(315 days)
The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position.
The RADspeed Pro uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals.
The device is not intended for mammographic applications.
The RADspeed Pro is an X-ray radiography system that is mainly used for the radiography of various regions of the patient's body in a standing or recumbent position. The RADspeed Pro can be used in a wide range of applications from general radiography using X-ray film or Computed Radiography (CR) cassettes, to digital radiography. The RADspeed Pro consists of an X-ray high voltage generator, X-ray tube unit, X-ray tube support and collimator. The system can be configured with radiographic table, radiographic stand and digital radiography system as well.
Optionally, the device is also used to perform tomosynthesis radiography by three different reconstruction modes. Filtered Back-Projection (FBP) mode is used to obtain a tomosynthesis image by performing back-projection after correcting the projection data. Shift Addition (SA) mode is used to obtain a tomosynthesis image at an arbitrary slice plane height by shifting each image according to projection angle of the tube based on the reconstruction height, and by applying image addition processing to them. Iteration (IR) mode is used to reduce metal artifact in tomosynthesis image. FBP mode is generally recommended for all body parts. In case an artifact is observed at joints and other similar places, SA mode may remedy this artifact. In case metal artifact is obviously displayed. IR mode is recommended to reduce metal artifact.
The provided text describes a 510(k) summary for the RADspeed Pro X-ray system. Here's a breakdown of the acceptance criteria and study details based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with quantitative performance metrics for disease detection or diagnostic accuracy. Instead, the performance studies primarily aim to demonstrate "substantial equivalence" to predicate devices, focusing on image quality and artifact reduction.
The closest to "acceptance criteria" are the conformities to various standards and the stated goal of achieving image quality comparable to predicate devices.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Conformity to IEC60601-1-2 (Electromagnetic compatibility) | Both the new device and primary predicate device are in conformity with IEC60601-1-2. (Difference in edition does not affect safety and effectiveness.) |
| Conformity to IEC10993-1 (Materials and Bio-compatibility) | All materials and components of the new device are in conformity with IEC10993-1 and are in clinical use by other commercial products. (Difference between new device and primary predicate device does not affect safety and effectiveness.) |
| Electrical, Mechanical, Chemical, and Thermal Safety | New device and primary predicate device are substantially equivalent by their conformed standards. |
| Software Risk Analysis, Verification, and Validation (as per FDA guidance) | Software Risk Analysis and Verification and Validation testing were conducted. The level of concern associated with the new device was identified, and documentation consistent with that level was provided. (Risk analysis indicates the difference does not affect safety and effectiveness.) |
| Image Quality - General Radiography (for software processing differences) | Non-clinical performance bench tests and image quality evaluations were conducted between the new device and predicate devices. The result demonstrates substantial equivalence in image acquisition, image processing, and image quality capabilities. |
| Image Quality - Dual Energy Subtraction (distinction of lesions) | An additional test report was submitted to assure the distinction of lesions overlying bones to calcification or tubercle shade. |
| Metal Artifact Reduction (Tomosynthesis option) | FBP (Filtered Back-Projection) mode and IR (Iteration) mode of the new device have the same level of metal artifact reduction capability as the mode with artifact reduction of its predicate device. |
| Image Quality - Tomosynthesis Reconstruction (anthropomorphic phantoms) | A U.S. board-certified radiologist review of anthropomorphic phantom images of tomosynthesis reconstruction covers an evaluation of each reconstruction mode for each anatomy. The review supports that the new device is substantially equivalent in its image quality of tomosynthesis reconstruction as compared to its predicate devices. SA (Shift Addition) mode is acceptable as a traditional image but is suggested for use with FBP or IR mode. |
| Radiation Safety (Conformity to CFR and IEC60601-1-3) | New device and primary predicate device are substantially equivalent by their conformities of CFR and IEC60601-1-3. |
| Overall Safety and Effectiveness | The non-clinical data support the safety of the device, and the hardware and software verification and validation demonstrate that the new device performs as intended. Differences do not affect clinical safety or effectiveness. The device is as safe, effective, and performs as well as or better than the predicate device(s). |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document mentions "several different anatomical body phantoms" for phantom tests, and an unspecified number of "anthropomorphic phantom images" for the tomosynthesis reconstruction review. Specific numerical sample sizes (e.g., number of phantoms, number of images reviewed) are not provided.
