K Number
K152251

Validate with FDA (Live)

Date Cleared
2016-06-10

(305 days)

Product Code
Regulation Number
868.5740
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Endobronchial Tube is used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

Device Description

The Endobronchial Tube is made of Polyvinylchloride and is available in sizes 26fr to 41fr. They are designated as double lumen tube with 2 cuffs and separate 15mm connectors for isolating and ventilating one lung during surgical procedures. The tubes contain an x-ray opaque line that runs through the tube making them detectable by x-ray. Environments of use: Hospital-OR and ICU

AI/ML Overview

The provided text describes the 510(k) submission for the "Endobronchial Tube" by Henan Tuoren Medical Device Co., Ltd. and its substantial equivalence to a predicate device (Well LEAD Endobronchial Tubes, K092886). The evaluation primarily relies on non-clinical testing (bench testing and biocompatibility assessments) to demonstrate that the proposed device meets established standards and performs comparably to the predicate.

Here's an analysis of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document provides a detailed table of performance testing for the Endobronchial Tube.

Type of Bench TestReferenced StandardAcceptable CriteriaReported Device Performance (Pass/fail)
Cuff diameterISO 5361: 2012The maximum cuff diameter shall be within ± 0.5mm of the nominal value when tested according to ISO 5361 Annex B.Pass
Cuffed tube collapse testISO 5361: 2012The steel ball shall pass freely through the tube when tested according to ISO 5361 Annex C.Pass
Cuff herniation testISO 5361: 2012No part of the inflated cuff shall reach beyond the nearest edge of the bevel when tested according to ISO 5361 Annex D.Pass
Seal testingISO 5361: 2012The leakage limit is ≤2.0ml/h at cuff pressures not to exceed 2.7 kPa (27 cmH2O) when tested according to ISO 5361 Annex G.Pass
Sealing of cuff inflating systemNot explicitly referencedNo air leakage happens under the condition of continuous 3kPa positive pressure imposition for 10 seconds.Pass
Sealing of connector assemblyNot explicitly referencedNo air leakage happens on any joints under the condition of continuous 6kPa gas pressure imposition to lumen of main tube (shaft) for 60 seconds.Pass
Inflating tube (outside diameter)ISO 5361: 2012The inflating tube shall have an outside diameter of not more than 3.0mm.Pass
Inflating tube (angle)ISO 5361: 2012The angle between the inflating tube and the Endobronchial tube at the point of separation shall not exceed 45°.Pass
Kink resistance testISO 5361: 2012The steel ball shall pass freely through the lumen of the tube when tested according to ISO 5361 Annex H.Pass
Gauging of One Way ValveISO 594/1 1986The plane of the maximum diameter at the opening of the female conical fitting of One Way Valve lie between the two limit planes of the gauge.Pass
Liquid leakage of One Way ValveISO 594/1 1986No leakage sufficient to form a falling drop of water.Pass
Air leakage of One Way ValveISO 594/1 1986Continued formation of air bubbles not be evident.Pass
Separation force of One Way ValveISO 594/1 1986The conical fitting under test remain attached to the test fixture.Pass
Stress cracking (One Way Valve)ISO 594/1 1986There shall be no evidence of stress cracking of the conical fitting.Pass
Security of construction of suction catheterISO 8836-2007The force required to detach any component permanently attached to the shaft shall be not less than that specified in standard.Pass
Shaft resistant to negative pressure (suction catheter)ISO 8836-2007A vacuum source at 40kPa below ambient pressure for 15s at a temperature of 23°C ±2°C with the patient end occluded, the shaft shall not collapse.Pass
15mm connectorISO5356-1: 2004Comply with the ISO5356-1: 2004.Pass
Burst Testing (Cuff burst)Not explicitly referencedThe volume of injected gas when cuff bursting happens is larger than 40ml during inflation of endobronchial tube.Pass
Burst Testing (Bursting between cuff and main tube)Not explicitly referencedThere should be no fracture on junctions while inflated 30kpa gas.Pass
Bond StrengthNot explicitly referencedThe joints of endobronchial tube should be firm bonding. When an axial force of 50±5N is applied at 50±5mm/min, the testing portion (between connector and...) [The sentence is cut off here, but "Pass" is indicated]Pass
Air flow resistanceNot explicitly referencedPressure increment should no more than 0.2Kpa/h when testing the endobronchial tube under the specified flow rate (3, 6, 9L/min).Pass
Radiopaque testNot explicitly referencedWhen exposing the Endobronchial tube with the low dose rays, the X-ray machine should have film development.Pass
Endotoxin testUSP36_NF31<85>0.25EU/ml, 20EU/Device.Pass
In vitro cytotoxicityISO10993-5:2009No potential toxicity to L-929 cells under specified test conditions (using MTT method MEM extract and MTT method MEM with 10% FBS extract).Pass
Skin sensitization Test (Guinea Pig)ISO10993-10:2010No significant evidence of causing skin sensitization in the guinea pig under specified test conditions (using 0.9% sodium chloride injection extract and sesame oil extract).Pass
Oral Mucosa Irritation testISO10993-10:2010No significant evidence of causing oral irritation in the hamster under specified test conditions (using 0.9% sodium chloride extract and sesame oil extract).Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample size for the mechanical performance tests, biocompatibility tests, or endotoxin tests. It mentions "non-clinical testing was performed" and provides "Test result" for each criterion. The data provenance is not explicitly stated in terms of country of origin for the testing, but the manufacturer is Henan Tuoren Medical Device Co., Ltd, China. The testing described is clearly prospective for the device being submitted, as it was performed to demonstrate compliance with standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This product is a medical device (Endobronchial Tube, Class II), not an AI or imaging diagnostic device. Therefore, the concept of "experts establishing ground truth for a test set" in the context of image interpretation or clinical diagnosis is not applicable here. The "ground truth" for this device's performance is established by adherence to engineering and biocompatibility standards and quantified through objective physical and chemical measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a study involving human interpretation or subjective adjudication. The tests involve objective measurements against predefined criteria in recognized standards (e.g., ISO).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI diagnostic device. No MRMC study was performed or required.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm or AI device. The performance is assessed based on the physical properties and biological compatibility of the device itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance assessment is primarily defined by:

