K Number
K152330

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-12-23

(127 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended for use by healthcare professionals.

The Expression MR400 MRI Patient Monitoring System (Model MR400) provides monitoring for the following vital sign parameters: ECG, pulse oximetry (SpO2), non-invasive blood pressure (NIBP), and optionally, invasive blood pressure (IBP), carbon dioxide (CO2) and respiration rate, anesthetic agents, nitrous oxide (N2O), oxygen (02), and/or temperature.

Device Description

The Expression MR400 is a multi-parameter patient monitoring system used to monitor the vital signs of patients in an MRI magnet room and throughout an MRI suite. The MR400's software simultaneously processes and displays multiple parameters, waveforms, measurement numeric values and alarms. All patient information is provided on a wheeled cart which consists of a patient monitor with a touch screen display. The device is powered by either AC line power or its internal battery. The standard and optional parameters which can be monitored by the MR400 include:

  • Electrocardiogram (ECG)
  • Heart rate (HR, ECG and SPO2 derived) ●
  • Blood oxygen saturation/pulse oximetry (SPO2)
  • End-tidal and fractional inspired CO2 (EtCO2 and FiCO2) ●
  • Invasive blood pressure (IBP) (P1 and P2) ●
  • Anesthetic agents (AGENT) ●
    • o Desflurane (DES)
    • o Enflurane (ENF)
    • o Halothane (HAL)
    • o Isoflurane (ISO)
    • o Sevoflurane (SEV)
  • Fractional inspired O2 (FiO2), and end-tidal and fractional inspired N2O (EtN2O ● and FiN2O)
  • Temperature (TEMP) ●
  • Non-invasive blood pressure (NIBP)
  • . Respiration rate (CO2-derived and/or bellows-derived)

The MR400 consists of the following key components:

  • Cart with touchscreen display ●
  • Wireless ECG module .
  • Wireless SpO2 module ●
  • Batteries for both the Cart and Wireless Modules ●
  • Battery charger ●
  • Power cord ●

Accessories to the MR400 are offered to accommodate various patient sizes-neonatal, infant, pediatric, adult, and large adult. Key accessories include:

  • . ECG lead cables and electrodes
  • SPO2 probes and clips ●
  • NIBP cuffs and hoses ●
  • IBP transducer
  • Temperature probe and sheath ●
  • Cannulas ●
  • Water traps ●
  • Chest pneumograph ●
  • Expression IP5 (optional secondary display for control room w/printer) ●
AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Expression MR400 MRI Patient Monitoring System

Page 1

DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Center - WO66-G609
Silver Spring, MD 20993-0002

Invivo Corporation
c/o Christine Trefethen
Regulatory Manager
Philips Medical Systems
3000 Minuteman Rd.
Andover, MA 01801

Re: K152330
Trade/Device Name: Expression MR400 MRI Patient Monitoring System (Model MR400)
Regulation Number: 21 CFR 870.2300
Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm)
Regulatory Class: Class II
Product Code: MWI
Dated: November 24, 2015
Received: November 25, 2015

Dear Christine Trefethen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

December 23, 2015

Page 2

DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service

Food and Drug Administration
10903 New Hampshire Avenue
Document Control Center - WO66-G609
Silver Spring, MD 20993-0002

December 23, 2015

Invivo Corporation
c/o Christine Trefethen
Regulatory Manager
Philips Medical Systems
3000 Minuteman Rd.
Andover, MA 01801

Re: K152330
Trade/Device Name: Expression MR400 MRI Patient Monitoring System (Model MR400)
Regulation Number: 21 CFR 870.2300
Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm)
Regulatory Class: Class II
Product Code: MWI
Dated: November 24, 2015
Received: November 25, 2015

Dear Christine Trefethen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

Page 2 - Christine Trefethen

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

Page 3

FORM FDA 3881 (8/14)

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.

510(k) Number (if known): K152330

Device Name: Expression MR400 MRI Patient Monitoring System (Model MR400)

Indications for Use (Describe)

The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended for use by healthcare professionals.

