K Number
K152322

Validate with FDA (Live)

Device Name
Resolve 2.0
Manufacturer
Date Cleared
2015-12-16

(121 days)

Product Code
Regulation Number
872.3260
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Resolve 2.0 is recommended for use on dental restorations, external to the mouth, prior to insertion into the mouth. After applying Resolve 2.0 to cleanse the restoration, the surface of the restoration should be washed thoroughly with water, and then air dried before placement in the mouth. For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.

Device Description

Resolve 2.0 is a thin, non-alcohol based gel containing chlorhexidine gluconate. This thin gel because of its water solubility, abrasive action and low viscosity when applied to dental restorations, acts as a cleaning agent. This gel is provided in easy to use syringes and floc-tipped applicator tips to precisely deliver the gel to the intended areas. For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.

AI/ML Overview

It appears the provided document is a 510(k) premarket notification for a dental device named "Resolve 2.0." This type of document is for demonstrating substantial equivalence to a predicate device, not for establishing novel acceptance criteria or conducting a comprehensive clinical study in the way one might for a new drug or a high-risk AI-powered diagnostic device.

Therefore, the requested information categories related to acceptance criteria, detailed study design, sample sizes for test/training sets, expert adjudication, MRMC studies, or specific ground truths for AI models do not directly apply to this document. This submission focuses on demonstrating similarity to an existing, legally marketed device.

However, I can extract the information relevant to the equivalence demonstration as presented in the document:

1. A table of acceptance criteria and the reported device performance:

The document doesn't present "acceptance criteria" in the traditional sense of a performance study with defined thresholds. Instead, it compares the technological characteristics of Resolve 2.0 to its predicate device, Consepsis. The "acceptance" is based on demonstrating that these characteristics are equivalent.

PropertyResolve 2.0Consepsis*Equivalence Conclusion
Cleansing ActivityYesYesEquivalent
Contains Chlorhexidine gluconateYesYesEquivalent
Water SolubleYesYesEquivalent
pH66Equivalent
Contains silicate (sand)YesYesEquivalent
ConsistencyThin liquidThin liquidEquivalent
PigmentedYesYesEquivalent
Ageing/Shelf Life Test - Accelerated Ageing at 37°CPASS, 2.0+ year shelf life establishedPASS, 2.0+ year shelf life establishedEquivalent
Biocompatibility (ISO 10993-5)"Moderate" cytotoxicity"Moderate" cytotoxicityEquivalent

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

This information is not applicable/not provided in this type of 510(k) submission. The evaluation is based on physicochemical and biocompatibility testing, not clinical data sets in the AI sense.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not applicable/not provided as there is no "test set" and "ground truth" in the context of an AI device. The evaluation is based on laboratory testing and comparison to the predicate.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable/not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable/not provided. This is not an AI-powered device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable/not provided. This is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

For the properties evaluated (e.g., pH, water solubility, cleansing activity, biocompatibility), the "ground truth" would be established through laboratory testing and established scientific methods/standards (e.g., pH meter readings, solubility tests, accelerated aging protocols, ISO 10993-5 standard for biocompatibility).

8. The sample size for the training set:

This information is not applicable/not provided. This is not an AI device with a training set.

9. How the ground truth for the training set was established:

This information is not applicable/not provided.

Summary of the Study:

The document describes a benchtop and biocompatibility study to demonstrate that Resolve 2.0 is substantially equivalent to the predicate device, Consepsis. The study involved:

  • Comparison of Technological Characteristics: Resolve 2.0 was compared to Consepsis across several physical and chemical properties including cleansing activity, active ingredient (chlorhexidine gluconate), water solubility, pH, presence of silicate (sand), consistency, pigmentation, and accelerated aging/shelf life.
  • Biocompatibility Testing: Resolve 2.0 was evaluated per ISO 10993-5 "Biological Evaluation of Medical Devices" for cytotoxicity. The results were compared to those of Consepsis under the same standard. Both devices exhibited "Moderate" cytotoxicity, showing equivalence.

