The FUJIFILM SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Ophthalmic Fetal – OB/GYN Abdominal Intra-operative (Abdominal organs and vascular) Pediatric Small Organ (breast, thyroid, testicles, prostate) Neonatal Cephalic Adult Cephalic Trans-vaginal Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Trans-esophageal (card.) Peripheral Vessel
The SonoSite X-Porte Ultrasound System is a highly mobile, full featured, general purpostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. X-Porte is a custom fabricated digital electronic design that readliy lends itself to be configured for specific ultrasound imaging applications through different system feature selections. The system interface can be customized for the user and controlled using a backlit touchscreen much like what is used in consumer tablet products. X-Porte can be operated in two different configurations, standbased with AC power or battery, and desktop-based with AC power only. In desktop configuration the ultrasound engine can be removed from the stand and used by itself with a single transducer and external monitor.
The provided text is a 510(k) Premarket Notification for the FUJIFILM SonoSite X-Porte Ultrasound System. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a clinical study for a novel AI/algorithm.
Therefore, much of the requested information regarding acceptance criteria, sample sizes for test/training sets, expert involvement, adjudication methods, MRMC studies, and standalone performance for an AI/algorithm is not applicable to this document. This document describes a traditional medical device (ultrasound system and transducers) and its substantial equivalence to previously cleared devices.
Here's a breakdown of the available and non-applicable information based on your request:
1. A table of acceptance criteria and the reported device performance
This document does not present specific acceptance criteria (e.g., sensitivity, specificity, or quantifiable performance metrics) and reported device performance in the context of an AI/algorithm for diagnostic accuracy. Instead, it relies on demonstrating substantial equivalence to predicate devices for its intended uses and technological characteristics.
The "performance" is implicitly the ability of the device to perform diagnostic ultrasound imaging and fluid flow analysis for the stated clinical applications, which is assumed to be equivalent to the predicate devices.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable. This is a traditional medical device submission, not an AI/algorithm submission requiring test set data for performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. Ground truth establishment by experts for a test set is typically for AI/algorithm performance validation, which is not described here. The "ground truth" for this device would be its validated capabilities for imaging and fluid flow analysis, established through technical testing and comparison to predicate devices, rather than being determined by experts for diagnostic accuracy in a study.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This document does not describe an adjudication method for a test set as it is not an AI/algorithm submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for an ultrasound system, not an AI diagnostic algorithm, and therefore does not include MRMC studies assessing human reader improvement with AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is an ultrasound system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For a traditional ultrasound system, the "ground truth" is its ability to produce images and perform measurements consistent with established medical and engineering standards, validated through a series of technical and safety tests referenced in the "Summary of Non-Clinical Tests" (e.g., electrical, thermal, mechanical, EMC safety, cleaning/disinfection, biocompatibility, acoustic output).
8. The sample size for the training set
Not applicable. There is no training set mentioned, as this is not an AI/algorithm submission.
9. How the ground truth for the training set was established
Not applicable. There is no training set and therefore no ground truth establishment for a training set described.
Summary of Information from the Document:
The document describes the FUJIFILM SonoSite X-Porte Ultrasound System as a general-purpose ultrasound system for diagnostic imaging and fluid flow analysis. It primarily demonstrates substantial equivalence to predicate devices (SonoSite X-Porte Ultrasound System K142017 and SonoSite Edge Ultrasound System K133454).
The "acceptance criteria" and "performance" described are related to the device's adherence to safety standards and its functional equivalence to predicate devices for specific clinical applications and imaging modes.
Non-Clinical Tests Summary:
The device was evaluated for:
- Electrical safety
- Thermal safety
- Mechanical safety
- EMC safety
- Cleaning/disinfection
- Biocompatibility
- Acoustic output
The device conforms to applicable mandatory medical device safety standards, including:
- AAMI / ANSI / ISO 10993-1 (Biological evaluation of medical devices)
- AAMI / ANSI ES60601-1 (Medical electrical equipment - General requirements)
- AAMI / ANSI / IEC 60601-1-2 (Electromagnetic compatibility)
- IEC 60601-2-37 (Particular requirements for ultrasonic medical diagnostic equipment)
- IEC 62359 (Ultrasonics - Field Characterization - Thermal and Mechanical Indices)
- ISO 14971 (Risk management to medical devices)
- NEMA UD 2-2004 (Acoustic Output Measurement Standard)
- NEMA PS 3.15 (DICOM - Security and System Management Profiles)
Clinical Tests Summary:
The document explicitly states: "The SonoSite X-Porte Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence."