- Data Provenance: The studies were non-clinical performance testing using phantoms. There is no mention of human patient data (retrospective or prospective) being used for the performance evaluation detailed in Section V.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: "A U.S. board-certified radiologist" (singular) was used for the review of the tomosynthesis anthropomorphic phantom images.
- Qualifications: "U.S. board-certified radiologist." No information about years of experience or subspecialty is provided.
4. Adjudication Method for the Test Set
The document mentions "a U.S. board-certified radiologist review" for the tomosynthesis images. This single-reader review implies no adjudication method (like 2+1 or 3+1 consensus) was conducted, as there was only one expert involved in that specific review. For other non-clinical tests (bench tests, image quality evaluations), the method of evaluation and any expert involvement or adjudication is not specified.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. The document does not describe an MRMC comparative effectiveness study involving human readers with and without AI assistance. The radiologist review mentioned was for phantom images to assess image quality for substantial equivalence, not a comparative reader study to measure improvement from AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, implicitly. The "non-clinical performance bench test and image quality evaluate substantial equivalence of image processing technique" were conducted comparing the new device's output to predicate devices. This implies evaluating the algorithm's performance (specifically, its image processing and reconstruction capabilities) on phantom data. The metal artifact reduction function of the tomosynthesis option was also "conducted performance testing" which would be a standalone evaluation of the algorithm's output. The radiologist review of phantom images could be seen as an 'expert-in-the-loop' check on the standalone output.
7. The Type of Ground Truth Used
- Phantom Gold Standard: For the non-clinical performance testing (including general image quality, tomosynthesis reconstruction, and metal artifact reduction), the ground truth was established by the physical properties and known characteristics of the anatomical body phantoms used. The 'truth' is inherent in the phantom's design and the controlled testing environment. For Dual Energy Subtraction, it was related to the "distinction of lesion overlying bones to calcification or tubercle shade" within a phantom context.
- Expert Review of Phantom Images: For the tomosynthesis reconstruction review, a U.S. board-certified radiologist provided an evaluation, which serves as an expert opinion on the image quality rendered from the phantom data against the expected 'truth' from the phantom.
8. The Sample Size for the Training Set
The document does not provide any information about a training set or its sample size. This submission focuses on the performance testing of the modified device against predicate devices for substantial equivalence, not on the development or training of an AI algorithm from a dataset in the typical sense. While software processing is mentioned, details about machine learning training are absent.
9. How the Ground Truth for the Training Set was Established
As no training set is mentioned, there is no information on how its ground truth might have been established.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three human profiles facing right, with a stylized design that resembles a caduceus or a symbol of health and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2016
Shimadzu Corporation % Toshio Kadowaki General Manager, Quality Assurance Dept. Shimadzu Medical Systems 1. Nishinokyo-Kuwabaracho. Nakagyo-ku. Kyoto 604-8511, Japan
Re: K152244
Trade/Device Name: RADspeed Pro Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR Dated: June 13, 2016 Received: June 16, 2016
Dear Toshio Kadowaki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152244
Device Name RADspeed Pro
Indications for Use (Describe)
The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position.
The RADspeed Pro uses portable or integrated flat panel detectors to generate images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/ trained professionals.
The device is not intended for mammographic applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
HIMAD
510(k) Notification Submission, RADspeed Pro
SECTION V:
510(k) Summary
The following information is submitted in accordance with the requirements of 21 CFR§807.92.
