K Number
K152269

Validate with FDA (Live)

Date Cleared
2015-09-09

(29 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Healgen Amphetamine Test is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Cocaine Test is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. The test is available in a Strip format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Methamphetamine Test is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Device Description

Healgen Amphetamine Test, Healgen Cocaine Test and Healgen Methamphetamine Test are immunochromatographic assays for Amphetamine, Cocaine and Methamphetamine. Each assay test is a lateral flow system for the qualitative detection of Amphetamine, Benzoylecgonine and Methamphetamine (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

The document describes several tests conducted to prove the performance of the Healgen Amphetamine Test, Healgen Cocaine Test, and Healgen Methamphetamine Test devices.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state formal "acceptance criteria" through a table or specific thresholds that define success. Instead, it presents performance characteristic studies (precision, cut-off verification, interference, specificity, and comparison to GC/MS) and a lay-user study.

Based on the presented data, the implicit acceptance criteria appear to be:

  • Precision: Consistent results across different lots, operators, and concentration levels, especially for samples near the cut-off.
  • Cut-off Verification: Accurate classification of samples at or near the specified cut-off concentrations for each drug.
  • Interference: No false positive or false negative results due to common interfering substances in urine.
  • Specificity: Low or no cross-reactivity with structurally similar but non-target compounds at expected physiological concentrations.
  • Method Comparison (with GC/MS): High concordance with GC/MS results, especially for clearly negative and clearly positive samples, with reasonable performance around the cut-off.
  • Lay-user usability: High percentage of correct results by lay users and clear instructions for use.

Here's a summary table of the reported device performance for key aspects:

Performance AspectAcceptance Criteria (Implicit, based on study)Reported Device Performance (Summary from tables)
PrecisionConsistent classification (positive/negative) at various concentrations, particularly 100% agreement at -100% to -25% cut-off (negative) and +25% to +100% cut-off (positive). Variability allowed at cut-off.Observed: For all three analytes (Amphetamine, Cocaine, Methamphetamine) across all formats (Strip, Cassette, Dip Card, Cup) and 3 lots: - 100% Negative results from -100% to -25% cut-off concentrations (e.g., 50-/0+ means 50 negative, 0 positive). - 100% Positive results from +25% to +100% cut-off concentrations (e.g., 50+/0- means 50 positive, 0 negative). - At Cut-off Concentration: Mixed results (e.g., Amphetamine Strip: Lot 1: 23-/27+, Lot 2: 26-/24+, Lot 3: 26-/24+). This variability at the cut-off is expected for qualitative tests.
Cut-offAccurate detection at specified cut-off concentrations.Observed: For Amphetamine (500 ng/mL), Cocaine (150 ng/mL), and Methamphetamine (500 ng/mL): - All samples at and above +25% cut-off were positive. - All samples at and below -25% cut-off were negative. The specified cut-off values were verified.
InterferenceNo interference (false positives or negatives) by common physiological substances or over-the-counter medications at specified concentrations (100 ug/mL).Observed: Numerous listed compounds (e.g., Acetophenetidin, Acetylsalicylic Acid, Caffeine, Ibuprofen, etc.) showed no interference at 100 ug/mL for all three drug tests and formats when tested with drug-free and +25% cut-off spiked samples.
SpecificityAcceptable levels of cross-reactivity with structurally similar compounds.Observed: Each drug test showed varying degrees of cross-reactivity with related substances. For example: - Amphetamine: D-Amphetamine (100%), D,L-Amphetamine (67%), L-Amphetamine (3%), Phentermine (77%). Very low cross-reactivity with d-Methamphetamine (<0.5%) - Cocaine: Benzoylecgonine (100%), Cocaethylene (6%), Cocaine (30%). - Methamphetamine: d-Methamphetamine (100%), MDA (50%), l-Methamphetamine (13%). Low cross-reactivity with d,l-Amphetamine (0.7%) and d-Amphetamine (1%).
Method Comparison (GC/MS)High agreement (low discordance) with GC/MS results for samples well above and well below the cut-off, with some expected discordance near the cut-off.Observed: For each drug and format (Strip, Cassette, Cup, Dip Card) from Viewer A, B, and C: - 100% agreement for negative samples (GC/MS Negative and Low Negative) and high positive samples (GC/MS High Positive). - Discordant results primarily occurred for samples with GC/MS concentrations near the cut-off (Between cut-off and +50% for false negatives; Between -50% and cut-off for false positives). For instance, for Amphetamine Strip, Viewer A had 1 false negative (GC/MS 531 ng/mL) out of 16 'Near Cutoff Positive' samples, indicating generally good but not perfect agreement at the critical threshold. These discordant results are explicitly listed.
Lay-user StudyHigh percentage of correct results for various concentrations by lay users, and clear, understandable instructions.Observed: For all analytes and formats, lay users achieved: - 100% correct results for concentrations -100% to -50% cut-off and +50% to +75% cut-off. - 90-95% correct results for concentrations at -25% and +25% cut-off. - All lay users indicated the device instructions were easily followed. Flesch-Kincaid Grade Level was 7, indicating good readability.