The Expression MR400 MRI Patient Monitoring System (Model MR400) provides monitoring for the following vital sign parameters: ECG, pulse oximetry (SpO2), non-invasive blood pressure (NIBP), and optionally, invasive blood pressure (IBP), carbon dioxide (CO2) and respiration rate, anesthetic agents, nitrous oxide (N2O), oxygen (O2), and/or temperature.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Appendix D Page 1 of 1

Page 4

K152330
Page 1 of 5

SPECIAL 510(K) SUMMARY

SUBMITTER INFORMATION

FieldInformation
Establishment Name:Invivo Corporation
Establishment Address:12151 Research ParkwayOrlando, FL 32826 USA
Company Phone:(407) 252-0414
Company Fax:(407) 249-2022
Person to contact regarding questions:Rusty KellySr. Manager Quality and Regulatory(407)-252-0414Rusty.Kelly@philips.com
Establishment Registration Number:1051786
Date Summary Prepared:August 11, 2015

DEVICE IDENTIFICATION

FieldInformation
Trade name:Expression MR400 MRI Patient Monitoring System (Model MR400)
Common name:MRI patient monitoring system
Classification name:Cardiac monitor (including cardiotachometer and rate alarm)(21 CFR 870.2300, Product Code MWI)

Page 5

K152330
Page 2 of 5

IDENTIFICATION OF LEGALLY MARKETED PREDICATE DEVICE

The Expression MR400 MRI Patient Monitoring System also referred to as the Expression MR400 or MR400, is substantially equivalent to the following predicate device:

Predicate DeviceManufacturer510(k) No.Clearance Date
MRI Patient Monitoring System (Model 865214)Invivo CorporationK124061Feb 22, 2013

DEVICE DESCRIPTION

The Expression MR400 is a multi-parameter patient monitoring system used to monitor the vital signs of patients in an MRI magnet room and throughout an MRI suite. The MR400's software simultaneously processes and displays multiple parameters, waveforms, measurement numeric values and alarms. All patient information is provided on a wheeled cart which consists of a patient monitor with a touch screen display. The device is powered by either AC line power or its internal battery. The standard and optional parameters which can be monitored by the MR400 include:

  • Electrocardiogram (ECG)
  • Heart rate (HR, ECG and SPO2 derived)
  • Blood oxygen saturation/pulse oximetry (SPO2)
  • End-tidal and fractional inspired CO2 (EtCO2 and FiCO2)
  • Invasive blood pressure (IBP) (P1 and P2)
  • Anesthetic agents (AGENT)
    • Desflurane (DES)
    • Enflurane (ENF)
    • Halothane (HAL)
    • Isoflurane (ISO)
    • Sevoflurane (SEV)
  • Fractional inspired O2 (FiO2), and end-tidal and fractional inspired N2O (EtN2O and FiN2O)
  • Temperature (TEMP)
  • Non-invasive blood pressure (NIBP)
  • Respiration rate (CO2-derived and/or bellows-derived)

Page 6

K152330
Page 3 of 5

The MR400 consists of the following key components:

  • Cart with touchscreen display
  • Wireless ECG module
  • Wireless SpO2 module
  • Batteries for both the Cart and Wireless Modules
  • Battery charger
  • Power cord

Accessories to the MR400 are offered to accommodate various patient sizes—neonatal, infant, pediatric, adult, and large adult. Key accessories include:

  • ECG lead cables and electrodes
  • SPO2 probes and clips
  • NIBP cuffs and hoses
  • IBP transducer
  • Temperature probe and sheath
  • Cannulas
  • Water traps
  • Chest pneumograph
  • Expression IP5 (optional secondary display for control room w/printer)

INDICATIONS FOR USE

The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended for use by healthcare professionals.

The Expression MR400 MRI Patient Monitoring System (Model MR400) provides monitoring for the following vital sign parameters: ECG, pulse oximetry (SpO2), non-invasive blood pressure (NIBP), and optionally, invasive blood pressure (IBP), carbon dioxide (CO2) and respiration rate, anesthetic agents, nitrous oxide (N2O), oxygen (O2), and/or temperature.