Transducer New Indications:
The submission does include "new indications" (marked with 'N') for certain transducers, where the device previously cleared ('P') for other indications. For example, the TEExp/8-3 MHz Transducer has "N" for all Cardiac Adult, Cardiac Pediatric, and Trans-esophageal (card.) modes, indicating these are new for this specific transducer. However, the document relies on comparisons to predicate devices that already have these capabilities, effectively claiming equivalence by association rather than presenting new performance data for these new indications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 19, 2015
Fujifilm Sonosite, Inc. % Mark Job Responsible Third Party Official 1394 25th Street, NW BUFFALO MN 55313
Re: K152209
Trade/Device Name: FUJIFILM Sonosite X-porte Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: July 21, 2015 Received: August 7, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara For
Robert Ochs Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) TBD
Device Name
FUJIFILM SonoSite X-Porte Ultrasound System
Indications for Use (Describe)
The FUJIFILM SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
Ophthalmic Fetal – OB/GYN Abdominal Intra-operative (Abdominal organs and vascular) Pediatric Small Organ (breast, thyroid, testicles, prostate) Neonatal Cephalic Adult Cephalic Trans-vaginal Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Trans-esophageal (card.) Peripheral Vessel
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Table 1.3-1: Diagnostic Ultrasound Indications for Use Form – FUJIFILM SonoSite X-Porte Ultrasound System
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | |||||||
|---|---|---|---|---|---|---|---|---|
| Transducer: | N/A | |||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the humanbody as follows: | |||||||
| Clinical Application | Mode of Operation | |||||||
| B | M | PWD | CWD | ColorDoppler | Combined(Spec.) | Other(Spec.) | ||
| Ophthalmic | P | P | P | P | B+M; B+CD | 1-5 | ||
| Fetal | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | ||
| Abdominal | P | P | P | P | P | B+M; B+PWD;B+CWD; B+CD | 1-5 | |
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1-5 | ||
| Intra-operative (Neuro.)Laparoscopic | ||||||||
| Pediatric | P | P | P | P | B+M; B+PWD;B+CWD; B+CD | 1-5 | ||
| Small Organ (breast, thyroid,testicles, prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1-5 | ||
| Neonatal Cephalic | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | ||
| Adult Cephalic | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | ||
| Trans-rectal | ||||||||
| Trans-vaginal | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1-5 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Intra-luminal | ||||||||
| Other (spec.) | ||||||||
| Cardiac Adult | P | P | P | P | P | B+M; B+PWD;B+CWD; B+CD | 1-5 | |
| Cardiac Pediatric | P | P | P | P | P | B+M; B+PWD;B+CWD; B+CD | 1-5 | |
| Trans-esophageal (card.) | N | N | N | N | N | B+M; B+PWD;B+CWD; B+CD | 1,3,5 | |
| Other (spec.) | ||||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1-5 | ||
| Other (spec.) |
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
- 2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017 and K133134.
{4}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | |||||||
|---|---|---|---|---|---|---|---|---|
| Transducer: | C11xp/8-5 MHz Transducer | |||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human | |||||||
| body as follows: | ||||||||
| Clinical Application | Mode of Operation | |||||||
| B | M | PWD | CWD | ColorDoppler | Combined(Spec.) | Other(Spec.) | ||
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Small Organ (breast, thyroid,testicles. prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Neonatal Cephalic | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (spec.) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Trans-esophageal (card.) | ||||||||
| Other (spec.) | ||||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | ||
| Other (spec.) |
Table 1.3-2: Diagnostic Ultrasound Indications for Use Form – C11xp/8-5 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017.