1) Date of Submission
July 24th, 2015
2) Submitter
SHIMADZU CORPORATION 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto, 604-8511, Japan Registration Number: 8030233 Phone: +81-75-823-1305 Fax : +81-75-823-1377
3) Primary Contact Person
Toshio Kadowaki Phone: +81-75-823-1920 E-mail: kadowaki@shimadzu.co.jp
4) Secondary Contact Person
Jeffrey Seiler SHIMADZU MEDICAL SYSTEMS USA 20101 South Vermont Ave., Torrance, CA 90502 USA Phone: 310-217-8855 ext 174 Email: seiler@shimadzu-usa.com
5) Device
| Name of Device | : RADspeed Pro |
|---|---|
| Common Name | : General Radiography System |
| Classification Name | : Stationary x-ray system (21 CFR§892.1680) |
| Classification Panel | : Radiology |
| Regulatory Class | : Class II |
| Product Code | : KPR |
6) Predicate Device
| Primary | K093427 | FDR ACSELERATE SYSTEM, DR-ID 200FLAT PANEL DETECTOR | FUJIFILM CORPORATION |
|---|---|---|---|
| Secondary | K121499 | FUJIFILM TOMOSYNTHESIS OPTIONFOR FDR ACSELERATE STATIONARYX-RAY SYSTEM | FUJIFILM CORPORATION |
| Tertiary | K101686 | FDR IMAGE STITCHING OPTION | FUJIFILM CORPORATION |
| Quaternary | K122454 | FUJIFILM DUAL ENERGY SUBTRACTION(DES) SOFTWARE OPTION | FUJIFILM CORPORATION |
No reference devices were used in this submission.
{4}------------------------------------------------
SECTION V:
510(k) Summary
7) Purpose of Submission
The purpose of this submission is to notify FDA of our intent to introduce a modified device of legally marketed devices, K093427 FDR ACSELERATE SYSTEM, DR-ID 200 FLAT PANEL DETECTOR, K121499 FUJIFILM Tomosynthesis Option for FDR ACSELERATE STATINOARY X-RAY SYSTEM, K101686 FDR IMAGE STITCHING OPTION and K122454 FUJIFILM DUAL ENERGY SUBTRACTION (DES) SOFTWARE OPTION. The last three predicate devices are image processing options for the primary predicate device. All predicate devices are manufactured by FUJIFILM CORPORATION, Japan.
8) Device Description
The RADspeed Pro is an X-ray radiography system that is mainly used for the radiography of various regions of the patient's body in a standing or recumbent position. The RADspeed Pro can be used in a wide range of applications from general radiography using X-ray film or Computed Radiography (CR) cassettes, to digital radiography. The RADspeed Pro consists of an X-ray high voltage generator, X-ray tube unit, X-ray tube support and collimator. The system can be configured with radiographic table, radiographic stand and digital radiography system as well.
Optionally, the device is also used to perform tomosynthesis radiography by three different reconstruction modes. Filtered Back-Projection (FBP) mode is used to obtain a tomosynthesis image by performing back-projection after correcting the projection data. Shift Addition (SA) mode is used to obtain a tomosynthesis image at an arbitrary slice plane height by shifting each image according to projection angle of the tube based on the reconstruction height, and by applying image addition processing to them. Iteration (IR) mode is used to reduce metal artifact in tomosynthesis image. FBP mode is generally recommended for all body parts. In case an artifact is observed at joints and other similar places, SA mode may remedy this artifact. In case metal artifact is obviously displayed. IR mode is recommended to reduce metal artifact.
9) Intended Use
The RADspeed Pro is a stationary X-ray system for taking general radiographic and tomographic images of the whole body. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals.
10) Indications for Use
The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position.
The RADspeed Pro uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option
{5}------------------------------------------------
SECTION V:
510(k) Summary
Page 5-3
is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals.
The device is not intended for mammographic applications.
