K Number
K152204

Validate with FDA (Live)

Device Name
VS3-IR system
Manufacturer
Date Cleared
2015-10-28

(83 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3D Endoscope Module of the VS3-IR System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures.

The MMS configuration of the Iridium Module of the VS3-IR System is intended for capturing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic. micro- and reconstructive surgical procedures.

The Endoscope configuration of the Iridium Module of the VS3-IR System is intended to provide real-time endoscopic visible and near infrared fluorescence imaging. Iridium Endoscope enables surgeons to perform routine visible light endoscopic procedures as well as further visually assess vessels, blood flow and related tissue perfusion with near infrared imaging during minimally invasive surgery.

Device Description

The VS3-IR System is made up of the following components:

The VS3 3D endoscopes (cleared in K123467, K131434, and K141002) are based on the proximal camera concept with a stereoscopic camera block on the proximal side of the endoscope (the handle). The stereoscopic images are transmitted from the visual field at the distal tip of the endoscope to the proximal camera block through an optical transmission system to PC workstations or external display monitors for enhanced viewing or capturing. VS3 includes reusable endoscopes which are cleaned and sterilized using the same methods and are indicated for the same lifecycle.

In addition to traditional endoscopic procedures, VS3-IR system includes support for Infrared (IR) Fluorescence visualization (hereafter referred to as Iridium utilizes the VS3 system with scopes that support IR fluorescence visualization. VS3 Iridium Miniature Microscope (VS3-IR-MMS) which was cleared in K150018 is positioned 20cm to 45cm above the patient during an open procedure. The proposed Iridium Endoscopes (VS3-R-Endoscope) is used during laparoscopic minimally invasive procedures. VS3 Iridium MMS and Endoscopes are designed to work with an approved IR fluorescence dve (Indocvanine Green or ICG), which has excitation at 805nm and emission band between 825nm and 850nm. VS3 Iridium (VS3-IR) provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye using an IR camera.

Both the Iridium MMS and Iridium Endoscope utilize the left channel of the VS3 imaging system (camera and image processor) for IR fluorescence imaging and the right channel for visible light imaging. This structure allows fluorescence imaging to use the same acquisition and processing pathways as the approved 3D endoscopes (cleared in K123467, K131434, and K141002).

AI/ML Overview

The provided document describes the acceptance criteria and study for the Visionsense VS3-IR System, specifically focusing on the VS3-IR-Endoscope component.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" with quantitative metrics alongside "reported device performance" in a separate column. Instead, it states that the device "met all performance acceptance criteria" during bench testing. The closest we can get to structured criteria are the functional requirements and the successful detection and visualization described.

Feature/ParameterAcceptance Criteria (Implied)Reported (Achieved) Performance
Able to function with no cameraNot applicable (requires camera for function)No
Working distance2cm – 7cm from tissueVerified to function within 2cm – 7cm from tissue
Direction Of View0°, 30°Verified to support 0°, 30°
Software FunctionalityPerforms as intended per pre-specified requirementsThe system software was validated and performs as intended per the pre-specified requirements.
Detection and VisualizationDetect and visualize anatomy and blood flow per intended usesBench testing verified that the subject device can detect and visualize anatomy and blood flow per the proposed indications for use of the Iridium Module. In all instances, the VS3-IR System using the VS3-IR-Endoscope functioned as intended and met all performance acceptance criteria (including laser excitation triggering IR fluorescence in ICG molecules, and fluorescence light being detected by an image sensor to produce an IR image, fused with a visible light image to produce a composite image showing blood flow).
Standard ConformanceConforms to IEC 60601-1-2, 60601-1, 60601-1-4, and Laser SafetyThe subject device conforms to these recognized standards.

2. Sample Size Used for the Test Set and the Data Provenance

The document describes bench testing, which typically does not involve a "test set" in the sense of clinical cases or data. The testing involved simulating a surgical field.

  • Sample Size: Not applicable in the context of clinical data. The testing was conducted on the device itself and its ability to perform certain functions in a simulated environment.
  • Data Provenance: Not applicable as it's bench testing, not clinical data. The testing was performed by Visionsense Ltd.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Number of Experts: Not mentioned. Bench testing usually relies on engineering and scientific personnel to verify device function against specifications.
  • Qualifications of Experts: Not specified.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. Bench testing for functional verification does not typically involve an adjudication process by human experts in the same way clinical data evaluation would.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. The document describes bench testing of the device's functional performance, not a study of human reader improvement with or without AI assistance. The device is for visualization, not an AI diagnostic tool that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: The study focuses on the standalone performance of the device's imaging capabilities. It verified that the device, without human interpretation for evaluation of its own core function, could detect and visualize fluorescence. However, the device itself is a "human-in-the-loop" tool, as it's an endoscope for surgeons to use for visualization. The bench testing verified the device's ability to produce the images that a human user would then interpret.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: For the bench testing, the "ground truth" was established by the physical and chemical principles of fluorescence (ICG excitation and emission) and the known properties of the simulated surgical field. The expectation was that the device would accurately capture and display this known phenomenon. It's a functional ground truth, verified by the correct operation of the system (laser excitation triggering fluorescence, detection by IR camera, and appropriate image fusion).