{5}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | |||||||
|---|---|---|---|---|---|---|---|---|
| Transducer: | C35xp/8-3 MHz Transducer | |||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the humanbody as follows: | |||||||
| Clinical Application | Mode of Operation | Other(Spec.) | ||||||
| B | M | PWD | CWD | ColorDoppler | Combined(Spec.) | |||
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | ||
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | ||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | ||
| Small Organ (breast, thyroid,testicles. prostate) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | ||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (spec.) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac Pediatric | ||||||||
| Trans-esophageal (card.) | ||||||||
| Other (spec.) | ||||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | ||
| Other (spec.) |
Table 1.3-3: Diagnostic Ultrasound Indications for Use Form – C35xp/8-3 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017.
{6}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | C60xp/5-2 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | ||||||
| Clinical Application | Mode of Operation | ||||||
| B | M | PWD | CWD | Color Doppler | Combined (Spec.) | Other (Spec.) | |
| Ophthalmic | P | P | P | P | B+M; B+PWD; B+CD | 1,2,5 | |
| Fetal | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | |
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | |
| Intra-operative (Abdominal organs and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | |
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | |
| Small Organ (breast, thyroid, testicles, prostate) | |||||||
| Neonatal Cephalic | |||||||
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | |
| Musculo-skel. (Superfic.) | |||||||
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | |||||||
| Cardiac Pediatric | |||||||
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,2,4,5 | |
| Other (spec.) |
Table 1.3-4: Diagnostic Ultrasound Indications for Use Form – C60xp/5-2 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
- 2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K133134.
{7}------------------------------------------------
Table 1.3-5: Diagnostic Ultrasound Indications for Use Form – HFL38xp/13-6 Transducer
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | |||||||
|---|---|---|---|---|---|---|---|---|
| Transducer: | HFL38xp/13-6 MHz Transducer | |||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human | |||||||
| body as follows: | ||||||||
| Clinical Application | Mode of Operation | |||||||
| ColorCombinedOther | ||||||||
| B | M | PWD | CWD | Doppler | (Spec.) | (Spec.) | ||
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Intra-operative (Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| organs and vascular) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Small Organ (breast, thyroid, | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| testicles. prostate) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Intra-luminal | ||||||||
| Other (spec.) | ||||||||
| Cardiac Adult | E | E | E | E | B+M; B+PWD; B+CD | 1,4,5 | ||
| Cardiac Pediatric | E | E | E | E | B+M; B+PWD; B+CD | 1,4,5 | ||
| Trans-esophageal (card.) | ||||||||
| Other (spec.) | ||||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Other (spec.) |
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017.
Prescription Use (Per 21 CFR 801.109)
Page 6 of 14
{8}------------------------------------------------
Table 1.3-6: Diagnostic Ultrasound Indications for Use Form – HFL50xp/15-6 Transducer
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | HFL50xp/15-6 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the humanbody as follows: | ||||||
| Clinical Application | B | M | PWD | CWD | ColorDoppler | Combined(Spec.) | Other(Spec.) |
| Ophthalmic | |||||||
| Fetal | |||||||
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Small Organ (breast, thyroid,testicles. prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Neonatal Cephalic | |||||||
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Musculo-skel. (Superfic.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | |||||||
| Cardiac Pediatric | |||||||
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K133134.
Prescription Use (Per 21 CFR 801.109)
Page 7 of 14
{9}------------------------------------------------
Table 1.3-7: Diagnostic Ultrasound Indications for Use Form – HSL25xp/13-6 Transducer
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | HSL25xp/13-6 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the humanbody as follows: | ||||||
| Clinical Application | Mode of Operation | ||||||
| B | M | PWD | CWD | ColorDoppler | Combined(Spec.) | Other(Spec.) | |
| Ophthalmic | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Fetal | |||||||
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Small Organ (breast, thyroid,testicles. prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Neonatal Cephalic | |||||||
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Musculo-skel. (Superfic.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Cardiac Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017.