2. Sample size used for the test set and the data provenance

  • Precision Study (Test Set):
    • Sample Size: For each drug (Amphetamine, Cocaine, Methamphetamine) and each device format (Strip, Cassette, Cup, Dip Card), there were 8 concentration levels tested. For each concentration, tests were performed 2 runs per day for 25 days by 3 different operators, for 3 different lots of devices. This means for one drug and one format: 8 concentrations * 50 results (2 runs/day * 25 days) * 3 lots = 1200 individual tests.
    • Data Provenance: Samples were "negative samples" spiked with drugs to specific concentrations. The drug concentrations were confirmed by GC/MS. The samples were "blind labeled and randomized" by a person not involved in testing. This indicates retrospective preparation of controlled samples, likely performed in a lab setting, not from actual patient samples from a specific country.
  • Cut-off Verification Study (Test Set):
    • Sample Size: A total of 150 samples (equally distributed at -50%, -25%, cut-off, +25%, +50% cut-off) were tested. Given 5 concentration levels, this suggests 30 samples per concentration. These were tested using 3 different lots of each device by 3 different operators. So, 150 samples * 3 lots * 3 operators = 1350 individual tests per drug and format.
    • Data Provenance: Samples were prepared by spiking drug in "negative samples." Concentrations confirmed by GC/MS. This indicates retrospective preparation of controlled samples.
  • Interference Study (Test Set):
    • Sample Size: Not explicitly stated as a total number of samples, but "Potential interfering substances" were tested by adding them to drug-free urine and target drug urine (+25% cut-off samples). These were tested using three batches of each device for all formats.
    • Data Provenance: Controlled laboratory-prepared urine samples with specified interfering substances and drug concentrations.
  • Specificity Study (Test Set):
    • Sample Size: Not explicitly stated as a total number of samples, but drug metabolites and other components were tested using three batches of each device for all formats.
    • Data Provenance: Controlled laboratory-prepared urine samples with specific drug metabolites or related compounds at various concentrations.
  • Method Comparison Study (Test Set):
    • Sample Size: For each drug (Amphetamine, Cocaine, Methamphetamine) and each device format, 80 unaltered clinical samples were used (40 negative and 40 positive). These were tested independently by 3 different laboratory assistants (viewers). For one drug and one format: 80 samples * 3 viewers = 240 individual tests.
    • Data Provenance: "unaltered clinical samples." The origin (e.g., country) is not specified. It is reasonable to assume these were retrospective clinical samples as they were "blind labeled."
  • Lay-user Study (Test Set):
    • Sample Size: 560 lay persons tested each of the Amphetamine, Cocaine, and Methamphetamine devices, resulting in a total of 1680 individuals who performed the study. Each participant received 1 blind labeled sample and a device.
    • Data Provenance: Urine samples were prepared by spiking drugs into "drug free-pooled urine specimens" to specific concentrations. Concentrations were confirmed by GC/MS. These were retrospectively prepared controlled samples, not actual patient samples. The study participants were "diverse educational and professional backgrounds and ranged in age from 21 to > 50 years."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Precision, Cut-off, Interference, Specificity, and Lay-user Studies: The ground truth for these studies was established by GC/MS (Gas Chromatography/Mass Spectrometry), which is a gold standard analytical method for drug concentration determination. The individuals performing the GC/MS analysis are typically trained analytical chemists or laboratory technicians, but their specific qualifications are not detailed in the document.
  • Method Comparison Study: The ground truth for the 80 unaltered clinical samples was established by GC/MS. As above, specific qualifications of the GC/MS operators are not provided. The device results were then compared to these GC/MS ground truth results.

