K Number
K152199

Validate with FDA (Live)

Device Name
RejuvatoneMD
Manufacturer
Date Cleared
2016-03-23

(230 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RejuvatoneMD is intended for facial stimulation and is indicated for over-the-counter cosmetic use.

Device Description

The RejuvatoneMD device is a non-invasive at-home facial stimulation device intended for cosmetic use. The device works by delivering low-level electrical microcurrent impulses through dual contact spheres to strategic locations on the face and allows users to adjust the output level for personalized comfort.

AI/ML Overview

Here's an analysis of the provided text regarding the RejuvatoneMD device's acceptance criteria and studies, organized according to your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in a quantitative format for performance of the device in its intended cosmetic use (e.g., "90% of users report improved skin tone"). Instead, the "acceptance criteria" for this 510(k) submission are based on demonstrating substantial equivalence to a legally marketed predicate device and compliance with relevant safety and manufacturing standards.

The table below summarizes the technical characteristics comparison between the RejuvatoneMD (Subject Device) and its predicate (NuFACE® Mini Device). For performance specifications like maximum output voltage/current, pulse duration, frequency, etc., the "acceptance" is implied by demonstrating values that are similar to and within an acceptable range of the predicate device, and crucially, comply with IEC 60601-2-10 requirements.

Element of ComparisonAcceptance Criteria (Implied by Substantial Equivalence and Standards)Reported RejuvatoneMD Performance
Intended UseFacial stimulation for over-the-counter cosmetic use, equivalent to predicate.Intended for facial stimulation and indicated for over-the-counter cosmetic use.
Electrical SafetyCompliance with IEC/EN 60601-1:2005 Edition 3/(R)2012 And A1:2012.Complies with IEC/EN 60601-1:2005 Edition 3/(R)2012 And A1:2012. Patient Leakage Current: NC 60μA, SFC 360μA. (Predicate: 56μA, 374μA)
Performance StandardCompliance with IEC 60601-2-10 Edition 2.0 2012-06.Complies with IEC 60601-2-10 Edition 2.0 2012-06.
EMCCompliance with IEC 60601-1-2 Edition 3: 2007-03.Complies with IEC 60601-1-2 Edition 3: 2007-03.
BiocompatibilityDirect patient contact materials comply with ISO10993-5 and ISO10993-10.All user directly contacting materials comply with ISO10993-5 and ISO10993-10 requirements.
Usability/Label ComprehensionAt least 95% of participants able to correctly self-select, complete comprehension, and perform device use based on labeling.Reported Performance: Over 95% of 30 participants were able to: Correctly Self-Select the device, Correctly complete the Comprehension portion, and Correctly perform the Device Use portion of the study.
Output CharacteristicsValues for maximum output voltage/current, pulse duration/frequency, net charge, phase charge, current density, power density, and burst mode must be comparable to the predicate and fall within the safety limits defined by IEC 60601-2-10.Maximum Output Voltage: 256mV@500Ω (Predicate: 222mV), 806mV@2KΩ (Predicate: 781mV), 4.02V@10KΩ (Predicate: 3.90V). Maximum Output Current: 512µA @ 500Ω (Predicate: 396µA), 403µA @ 2KΩ (Predicate: 395µA), 403µA @ 10KΩ (Predicate: 391µA). Pulse Duration: ON phase: 60 ms (Predicate: 60.4 ms), OFF phase: 60 ms (Predicate: 60.4 ms), Total Pulse Width: 120 ms (Predicate: 120.8 ms). Pulse Frequency: 8.33Hz (Predicate: 8.28 Hz). Net Charge (per pulse): 0µC @ 500Ω (Predicate: 1.43µC). Maximum Phase Charge: 24.18µC @ 500Ω (Predicate: 23.7µC). Maximum Current Density: 0.498mA/cm²@500Ω (Predicate: 0.514mA/cm²). Maximum Average Power Density: 32.77µW/cm² @500Ω (Predicate: 99.05µW/cm²). Burst Mode (Pulses per burst): 20 (Predicate: 20). Bursts per second: 8.33 (Predicate: 8.28). Burst duration (seconds): 2.4 (Predicate: 2.42). Duty Cycle: 20 (Predicate: 20). All deemed "very close to the predicate device in these areas and within an acceptable range" and comply with IEC 60601-2-10.