Prescription Use (Per 21 CFR 801.109)
Page 8 of 14
{10}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | ICTxp/9-5 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | ||||||
| Clinical Application | B | M | PWD | CWD | Color Doppler | Combined (Spec.) | Other (Spec.) |
| Ophthalmic | |||||||
| Fetal | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Abdominal | |||||||
| Intra-operative (Abdominal organs and vascular) | |||||||
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | |||||||
| Small Organ (breast, thyroid, testicles, prostate) | |||||||
| Neonatal Cephalic | |||||||
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | |||||||
| Musculo-skel. (Superfic.) | |||||||
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | |||||||
| Cardiac Pediatric | |||||||
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | |||||||
| Other (spec.) |
Table 1.3-8: Diagnostic Ultrasound Indications for Use Form – ICTxp/9-5Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
- 2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K133134.
Prescription Use (Per 21 CFR 801.109)
Page 9 of 14
{11}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | L25xp/13-6 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the humanbody as follows: | ||||||
| Clinical Application | Mode of Operation | ||||||
| B | M | PWD | CWD | ColorDoppler | Combined(Spec.) | Other(Spec.) | |
| Ophthalmic | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Fetal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Small Organ (breast, thyroid,testicles, prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Neonatal Cephalic | |||||||
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Musculo-skel. (Superfic.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Cardiac Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | |
| Other (spec.) |
Table 1.3-9: Diagnostic Ultrasound Indications for Use Form – L25xp/13-6 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
- 2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017 and K133134.
Prescription Use (Per 21 CFR 801.109)
Page 10 of 14
{12}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | |||||||
|---|---|---|---|---|---|---|---|---|
| Transducer: | L38xp/10-5 MHz Transducer | |||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | |||||||
| Clinical Application | Mode of Operation | |||||||
| B | M | PWD | CWD | Color Doppler | Combined (Spec.) | Other (Spec.) | ||
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Intra-operative (Abdominal organs and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Small Organ (breast, thyroid, testicles. prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Intra-luminal | ||||||||
| Other (spec.) | ||||||||
| Cardiac Adult | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Cardiac Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Trans-esophageal (card.) | ||||||||
| Other (spec.) | ||||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,4,5 | ||
| Other (spec.) |
Table 1.3-10: Diagnostic Ultrasound Indications for Use Form – L38xp/10-5 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017 and K133134.
Prescription Use (Per 21 CFR 801.109)
Page 11 of 14
{13}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | P10xp/8-4 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the humanbody as follows: | ||||||
| Clinical Application | Mode of Operation | ||||||
| B | M | PWD | CWD | Color Doppler | Combined (Spec.) | Other (Spec.) | |
| Ophthalmic | |||||||
| Fetal | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Abdominal | P | P | P | P | P | B+M; B+PWD; B+CWD; B+CD | 1,5 |
| Intra-operative (Abdominalorgans and vascular) | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Small Organ (breast, thyroid,testicles. prostate) | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Neonatal Cephalic | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Musculo-skel. (Superfic.) | |||||||
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | P | P | P | P | P | B+M; B+PWD; B+CWD; B+CD | 1,3,5 |
| Cardiac Pediatric | P | P | P | P | P | B+M; B+PWD; B+CWD; B+CD | 1,3,5 |
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1,5 | |
| Other (spec.) |
Table 1.3-11: Diagnostic Ultrasound Indications for Use Form – P10xp/8-4 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017.
{14}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | P21xp/5-1 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human | ||||||
| body as follows: | |||||||
| Clinical Application | Mode of Operation | ||||||
| B | M | PWD | CWD | ColorDoppler | Combined (Spec.) | Other (Spec.) | |
| Ophthalmic | |||||||
| Fetal | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| Abdominal | P | P | P | P | P | B+M; B+PWD; B+CWD; B+CD | 1-3,5 |
| Intra-operative (Abdominal | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| organs and vascular) | |||||||
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | P | P | P | P | B+M; B+PWD | 1-3,5 | |
| Small Organ (breast, thyroid, | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| testicles, prostate) | |||||||
| Neonatal Cephalic | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| Adult Cephalic | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| Musculo-skel. (Superfic.) | |||||||
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | P | P | P | P | P | B+M; B+PWD; B+CWD; B+CD | 1-3,5 |
| Cardiac Pediatric | P | P | P | P | P | B+M; B+PWD; B+CWD; B+CD | 1-3,5 |
| Trans-esophageal (card.) | |||||||
| Other (spec.) | |||||||
| Peripheral vessel | P | P | P | P | B+M; B+PWD; B+CD | 1-3,5 | |
| Other (spec.) |
Table 1.3-12: Diagnostic Ultrasound Indications for Use Form – P21xp/5-1 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
- 2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
All items marked "P" were previously cleared in 510(k) K142017 and K133134.
