(337 days)
The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The FDR Visionary Suite enables radiographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The FDR Visionary Suite uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes.
The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/ trained professionals.
The device is not intended for mammographic applications.
The FDR Visionary Suite is an X-ray radiography system that is mainly used for the radiography of various regions of the patient's body in a standing or recumbent position. The device can be used in a wide range of applications of digital radiography. The device consists of an X-ray high voltage generator, X-ray tube unit, X-ray tube support, collimator and digital radiography system. The system can be configured with radiographic table and/or radiographic stand as well.
Optionally, the device is also used to perform tomosynthesis radiography by two different reconstruction modes, e.g. Filtered Back-Projection (FBP) mode and Shift-and-Add (SA) mode. SA reconstructed images are intended to be used alongside FBP reconstructed images and not for stand-alone diagnostic purpose.
This document pertains to the FDR Visionary Suite, a stationary X-ray system. The acceptance criteria and the study proving the device meets these criteria are primarily based on non-clinical performance testing and substantial equivalence to predicate devices, rather than a clinical study with detailed performance metrics.
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Aspect | Acceptance Criteria (Implied by Substantial Equivalence to Predicate) | Reported Device Performance |
|---|---|---|
| Image Acquisition & Processing | Functionality and image quality comparable to predicate devices (K093427, K121499, K101686, K122454) for general radiography and tomosynthesis, including options for image stitching and dual-energy subtraction. | Non-clinical performance bench tests and image quality evaluations were conducted to demonstrate substantial equivalence of image processing techniques. This included phantom tests with various anatomical body phantoms. Specifically, for Dual Energy Subtraction, an additional test assured distinction of lesions overlying bones to calcification/tubercle shade. For metal artifact reduction in tomosynthesis, the FBP mode of the new device showed a "same level of metal artifact reduction capability as the mode with artifact reduction of its predicate device." |
| Tomosynthesis Specifics | Shot Interval: 200msec; Exposure number: Max 60; Angulation: Table Max 60 deg, Wall Stand Max 40 deg; Slice Interval: Min 1mm; Pixel Size: 150um; Processing: FUJIFILM Tomosynthesis processing (DR-ID 300CL/900CL) with PPU. | The new device matches these specifications, using FUJIFILM Tomosynthesis processing (DR-ID 900CL) with Pre-Processing Unit (PPU). Optional FBP and Shift-and-Add (SA) modes are available, with SA images intended for use alongside FBP, not for stand-alone diagnostic purposes. |
| Electromagnetic Compatibility | Conformity with IEC60601-1-2. | In conformity with IEC60601-1-2 (difference in edition, but deemed not to affect safety/effectiveness). |
| Materials & Biocompatibility | Conformity with IEC10993-1. | All materials and components conform to IEC10993-1 and are in clinical use by other commercial products. Difference from predicate device deemed not to affect safety/effectiveness. |
| Sterility | Not sterilized. | Not sterilized (same as predicate). |
| Electrical, Mechanical, Chemical, Thermal Safety | Substantial equivalence to predicate by conformed standards. | Substantially equivalent by their conformed standards. |
| Software Verification & Validation | Adherence to FDA guidance for medical device software and cybersecurity. Assurance of safety and effectiveness despite software processing differences. | Software Risk Analysis and Verification and Validation testing were conducted per FDA guidance. "Based on our risk analysis of software, the difference does not affect its safety and effectiveness." |
| Radiation Safety | Conformity with CFR and IEC60601-1-3. | Substantially equivalent by conformities of CFR and IEC60601-1-3. |
2. Sample Size Used for the Test Set and Data Provenance
- The document does not specify a numerical sample size for the test set. It mentions "phantom tests using several different anatomical body phantoms."
- Data Provenance: Not explicitly stated, but given the manufacturer (SHIMADZU CORPORATION, Japan) and the predicate devices being from FUJIFILM CORPORATION, Japan, it is highly likely the phantom tests and image quality evaluations were conducted in a controlled, non-clinical laboratory setting, likely in Japan. This was a retrospective evaluation against existing predicate device performance, not a prospective clinical trial.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
- The document does not specify the number of experts or their qualifications for establishing ground truth for the test set.
- Given that the testing involved "phantom tests" and "image quality evaluation," the ground truth would likely be established based on known characteristics of the phantoms and objective image quality metrics, potentially interpreted by imaging scientists or engineers rather than clinical experts.
4. Adjudication Method for the Test Set
- The document does not specify any adjudication method for the test set. As the testing appears to be non-clinical (phantom-based), a formal adjudication process involving multiple human readers for diagnostic interpretation would not typically be required or relevant in the same manner as a clinical study.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The result of non-clinical bench testing demonstrates substantial equivalence to the predicate devices. Therefore, it is our understanding that clinical testing is not required in this case."
