K Number
K152248

Validate with FDA (Live)

Date Cleared
2015-11-05

(87 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/umbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric diagnosed with the following conditions: spondylolisthesis/ spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation' attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

Device Description

The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The subject devices include: Rods Extenders Sequential Reducer Inserters Modified Labeling (surgical techniques and disassembly/assembly instructions) Accessories, cases, caddies, trays, and lids which may be used to transport and sterilize the subject implants and instruments. The subject CD HORIZON® Spinal System devices will be available in similar sizes as the predicate systems.

AI/ML Overview

The provided text describes specific details about a medical device (CD HORIZON® Spinal System) and its regulatory approval process, focusing on substantial equivalence to predicate devices, rather than an AI/ML device or its associated studies for performance. Therefore, I cannot extract the information requested for a device that uses AI/ML.

However, I can provide a summary based on the information available about the CD HORIZON® Spinal System's performance validation from pages 4 and 5 of the document, which pertains to medical device testing for substantial equivalence, not AI/ML performance.

Acceptance Criteria and Study for CD HORIZON® Spinal System (Non-AI/ML Device)

The provided document details the substantial equivalence review for the CD HORIZON® Spinal System, a non-AI/ML device. The "study" mentioned refers to engineering rationales and bench testing conducted to demonstrate that the new device is as safe and effective as previously approved predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance/Conclusion
BiocompatibilityBiological Evaluation (ISO-10993 Part 1)Materials must be safe for body contact based on contact duration and type.Subject rods are manufactured from identical materials (Cobalt-28-Chromium-6-Molybdenum, Titanium-6 Aluminum-4 Vanadium ELI) as predicate devices, adhering to ASTM standards. Biocompatibility testing was not required as these materials have a long history of safe and effective use.
Instrument BiocompatibilityExternal communicating instruments (extenders, reducers, inserters) and their materials must be safe for limited body contact.Instruments are manufactured from medical grade stainless steel (ASTM F899, A564, A693, A276) and acetal (ASTM F1855) which have a long history of safe and effective use. Biocompatibility testing was not required.
Mechanical TestingStatic Reduction TestingPerformance must meet predetermined acceptance criteria based on ASTM F1717 standards.The tests met the predetermined acceptance criteria.
Cantilever Load TestingPerformance must meet predetermined acceptance criteria.The tests met the predetermined acceptance criteria.
Torsional Load TestingPerformance must meet predetermined acceptance criteria.The tests met the predetermined acceptance criteria.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Mechanical Testing: Sample size was determined based on ASTM F1717 for Static Reduction Testing. Specific numerical sample sizes are not provided in this document.
  • Data Provenance: The studies were bench tests conducted by Medtronic Sofamor Danek USA, Inc. The data is generated internally to support a 510(k) submission, therefore it is prospective in nature with respect to the regulatory submission, but the tests simulate conditions. There is no information about country of origin of data as it's not a clinical study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the document is about a hardware medical device and its mechanical/biocompatibility testing for substantial equivalence, not an AI/ML device requiring expert ground truth for diagnostic or prognostic purposes. The "ground truth" for mechanical tests is established by published ASTM standards.

4. Adjudication Method for the Test Set

This information is not applicable for the type of testing described (mechanical and biocompatibility bench testing). Adjudication typically refers to expert review of clinical cases. The fulfillment of acceptance criteria for bench tests is usually objective and measurable, not subject to adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This information is not applicable. The device is a spinal implant system, not an imaging or diagnostic AI/ML tool. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed or detailed.

6. Standalone (Algorithm Only) Performance

This information is not applicable as the device is a spinal implant system and does not involve an algorithm.

7. Type of Ground Truth Used

  • For Biocompatibility: The "ground truth" is established by adherence to recognized standards (ISO-10993) and the long history of safe and effective use of the materials (Cobalt Chrome, Titanium, Stainless Steel, Acetal) in spinal surgery, eliminating the need for novel biocompatibility testing.
  • For Mechanical Testing: The "ground truth" or reference for acceptable performance is established by ASTM F1717 and other relevant ASTM standards for materials and mechanical properties of spinal implants.