TECHNOLOGICAL CHARACTERISTICS

The modified device has the same fundamental scientific technology as the predicate device.

Page 7

K152330
Page 4 of 5

Technological CharacteristicPredicate: MRI Patient Monitoring System (K124061)Modified Device: Expression MR400 MRI Patient Monitoring SystemComparison
Cart
DisplayDetachable, located on top front of displayPermanently installed, located on the rear of the displaySubstantially Equivalent
Expression IP5Software is compatible with the predicate deviceSoftware was updated to be compatible with the predicate device and the Expression MR400Substantially Equivalent
PrinterIntegrated on side of displayNo printer permanently installed on display.Substantially Equivalent
Alarms Set Up, Printer Set Up and Monitor Set UpSeparate key on the keypad which is visible to the operator at all timesCollapsed within one Setup key on the virtual keypad that is visible to the operator at all timesSubstantially Equivalent
Power SupplyExternalIntegratedSubstantially Equivalent
Internal BatteriesFour User-replaceableTwo Permanently installed and two User-replaceableSubstantially Equivalent
AlarmsVital Sign AlarmsAdded user adjustable alarms for extreme bradycardia, extreme tachycardia, apnea and desaturationSubstantially Equivalent
Non-Invasive Blood Pressure
NIBP ConnectorConnector ports are for unique predicate accessoriesAccommodate locking connectors unique to the modified device accessoriesSubstantially Equivalent
NIBP Hose3 meters in length5 meters in lengthSubstantially Equivalent
NIBP Data Collection & MeasurementPerformed in the processing unit within the cart.Performed in the processing unit within the cart using the picoNIBP OEM ModuleSubstantially Equivalent
CO2
CO2 ConnectorConnector ports are for unique predicate accessoriesAccommodate locking connectors unique to the modified device accessoriesSubstantially Equivalent
CO2 Collection & MeasurementCompleted using Excelitas CO2 SensorCompleted using LoFloC5 Co2 SensorSubstantially Equivalent
CO2 cannula22 ft. in length17ft. in lengthSubstantially Equivalent
ECG
ECG Cable CoverWire bundle enclosed in a silicone foam trunkFlat wire assembly enclosed in Azote foamSubstantially Equivalent
Wireless ECG Module- communication5 user-selectable networks10 user-selectable networks. (only 5 per module, either 1-5 or 6-10)Substantially Equivalent

Page 8

K152330
Page 5 of 5

Technological CharacteristicPredicate: MRI Patient Monitoring System (K124061)Modified Device: Expression MR400 MRI Patient Monitoring SystemComparison
Wireless ECG Module- batteryOperates on one (1) 3.7 VDC user-replaceable lithium batteryOperates on two (2) 3.7 VDC user-replaceable lithium batteriesSubstantially Equivalent
Wireless ECG Module- housingPlastic covering, single battery slot, Ingress Protection IPX1Plastic housing made of same materials as predicate, supports two batteries, has eject buttons for batteries, improved Ingress Protection rating of IP21Substantially Equivalent
SPO2
Wireless SPO2 Module- communication5 user-selectable networks10 user-selectable networks. (only 5 per module, either 1-5 or 6-10)Substantially Equivalent
Wireless SPO2 Module- housingPlastic covering, single battery slot, Ingress Protection IPX1Plastic housing made of same materials as predicate, has eject button for battery, improved Ingress Protection rating of IP21Substantially Equivalent

SUMMARY OF NON-CLINICAL PERFORMANCE DATA

Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and the specifications of the modified device. Results of the non-clinical testing demonstrate that the modified device operates as intended within the performance specifications and is substantially equivalent to the cleared device. The results do not raise issues regarding the safety and effectiveness of the device. Clinical data was not required to substantiate claims of safety and effectiveness. The modified device, the Expression MR400 MRI Patient Monitoring System, described in this submission is substantially equivalent to the cleared device, the MRI Patient Monitoring System (Model 865214).

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).