolation Plan)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date of submission prepared: 24 December 2015

Submitter:Fuji Dynamics Ltd.
Address :Unit 1-3, 23/F., Laws Commercial Plaza
788 Cheung Sha Wan Road, Kowloon, Hong Kong
Tel :(852) 2786 4218
Fax :(852) 2744 6775
E-Mail :fdman@fuji-dynamics.com.hk
Official Contact:Man, Man Chung
Address of the manufacturing facility :No. A46, Gao Er South Road,
Pin Shan, Tangxia Town,
Dong Guan City,
Guang Dong Province, China
SUBMITTED DEVICE :
Generic Name:Transcutaneous Electrical Nerve Stimulator
(T.E.N.S)
Proprietary or Trade Name:LL TENS 160A
LL TENS 160B
Common / Usual Name:Stimulator, nerve, transcutaneous, over-the counter
Classification Name:Stimulator, nerve, transcutaneous, over-the counter21 CFR 882.5890
Product Code:NUH
Device Panel:Neurology
Device Classification:Class II

{4}------------------------------------------------

PREDICATE DEVICES :

Device Name: Pain Buddy Applicant: Biomedical Life Systems. Inc. 510(k) Number: K102051 Product Code: NUH

INDICATIONS FOR USE :

LL TENS 160A and LL TENS 160B are used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

DEVICE DESCRIPTION :

LL TENS 160A and LL TENS 160B are handheld battery powered TENS devices for pain relief intended for over-the-counter use. The device would generate electrical pulses and transmit it to the electrodes, which are attached to the patient's (user's) skin. Consequently, the electrical pulses would then pass through the skin to the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain.

LL TENS 160A and LL TENS 160B has two output channels and four preset programs. The program mode is displayed on LCD. The user can adjust the output intensity by 14 steps.

LL TENS 160A and LL TENS 160B have same hardware, software and mechanical structure. The difference is that LL TENS 160A's LCD display is placed on the bottom of keys while LL TENS 160B's LCD display is on the top of keys.

Since the device is battery powered, there is no connection to AC mains supply.

{5}------------------------------------------------

SUMMARY OF SUBSTANTIAL EQUIVALENCE

AttributeApplicant DevicePredicate Device
Product NameLL TENS 160ALL TENS 160BPain Buddy
510(K) numberK152374K102051
Product CodeNUHNUH
Indications for useThis device is used for thesymptomatic relief andmanagement of chronic intractablepain and/or as an adjunctivetreatment in the management ofpost-surgical and post-traumaticacute pain.Same as applicant device
Patient PopulationAdultAdult
Prescriptive or OTCOTCOTC
Environment of useClinics, hospital and homeenvironmentClinics, hospital and homeenvironment
Number of outputmodes41
Number of outputchannels21
WaveformAlternating Bi-phasic symmetricalRectangular waveformAsymmetrical BiphasicRectangular waveform
Maximum outputVoltage (max)
500 Ω70V30V
2K Ω98V36.8V
10K Ω108V38.4V
Maximum outputCurrent (max)
500 Ω140 mA60 mA
2K Ω49 mA18.4 mA
10K Ω10.8 mA3.84 mA
Maximum Phasecharge (500 Ω)18.1 μC13.3 μC
Maximum AverageCurrent (500 Ω)0.9 mA0.2 mA
Maximum Currentdensity (500 Ω)0.062 mA/cm²0.066 mA
Maximum PowerDensity (500 Ω)2.25 mW/cm²2.28 mW / cm²
Frequency (Hz)90 Hz80 Hz
Pulse Duration (μs)208 μs160 μs
Burst ModeYesNo
Timer range (min)Continuous, 15 min, 30 min, 45 minand 60 min selectable.Continuous
Indication display-On/Off statusYesYes
-Low BatteryYesYes
-Voltage/Current levelYesYes
-Output modeYesYes
-Time to cut-offNoNo
Power Source2 AAA Batteries2 AAA Batteries
Dimensions (mm)38 (W) x 127 (L) x 20 (H) mm35 (W) x 125(L) x 16(H) mm
Weight70 g60 g
Housing materialABSABS
Microprocessor controlYesYes
Automatic OverloadtripNoNo
Automatic no-load tripYesYes
Automatic shut-offYesYes
Electrode compliancewith 21 CFR 898YesYes
Electrode cableYesYes

{6}------------------------------------------------

E. 510(k) Summary

{7}------------------------------------------------

DIFFERENCES BETWEEN NEW DEVICE AND PREDICATE DEVICE:

The technical characteristics of LL TENS 160A / LL TENS 160B are similar to those of the predicate devices in design, energy source, intended use and function. Like the predicate device Pain Buddy, the LL TENS 160A / LL TENS 160B is a device used to apply an electrical current to electrodes on a patient's skin to relieve pain.

LL TENS 160A and LL TENS 160B are developed on the similar platform as the predicate device Pain Buddy. On hardware, the schematic and the use of electronics components are similar. The software uses same mechanism as Pain Buddy. Therefore the basic timing, key scanning and generation of pulse are the same.

Some output parameters, including maximum output voltage, maximum output current, phase charge, maximum average current, maximum frequency and pulse duration are different from those of Pain Buddy. However, the maximum phase charge of LL TENS is 18.1uC, which is less than the limit of the safety standard. The Maximum average current is 0.9mA, which is also below the 10mA limit. The maximum average power density is 0.0025 W/cm², which is also less than 0.25W/cm². The maximum frequency of LL TENS is 90Hz. It is slightly higher than those of predicate device and thus does not raise question of safety and effectiveness. Finally, the maximum pulse duration is 208us and higher than of predicate device. Nevertheless, the maximum phase charge of LL TENS is still less than safety limit in this pulse duration. In conclusion, these differences between LL TENS and predicate device do not raise a question in safety and effectiveness.

The LL TENS 160A and LL TENS 160B are views as substantially equivalent to the predicate devices because the electrical stimulation provided by the LL TENS 160A and LL TENS 160B is substantially equivalent to that commonly employed by TENS devices that have been cleared for marketing without prescription labeling, i.e. for OTC use.

The differences between LL TENS and the predicate device do not raise questions on safety and effectiveness. In other word, LL TENS 160A and LL TENS 160B are as safe as the predicate device.

PERFORMANCE TESTS:

The relevant standards including :

IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance.

{8}------------------------------------------------

IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance. Collateral Standard: Electromagnetic Compatibility. (Edition 3).

IEC 60601-1-11:2010 Medical electrical equipment - Part 1-11 General requirement for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

IEC 60601-1-6:2010 Medical electrical equipment - Part 1-6 General requirements for basic safety and essential performance – Collateral standard: Usability.

USABILITY STUDY:

A usability study has been performed and showed that users were able to use the device correctly and safety.

CONCLUSION:

The electrical stimulation provided by the LL TENS 160A and LL TENS 160B is similar to that commonly employed by TENS devices that have been cleared for marketing without prescription labeling (i.e. OTC).

LL TENS 160A and LL TENS 160B have the same intended use and the same technical characteristics as the OTC cleared predicate device, Pain Buddy [510(k) No.: K102051].

The differences between LLTENS and the predicate device do not raise questions on safety and effectiveness. In other words, LL TENS 160A and LL TENS 160B are as safe as the predicate device.

Thus, the new device LL TENS 160A and LL TENS 160B are substantially equivalent to the predicate device.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).