  • Engineering Standards: Specific requirements and limits outlined in international standards like ISO 5361, ISO 594/1, ISO 8836, and ISO 5356-1.
  • Biocompatibility Standards: Criteria for no toxicity, sensitization, or irritation as per ISO 10993 series.
  • Safety Standards: Endotoxin limits as per USP.

These standards provide objective, measurable criteria which serve as the ground truth against which the device's performance is evaluated.

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model. There is no "training set" in the context of data for an algorithm.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth established for it. The standards mentioned in point 7 serve as the universally accepted benchmarks for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol composed of three overlapping profiles facing right. The profiles are rendered in black and create a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 10, 2016

Henan Tuoren Medical Device Co., Ltd C/O Long Yang COO Shenzhen Hlongmed Biotech Company R15-08, East Building, Yihai Plaza, Chuangye Road, Nanshan District, Shenzhen, Guangdong, 518054 China

Re: K152251 Trade/Device Name: Endobronchial Tube Models: Left: Fr26, Fr28, Fr31, Fr32, Fr33, Fr35, Fr37, Fr39, Fr41; Right: Fr26, Fr28, Fr31, Fr32, Fr33, Fr35, Fr37, Fr39, Fr41 Regulation Number: 21 CFR 868.5740 Regulation Name: Tracheal/Bronchial Differential Ventilation Tube Regulatory Class: Class II Product Code: CBI Dated: May 6, 2016 Received: May 12, 2016

Dear Long Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K152251

Device Name

Endobronchial Tube

Models: Left: Fr26, Fr31, Fr32, Fr33, Fr35, Fr37, Fr39, Fr41; Right: Fr26, Fr28, Fr33, Fr33, Fr33, Fr37, Fr39, Fr41

Indications for Use (Describe)

The Endobronchial Tube is used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

TUOREN Henan Tuoren Medical Device Co., Ltd

510(k) Summary

(as required by 807.92(c))