4. Adjudication method for the test set

  • Precision, Cut-off, Interference, Specificity, and Lay-user Studies: No explicit adjudication method is mentioned for the ground truth. The ground truth was set by scientific preparation and GC/MS confirmation of samples. For the device results, multiple operators conducted the tests (3 operators for precision, cut-off; 3 batches for interference/specificity), but it's not stated that their results were adjudicated against each other; they are presented individually or as aggregate performance.
  • Method Comparison Study: The document lists results for Viewer A, Viewer B, and Viewer C separately. There is no explicit mention of an adjudication method for comparing interpretations between these viewers or for resolving discordant results by the device itself. The device results from each viewer are directly compared to the GC/MS ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was NOT done. This document describes the performance of in-vitro diagnostic test kits (dipstick, cassette, cup formats) for qualitative drug detection by visual interpretation. These are standalone devices, not AI-assisted reading systems.
  • Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, in essence, the "Analytical Performance" section (Precision, Cut-off, Interference, Specificity) and elements of the "Method Comparison Studies" (comparison to GC/MS ground truth) represent standalone performance of the device itself (the immunochromatographic assay reaction and visual line formation), without explicit human 'interpretation' as the primary variable. The visual reading done by operators/viewers in the precision and method comparison studies still involves human interpretation of the color lines.
  • However, the Lay-user study specifically evaluated human-in-the-loop performance (lay users interpreting the device according to instructions). Other sections primarily focus on the inherent analytical performance of the test strips/cassettes/cups.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The primary ground truth used throughout these studies is GC/MS (Gas Chromatography/Mass Spectrometry), which is an objective chemical analytical method considered the gold standard for confirming drug concentrations in toxicology.
  • For the lay user study, the ground truth was also established by GC/MS for the prepared spiked samples.

8. The sample size for the training set

  • The document describes premarket notification studies for an in-vitro diagnostic device. It does not mention a "training set" in the context of machine learning or AI models.
  • The device is a laboratory assay; it does not involve algorithms that require training data. The "training" in this context would refer to the development and optimization of the chemical reagents and manufacturing process, which are not detailed as a "training set" in a data science sense.

9. How the ground truth for the training set was established

  • N/A, as there is no "training set" in the context of data for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 9, 2015

HEALGEN SCIENTIFIC LLC C/O JOE SHIA LSI INTERNATIONAL 504 EAST DIAMOND AVE. SUITE F GAITHERSBURG MD 20877

Re: K152269

Trade/Device Name: Healgen Amphetamine Test (Strip, Cassette, Cup, Dip Card), Healgen Cocaine Test (Strip, Cassette, Cup, Dip Card), Healgen Methamphetamine Test (Strip, Cassette, Cup, Dip Card) Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: II Product Code: DKZ, DIO, LAF Dated: July 26, 2015

Received: August 11, 2015

Dear Mr. Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152269

Device Name

Healgen Amphetamine Test (Strip, Cassette, Cup, Dip Card) Healgen Cocaine Test (Strip, Cassette, Cup, Dip Card) Healgen Methamphetamine Test (Strip, Cassette, Cup, Dip Card)

Indications for Use (Describe)

Healgen Amphetamine Test is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Cocaine Test is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. The test is available in a Strip format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Methamphetamine Test is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

    1. Date: September 1, 2015
    1. Submitter: HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401
    1. Contact person: Jiangiu Fang HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401 Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com
    1. Device Name: Healgen Amphetamine Test (Strip, Cassette, Cup, Dip Card) Healgen Cocaine Test (Strip, Cassette, Cup, Dip Card) Healgen Methamphetamine Test (Strip, Cassette, Cup, Dip Card)

Classification: Class II

Product CodeCFR #Panel
DKZ21 CFR, 862.3100 Amphetamine Test SystemToxicology
DIO21 CFR, 862.3250 Cocaine Test SystemToxicology
LAF21 CFR, 862.3610 Methamphetamine Test SystemToxicology

Advin Multi-Drug Screen Test.

