(87 days)
The implants are intended to support normal bone healing for osteotomies, fractures, and reconstructions.
The RECON system consists of various system components and it is indicated for the treatment of fracture fixation, joint decompression and fusion, reconstruction or arthrodeses of small bones. The system can be used in both adult and pediatric patients. Additional information is provided in the corresponding surgical techniques.
Gruber Lapidus
The Gruber Lapidus Plates 2.7 combined with the Standard/Locking/ MiniCAN screws 2.7 are indicated for TMT I joint arthrodesis with cuneiforme transfixation for correction of hallux valgus deformity.
Plantar Lapidus
The Plantar Lapidus Plates 2.7 combined with the Standard/Locking Screws 2.7 and the MiniCAN screws 3.5 are indicated for TMT I joint arthrodesis.
Tarsalis
The Tarsalis Plates 2.7 combined with the Standard/Locking Screws 2.7 are indicated for Lisfranc fusions and fracture fixation of the forefoot and midfoot.
The RECON system – Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws is a plate and screw system intended for internal fixation. The system consists of the following plates and screws:
- Gruber Lapidus Plates 2.7, .
- . Plantar Lapidus Plates 2.7,
- Tarsalis Plates 2.7, ●
- Locking Screws 2.7 ●
- Standard Screws 2.7
- Minican Screws 2.7/3.5
of different sizes and designs.
The plates and screws are either made of titanium alloy Ti-6AI-4V (ASTM F136) or commercially pure Titanium (C.P. Titanium; ASTM F67).
The implants are offered in various sizes to accommodate the variations of bone size and geometry. The implants are provided non-sterile and single-use only. The instruments are non-sterile and reusable or for single use.
This document describes a 510(k) premarket notification for the RECON system - Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws. As such, it focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a standalone study with defined performance metrics. Therefore, many of the requested elements (like sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, and detailed ground truth establishment) are not applicable to this type of regulatory submission.
However, based on the provided text, here's an attempt to address your request:
Acceptance Criteria and Study for the RECON system - Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a 510(k) submission and not a standalone performance study with predefined clinical acceptance criteria, the "acceptance criteria" are implied by the performance characteristics demonstrated to be "similar" or "substantially equivalent" to legally marketed predicate devices. The "reported device performance" is characterized by meeting these similarity benchmarks.
The non-clinical performance data aimed to demonstrate that the subject device's mechanical properties and biocompatibility are comparable to those of the predicate devices, thereby ensuring similar safety and effectiveness.
| Performance Metric | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance (Summary) |
|---|---|---|
| Biocompatibility | Compliance with ISO 10993-1 | All testing passed (per ISO 10993-1). |
| Cross-sectional engineering analysis (Plates) | Similar mechanical properties and performance to predicate devices (Modular Foot System – 2.7 mm Module (K010321), Zimmer Universal Locking System (ULS) Plates and Screws (K063303 and K060710), Zimmer Plates and Screws System (ZPS) (K143066)) in terms of proof load, bending strength, bending stiffness, equivalent bending stiffness, and fatigue strength. | Resulted in similar mechanical properties and performance. |
| Bending strength engineering analysis (MiniCAN Screws 2.7mm & 3.5mm) | Similar mechanical properties and performance to predicate device (ZimmerBiomet Herbert Mini Bone screw (K143165)). | Resulted in similar mechanical properties and performance. |
| Screw Performance (Locking, Standard, MiniCAN Screws) | Similar mechanical properties and performance to predicate devices (Zimmer ULS Plates and Screws (K063303 and K060710), Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, Herbert Mini Bone Screw (K143165)) in terms of self-tapping performance, driving and removal torque, to failure, and axial pullout. | Resulted in similar mechanical properties and performance. |
2. Sample Size Used for the Test Set and Data Provenance
This is a non-clinical (bench) testing study, not a clinical trial. Therefore, the concept of a "test set" for clinical performance, with associated data provenance, is not applicable. The non-clinical tests involved physical samples of the devices. The document does not specify the exact number of physical samples used for each mechanical test. Data provenance is internal to Normed Medizin-Technik GmbH in Germany, as described in the "Submission Sponsor and Correspondent" section. The data is prospective, generated specifically for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. Ground truth for non-clinical engineering tests is established by standardized testing methods and measurement instruments, not by expert consensus.
4. Adjudication Method for the Test Set
Not applicable, as it's a non-clinical bench testing study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a MRMC comparative effectiveness study was not done. This device is not an AI/software device that would typically undergo such a study.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This is a physical medical device (plates and screws), not an algorithm or AI.