2. Sample Size Used for the Test Set and Data Provenance

  • Usability/Label Comprehension Study:
    • Sample Size: 30 participants
    • Data Provenance: Not explicitly stated (e.g., country of origin). The document implies it was a prospective study conducted specifically for this device validation.
  • Other Testing (Electrical Safety, Performance, EMC, Biocompatibility): These tests are typically laboratory-based and use technical samples of the device and its components, not human subjects, except for biocompatibility which is tested on materials. No sample size of specific user data is applicable here. The data provenance is laboratory testing against recognized standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • For the Usability/Label Comprehension Study, the ground truth was based on the participants' ability to correctly self-select, comprehend instructions, and use the device. This "ground truth" is a measure of user understanding and performance, not a clinical diagnosis or expert consensus on clinical outcomes. No external "experts" (like radiologists) are mentioned as establishing ground truth for this usability study. The study itself assesses whether lay users can achieve the intended "truth" of understanding and proper use.
  • For the technical standards compliance (Electrical Safety, Performance, EMC, Biocompatibility), the "ground truth" is the established requirements and limits defined by the international consensus standards (IEC, ISO). The experts involved would be the engineers and technicians performing the tests and the standard-setting bodies, not individual experts assessing specific cases.

4. Adjudication Method for the Test Set

  • For the Usability/Label Comprehension Study, the "adjudication method" was based on whether participants could "correctly" perform specific tasks. This would have been predetermined criteria defined in the study protocol (e.g., "User correctly identifies appropriate use for the device" or "User correctly selects power level"). It's not an expert adjudication in the sense of clinical disagreement.
  • For technical testing, compliance is typically determined by measurements falling within specified ranges or passing defined test procedures. There is no "adjudication" in the clinical sense.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done. This device is a direct-to-consumer microcurrent facial stimulator, not an AI-powered diagnostic imaging tool. Therefore, the concept of "human readers improving with AI vs. without AI assistance" is not applicable.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • A standalone performance was done, but not in the context of an "algorithm." The core of the submission relies on the standalone electrical and mechanical performance of the device as measured against international safety and performance standards (IEC, ISO). The device's technological characteristics are compared directly to the predicate device. The usability study is also a form of standalone performance assessment – how well does a human user interact with the device as designed.

7. The Type of Ground Truth Used

  • Usability/Label Comprehension Study: User performance and comprehension against pre-defined correct actions and understanding gleaned from the labeling.
  • Technical Performance (Electrical Safety, EMC, Biocompatibility): Compliance with international consensus standards (IEC, ISO) and comparison of technical specifications to the legally marketed predicate device. This is primarily objective, measurable physical parameters.
  • Substantial Equivalence: Comparison to the technological characteristics, intended use, and safety profile of a legally marketed predicate device (K133823 – NuFace Mini).

8. The Sample Size for the Training Set

  • Not applicable. This device is hardware for electrical stimulation. There is no "training set" in the machine learning sense. The device's design and parameters are based on established electrical engineering principles and the specifications of predicate devices.

9. How the Ground Truth for the Training Set was Established

  • Not applicable, as there is no training set. The "ground truth" for the device's design parameters is derived from engineering principles, safety standards, and established performance of similar predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be emerging from a single form. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 23, 2016

Trophy Skin, Inc. % Sue Anthoney-Dewet FDA Consultant Aegis Regulatory, Inc. 2424 Dempster Drive Coralville, IA 52241

Re: K152199

Trade/Device Name: RejuvatoneMD Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: February 17, 2016 Received: February 23, 2016

Dear Ms. Anthoney-Dewet:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael J. Hoffmann -A

  • for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152199

Device Name RejuvatoneMD

Indications for Use (Describe)

The RejuvatoneMD is intended for facial stimulation and is indicated for over-the-counter cosmetic use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K152199

This 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 807.92

Date Prepared: March 23, 2016

    1. Submitter Information: AEGIS Regulatory, Inc. – Susan Anthoney-DeWet 2424 Dempster Drive Coralville, IA 52241 Tel.: 865-982-5552 Email: bob@fdalistingconsultants.com

On Behalf of Sponsor: Trophy Skin, Inc.

Attn: Imran Karim, President 4372 Kenmare Trail Frisco, TX 75034 Phone: (469) 233-1768 Email: imran@trophyskin.com

2. General Information

2.1 Classification Name: Transcutaneous Electrical Nerve Stimulator for Aesthetic Purposes

  • 2.2 Common/Usual Name: TENS Device
  • 2.3 Proprietary Names: RejuvatoneMD
  • 2.4 Classification: Class II
  • 2.5 Classification Number: 882.5890
  • 2.6 Review Panel: Neurology
  • 2.6 Product Code: NFO

{4}------------------------------------------------

3. Device Description:

The RejuvatoneMD device is a non-invasive at-home facial stimulation device intended for cosmetic use. The device works by delivering low-level electrical microcurrent impulses through dual contact spheres to strategic locations on the face and allows users to adjust the output level for personalized comfort.