    1. Intended Use / Indications for Use
      Healgen Amphetamine Test is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

{4}------------------------------------------------

Healgen Cocaine Test is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Methamphetamine Test is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

    1. Device Description
      Healgen Amphetamine Test, Healgen Cocaine Test and Healgen Methamphetamine Test are immunochromatographic assays for Amphetamine, Cocaine and Methamphetamine. Each assay test is a flow system for the qualitative detection of Amphetamine, Benzoylecgonine lateral and Methamphetamine (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.
    1. Substantial Equivalence Information
      A summary comparison of features of the Healgen Amphetamine Test, Healgen Cocaine Test and Healgen Methamphetamine Test and the predicate device is provided in Table 1, Table 2 and Table 3.
ItemDevicePredicate -K122809
Intended UseFor the qualitative determination of drugsof abuse in human urine.Same
Drug AnalyteAmphetamineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onSame

Table 1: Features Comparison of Healgen Amphetamine Test and the Predicate Device

{5}------------------------------------------------

ItemDevicePredicate –K122809
the principle of antigen antibodyimmunochemistry.
Specimen TypeHuman UrineSame
Cut-Off Values500 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsStrip, Cassette, Cup, Dip CardCassette, Cup,DipCard

Table 2: Features Comparison of Healgen Cocaine Test and the Predicate Device

ItemDevicePredicate -K122809
Intended UseFor the qualitative determination ofdrugs of abuse in human urine.Same
Drug AnalyteBenzoylecgonineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values150 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsStrip, Cassette, Cup, Dip CardCassette, Cup,DipCard

Table 3: Features Comparison of Healgen Methamphetamine Test and the Predicate Device

ItemDevicePredicate -K122809
Intended UseFor the qualitative determination of drugsof abuse in human urine.Same
Drug AnalyteMETHAMPHETAMINESame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodySame

{6}------------------------------------------------

ItemDevicePredicate –K122809
immunochemistry.
Specimen TypeHuman UrineSame
Cut-Off Values500 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsStrip, Cassette, Cup, Dip CardCassette, Cup,DipCard

9. Test Principle

Healgen Amphetamine Test, Healgen Cocaine Test and Healgen Methamphetamine Test are rapid tests for the qualitative detection of Amphetamine, Benzoylecgonine and Methamphetamine in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cut-off concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the test has been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +75% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blind labeled and randomized by the person who prepared samples and did not take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days by three different operators for each format of devices. Different set of operators tested each format. The results obtained are summarized in the following tables:

Amphe tamine

Strip Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
--------------------------------------------------------------------------------------------------------------------------------------------------------

{7}------------------------------------------------

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot 150-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Cassette Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+33-/17+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+31-/19+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-

CUP Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+50-/0+32-/18+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+50-/0+32-/18+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+50-/0+30-/20+50+/0-50+/0-50+/0-50+/0-

Cocaine

Strip Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot 150-/0+50-/0+50-/0+50-/0+17-/33+50+/0-50+/0-50+/0-50+/0-
DrugLot 250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
DrugLot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Cassette Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+32-/18+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

{8}------------------------------------------------

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 350-/0+50-/0+50-/0+50-/0+32-/18+50+/0-50+/0-50+/0-

Dip Card Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+29-/21+50+/0-50+/0-50+/0-50+/0-

CUP Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+31-/19+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+30-/20+50+/0-50+/0-50+/0-50+/0-

METHAMPHETAMINE

Strip Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+34-/16+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+33-/17+50+/0-50+/0-50+/0-50+/0-

Cassette Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+19-/31+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+21-/29+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

{9}------------------------------------------------

CUP Format

ResultCut-offCut-off
Drug-100%-75%-50%-25%+25%+50%+75%+100%
Lot 150-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-

b. Linearity

Not applicable.

c. Stability

The devices are stable at 4-30°C for 24 months based on the accelerated stability study at 45°C and real time stability determination at both 4 ℃ and 30℃.