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. This document describes the functional testing of a medical device (endoscope), not the development or training of an AI algorithm from a dataset.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable as no training set for an AI algorithm is mentioned or implied.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, formed by three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Visionsense Ltd. % Mr. Raymond Kelly Licensale Incorporated 57 Lazy Brook Road Monroe, Connecticut 06468

October 28, 2015

Re: K152204

Trade/Device Name: VS3-IR Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OWN, GCJ, HRX, GWG Dated: August 5, 2015 Received: August 6, 2015

Dear Mr. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if

{1}------------------------------------------------

applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K152204

Device Name

VS3-IR

Indications for Use (Describe)

The 3D Endoscope Module of the VS3-IR System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures.

The MMS configuration of the Iridium Module of the VS3-IR System is intended for capturing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic. micro- and reconstructive surgical procedures.

The Endoscope configuration of the Iridium Module of the VS3-IR System is intended to provide real-time endoscopic visible and near infrared fluorescence imaging. Iridium Endoscope enables surgeons to perform routine visible light endoscopic procedures as well as further visually assess vessels, blood flow and related tissue perfusion with near infrared imaging during minimally invasive surgery.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Visionsense VS3-IR System

Submitter's Name, Address, Telephone Number,

Visionsense Ltd. 20 Hamagshimim Street, Petach Tikva 49348 Israel Phone: 972-3-9244339 Facsimile: +972-3-9244-335

Contact Person and Date Prepared

Raymond Kelly 57 Lazy Brook Rd Monroe, CT 06468 USA Phone: (203) 880-4091

Date Prepared: August 11, 2015

Name of Device

VS3-IR System

Common or Usual Name / Classification Name

Endoscopes and Accessories / Infrared Microscope / Angiographic X-ray System

Predicate Device

Novadaq Technologies SPY Imaging System (K063345 / K091515)

Intended Use / Indications for Use

The 3D Endoscope Module of the VS3-IR System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, and foraminotomy, and shoulder and knee arthroscopic procedures.

The MMS configuration of the Iridium Module of the VS3-IR System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.

The Endoscope configuration of the Iridium Module of the VS3-IR System is intended to provide real-time endoscopic visible and near infrared fluorescence imaging. Iridium Endoscope enables surgeons to perform routine visible light endoscopic procedures as well as further visually assess yessels, blood flow and related tissue perfusion with near infrared imaging during minimally invasive surgery.

{4}------------------------------------------------

Principles of Operation / Conditions of Use

The VS3-IR System is made up of the following components:

The following table summarizes FDA clearance history for the VS3-IR system:

ComponentFDA Clearance #Indication
VS3 3DEndoscopesK123467 (Laparoscope)K141002 (Arthroscope)K131434 (Neurological)Intended for viewing internal surgicalsites during general surgicalprocedures, for use in visualization ofventricles and structures within thebrain during neurological surgicalprocedures, viewing internal surgicalsites during anterior and posteriorspinal procedures, such asnucleotomy, discectomy, andforaminotomy, and shoulder and kneearthroscopic procedures.
VS3-IR-MMS& ICG KitK150018Intended for capturing and viewingfluorescent images for the visualassessment of blood flow, as anadjunctive method for the evaluationof tissue perfusion, and related tissue-transfer circulation in tissue and freeflaps used in plastic, micro- andreconstructive surgical procedures.
VS3-IR-Endoscope& ICG KitNot yet assignedIntended to provide real-timeendoscopic visible and near infraredfluorescence imaging. IridiumEndoscope enables surgeons toperform routine visible lightendoscopic procedures as well asfurther visually assess vessels, bloodflow and related tissue perfusion withnear infra-red imaging duringminimally invasive surgery.

This premarket notification adds the IR endoscope to the FDA cleared devices in K123467 (Laparoscope), K141002 (Arthroscope), K131434 (Neurological), and K150018 (MMS), and to change the family brand name to VS3-IR system.

{5}------------------------------------------------

The VS3-IR system includes:

The VS3 3D endoscopes (cleared in K123467, K131434, and K141002) are based on the proximal camera concept with a stereoscopic camera block on the proximal side of the endoscope (the handle). The stereoscopic images are transmitted from the visual field at the distal tip of the endoscope to the proximal camera block through an optical transmission system to PC workstations or external display monitors for enhanced viewing or capturing. VS3 includes reusable endoscopes which are cleaned and sterilized using the same methods and are indicated for the same lifecycle.