8. Sample Size for the Training Set

This information is not applicable as there is no AI/ML algorithm or training set involved in the development or testing of this hardware device.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no AI/ML algorithm or training set involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a stylized representation of three human profiles. The profiles are stacked on top of each other, creating a sense of unity and connection. The seal is simple, yet recognizable, and represents the department's mission to protect the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Medtronic Sofamor Danek USA, Incorporated Ms. Becky Ronner Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

November 5, 2015

Re: K152248

Trade/Device Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP, KWQ Dated: August 7, 2015 Received: August 10, 2015

Dear Ms. Ronner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and 8091); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if

{1}------------------------------------------------

Page 2 - Ms. Becky Ronner

applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K152248

Device Name CD HORIZON® Spinal System

Indications for Use (Describe)

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/umbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric diagnosed with the following conditions: spondylolisthesis/ spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation' attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

MEDTRONIC Sofamor Danek CD HORIZON® Spinal System

August 2015

Submitter:Medtronic Sofamor Danek USA, Inc.1800 Pyramid PlaceMemphis, Tennessee 38132Telephone: (901)396-3133Fax: (901) 346-9738
Contact PersonBecky RonnerSenior Regulatory Affairs SpecialistDirect Telephone: (901)399-2757
Date PreparedAugust 7, 2015
Name of DeviceCD HORIZON® Spinal System
Common NameRods, Extenders, Reducers and Inserters
Trade NameCD HORIZON® Spinal System
Regulatory Class,Regulation Number,Regulation Name, andDevice Product CodeClass III 21 CFR 888.3050 Spinal Interlaminal Fixation Orthosis; KWP 21 CFR 888. 3060 Spinal Intervertebral Body Fixation Orthosis; KWQ 21 CFR 888.3070 Pedicle Screw System; MNH, MNI, NKB, OSH
Predicate DevicesK143375 CD HORIZON® Spinal System (S.E 2/13/2015) – Primary PredicateAdditional Predicates:K113174 CD HORIZON® Spinal System (S.E 11/21/2011)K122862 CD HORIZON® Spinal System (S.E 10/03/2012)K132639 CD HORIZON® Spinal System (S.E 11/25/2013)K102807 CD HORIZON® Spinal System (S.E 11/17/2010)The predicates have not been subject to a design related recall.
Description of DevicesThe CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The subject devices include:
Rods Extenders Sequential Reducer Inserters Modified Labeling (surgical techniques and disassembly/assembly instructions) Accessories, cases, caddies, trays, and lids which may be used to transport and sterilize the subject implants and instruments. The subject CD HORIZON® Spinal System devices will be available in similar sizes as the predicate systems.The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
Indications for UseWith the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by
tumor and/or trauma, pseudarthrosis, and/or failed previous fusion.
These devices are to be used with autograft and/or allograft.
Pediatric pedicle screw fixation is limited to a posterior approach.
The CD HORIZON® SPIRE™ Plate is a posterior, single-level,
non-pedicle supplemental fixation device intended for use in the
non-cervical spine (T1-S1) as an adjunct to fusion in skeletally
mature patients. It is intended for plate fixation/attachment to
spinous processes for the purpose of achieving supplemental
fixation in the following conditions: degenerative disc disease (as
previously defined), spondylolisthesis, trauma, and/or tumor.
In order to achieve additional levels of fixation, the CD
HORIZON® Spinal System rods may be connected to the
VERTEX® Reconstruction System with the VERTEX® rod
connector. Refer to the VERTEX® Reconstruction System
Package Insert for a list of the VERTEX® indications of use.
The CD HORIZON® Spinal System has the same fundamental
technology and the same cobalt chrome, titanium and stainless
steel materials as the predicate devices. Additionally, acetal is
commonly used for handles of instruments used in spinal surgery.