Control materials are not provided with the device. The labeling provides information on how to obtain control materials.

d. Cut-off

A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% cut-off and all negative at and below -25% cut-off for Amphetamine, Cocaine and Methamphetamine. The following cut-off values for the test devices have been verified.

TestCalibratorCut-off(ng/mL)
Amphetamine TestAmphetamine500
Cocaine TestBenzoylecgonine150
Methamphetamine TestMethamphetamine500
  • e. Interference
    Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.

Compounds that showed no interference at a concentration of 100ug/mL are summarized in the following tables. There were no differences observed for different formats.

Amphe tamine

AcetophenetidinR(-)-EpinephrinePentobarbital
--------------------------------------------------

{10}------------------------------------------------

N-AcetylprocainamideErythromycinPerphenazine
Acetylsalicylic Acid(Aspirin)Estrone-3-sulfatePhencyclidine
AlphenalEthyl-p-aminobenzoatePhenelzine
AmobarbitalFenoprofenPhenobarbital
AminopyrineFurosemidePhenylephrine-L
AmitriptylineGentisic acidPhenylethylamine
AmoxicillinHemoglobinPhenylpropanolamine
Ampicinine(Ampicillin)HydralazinePrednisolone Acetate
Aprobarbital(+/-)-4-HydroxyamphetamineHCLPrednisone
ApomorphineHydrochlorothiazideProcaine (Novocaine)
L-Ascorbic AcidHydrocodonePromazine
AspartameHydrocortisonePromethazine
Atropinea -Hydroxyhippuric acidPropoxyphene,d-
Butabarbitalp-HydroxymethamphetaminePropranolol
ButalbitalIbuprofenPseudoephedrine HCL
ButethalImipramineQuinidine
Benzilic acidIsoxsuprineQuinine
BenzphetamineIsoproterenol-(+/-)Ranitidine(Zantac)
Bezoic AcidKetamineSalicylic Acid
BilirubinLabetalolSecobarbital
CaffeineLevorphanolSerotonin
ChloramphenicolLoperamideSulfamethazine
Chlordiazepoxide HCLMaprotilineSulindac
ChloroquineMeprobamateTemazepam
ChlorothiazideMethadone11-Nor-△9-Tetrahydrocannabinol
ChlorpheniramineMethoxyphenamineTetracycline
ChlorpromazineMethylphenidateTetrahydrozoline
CholesterolNalbuphineThiamine
ClomipramineNalidixic acidL-Thyroxine
Clonidine hydrochlorideNaloxone hydrochlorideThioridazineHydrochloride
CortisoneNaltrexone hydrochlorideTriamterene
Cotinine(-)NaproxenTriflupromazineHydrochloride
CreatinineNiacinamideTrimethoprim
Cyclopentobarbita1NifedipineTrimipramine
DeoxyepinephrineNorethindroneTryptamine
DextromethorphanNorpropoxypheneDL-Tryptophan
DiazepamNoscapineTyramine
DiflunisalOxazepamD/L-Tyrosine
DigoxinOxymetazolineUric Acid
DoxylaminePapaverineVerapamil
Ecgonine methylesterPenicillinZomepirac

{11}------------------------------------------------

Cocaine

AcetophenetidinDoxylaminePhentermine
N-AcetylprocainamideR(-)-EpinephrinePentobarbital
Acetylsalicylic Acid(Aspirin)ErythromycinPerphenazine
AlphenalEstrone-3-sulfatePhencyclidine
AmobarbitalEthyl-p-aminobenzoatePhenelzine
AminopyrineFenoprofenPhenobarbital
AmitriptylineFurosemidePhenylephrine-L
AmoxicillinGentisic acidPhenylethylamine
Amphetamine SulfateHemoglobinPhenylpropanolamine
Ampicinine(Ampicillin)HydralazinePrednisolone Acetate
Aprobarbital(+/-)-4-HydroxyamphetamineHCLPrednisone
ApomorphineHydrochlorothiazideProcaine (Novocaine)
d-amphetamineHydrocodonePromazine
D,l-amphetamineHydrocortisonePromethazine
l-amphetaminea -Hydroxyhippuric acidPropoxyphene,d-
L-Ascorbic Acidp-HydroxymethamphetaminePropranolol
AspartameIbuprofenPseudoephedrine HCL
AtropineImipramineQuinidine
ButabarbitalIsoxsuprineQuinine
ButalbitalIsoproterenol-(+/-)Ranitidine (Zantac)
ButethalKetamineSalicylic Acid
Benzilic acidLabetalolSecobarbital
BenzphetamineLevorphanolSerotonin
Bezoic AcidLoperamideSulfamethazine
BilirubinMaprotilineSulindac
CaffeineMeprobamateTemazepam
ChloramphenicolMethadone11-Nor-△9-Tetrahydrocannabinol
Chlordiazepoxide HCLMethoxyphenamineTetracycline
ChloroquineMethylphenidateTetrahydrozoline