The assigned 510(K) number is: K152251

Date of Summary: June 9, 2016

1. Submitter information

Manufacturer Name: Henan Tuoren Medical Device Co., Ltd

Address: Weiyuan Industrial Zone, Menggang ,Changyuan Country, Henan, China

Establishment Registration Number: 3006984712

Contact Person and Title: Meizi Zhang/Manager of Registration Department

Tel: 0086-13633736073

Fax: 0086-373-8605321

Email: 13633736073@163.com

2. Contact person

2.1 Primary Contact Person

Long Yang (COO) Shenzhen Hlongmed Biotech Company R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District,Shenzhen, P.R. China Tel: 0086-755-86664986 Fax: 0086-755-86664933 E-mail: yanglong@hlongmed.com

2.2 Secondary Contact Person

Meizi Zhang(Manager of Registration Department) Henan Tuoren Medical Device Co., Ltd Tel: 0086-13633736073

3. Device Classification

Device SponsorHenan Tuoren Medical Device Co., Ltd
------------------------------------------------------

{4}------------------------------------------------

Trade/Device NameEndobronchial Tube
ModelLeft: Fr26, Fr28, Fr31, Fr32, Fr33, Fr35, Fr37, Fr39, Fr41Right: Fr26, Fr28, Fr31, Fr32, Fr33, Fr35, Fr37, Fr39, Fr41
Common NameEndobronchial Double Lumen Tube
Classification NameTube, Tracheal/Bronchial, Differential Ventilation (W/Wo Connector)
Regulatory ClassClass II
Classification regulation21CFR 868.5740
Review PanelAnesthesiology
Regulation Medical SpecialtyAnesthesiology
Regulation NameTracheal/bronchial differential ventilation tube
Product CodeCBI

OREN Henan Tuoren Medical Device Co., Ltd

4. Intended Use/ Indications for Use

The Endobronchial Tube is used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

5. Predicate Device

Well LEAD Endobronchial Tubes(K092886)

6. Device Description

The Endobronchial Tube is made of Polyvinylchloride and is available in sizes 26fr to 41fr. They are designated as double lumen tube with 2 cuffs and separate 15mm connectors for isolating and ventilating one lung during surgical procedures. The tubes contain an x-ray opaque line that runs through the tube making them detectable by x-ray. Environments of use: Hospital-OR and ICU

7. Substantial Equivalence

Henan Tuoren Medical Device Co., Ltd claims the proposed devices to be substantially equivalent to the devices previously cleared by FDA in K092886. Henan Tuoren Medical

{5}------------------------------------------------

TUOREN Henan Tuoren Medical Device Co., Ltd

Device Co.. Ltd claims this equivalence because the proposed devices have an equivalent intended use, manufacturing materials, operating principles and physical operational specifications as compared to the predicate devices. The similarities and differences between the proposed and predicate devices have been identified and explained in the comparison matrix, the comparison matrix can be referred to the below table. The differences between subject device and the predicate device is just the size and packaging, these differences do not effect substantial equivalence, the analysis and justification is: For the size differences, all models of proposed devices comply to the same standard ISO 5361 compared to predicate devices. For the packaging differences, the multi-functional joint of proposed device comply to the same standard ISO 5356-1 compared to predicate devices, and the suction catheters of proposed devices comply to the same standard ISO 8836 compared to predicate devices..