{15}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | ||||||
|---|---|---|---|---|---|---|---|
| Transducer: | TEExp/8-3 MHz Transducer | ||||||
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | ||||||
| Clinical Application | B | M | PWD | CWD | Color Doppler | Combined (Spec.) | Other (Spec.) |
| Ophthalmic | |||||||
| Fetal | |||||||
| Abdominal | |||||||
| Intra-operative (Abdominal organs and vascular) | |||||||
| Intra-operative (Neuro.) | |||||||
| Laparoscopic | |||||||
| Pediatric | |||||||
| Small Organ (breast, thyroid, testicles, prostate) | |||||||
| Neonatal Cephalic | |||||||
| Adult Cephalic | |||||||
| Trans-rectal | |||||||
| Trans-vaginal | |||||||
| Trans-urethral | |||||||
| Trans-esoph. (non-Card.) | |||||||
| Musculo-skel. (Convent.) | |||||||
| Musculo-skel. (Superfic.) | |||||||
| Intra-luminal | |||||||
| Other (spec.) | |||||||
| Cardiac Adult | |||||||
| Cardiac Pediatric | |||||||
| Trans-esophageal (card.) | N | N | N | N | N | B+M; B+PWD; B+CWD; B+CD | 1,3,5 |
| Other (spec.) | |||||||
| Peripheral vessel | |||||||
| Other (spec.) |
Table 1.3-13: Diagnostic Ultrasound Indications for Use Form – TEExp/8-3 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance. M-Mode includes Simultaneous M-Mode.
- 2: Tissue Harmonic Imaging (THI)
- 3: Tissue Doppler Imaging (TDI)
- 4: Steep Needle Profiling (Sono MBe)
- 5: Multi-beam Imaging (SonoMB) in B-Mode
Prescription Use (Per 21 CFR 801.109)
Page 14 of 14
{16}------------------------------------------------
510(K) Summary
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
1) Submitter's name, address, telephone number, contact person:
FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904
| Corresponding Official: | Patricia Liau |
|---|---|
| Supervisor, Regulatory Affairs | |
| E-mail: | Patricia.Liau@sonosite.com |
| Telephone: | (425) 951-6870 |
| Facsimile: | (425) 951-1201 |
| Date prepared: | July 20, 2015 |
2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
Proprietary Name
SonoSite X-Porte™ Ultrasound System (subject to change)
Classification Names
| Name | FR Number | Product Code |
|---|---|---|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
3) Identification of the predicate or leqally marketed device:
SonoSite X-Porte Ultrasound System K142017 SonoSite Edge Ultrasound System K133454
4) Device Description:
The SonoSite X-Porte Ultrasound System is a highly mobile, full featured, general purpostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. X-Porte is a custom fabricated digital electronic design that readliy lends itself to be configured for specific ultrasound imaging applications through different system feature selections. The system interface can be customized for the user and controlled using a backlit touchscreen much like what is used in consumer tablet products. X-Porte can be operated in two different configurations, standbased with AC power or battery, and desktop-based with AC power only. In desktop configuration the ultrasound engine can be removed from the stand and used by itself with a single transducer and external monitor.
{17}------------------------------------------------
5) Intended Use:
The FUJIFILM SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
Ophthalmic Fetal - OB/GYN Abdominal Intra-operative (Abdominal organs and vascular) Pediatric Small Organ (breast, thyroid, testicles, prostate) Neonatal Cephalic Adult Cephalic Trans-vaginal Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Trans-esophageal (card.) Peripheral Vessel
6) Technological Characteristics:
SonoSite X-Porte and Edge Ultrasound Systems are both Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.