{12}------------------------------------------------

Chlorothiazide(+/-)-Methylenedioxyamphetamine (MDA)Thiamine
ChlorpheniramineNalbuphineL-Thyroxine
ChlorpromazineNalidixic acidThioridazineHydrochloride
CholesterolNaloxone hydrochlorideTriamterene
ClomipramineNaltrexone hydrochlorideTriflupromazineHydrochloride
Clonidine hydrochlorideNaproxenTrimethoprim
CortisoneNiacinamideTrimipramine
Cotinine(-)NifedipineTryptamine
CreatinineNorethindroneDL-Tryptophan
CyclopentobarbitalNorpropoxypheneTyramine
DeoxyepinephrineNoscapineD/L-Tyrosine
DextromethorphanOxazepamUric Acid
DiazepamOxymetazolineVerapamil
DiflunisalPapaverineZomepirac
DigoxinPenicillin

METHAMPHETAMINE

AcetophenetidinR(-)-EpinephrinePentobarbital
N-AcetylprocainamideErythromycinPerphenazine
Acetylsalicylic Acid(Aspirin)Estrone-3-sulfatePhencyclidine
AlphenalEthyl-p-aminobenzoatePhenelzine
AmobarbitalFenoprofenPhenobarbital
AminopyrineFurosemidePhenylpropanolamine
AmitriptylineGentisic acidPrednisolone Acetate
AmoxicillinHemoglobinPrednisone
Ampicinine(Ampicillin)HydralazineProcaine(Novocaine)
Aprobarbital(+/-)-4-HydroxyamphetamineHCLPromazine
ApomorphineHydrochlorothiazidePromethazine
L-Ascorbic AcidHydrocodonePropoxyphene,d-
AspartameHydrocortisonePropranolol
Atropinea -Hydroxyhippuric acidPseudoephedrine HCL
ButabarbitalIbuprofenQuinidine
ButalbitalImipramineQuinine
ButethalIsoxsuprineRanitidine(Zantac)
Benzilic acidIsoproterenol-(+/-)Salicylic Acid
BenzphetamineKetamineSecobarbital

{13}------------------------------------------------

Bezoic AcidLabetalolSerotonin
BilirubinLevorphanolSulfamethazine
CaffeineLoperamideSulindac
ChloramphenicolMaprotilineTemazepam
Chlordiazepoxide HCLMeprobamate11-Nor-Δ9-Tetrahydrocannabinol
ChlorothiazideMethadoneTetracycline
ChlorpheniramineMethoxyphenamineTetrahydrozoline
ChlorpromazineMethylphenidateThiamine
CholesterolNalbuphineL-Thyroxine
ClomipramineNalidixic acidThioridazine Hydrochloride
Clonidine hydrochlorideNaloxone hydrochlorideTriamterene
CortisoneNaltrexone hydrochlorideTriflupromazine Hydrochloride
Cotinine(-)NaproxenTrimethoprim
CreatinineNiacinamideTrimipramine
CyclopentobarbitalNifedipineTryptamine
DeoxyepinephrineNorethindroneDL-Tryptophan
DextromethorphanNorpropoxypheneTyramine
DiazepamNoscapineD/L-Tyrosine
DiflunisalOxazepamUric Acid
DigoxinOxymetazolineVerapamil
DoxylaminePapaverineZomepirac
Ecgonine methylesterPenicillin

f. Specificity

To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.