4. Indications / Intended Use:

The RejuvatoneMD is intended for facial stimulation and is indicated for over-the-counter cosmetic use.

Rx or OTC:

The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate devices are indicated for (OTC) use.

5. Predicate Device:

The RejuvatoneMD device is substantially equivalent in design, function and intended use to the following predicate device:

  1. K133823 – NuFace Mini (Carol Cole Company)

6. Technological Characteristics and Substantial Equivalence

The technological characteristics of the RejuvatoneMD device and the predicate device, are substantially equivalent in that both devices are designed to provide electrical stimulation to the facial skin for cosmetic use. The RejuvatoneMD and the predicate device function using the same mode of action by producing microcurrent that is discharged through electrodes that are in contact with the patient's skin, and both have user adjustable settings . The technical characteristics and physical presented in this submission demonstrate that the RejuvatoneMD device is substantially equivalent in design, function and intended use to the predicate device.

Sections below compare the characteristics of the RejuvatoneMD and the predicate device:

{5}------------------------------------------------

Elements of ComparisonSubject DevicePredicate Device
Device Name and ModelRejuvatoneMD, Model: TSTONEMDNuFACE® Mini Device
510(k) NumberApplyingK133823
Product CodeNFONFO
ManufacturerTrophy Skin, Inc.Carol Cole Company
Intended UseThis device is intended for facial stimulation and is indicated for over-the-counter cosmetic use.The NuFACE® Mini Device is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
Power Source(s)9V battery (Type: 6LR61)2 rechargeable AA NiMH batteries
Applied PartType BFType BF
Patient Leakage CurrentNC 60μAComply with IEC 60601-1 and IEC 60601-2-1056μAComply with IEC 60601-1 and IEC 60601-2-10
SFC 360μAComply with IEC 60601-1 and IEC 60601-2-10374μAComply with IEC 60601-1 and IEC 60601-2-10
Average DC current through electrodes when device is on but no pulses are being applied0A--
Number of Output ChannelsOne channelOne channel
Number of Output Modes1 mode1 mode
Output Intensity Level5 steps3 steps
Synchronous or Alternating?N/A - 1 Output ChannelN/A - 1 Output Channel
Method of Channel IsolationN/A - 1 Output ChannelN/A - 1 Output Channel
Regulated Current or Regulated Voltage?BothBoth
Elements of ComparisonSubject DevicePredicate Device
Software/Firmware/Microprocessor Control?YesYes
Automatic Overload TripNot required due to circuit designNot required due to circuit design
Automatic No-Load TripNoYes
Automatic Shut OffYesYes
Patient Override ControlYesYes
Indicator On/Off StatusYesYes
Display Low BatteryNoYes
Voltage/ Current LevelYesYes
Timer RangeYes (20 minutes)Yes (21 minutes)
LCD DisplayNoNo
Compliance with Voluntary StandardsYesComply with IEC 60601-1 and IEC 60601-2-10, IEC 60601-1-2YesComply with IEC 60601-1 and IEC 60601-1-2
BiocompatibilityAll user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements.All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements.
Compliance* with 21 CFR 898YesYes
Main Unit Weight248g816g
Dimensions55.0mm x 35.0mm x 177.2mm (L x W x H)2.5" W x 4.2" L x 1.2" D
Electrode Size0.81cm²--
Housing Materials and ConstructionHousing made from ABS plastic and output contacts made from ACD12 Chrome.Thermo Plastic
Environment for operatingTemperature: 5 ~ 45°Temperature: 5 ~ 40° C
Elements of ComparisonSubject DevicePredicate Device
Humidity: 20 ~ 65% RHHumidity: 20 ~ 65% RH
Environment for storageTemperature: 0 ~ 45°Temperature: 0 ~ 40° C
Humidity: 10 ~ 90% RHHumidity: 10 ~ 90% RH
WaveformPulsed MonophasicPulsed Monophasic
ShapeModulated SquareModulated Square
Maximum OutputVoltage(+/- 10%)256mV@500Ω222mV@ 500Ω
806mV@2KΩ781mV@ 2KΩ
4.02V@10KΩ3.90V@ 10KΩ
Maximum OutputCurrent(+/- 10%)512µA @ 500Ω396µA @ 500Ω
403µA @ 2KΩ395µA @ 2KΩ
403µA @ 10KΩ391µA @10KΩ
Pulse DurationON phase: 60 msON phase: 60.4 ms
OFF phase: 60 msOFF phase: 60.4 ms
Total Pulse Width: 120 msTotal Pulse Width: 120.8ms
Pulse Frequency8.33Hz8.28 Hz
Net Charge (per pulse)0µC @ 500Ω1.43µC @ 500Ω
Maximum Phase Charge24.18µC @ 500Ω23.7µC @ 500Ω
Maximum Current Density0.498mA/cm²@500Ω(Minimum Electrode ContactArea:0.81cm²)0.514mA/cm²@500Ω(Minimum ElectrodeContact Area:0.771cm²)
Maximum Average PowerDensity32.77µW/cm² @500Ω(Minimum Electrode ContactArea:0.81cm²)99.05µW/cm² @500Ω(Minimum ElectrodeContact Area:0.771cm²)
Maximum Power Density2005µW/cm²@10kΩ(MinimumElectrode Contact Area:0.81cm²)1981µW/cm²@10kΩ(MinimumElectrode Contact Area:0.771cm²)
Elements of ComparisonSubject DevicePredicate Device
Burst Mode (i.e., pulse trains)a. Pulses per burst2020
b. Pulses per second8.338.28
c. Burst duration (seconds)2.42.42
d. Duty Cycle[line (b)xline (c)]2020
Contraction and Relaxation TimeAdjustable, due to different modes.Adjustable, due to different modes.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