- Consequently, there is no reported effect size of human reader improvement with or without AI assistance, as this type of study was not conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, standalone performance testing was implicitly performed as part of the "non-clinical performance bench test and image quality evaluate substantial equivalence of image processing technique to each predicate device". These tests assess the device's image acquisition and processing capabilities independent of human interpretation for diagnostic purposes, focusing on technical image quality metrics and functional equivalence to the predicate. The "metal artifact reduction function of tomosynthesis option" testing is an example of an algorithm-only performance assessment.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- For the non-clinical performance testing, the ground truth was based on known characteristics of anatomical phantoms and objective image quality metrics. For example, in metal artifact reduction, the ground truth would be the known presence and characteristics of metal in the phantom and objective metrics of how well the artifact is reduced. For "distinction of lesion overlying bones," the ground truth would be the known presence/absence and characteristics of these features in the phantom.
8. The Sample Size for the Training Set
- The document does not provide information regarding a "training set" or its sample size. This is because the submission primarily focuses on demonstrating substantial equivalence to physically existing predicate devices through engineering and quality testing, rather than an AI/machine learning model that typically requires a large training dataset. The "software processing methods" mentioned are likely traditional image processing algorithms, not deep learning models.
9. How the Ground Truth for the Training Set was Established
- As no training set is described, this information is not applicable and not provided in the document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2016
SHIMADZU CORPORATION % Mr. Don Karle Manager, Customer Service Shimadzu Medical Systems 20101 South Vermont Avenue TORRANCE CA 90502-1328
Re: K152294
Trade/Device Name: FDR Visionary Suite Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: June 10, 2016 Received: June 16, 2016
Dear Mr. Karle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152294
Device Name FDR Visionary Suite
Indications for Use (Describe)
The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The FDR Visionary Suite enables radiographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The FDR Visionary Suite uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes.
The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/ trained professionals.
The device is not intended for mammographic applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
The following information is submitted in accordance with the requirements of 21 CFR§807.92.
1) Date of Submission
August 5th, 2015
2) Submitter
SHIMADZU CORPORATION 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto, 604-8511, Japan Registration Number: 8030233 Phone: +81-75-823-1305 Fax : +81-75-823-1377
3) Primary Contact Person
Toshio Kadowaki Phone: +81-75-823-1920 E-mail: kadowaki@shimadzu.co.jp
4) Secondary Contact Person
Jeffrey Seiler SHIMADZU MEDICAL SYSTEMS USA 20101 South Vermont Ave., Torrance, CA 90502 USA Phone: 310-217-8855 ext 174 Email: seiler@shimadzu-usa.com
5) Device
| Name of Device | : FDR Visionary Suite |
|---|---|
| Common Name | : Digital Radiography System |
| Classification Name | : Stationary x-ray system (21 CFR§892.1680) |
| Classification Panel | : Radiology |
| Regulatory Class | : Class II |
| Product Code | : KPR |
6) Predicate Device
| Primary | K093427 | FDR ACSELERATE SYSTEM, DR-ID 200FLAT PANEL DETECTOR | FUJIFILMCORPORATION |
|---|---|---|---|
| Secondary | K121499 | FUJIFILM TOMOSYNTHESIS OPTIONFOR FDR ACSELERATE STATIONARYX-RAY SYSTEM | FUJIFILMCORPORATION |
| Tertiary | K101686 | FDR IMAGE STITCHING OPTION | FUJIFILMCORPORATION |
| Quaternary | K122454 | FUJIFILM DUAL ENERGY SUBTRACTION(DES) SOFTWARE OPTION | FUJIFILMCORPORATION |
No reference devices were used in this submission.
{4}------------------------------------------------
7) Purpose of Submission
The purpose of this submission is to notify FDA of our intent to introduce a modified device of legally marketed devices, K093427 FDR ACSELERATE SYSTEM, DR-ID 200 FLAT PANEL DETECTOR, K121499 FUJIFILM Tomosynthesis Option for FDR ACSELERATE STATIONARY X-RAY SYSTEM, K101686 FDR IMAGE STITCHING OPTION and K122454 FUJIFILM DUAL ENERGY SUBTRACTION (DES) SOFTWARE OPTION. The last three predicate devices are image processing options for the primary predicate device. All predicate devices are manufactured by FUJIFILM CORPORATION, Japan.
8) Device Description
The FDR Visionary Suite is an X-ray radiography system that is mainly used for the radiography of various regions of the patient's body in a standing or recumbent position. The device can be used in a wide range of applications of digital radiography. The device consists of an X-ray high voltage generator, X-ray tube unit, X-ray tube support, collimator and digital radiography system. The system can be configured with radiographic table and/or radiographic stand as well.
Major hardware components of new device are substantially equivalent to those of primary predicate device as below, even though some of components are optional in new device.