11) Comparison of Technological Characteristics with the predicate devices
At a high level, new device and its predicate devices are based on the following same technological elements:
- Energy emission to the patient - X-ray
- Power requirement, Environmental requirement
- Mechanism to generate X-ray o
- Mechanism to acquire, process and store image data O
- Use of the hardware components O
The following technological differences exist between new and its predicate devices.
- Use of software processing
- Addition of iteration reconstruction
12) Performance Data
The following performance data were provided in support of the substantial equivalence determination.
Electromaqnetic compatibility
Both new device and primary predicate device are in conformity with IEC60601-1-2. The difference is edition of the standard. Difference of edition of the standard does not affect safety and effectiveness.
Materials and Bio-compatibility testing
All materials and components of new device are in conformity with IEC10993-1. All materials and components of new device are in clinical use by other commercial products. Difference between new device and primary predicate device does not affect safety and effectiveness.
Sterility
Both new device and primary predicate device are not sterilized.
Electrical, Mechanical, Chemical and Thermal safety
As to these safeties, new device and primary predicate device are substantially equivalent by their conformed standards.
{6}------------------------------------------------
SECTION V:
510(k) Summary
Software Verification and Validation Testing
Digital radiography system is using similar, but different software processing method. Therefore, Software Risk Analysis and Verification and Validation testing were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" to assure substantial equivalence.
As a result, we identified the level of concern associated with new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.
Non-clinical Performance Testing
For the difference of software processing methods which is different from primary predicate device, we performed non clinical performance testing between new device and predicate devices, in accordance with the FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging(SSXI) Devices" issued on Aug. 6, 1999.
We conducted performance bench test and image quality evaluate substantial equivalence of image processing technique to each predicate device respectively.
We also performed phantom tests using several different anatomical body phantoms.
For Dual Energy Subtraction, additional test report was submitted to assure the distinction of lesion overlying bones to calcification or tubercle shade.
For metal artifact reduction function of tomosynthesis option, we conducted performance testing. The result indicates that both FBP(filtered back-projection) mode and IR(Iteration) mode of new device has a same level of metal artifact reduction capability as the mode with artifact reduction of its predicate device.
We performed a U.S. board-certified radiologist review of the anthropomorphic phantom images of tomosynthesis reconstruction. The review covers an evaluation of each reconstruction mode of new device for each anatomies. As a result, the review supports that new device is substantially equivalent in its image quality of tomosynthesis reconstruction as compared to its predicate devices. SA (Shift Addition) mode is acceptable as a traditional image, however, it is suggested that SA mode is used together with FBP or IR mode, whenever available.
Considering all these aspects, the result of non-clinical performance data demonstrates substantial equivalence of new device to the predicate devices in aspect of image acquisition, image processing as well as image quality capabilities.
{7}------------------------------------------------
SECTION V:
510(k) Summary
Clinical Performance Testing
The result of non-clinical bench testing demonstrates substantial equivalence to the predicate devices. Therefore, it is our understanding that clinical testing is not required in this case.
Radiation safety
New device and primary predicate device is substantially equivalent by their conformities of CFR and IEC60601-1-3.
13) Conclusion
The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that new device should perform as intended in the specified use. Based on our risk analysis and bench testing, the differences do not affect its clinical safety or effectiveness.
From the result of our risk analysis, software verification and validation, and nonclinical testing discussed above, it is our conclusion that,
- -New device is substantially equivalent to the legally marketed predicate devices, K093427 FDR ACSELERATE SYSTEM, DR-ID 200 FLAT PANEL DETECTOR, K121499 FUJIFILM Tomosynthesis Option for FDR ACSELERATE STATINOARY X-RAY SYSTEM, K101686 FDR IMAGE STITCHING OPTION and K122454 FUJIFILM DUAL ENERGY SUBTRACTION (DES) SOFTWARE OPTION.
- Additional Indication for Use of new device does not introduce any new safety and effectiveness concern.
- Therefore, new device is as safe, as effective, and performs as well as or better than the predicate device(s).
N/A