The RejuvatoneMD has similar technological characteristics with the predicate device in the product design, material, program mode and output waveform.

There are minor differences between the RejuvatoneMD device and the predicate device in the electrode size and subsequently the output characteristics including "Maximum Output Voltage", "Maximum Output Current", "Pulse Duration", "Pulse Frequency", "Net Charge (per pulse)", "Maximum Average Current", "Maximum Phase Charge", "Maximum Current Density", "Maximum Average Power Density", and "Burst Mode (i.e., pulse trains)." The output specification of the RejuvatoneMD device is very close to the predicate device in these areas and within an acceptable range. Both the predicate device and the RejuvatoneMD device comply with IEC 60601-2-10 requirement. So the differences of function specification do not raise any safety or effectiveness issue and does not effect the determination of substantial equivalence.

7. Performance Testing and Standards:

Safety & performance testing demonstrates that the RejuvatoneMD is at least as safe and effective as the legally marketed predicate device.

The RejuvatoneMD device has been tested and conforms to international consensus standards:

ELECTRICAL SAFETY:

Recognition Number 19-4:

  • . IEC/EN 60601-1:2005 Edition 3/(R)2012 And A1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (lec 60601-1:2005, Mod). (General II (ES/EMC))
    PERFORMANCE: Recognition Number 17-11:

{9}------------------------------------------------

  • IEC 60601-2-10 Edition 2.0 2012-06, Medical Electrical Equipment -- Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators. (Neurology)
    EMC:

Recognition Number 19-1:

  • IEC 60601-1-2 Edition 3: 2007-03, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. (General II (ES/EMC))

LABEL VALIDATION:

A Usability/Label Comprehension Study was conducted in order to validate the device labeling with 30 participants.

The results of the final study group found that over 95% of the participants were able to:

  • Correctly Self-Select the device as the correct or incorrect device for themselves
  • Correctly complete the Comprehension portion of the study
  • Correctly perform the Device Use portion of the study

In addition to the testing above, the RejuvatoneMD was designed referencing the following standards and FDA Guidance documents:

USABILITY:

Recognition Number 5-85:

  • . IEC 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for safety – Collateral Standard: Usability Edition 3.0, 2010
    Recognition Number 5-67:

  • . IEC 62366:2007/(R)2013, Medical Devices - Application Of Usability Engineering To Medical Devices. (General I (QS/RM))

RISK MANAGEMENT:

Recognition Number 5-40:

  • ISO 14971:2007, ed.2 + MDD 93/42/EEC Annex I, IEC 60601-1, ed. 3+A1, Medical Devices -. Application Of Risk Management To Medical Devices. (General I (QS/RM))

SOFTWARE DESIGN:

Recognition Number 13-8:

  • . IEC 62304 First Edition 2006-05, Medical Device Software - Software Life Cycle Processes. (Software/Informatics)
  • . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005; and
  • General Principles of Software Validation: Final Guidance for Industry and FDA Staff, January 11, 2002

8. Substantial Equivalence Conclusion

After an analysis of the safety, indications, intended uses, performance features, design materials,

biocompatibility, power output, technological properties, treatment regimes and

{10}------------------------------------------------

methods of operation, the Sponsor believes that no significant differences exist between the device and the predicate device. Minor differences in the technological characteristics of the RejuvatoneMD device and the predicate devices do not raise any issues of safety or effectiveness Therefore substantial equivalency is requested.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).