7. Type of Ground Truth Used
For non-clinical performance, the "ground truth" is defined by the physical properties of the materials and designs, measured through standardized mechanical tests. For biocompatibility, it's adherence to international standards (ISO 10993-1).
8. Sample Size for the Training Set
Not applicable. This is a physical medical device and does not involve AI or machine learning models that require a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, resembling a bird in flight.
November 5, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
NORMED Medizin-Tecknik GmbH Mr. Arne Briest Regulatory Affairs Manager Ulrichstrasse 7 D-78532 Tuttlingen GERMANY
Re: K152256 Trade/Device Name: RECON system- Grubber Lapidus-, Platar Lapidus-, Tarsalis Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: August 7, 2015 Received: August 10, 2015
Dear Mr. Briest:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Arne Briest
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152256
Device Name
RECON system - Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws
Indications for Use (Describe)
The implants are intended to support normal bone healing for osteotomies, fractures, and reconstructions.
The RECON system consists of various system components and it is indicated for the treatment of fracture fixation, joint decompression and fusion, reconstruction or arthrodeses of small bones. The system can be used in both adult and pediatric patients. Additional information is provided in the corresponding surgical techniques.
Gruber Lapidus
The Gruber Lapidus Plates 2.7 combined with the Standard/Locking/ MiniCAN screws 2.7 are indicated for TMT I joint arthrodesis with cuneiforme transfixation for correction of hallux valgus deformity.
Plantar Lapidus
The Plantar Lapidus Plates 2.7 combined with the Standard/Locking Screws 2.7 and the MiniCAN screws 3.5 are indicated for TMT I joint arthrodesis.
Tarsalis
The Tarsalis Plates 2.7 combined with the Standard/Locking Screws 2.7 are indicated for Lisfranc fusions and fracture fixation of the forefoot and midfoot.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
ال Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a logo for NORMED, a Zimmer company. The logo features the text "The specialist for small bones" above the word "NORMED" in a stylized font. Below the word "NORMED" is the text "A ZIMMER COMPANY" with a Zimmer logo.
510(k) Summary
1. Submission Sponsor and Correspondent
| Normed Medizin-Technik GmbH |
|---|
| Ulrichstrasse 7 |
| D-78532 Tuttlingen |
| Germany |
| Phone: | + 49 7461 93 43 0 |
|---|---|
| Fax: | + 49 7461 93 43 20 |
| Contact: | Mr. Arne Briest |
FDA Establishment Registration #:
Date Prepared 2.
Date Prepared:
November 2, 2015
9611091
Device Identification 3.
| Trade/Proprietary Name: | RECON system - Gruber Lapidus-, Plantar Lapidus-,Tarsalis Plates and Screws |
|---|---|
| Common/Usual Name: | bone plate and screw |
| Classification Name: | Plate, Fixation, Bone |
| Screw, Fixation, Bone | |
| Classification Regulation | 21CFR 888.3030 |
| 21CFR 888.3040 | |
| Product Code: | HRS |
| HWC | |
| Device Class: | Class II |
| Classification Panel | Orthopedic |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image is a logo for NORMED, a Zimmer company. The logo features the text "The specialist for small bones" above the NORMED logo. The NORMED logo is in a bold, sans-serif font, with a globe in the center of the "O". Below the NORMED logo is the text "A ZIMMER COMPANY" with a blue Z logo.
4. Legally Marketed Predicate Device
K010321 - Modular Foot System - 2.7 mm Module, manufactured by Synthes Inc., cleared May 2, 2001
K063303 - Universal Locking Plate System 2.7, Plates and Screws, manufactured by Zimmer Inc., cleared November 22, 2006
K060710 - Universal Locking Plate System 3.5, Plates and Screws manufactured by Zimmer Inc., cleared April 26, 2006
K143066 - Zimmer Plates and Screws System (ZPS) - Non-sterile ZPS Plate Line Extensions, Sterile/Non-sterile ZPS Screws and Washers, manufactured by Zimmer Inc., cleared November 28, 2014
K143165 - Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw, manufactured by Zimmer Inc., cleared March 31, 2015
K151407 - Recon System, manufactured by Normed Medizin-Technik GmbH, cleared August 12, 2015
K152142 - Recon System-MPJ-Plates, manufactured by Normed Medizin-Technik GmbH, cleared September 16, 2015
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image is a logo for NORMED, "The specialist for small bones", which is "A ZIMMER COMPANY". The logo is black and white, with the word "NORMED" in large, bold letters. The "O" in "NORMED" is replaced by a target symbol with a world map in the center. The words "The specialist for small bones" are above the word "NORMED", and the words "A ZIMMER COMPANY" are below the word "NORMED".