| Feature | SonoSite X-PorteUltrasound System(This submission) | SonoSite X-PorteUltrasound System(K142017) | SonoSite EdgeUltrasound System(K133454) |
|---|---|---|---|
| Intended Use | Diagnostic ultrasound imaging orfluid flow analysis of the humanbody | Diagnostic ultrasound imaging orfluid flow analysis of the humanbody | Diagnostic ultrasound imaging orfluid flow analysis of the humanbody |
| Indications forUse | OpthalmicFetal - OB/GYNAbdominalIntraoperative (abdominal organsand vascular)PediatricSmall Organ (breast, thyroid,testicle, prostate)Neonatal CephalicAdult CephalicTrans-VaginalMusculo-skeletal (Conventional)Musculo-skeletal (Superficial)Cardiac AdultCardiac PediatricTrans-esophageal (cardiac)Peripheral VesselNeedle guidance | OpthalmicFetal - OB/GYNAbdominalIntraoperative (abdominal organsand vascular)PediatricSmall Organ (breast, thyroid,testicle, prostate)Neonatal CephalicAdult CephalicTrans-VaginalMusculo-skeletal (Conventional)Musculo-skeletal (Superficial)Cardiac AdultCardiac PediatricPeripheral VesselNeedle guidance | OpthalmicFetal - OB/GYNAbdominalIntraoperative (abdominal organsand vascular)Intra-operative (Neuro.)PediatricSmall Organ (breast, thyroid,testicle, prostate)Neonatal CephalicAdult CephalicTrans-RectalTrans-VaginalMusculo-skeletal (Conventional)Musculo-skeletal (Superficial)Cardiac AdultCardiac PediatricTrans-esophageal (cardiac)Peripheral VesselNeedle guidance |
| TransducerTypes | Linear ArrayCurved Linear ArrayIntracavitaryPhased Array | Linear ArrayCurved Linear ArrayIntracavitaryPhased Array | Linear ArrayCurved Linear ArrayIntracavitaryPhased ArrayStatic ProbesTrans-esophageal |
| Transducer | Trans-esophageal1.0 - 15.0 MHz | 1.0 — 15.0 MHz | 1.0 — 15.0 MHz |
| Feature | SonoSite X-PorteUltrasound System(This submission) | SonoSite X-PorteUltrasound System(K142017) | SonoSite EdgeUltrasound System(K133454) |
| FrequencyAcousticOutput Display& FDA Limits | ISTPA.3 ≤ 720 (mW/cm²)TI ≤ 4.0MI ≤1.9Display Feature for HigherOutputsMI Output DisplayTI Output Display | ISTPA.3 ≤ 720 (mW/cm²)TI ≤ 4.0MI ≤1.9Display Feature for HigherOutputsMI Output DisplayTI Output Display | ISTPA.3 ≤ 720 (mW/cm²)TI ≤ 4.0MI ≤1.9Display Feature for HigherOutputsMI Output DisplayTI Output Display |
| Modes ofOperation | B-mode Grayscale ImagingTissue Harmonic ImagingM-modeSimultaneous M-ModeColor Power DopplerZoomCombination ModesPulsed Wave (PW) DopplerContinuous Wave (CW) DopplerSonoHD2 Noise ReductionSonoMB/MBe ImageCompoundingSteered CW DopplerVelocity Color DopplerTissue Doppler Imaging (TDI) | B-mode Grayscale ImagingTissue Harmonic ImagingM-modeSimultaneous M-ModeColor Power DopplerZoomCombination ModesPulsed Wave (PW) DopplerContinuous Wave (CW) DopplerSonoHD2 Noise ReductionSonoMB/MBe ImageCompoundingSteered CW DopplerVelocity Color DopplerTissue Doppler Imaging (TDI) | B-mode Grayscale ImagingTissue Harmonic ImagingM-modeColor M-ModeColor Power DopplerZoomCombination ModesPulsed Wave (PW) DopplerContinuous Wave (CW) DopplerSonoHD2 Noise ReductionSonoMB/MBe ImageCompoundingSteered CW DopplerVelocity Color DopplerTissue Doppler Imaging (TDI) |
| PW Doppler | Available | Available | Available |
| CW Doppler | Available | Available | Available |
| Velocity ColorDoppler | Available | Available | Available |
| Elastography(Strain), andStrain RateImaging | Not available | Not available | Available |
| ECG Feature | 3-lead ECG input | 3-lead ECG input | 3-lead ECG input |
| DICOM | DICOM 3.0 | DICOM 3.0 | DICOM 3.0 |
| IMTMeasurement | Not available | Not available | Available |
| #TransmitChannels | 128 digital channels | 128 digital channels | 128 digital channels |