Amphetamine(Cut-off=500 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
D-Amphetamine500100%
D,L-Amphetamine75067%
L-Amphetamine160003%
Phentermine65077%

{14}------------------------------------------------

(+/-)-Methylenedioxyamphetamine(MDA)80063%
d-Methamphetamine>100,000<0.5%
1-Methamphetamine>100,000<0.5%
ephedrine>100,000<0.5%
3,4-Methylenedioxyethylamphetamine(MDE)>100,000<0.5%
3,4-methylenedioxy-methamphetamine(MDMA)>100,000<0.5%
Cocaine(Cut-off=150 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
Benzoylecogonine150100%
Cocaethylene25006%
Cocaine50030%
Ecgonine12,5001.2%
Ecgonine methylester50,0000.3%
Methamphetamine(Cut-off=500 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
d-Methamphetamine500100%
Methylenedioxyamphetamine(MDA)100050%
p-Hydroxymethamphetamine150003%
l-Methamphetamine400013%
Mephentermine250002%
d,l-Amphetamine750000.7%
(1R,2S)-(-)-Ephedrine500001%
β-Phenylethylamine750000.7%
Methylenedioxymethamphetamine(MDMA)100050%
Methylenedioxy-n-ethylamphetamine(MDEA)200002.5%
d-Amphetamine500001%
Chloroquine125004%
l-Phenylephrine1000000.5%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25%

{15}------------------------------------------------

below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

    1. Comparison Studies
      The method comparison studies for the Amphetamine Test, the Cocaine Test and the Methamphetamine Test were performed in-house with three different laboratory assistants for each format of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:
Amphetamine
StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0001524
Negative10151510
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001424
Negative10151520

Discordant Results of Amphetamine Strip

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer AAMP68531Negative
Viewer BAMP68531Negative
Viewer CAMP68531Negative
Viewer CAMP63504Negative
CassetteformatNegativeLowNegativeby GC/MS(less than-50%)NearCutoffNegativeby GC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424

{16}------------------------------------------------

Negative10151520
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001524
Negative10151510

Discordant Results of Amphetamine Cassette

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer AAMP68531Negative
Viewer AAMP63504Negative
Viewer BAMP68531Negative
Viewer CAMP68531Negative
Cup formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001524
Negative10151510

Discordant Results of Amphetamine Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer AAMP68531Negative
Viewer AAMP63504Negative
Viewer BAMP68531Negative
Viewer BAMP63504Negative
Viewer CAMP68531Negative
Dip CardformatLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Negative

{17}------------------------------------------------

Viewer APositive0001424
Negative10151520
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001424
Negative10151520

Discordant Results of Amphetamine Dip Card

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer AAMP68531Negative
Viewer AAMP63504Negative
Viewer BAMP68531Negative
Viewer CAMP68531Negative
Viewer CAMP63504Negative
StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001524
Negative10151510

Discordant Results of Cocaine Strip

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer ACOC68171Negative
Viewer ACOC69164Negative
Viewer BCOC68171Negative
Viewer BCOC71187Negative
Viewer CCOC69164Negative

{18}------------------------------------------------

CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001524
Negative10151510
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001424
Negative10151520

Discordant Results of Cocaine Cassette

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer ACOC69164Negative
Viewer BCOC68171Negative
Viewer CCOC68171Negative
Viewer CCOC69164Negative
Dip Card formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520

Discordant Results of Cocaine Dip Card

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer ACOC68171Negative
Viewer ACOC69164Negative
Viewer BCOC68171Negative
Viewer BCOC03182Negative

{19}------------------------------------------------

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer CCOC68171Negative
Viewer CCOC69164Negative
Cup formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001524
Negative10151510
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001524
Negative10151510

Discordant Results of Cocaine Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer ACOC68171Negative
Viewer BCOC68171Negative
Viewer CCOC69164Negative

Me thamphe tamine

StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001524
Negative10151510
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001424
Negative10151520

Discordant Results of Methamphetamine Strip

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
---------------------------------------------------------------------

{20}------------------------------------------------

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer AMET64501Negative
Viewer BMET64501Negative
Viewer CMET64501Negative
Viewer CMET62504Negative
CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001524
Negative10151510
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001424
Negative10151520