The difference is seen in below tables.
| FDR Visionary Suite(New Device) | AcSelerate Tomosynthesis(K121499) | |
|---|---|---|
| Available type | Stand, Table | Stand, Table |
| Shot Interval | 200msec | 200msec |
| Exposure number | Maximum 60 exposures | Maximum 60 exposures |
| Angulation | Table : Maximum 60 degreeWall stand : Maximum 40 degree | Table : Maximum 60 degreeWall stand : Maximum 40 degree |
| Slice Interval | Minimum 1mm | Minimum 1mm |
| Pixel Size | 150um | 150um |
| Processing | FUJIFILM Tomosynthesis processing (DR-ID 900CL) With Pre-Processing Unit (PPU) | FUJIFILM Tomosynthesis processing (DR-ID 300CL same as DR-ID 900CL) With Pre-Processing Unit (PPU) |
Tomosynthesis Comparison Table
Optionally, the device is also used to perform tomosynthesis radiography by two different reconstruction modes, e.g. Filtered Back-Projection (FBP) mode and Shift-and-Add (SA) mode. SA reconstructed images are intended to be used alongside FBP reconstructed images and not for stand-alone diagnostic purpose.
9) Intended Use
{5}------------------------------------------------
SECTION V:
The FDR Visionary Suite is a stationary X-ray system for taking general radiographic and tomographic images of the whole body. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals.
10) Indications for Use
- The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.
- . The FDR Visionary Suite enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position.
- · The FDR Visionary Suite uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes.
- The Tomosynthesis option is intended to generate tomographic images of human anatomies. . Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
- · The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals.
- . The device is not intended for mammographic applications.
11) Comparison of Technological Characteristics with the predicate devices
At a high level, new device and its predicate devices are based on the following same technological elements:
- · Energy emission to the patient X-ray
- . Power requirement, Environmental requirement
- Mechanism to generate X-ray ●
- Mechanism to acquire, process and store image data
- Use of the hardware components
The following technological differences exist between new and its predicate devices.
- Use of software processing
12) Performance Data
The following performance data were provided in support of the substantial equivalence determination.
Electromagnetic compatibility
Both new device and primary predicate device are in conformity with IEC60601-1-2. The
{6}------------------------------------------------
SECTION V:
510(k) Summary
difference is edition of the standard. Difference of edition of the standard does not affect safety and effectiveness.
Materials and Bio-compatibility testing
All materials and components of new device are in conformity with IEC10993-1. All materials and components of new device are in clinical use by other commercial products. Difference between new device and primary predicate device does not affect safety and effectiveness.
Sterility
Both new device and primary predicate device are not sterilized.
Electrical, Mechanical, Chemical and Thermal safety
As to these safeties, new device and primary predicate device are substantially equivalent by their conformed standards.
Software Verification and Validation Testing
Digital radiography system is using similar, but different software processing method. Therefore, Software Risk Analysis and Verification and Validation testing were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" to assure substantial equivalence.
As a result, we identified the level of concern associated with new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.
Non-clinical Performance Testing
For the difference of software processing methods which is different from primary predicate device, we performed non clinical performance testing between new device and predicate devices, in accordance with the FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging(SSXI) Devices" issued on Aug. 6, 1999.
We conducted performance bench test and image quality evaluate substantial equivalence of image processing technique to each predicate device respectively.
We also performed phantom tests using several different anatomical body phantoms.
For Dual Energy Subtraction, additional test report was submitted to assure the distinction of lesion overlying bones to calcification or tubercle shade.
For metal artifact reduction function of tomosynthesis option, we conducted performance testing.
{7}------------------------------------------------
SECTION V: 510(k) Summary
The result indicates that FBP(filtered back-projection) mode of new device has a same level of metal artifact reduction capability as the mode with artifact reduction of its predicate device.
Considering all these aspects, the result of non-clinical performance data demonstrates substantial equivalence of new device to the predicate devices in aspect of image acquisition, image processing as well as image quality capabilities.
Clinical Performance Testing
The result of non-clinical bench testing demonstrates substantial equivalence to the predicate devices. Therefore, it is our understanding that clinical testing is not required in this case.
Radiation safety
New device and primary predicate device is substantially equivalent by their conformities of CFR and IEC60601-1-3.
13) Conclusion
The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that new device should perform as intended in the specified use. Based on our risk analysis and bench testing, the differences do not affect its clinical safety or effectiveness.
From the result of our risk analysis, software verification, and nonclinical testing discussed above, it is our conclusion that,
- -New device is substantially equivalent to the legally marketed predicate devices, K093427 FDR ACSELERATE SYSTEM, DR-ID 200 FLAT PANEL DETECTOR, K121499 FUJIFILM Tomosynthesis Option for FDR ACSELERATE STATINOARY X-RAY SYSTEM, K101686 IMAGE STITCHING OPTION and K122454 FUJIFILM DUAL ENERGY FDR SUBTRACTION (DES) SOFTWARE OPTION.
- -Additional Indication for Use of new device does not introduce any new safety and effectiveness concern.
- -Therefore, new device is as safe, as effective, and performs as well as or better than the predicate device(s).
N/A