The conclusion drawn from this "study" (which is more accurately a set of comparative tests) is that Resolve 2.0 is substantially equivalent to Consepsis in design, composition, biocompatibility, performance, aging, limited intended uses, and safety and effectiveness. This equivalence allows the device to be marketed without requiring a full Premarket Approval (PMA) process.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a central emblem. The emblem consists of a stylized representation of three human profiles facing right, stacked on top of each other, symbolizing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 16, 2015

Denali Corporation Jan G. Stannard President 134 Old Washington Street Hanover, Massachusetts 02339

Re: K152322

Trade/Device Name: Resolve 2.0 Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity varnish Regulatory Class: Class II Product Code: LBH Dated: November 4, 2015 Received: November 9, 2015

Dear Ms. Jan G. Stannard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erin I. Keith -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510 (k) Number 152322

Device Name

Resolve 2.0

Indications for Use:

Resolve 2.0 is recommended for use on dental restorations, external to the mouth, prior to insertion into the mouth. After applying Resolve 2.0 to cleanse the restoration, the surface of the restoration should be washed thoroughly with water, and then air dried before placement in the mouth.

For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.

Please do not write below this line. Continue on another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

or

Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use

{3}------------------------------------------------

denali corporation 134 Old Washington Street

Hanover, MA 023339 USA

T 781.826.9190

F 781.826.4465
denalicorporation.com

510 (k) Summary

December 10, 2015

ADDRESSDENALI CORPORATION134 Old Washington StreetHanover, MA 02339-1629
OWNER/CONTACT PERSONDr. Jan G. StannardTEL:FAX:781-826-9190781-826-4465j.stannard@denalicorporation.com
TRADE NAMEResolve 2.0
COMMON NAMEExternal Cleansing Solution, Cavity Varnish
CLASSIFICATION NAMECavity Varnish, Class II (21 CFR 872.3260, Product Code PME)
REGISTRATION3006367836
PRIMARY PREDICATE DEVICEConsepsis, Ultradent Products (K925375)
EQUIVALENCEThe predicate product has been found substantially equivalentunder the 510(k) premarket notification process as Class II DentalDevices under CFR PME 872.3260.
DEVICE DESCRIPTIONResolve 2.0 is a thin, non-alcohol based gel containing chlorhexidine gluconate.This thin gel because of its water solubility, abrasive action and low viscosity whenapplied to dental restorations, acts as a cleaning agent. This gel is provided in easyto use syringes and floc-tipped applicator tips to precisely deliver the gelto the intended areas.For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.
INTENDED USEResolve 2.0 is recommended for use on dental restorations, external to themouth, prior to insertion into the mouth. After applying Resolve 2.0 to cleanse therestoration, the surface of the restoration should be washed thoroughly withwater, and then air dried before placement in the mouth.For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Denali Corporation. The logo features a blue and black geometric shape with a blue sphere and a curved line around it. Next to the logo is the text "denali corporation" in gray. Below the company name is the address "134 Old Washington Street, Hanover, MA 02339 USA".

T 781.826.9190 F 781.826.4465 denalicorporation.com

TECHNOLOGICAL CHARACTERISTICS SUMMARY

Resolve 2.0 has the same technological characteristics as Consepsis including: design, composition, biocompatibility, performance, ageing, intended use, physical characteristics, including pH, shelf life, water solubility, and cleansing activity, as summarized in the Table below.

PROPERTYResolve 2.0Consepsis*
Cleansing ActivityYesYes
Contains Chlorhexidine gluconateYesYes
Water SolubleYesYes
pH66
Contains silicate (sand)YesYes
ConsistencyThin liquidThin liquid
PigmentedYesYes
Ageing/Shelf Life Test - Accelerated Ageing at 37°CPASS, 2.0+ year shelf life establishedPASS, 2.0+ year shelf life established

BIOCOMPATIBILITY

Resolve 2.0 was found to be biocompatible, and is consistent in formulation to Consepsis, in its limited intended uses, biocompatibility, and properties, Resolve 2.0 was evaluated per ISO 10993-5 "Biological Evaluation of Medical Devices". Resolve 2.0 and Consepsis according to ISO 10993-5 both had cytotoxicity values of "Moderate", exhibiting a reactivity zone extending from the specimen up to 1.0 cm. According to these results Resolve 2.0 was found to be equivalent in biocompatibilty compared to the predicate product Consepsis.

SUBSTANTIAL Resolve 2.0 is substantially equivalent in design, composition, biocompatibility EQUIVALENCE according to ISO 10993, performance, ageing, limited intended uses, and safety DETERMINATION and effectiveness to Consepsis. This assessment is based upon a comparison AND SUMMARY of the composition and physical properties of water solubility, pH, shelf-life and cleansing activity to Consepsis, as well as Biocompatibility evaluation. The results of these tests are consistent with values of the predicate product.

CONCLUSIONS Resolve 2.0 has been found to be substantially equivalent in design, composition, ageing, biocompatibility, performance, with limited intended uses to Consepsis.

§ 872.3260 Cavity varnish.

(a)
Identification. Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.(b)
Classification. Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.