Discordant Results of Methamphetamine Cassette

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer AMET64501Negative
Viewer BMET64501Negative
Viewer CMET64501Negative
Viewer CMET62504Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001524
Negative10151510

{21}------------------------------------------------

Discordant Results of Methamphetamine Dip Card
ViewerSample NumberGC/MS ResultDip Card Format Viewer Results
Viewer AMET64501Negative
Viewer AMET70517Negative
Viewer BMET64501Negative
Viewer BMET62504Negative
Viewer CMET64501Negative

Discordant Results of Methamphetamine Din Card

CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001524
Negative10151510

Discordant Results of Methamphetamine Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer AMET25509Negative
Viewer AMET64501Negative
Viewer BMET64501Negative
Viewer CMET64501Negative

Lay-user study

A lay user study was performed at three intended user sites with 560 lay persons testing each of the Amphetamine, the Cocaine and the Methamphetamine devices. Total of 1680 individuals performed the study. A total of 214 females and 346 males tested the Amples, 196 females and 364 males tested the Cocaine samples and 205 females and 355 males tested the Methamphetamine samples. They had diverse educational and professional backgrounds and ranged in age from 21 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

{22}------------------------------------------------

NumberofsamplesAMP Concentration byGC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037511995%
+25% Cutoff2062519195%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Amphetamine Strip)

Comparison between GC/MS and Lay Person Results (Amphetamine Cassette)

% of CutoffNumberofsamplesAMP Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037521890%
+25% Cutoff2062518290%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Amphetamine DipCard)

% of CutoffNumberofsamplesAMP Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037511995%
+25% Cutoff2062519195%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Amphetamine Cup)

% of CutoffNumberAMP Concentration byLay person resultsThe
--------------------------------------------------------------------

{23}------------------------------------------------

ofsamplesGC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037511995%
+25% Cutoff2062519195%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Cocaine Strip)

% of CutoffNumberofsamplesCOC Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff2037.5020100%
-50% Cutoff2075020100%
-25% Cutoff20112.511995%
+25% Cutoff20187.519195%
+50% Cutoff20225200100%
+75% Cutoff20262.5200100%

Comparison between GC/MS and Lay Person Results (Cocaine Cassette)

% of CutoffNumberofsamplesCOC Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100%Cutoff200020100%
-75% Cutoff2037.5020100%
-50% Cutoff2075020100%
-25% Cutoff20112.521890%
+25% Cutoff20187.519195%
+50% Cutoff20225200100%
+75% Cutoff20262.5200100%

Comparison between GC/MS and Lay Person Results (Cocaine DipCard)

% of CutoffNumberCOC Concentration byLay person resultsThe
--------------------------------------------------------------------

{24}------------------------------------------------

ofsamplesGC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff2037.5020100%
-50% Cutoff2075020100%
-25% Cutoff20112.521890%
+25% Cutoff20187.518290%
+50% Cutoff20225200100%
+75% Cutoff20262.5200100%

Comparison between GC/MS and Lay Person Results (Cocaine Cup)

NumberofsamplesCOC Concentration byGC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff2037.5020100%
-50% Cutoff2075020100%
-25% Cutoff20112.511995%
+25% Cutoff20187.519195%
+50% Cutoff20225200100%
+75% Cutoff20262.5200100%

Comparison between GC/MS and Lay Person Results (Methamphetamine Strip)

NumberofsamplesMET Concentration byGC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037521890%
+25% Cutoff2062518290%
+50% Cutoff20750200100%
+75% Cutoff20875200100%
% of CutoffNumberMET Concentration byLay person resultsThe
--------------------------------------------------------------------

{25}------------------------------------------------

ofsamplesGC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037511995%
+25% Cutoff2062519195%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Methamphetamine DipCard)

% of CutoffNumber of samplesMET Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100%Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037521890%
+25% Cutoff2062518290%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Methamphetamine Cup)

NumberofsamplesMET Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
% of CutoffNo. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff2037521890%
+25% Cutoff2062518290%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

{26}------------------------------------------------

3. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the Healgen Amphetamine Test, Healgen Cocaine Test and Healgen Methamphetamine Test are substantially equivalent to the predicate.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).