5. Device Description
The RECON system – Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws is a plate and screw system intended for internal fixation. The system consists of the following plates and screws:
- Gruber Lapidus Plates 2.7, .
- . Plantar Lapidus Plates 2.7,
- Tarsalis Plates 2.7, ●
- Locking Screws 2.7 ●
- Standard Screws 2.7
- Minican Screws 2.7/3.5
of different sizes and designs.
The plates and screws are either made of titanium alloy Ti-6AI-4V (ASTM F136) or commercially pure Titanium (C.P. Titanium; ASTM F67).
The implants are offered in various sizes to accommodate the variations of bone size and geometry. The implants are provided non-sterile and single-use only. The instruments are non-sterile and reusable or for single use.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image is a logo for Normed, a Zimmer company. The logo is black and white and features the word "NORMED" in a bold, sans-serif font. Above the word "NORMED" is the tagline "The specialist for small bones". Below the word "NORMED" is the text "A ZIMMER COMPANY".
6. Indications for Use
The implants are intended to support normal bone healing for osteotomies, fractures, and reconstructions.
The RECON system - Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws consists of various system components and it is indicated for the treatment of fracture fixation, joint decompression and fusion, reconstruction or arthrodeses of small bones. The system can be used in both adult and pediatric patients. Additional information is provided in the corresponding surgical techniques.
Gruber Lapidus
The Gruber Lapidus Plates 2.7 combined with the Standard/Locking/ MiniCAN screws 2.7 are indicated for TMT I joint arthrodesis with cuneiforme transfixation for correction of hallux valgus deformity.
Plantar Lapidus
The Plantar Lapidus Plates 2.7 combined with the Standard/Locking Screws 2.7 and the MiniCAN screws 3.5 are indicated for TMT I joint arthrodesis.
Tarsalis
The Tarsalis Plates 2.7 combined with the Standard/Locking Screws 2.7 are indicated for Lisfranc fusions and fracture fixation of the forefoot and midfoot.
7. Substantial Equivalence Discussion
The RECON system - Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws have the same intended use, similar performance characteristics, is manufactured from similar materials and is similar in design to the predicate devices.
{8}------------------------------------------------
Traditional 510(k)
Image /page/8/Picture/2 description: The image is a logo for NORMED, a Zimmer company. The logo features the word "NORMED" in a bold, stylized font, with an image of the earth inside the letter "O". Above the word "NORMED" is the text "The specialist for small bones". Below the word "NORMED" is the text "A ZIMMER COMPANY" with a stylized Z to the left.
8. Non-Clinical Performance Data
- -Biocompatibility - Biocompatibility testing on the plates was conducted and evaluated per ISO 10993-1. All testing passed.
- -Cross sectional engineering analysis of the plates, proof load, bending strength, bending stiffness, equivalent bending stiffness and fatique strength of the RECON System-Tarsalis Plates, the Lapidus Gruber Plates and the Plantar Lapidus Plates and the predicate devices, the Modular Foot System – 2.7 mm Module (K010321), Zimmer Universal Locking System (ULS) Plates and Screws (K063303 and K060710), and Zimmer Plates and Screws System (ZPS) (K143066) resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.
- Bending strength engineering analysis for Minican Screws 2.7 mm and Minican Screws 3.5 mm (Partially-Treaded and Fully Threaded) and the predicate devices, the ZimmerBiomet Herbert Mini Bone screw (K143165) resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.
- Self-tapping performance, driving and removal torque, to failure and axial pullout of the -RECON System - Locking Screws 2.7mm, Standard Screws 2.7mm, MiniCAN Screws 2.7mm and MiniCAN Screws 3.5mm (Partial Thread) and the predicate devices, the Zimmer Universal Locking System (ULS) Plates and Screws (K063303 and K060710) and Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw (K143165), resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.
9. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device.
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image is a logo for NORMED, "The specialist for small bones", which is a Zimmer company. The logo is black and white, with the word "NORMED" in large, bold letters. The "O" in NORMED is a globe with a target over it. Below the word "NORMED" is the text "A ZIMMER COMPANY".
10. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the RECON system -Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and the predicate devices do not raise any questions regarding its safety and effectiveness.
Performance testing and compliance with voluntary standards, demonstrate that the RECON system – Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.
The RECON system - Gruber Lapidus-, Plantar Lapidus-, Tarsalis Plates and Screws are determined to be substantially equivalent to the referenced predicate devices.
N/A