| #ReceiveChannels | 64 digital channels(128 digital channels usingSynthetic Aperture) | 64 digital channels(128 digital channels usingSynthetic Aperture) | 64 digital channels(128 digital channels usingSynthetic Aperture) |
| PatientContactMaterials | Transducers:Acrylonitrile-butadien-styrene(ABS)CycoloyDow Medical Adhesive, Type AEpoxy paste adhesivePolyethylene (PE) IonomerPolyetheretherketone (PEEK)Polysulfone UDEL P1700PolyurethanePoly-Vinyl-Chloride (PVC)Silicone RTV AdhesiveSilicone RubberUrethaneNeedle Guides:Acetal copolymerAcrylonitrile-butadien-styrene(ABS) | Transducers:Acrylonitrile-butadien-styrene(ABS)CycoloyDow Medical Adhesive, Type AEpoxy paste adhesivePolyethylene (PE) IonomerPolyetheretherketone (PEEK)Polysulfone UDEL P1700PolyurethanePoly-Vinyl-Chloride (PVC)Silicone RTV AdhesiveSilicone RubberUrethaneNeedle Guides:Acetal copolymerAcrylonitrile-butadien-styrene(ABS) | Transducers:Acrylonitrile-butadien-styrene(ABS)CycoloyDow Medical Adhesive, Type AEpoxy paste adhesivePolyethylene (PE) IonomerPolyetheretherketone (PEEK)Polysulfone UDEL P1700PolyurethanePoly-Vinyl-Chloride (PVC)Silicone RTV AdhesiveSilicone RubberUrethaneNeedle Guides:Acetal copolymerAcrylonitrile-butadien-styrene(ABS) |
| System | X-Porte (stand configuration):Beamformer 128/128 using SA | X-Porte (stand configuration):Beamformer 128/128 using SA | Edge:Beamformer 128/128 using SA |
| Feature | SonoSite X-PorteUltrasound System(This submission) | SonoSite X-PorteUltrasound System(K142017) | SonoSite EdgeUltrasound System |
| (K133454) | |||
| Characteristics | (configurable)12.1" Capacitive touch screeninterface19" LED LCD HD monitor256 gray shades on LED LCD | (configurable)12.1" Capacitive touch screeninterface19" LED LCD HD monitor256 gray shades on LED LCD | (configurable)Hand held display and controlSingle 12.1" Liquid Crystal Display(LCD)256 gray shades on LCD |
| 6 USB 2.0 ports | 6 USB 2.0 ports | 2 USB ports | |
| Stand Base Dimensions: 26.4" L x21.2" WStand Height (max): 64" (monitorup)Stand Height (min): 42.2" (monitordown) | Stand Base Dimensions: 26.4" L x21.2" WStand Height (max): 64" (monitorup)Stand Height (min): 42.2" (monitordown) | Dimensions: 12.9"(W) x 12.4 (L) x2.5"(H) | |
| Weight: 149.35 lbs (fullyconfigured w/ 3 transducers | Weight: 149.35 lbs (fullyconfigured w/ 3 transducers | Weight: 8.5 lbsBattery operated (1.5 - 4 houroperation per charge) | |
| System operates via battery or ACpower | System operates via battery or ACpower | System operates via battery or ACpower | |
| Battery life: 1 hour operational - 3days idle | Battery life: 1 hour operational - 3days idle | 100 - 240V options, 50/60 Hz,15VDC output | |
| Input: 100 – 240 VAC, 50/60 HzOutput 1: 24VDC output, 275 WmaxOutput 2: 100-240VAC, 50-60 Hz(AC Printer) | Input: 100 – 240 VAC, 50/60 HzOutput 1: 24VDC output, 275 WmaxOutput 2: 100-240VAC, 50-60 Hz(AC Printer) | Various obstetrical, cardiac,volume, M-mode, PW and CWDoppler measurement andcalculation packages | |
| Various obstetrical, cardiac,volume, M-mode, PW and CWDoppler measurement andcalculation packages | Various obstetrical, cardiac,volume, M-mode, PW and CWDoppler measurement andcalculation packages | ECG acquisition and displaycapabilitiesCW/PW Doppler Audio | |
| ECG acquisition and displaycapabilitiesCW/PW Doppler Audio | ECG acquisition and displaycapabilitiesCW/PW Doppler Audio | Spectral Doppler Audio and imagestorage on removable media | |
| Spectral Doppler Audio and imagestorage on removable mediaMeasurement on RecalledImages. | Spectral Doppler Audio and imagestorage on removable mediaMeasurement on RecalledImages. | Wireless 802.11 (a/b/g) support forimage transfer | |