{6}------------------------------------------------

ItemElement Of ComparisonProposed DevicePredicate DeviceS/D*
1Classification NameTube, Tracheal/Bronchial, Differential Ventilation(W/Wo Connector)Tube, Tracheal/Bronchial, Differential Ventilation(W/Wo Connector)S
2Regulatory ClassClass IIClass IIS
3Classification regulation21CFR 868.574021CFR 868.5740S
4Classification PanelAnesthesiologyAnesthesiologyS
5Product CodeCBICBIS
6Regulation NameTracheal/bronchial differential ventilation tubeTracheal/bronchial differential ventilation tubeS
7IndicationsforUse/Intended UseEndobronchial Tubes is used to isolate the left or theright lung of a patient for surgery,one lung ventilationor one lung anesthesia.The Well Lead Endobronchial Tubes is used toisolate the left or the right lung of a patient forsurgery,one lung ventilation or one lung anesthesia.S
8Directions for UsePrescription UsePrescription UseS
9Patient PopulationPatients undergoing surgical procedure requiringisolation of one lungPatients undergoing surgical procedure requiringisolation of one lungS
10Patient contact MaterialTube-PVC and Cuff-PVCTube-PVC and Cuff-PVCS
11Size(Fr)26 to 41 French28 to 41 FrenchD, does not affect theproduct performance
12Design FeaturesDouble lumen shaft, 2 cuffs, Stylet, Carlens adapterDouble lumen shaft, 2 cuffs, Stylet, Carlens adapterS
13CompositionofEndobronchial Tubesmain tube, cuff, connector, inflating tube, valve, pilotballoonmain tube, cuff, connector, inflating tube, valve,pilot balloonS
14Angle of bevel70°70°S
15Radius of curvature140mm140mmS
ItemElement Of ComparisonProposed DevicePredicate DeviceS/D*
16Radiopaque lineYesYesS
17Connection to ventilationsource15mm connector15mm connectorS
18Shelf-life5 years5 yearsS
19Single UseSingle patient, disposableYesYesS
20Method of sterilizationEthylene Oxide Sterilized per ISO 11135Ethylene Oxide Sterilized per ISO 11135S
21The Sterility AssuranceLevelSAL 10-6SAL 10-6S
22EO and ECH residualThe ethylene oxide residual is conform to ISO10993:7 for Limited Exposure Devices of 4mg/dayfor EO and 9/mg/day for ECH.The ethylene oxide residual is conform to ISO10993:7 for Limited Exposure Devices of 4mg/dayfor EO and 9/mg/day for ECH.S
23Biocompatibility1.Cytotoxicity testing per 10993-52. Sensitization testing per ISO 10993-103.Irritation Test per ISO 10993-10All the test passed.1. Cytotoxicity testing per 10993-52. Sensitization testing per ISO 10993-103. Irritation Test per ISO 10993-10All the test passed.S
24Applied StandardsISO 5361ISO 10993-1ISO 10993-5ISO 10993-10ISO 11135ISO 10993-7ISO 5361ISO 10993-1ISO 10993-5ISO 10993-10ISO 11135ISO 10993-7S
25PackagingSterile, packed with a multi-functional joint and threeSuction CathetersSterile, packed with switch connector and twosuction cathetersD, does not affect the product performance
26Environments of useHospital-OR and ICUHospital-OR and ICUS

{7}------------------------------------------------

{8}------------------------------------------------

TUORen Henan Tuoren Medical Device Co., Ltd

8. Product Performance Testing

All testing that is required by the required standards has been performed. Non-clinical testing was performed and included standards such as ISO 5361 and ISO 10993-1. The Endobronchial Tube have been found to fall within the required limits of the testing. The test results can be found in both the Biocompatibility (Section16) and the Performance Testing_Bench (Section19) of this submission. Therefore we have concluded that the Endobronchial Tube are substantially equivalent.

A brief summary of tests relied upon to demonstrate substantial equivalence to the predicate can be found in the table below:

Type of TestReferenced StandardTest result
in vitro cytotoxicityISO10993-5:2009Under the conditions of using ISO10993-5: 2009test method MTT method MEM extract, the testarticle Endobronchial Tube extract showed nopotential toxicity to L-929 cells.
Under the conditions of using ISO10993-5: 2009test method MTT method MEM with 10% FBSextract, the test article Endobronchial Tube extractshowed no potential toxicity to L-929 cells.
Skin sensitizationTest using GuineaPig MaximizationtestISO10993-10:2010Under the conditions of using ISO10993-10: 2010test methods guinea pig maximization test 0.9%sodium chloride injection extract, the test articleEndobronchial Tube extract showed no significantevidence of causing skin sensitization in the guineapig;
Under the conditions of using ISO10993-10:2010test methods guinea pig maximization test sesameoil extract, the test article Endobronchial Tubeextract showed no significant evidence of causingskin sensitization in the guinea pig;

1) Biocompatibility testing

{9}------------------------------------------------

Type of TestReferenced StandardTest result
Oral MucosaIrritation testISO10993-10:2010Under the conditions of using ISO10993-10:2010test methods 0.9% sodium chloride extract, the testresult showed that the test article EndobronchialTube extract show no significant evidence ofcausing oral irritation in the hamster.Under the conditions of using ISO10993-10:2010test methods sesame oil extract, the test resultshowed that the test article Endobronchial Tubeextract show no significant evidence of causingoral irritation in the hamster.