The predicate and subject devices are intended to help provide
Comparison ofTechnologicalCharacteristics with thePredicate Devices:immobilization and stabilization of spinal segments as an adjunct
to fusion of the thoracic lumbar and/or sacral spine.
K143375 CD HORIZON® Spinal System (S.E 2/13/2015) –Primary Predicate
K113174 CD HORIZON® Spinal System (S.E 11/21/2011)
K122862 CD HORIZON® Spinal System (S.E 10/03/2012)
K132639 CD HORIZON® Spinal System (S.E 11/25/2013)
K102807 CD HORIZON® Spinal System (S.E 11/17/2010)
Performance Data:The following performance data were provided in support of
substantial equivalence.
Biocompatibility
The biocompatibility evaluation for the CD HORIZON® Spinal
System devices was conducted in accordance with FDA's Draft
Guidance for Industry and FDA Staff "Use of International
Standard ISO-10993, Biological Evaluation of Medical Devices
Part 1: Evaluation and Testing" issued April, 23, 2013.
The subject CD HORIZON® Spinal System rods are permanent
implants and are classified as permanent , >30 day body contact
according to with FDA's Draft Guidance for Industry and FDA
Staff "Use of International Standard ISO-10993, Biological
Evaluation of Medical Devices Part 1: Evaluation and Testing".
The subject rods are manufactured from identical materials as the
predicate devices, in accordance with the following ASTM
standards:
ASTM F1537 - Standard Specification for Wrought●
Cobalt-28-Chromium-6-Molybdenum Alloys for Surgical
Implants
ASTM F136 - Standard Specification for Wrought●
Titanium-6 Aluminum-4 Vanadium ELI (Extra LowInterstitial) Alloy for Surgical Implant Applications
The CD HORIZON® Spinal System extender, sequential reducer,
and inserters are external communicating devices and are
classified as limited, up to 24 hours of body contact according to
with FDA's Draft Guidance for Industry and FDA Staff "Use of
International Standard ISO-10993, Biological Evaluation of
Medical Devices Part 1: Evaluation and Testing". These
instruments are manufactured from the same medical grade
stainless steel as the predicate devices in accordance with the
following ASTM standards:
ASTM F899 - Standard Specification for Wrought●
Stainless Steel for Surgical Instruments
ASTM A564 - Standard Specification for Hot-Rolled and●
Cold Finished Age-Hardening Stainless Steel Bars and
Shapes
ASTM A693 - Standard Specification for Precipitation-●
Hardening Stainless and Heat-Resisting Steel Plate, Sheet
and Stripe
ASTM A276 - Standard Specification for Stainless Bars●
and Shapes
Additionally, the handle of subject inserter which has indirect
patient contact is manufactured from acetal in accordance with
ASTM F1855 - "Standard Specification for Polyoxymethylene
(Acetal) for Medical Applications” with Celcon M25 Dark Bluecolorant. This material is surface material and is classified aslimited, up to 24 hours of body contact according to the FDA'sDraft Guidance for Industry and FDA Staff "Use of InternationalStandard ISO-10993, Biological Evaluation of Medical DevicesPart 1: Evaluation and Testing" issued April 23, 2013.
Cobalt Chrome, Commercially Pure Titanium, Titanium Alloy,medical grade stainless steel and acetal have a long history of safeand effective use in spinal surgery, and biocompatibility testing isnot required and no testing was conducted.
Mechanical TestingIn accordance with, Guidance for Industry and FDA Staff - SpinalSystem 510(k)'s", Medtronic has evaluated the subject devices todemonstrate substantial equivalence to the predicate devices. Acombination of engineering rationales and testing were used toestablish substantial equivalence. The following bench testingwas completed and the tests met the predetermined acceptancecriteria:
Static Reduction Testing Sample size was determined based on ASTM F1717 Cantilever Load testing Torsional Load Testing
Based on the risk analysis, test results, and additional supportingdocumentation provided in the pre-market notification, thesubject CD HORIZON® Spinal System are as safe and effectiveas the following predicates:
Conclusion:K143375 CD HORIZON® Spinal System (S.E 2/13/2015) -(Primary Predicate) K113174 CD HORIZON® Spinal System (S.E 11/21/2011) K122862 CD HORIZON® Spinal System (S.E 10/03/2012) K132639 CD HORIZON® Spinal System (S.E 11/25/2013) K102807 CD HORIZON® Spinal System (S.E 11/17/2010)

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.