| Wireless 802.11 (a/b/g/n) supportfor image transfer | Wireless 802.11 (a/b/g/n) supportfor image transfer | ||
| X-Porte (desktop configuration): | X-Porte (desktop configuration): | ||
| Same softwarefeatures/capabilities as the standconfiguration. Does not have thestand, touch panel interface, DVR,and mobile power unit. | Same softwarefeatures/capabilities as the standconfiguration. Does not have thestand, touch panel interface, DVR,and mobile power unit. | ||
| Weight: 32.80 lbs (w/ 1transducer) | Weight: 32.80 lbs (w/ 1transducer) | ||
| 510(k) Track | AC power only.100 - 240V options, 50/60 HzTrack 3 | AC power only.100 - 240V options, 50/60 HzTrack 3 | Track 3 |
{18}------------------------------------------------
{19}------------------------------------------------
{20}------------------------------------------------
7) Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
The X-Porte Ultrasound System has been evaluated for electrical, thermal, mechanical and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated, and the device has been found to conform to applicable mandatory medical device safety standards. Assurance of quality was established by employing the following elements of product development: Design Phase Reviews, Risk Assessment, Requirement, System and Software Verification, Hardware Verification, Safety Compliance Verification, Clinical Validation. All patient contact materials are biocompatible. Reports for these elements of product development are referenced in Attachment 6.
| Reference No. | Title |
|---|---|
| AAMI / ANSI / ISO10993-1 | ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices -- Part 1:Evaluation and testing within a risk management process |
| IEC 60601-1 | AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 andA2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: Generalrequirements for basic safety and essential performance (IEC 60601-1:2005, MOD) |
| IEC 60601-1-2 | AAMI / ANSI / IEC 60601-1-2:2007(R)2012, Medical electrical equipment - Part 1-2:General requirements for basic safety and essential performance - Collateralstandard: Electromagnetic compatibility - Requirements and tests (Edition 3) |
| IEC 60601-2-37 | IEC 60601-2-37:2007, Particular Requirements for the basic safety and essentialperformance of ultrasonic medical diagnostic and monitoring equipment |
| IEC 62359 | IEC 62359:2010, Ultrasonics - Field Characterization - Test Methods For TheDetermination Of Thermal And Mechanical Indices Related To Medical DiagnosticUltrasonic Fields [Including: Technical Corrigendum 1 (2011)] (Edition 2) |
| ISO 14971 | ISO 14971: 2007, Medical devices - Application of risk management to medicaldevices |
| NEMA UD 2-2004 | Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment |
| NEMA PS 3.15 | NEMA Ps 3.15:2011, Digital Imaging and Communications in Medicine (DICOM),Part 15: Security and System Management Profiles |
The X-Porte Ultrasound System is designed to comply with the following voluntary standards.
Summary of Clinical Tests:
The SonoSite X-Porte Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.
8) Conclusion:
Intended uses and other key features are consistent with traditional clinical practice and FDA guidance. The X-Porte device and predicates conform to applicable electromedical device safety standards with compliance verified through independent evaluation. The X-Porte device and predicates meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence and are manufactured using the same ISO 13485 quality system. FUJIFILM SonoSite, Inc. believes that the X-Porte Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate devices.
N/A