TUOREN Henan Tuoren Medical Device Co., Ltd

2) Performance Testing

Type of Bench TestReference standardsAcceptable CriteriaPass/fail
cuff diameterISO 5361: 2012The maximum cuff diameter shall be within ± 0.5mm of the nominal value when tested according to ISO 5361 Annex BPass
cuffed tube collapse testISO 5361: 2012the steel ball shall pass freely through the tube when tested according to ISO 5361 Annex CPass
Cuff herniation testISO 5361: 2012no part of the inflated cuff shall reach beyond the nearest edge of the bevel when tested according to ISO 5361 Annex DPass
Seal testingISO 5361: 2012The leakage limit is ≤2.0ml/h at cuff pressures not to exceed 2,7 kPa (27 cmH2O) when tested according to ISO 5361 Annex GPass
/Sealing of cuff inflating system:No air leakage happens under the condition of continuous 3kPa positive pressure imposition for 10 secondsPass
/Sealing of connector assembly:No air leakage happens on any joints under the condition of continuous 6kPa gas pressure imposition to lumen of main tube(shaft) for 60 seconds.Pass
inflating tubeISO 5361: 2012The inflating tube shall have an outside diameter of not more than 3.0mmPass
ISO 5361: 2012The angle between the inflating tube and the Endobronchial tube at the point of separation shall not exceed 45°.Pass
Kink resistance testISO 5361: 2012the steel ball shall pass freely through the lumen of the tube when tested according to ISO 5361 Annex HPass
Gauging of One Way ValveISO 594/1 1986The plane of the maximum diameter at the opening of the female conical fitting of One Way Valve lie between the two limit planes of the gauge.Pass
Liquid leakage of One Way ValveISO 594/1 1986no leakage sufficient to form a falling drop of waterPass
Air leakage of One Way ValveISO 594/1 1986Continued formation of air bubbles not be evidentPass
Separation force of One Way ValveISO 594/1 1986The conical fitting under test remain attached to the test fixturePass
Stress crackingISO 594/1 1986There shall be no evidence of stress cracking of the conical fitting
Security of construction of suction catheterISO 8836-2007the force required to detach any component permanently attached to the shaft shall be not less than that specified in standardPass
Shaft resistant to negative pressure of suction catheterISO 8836-2007a vacuum source at 40kPa below ambient pressure for 15s at a temperature of 23°C ±2°C with the patient end occluded, the shaft shall not collapsePass
15mm connectorISO5356-1: 2004Comply with the ISO5356-1: 2004Pass
Burst Testing/Cuff burst: The volume of injected gas when cuff bursting happens is larger than 40ml during inflation of endobronchial tubebursting between cuff and main tube:There should be no fracture on junctions while inflated 30kpa gas.Pass
Bond Strength/The joints of endobronchial tube should be firm bonding.When an axial force of 50±5N is applied at 50±5mm/min, the testing portion(between connector andPass
Henan Tuoren Medical Device Co., Ltd
Air flow resistance/Pressure increment should no more than 0.2Kpa/h when testing the endobronchial tube under the specified flow rate(3, 6, 9L/min)Pass
Radiopaque test/When exposing the Endobronchial tube with the low dose rays, the X-ray machine should have film development.Pass
Endotoxin testUSP36_NF31<85>0.25EU/ml, 20EU/DevicePass

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "TUORen" in blue font. The "O" in the word is a circle filled with a pattern of small dots. The font is sans-serif and the letters are connected.

TUORen Henan Tuoren Medical Device Co., Ltd

{11}------------------------------------------------

TU●Ren Henan Tuoren Medical Device Co., Ltd

9. Clinical Testing and animal testing

Clinical and animal testing were not performed for Endobronchial Tube as part of the premarket Notification requirements for this 510(k) submission and the subject of this premarket submission, Endobronchial Tube, did not require clinical and animal studies to support substantial equivalence.

10. Substantial Equivalence Conclusion

The information provided in the 510(k) submission is sufficient to demonstrate the substantial equivalence of subject device to the predicate device.

§ 868.5740 Tracheal/bronchial differential ventilation tube.

(a)
Identification. A tracheal/bronchial differential ventilation tube is a device used to isolate the left or the right lung of a patient for anesthesia or pulmonary function testing.(b)
Classification. Class II (performance standards).