In addition to traditional endoscopic procedures, VS3-IR system includes support for Infrared (IR) Fluorescence visualization (hereafter referred to as Iridium utilizes the VS3 system with scopes that support IR fluorescence visualization. VS3 Iridium Miniature Microscope (VS3-IR-MMS) which was cleared in K150018 is positioned 20cm to 45cm above the patient during an open procedure. The proposed Iridium Endoscopes (VS3-R-Endoscope) is used during laparoscopic minimally invasive procedures. VS3 Iridium MMS and Endoscopes are designed to work with an approved IR fluorescence dve (Indocvanine Green or ICG), which has excitation at 805nm and emission band between 825nm and 850nm. VS3 Iridium (VS3-IR) provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye using an IR camera.

Both the Iridium MMS and Iridium Endoscope utilize the left channel of the VS3 imaging system (camera and image processor) for IR fluorescence imaging and the right channel for visible light imaging. This structure allows fluorescence imaging to use the same acquisition and processing pathways as the approved 3D endoscopes (cleared in K123467, K131434, and K141002).

{6}------------------------------------------------

Technological Characteristics

Requirements for the Iridium Endoscope assume that the Iridium Scope is used with the Iridium Camera, CCU, and Laser Light source. Stand-alone functionality of the Iridium Scope is not supported.

#Feature/ParameterValue/Description
1Able to function with nocameraNo
2Working distance2cm – 7cm from tissue
3Direction Of View0°, 30°

Performance Data

The subject device conforms to the following recognized standards:

StandardsNo.StandardsOrganizationStandards Title
60601-1-2IECMedical electrical equipment - Part 1-2: Generalrequirements for basic safety and essential performance -collateral standard: electromagnetic compatibility -requirements and tests (Edition 3).
60601-1IECMedical electrical equipment - Part 1: General requirementsfor basic safety and essential performance.
60601-1-4IECMedical electrical equipment - Part 1: General requirementsfor safety, Collateral standard: Programmable electricalmedical systems.
60601-1-4IECSafety of laser products - Part 1: equipment classification,and requirements.

The system software was validated and performs as intended per the pre-specified requirements.

In addition, bench testing was conducted to verify that the subject device can detect and visualize anatomy and blood flow per the proposed indications for use of the Iridium Module.

The Visionsense VS3 system with ICG fluorescence endoscope (VS3 Iridium Endoscope) was bench tested in fluorescence mode so that the endoscope light sources provided both IR (laser light source) and Visible (xenon light source) light to illuminate the simulated surgical field. The left channel of the 3D endoscope was dedicated to visible light imaging, while the right channel of the scope was dedicated to IR imaging during the testing the laser excitation light triggered IR fluoresce in ICG molecules as intended. The fluorescence light was detected by one of the image sensors in the PHD camera to produce an IR image. The other sensor was used to capture a visible light image of the same scene. The two images were fused together to produce a composite image showing blood flow.

In all instances, the VS3-IR System using the VS3-IR-Endoscope functioned as intended and met all performance acceptance criteria.

{7}------------------------------------------------

Substantial Equivalence

The VS3-IR-Endoscope as part of the VS3-IR System is substantially equivalent to other legally marketed surgical endoscopes. Specifically, the VS3-IR-Endoscope is substantially equivalent to the cleared Novadaq SPY Imaging device (K091515). As explained in more detail below, the subject VS3-IR-Endoscope has been added to the cleared VS3 base system and VS3-IR-MMS to create the VS3-IR-Endoscope and has the same indications for use and substantially similar technological characteristics and principles of operation as the identified predicate device. A substantial equivalence chart comparing the similarities and differences between the subject and predicate device is provided below

All the components of the subject device and associated software have been cleared in the original VS3 base system 510(k) notices (K123467, K131434, K141002) and the Infrared Iridium VS3-IR-MMS submissions (K150018) including the ICG Kit (K150018) and are included in this submission by reference for purposes of completeness in describing the updated VS3-IR System that the company plans to market.

VS3-IR IR Endoscope(proposed)Novadaq Endoscopic SPY(K091515)
ManufacturerVisionsense Ltd.Novadaq Technologies Inc
SterilizationSteam AutoclaveSteam Autoclave
Imaging AgentIR fluorescence dye (i.e.Indocyanine Green-ICG)IR fluorescence dye (i.e.Indocyanine Green-ICG)
Imaging HeadSilicon Image Sensor in thecameraSilicon Image Sensor in thecamera
Light SourceInfrared LaserInfrared Laser
Excitation Wavelength805nm805nm
Scope Diameter5.5mm10mm
Angles of View0°, 30°0°, 30°
ImagingFluorescent and White LightImagingFluorescent and White LightImaging
Emission Band825nm to 850nm825nm to 850nm
Emission CaptureIR cameraIR camera
Display both Visibleand IR imagesYesYes

Conclusions

The VS3-IR System is substantially equivalent to the predicate device. Performance testing demonstrates that the newly added components to the VS3-IR-System performs substantially equivalent to the predicate device, and any differences in technological characteristics do not raise different questions of safety or efficacy compared to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.