K Number
K152176

Validate with FDA (Live)

Date Cleared
2015-09-04

(31 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Immunalysis PCP Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 25ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of PCP in human urine with automated clinical chemistry analyzers. This assay is calibrated against PCP. This in-vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or permitting laboratories to establish quality control procedures.

The Immunalysis PCP Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS or Liquid Chromatography/Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The Immunalysis Multi-Drug Calibrators are intended for in vitro diagnostic use for the calibration of assays for the analytes currently listed in the package insert: Benzoylecgonine, Morphine and PCP. The calibrators are designed for prescription use with immunoassays.

Device Description
  1. The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes recombinant antibodies to Phencyclidine. glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes phencyclidine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide as a preservative.
  2. All of the Immunalysis Multi-Drug Calibrators are liquid and ready to use. Each contains a known concentration of a specific drug analyte as a mixture. The negative calibrator is a processed, drug-free synthetic urine matrix with sodium azide as a preservative. The Level 1, 2, 3 and 4 calibrators are prepared by spiking known concentrations of drug analyte into the negative calibrator matrix. These five calibrators (negative, Level 1, 2, 3 and 4) are sold as individual bottles. The concentration of drug analyte in the corresponding calibrators are summarized as follows:

Table 1 Immunalysis Multi-Drug Calibrators
Analyte, Multi-Drug Calibrators, Level 1, Level 2, Level 3, Level 4
Benzoylecgonine, 150ng/mL, 300ng/mL, 500ng/mL, 1000ng/mL
Morphine, 100ng/mL, 300ng/mL, 500ng/mL, 1000ng/mL
PCP, 12.5ng/mL, 25ng/mL, 50ng/mL, 100ng/mL

AI/ML Overview

Acceptance Criteria and Device Performance for Immunalysis PCP Urine Enzyme Immunoassay

This document describes the acceptance criteria and study results for the Immunalysis PCP Urine Enzyme Immunoassay, an in-vitro diagnostic device intended for the qualitative and semi-quantitative analysis of PCP in human urine. All reported testing used a Beckman Coulter AU 400e instrument unless otherwise specified.

1. Table of Acceptance Criteria & Reported Device Performance

Study/TestAcceptance CriteriaReported Device Performance (Summary)
Qualitative Analysis (25 ng/mL cutoff)Correct classification of samples at various concentrations around the cutoff.At -100% to -25% of cutoff (0-19 ng/mL): 80/80 Negative results for each concentration. At cutoff (25 ng/mL): 35 Negative, 45 Positive. At +25% to +100% of cutoff (31-50 ng/mL): 80/80 Positive results for each concentration.
Semi-Quantitative Analysis (25 ng/mL cutoff)Correct classification of samples at various concentrations around the cutoff.At -100% to -25% of cutoff (0-19 ng/mL): 80/80 Negative results for each concentration. At cutoff (25 ng/mL): 30 Negative, 50 Positive. At +25% to +100% of cutoff (31-50 ng/mL): 80/80 Positive results for each concentration.
Specificity and Cross-Reactivity (Qualitative)Correct classification (Positive result at specified concentration) and low cross-reactivity with structurally similar compounds.PCP: 100% cross-reactivity. Other structurally similar compounds showed very low cross-reactivity (0.00250% to 0.71429%).
Specificity and Cross-Reactivity (Semi-Quantitative)Correct classification and low cross-reactivity with structurally similar compounds, with mean values near the cutoff for tested compounds.PCP: Mean Value 26.7 ng/mL, 100% cross-reactivity. Other structurally similar compounds showed very low cross-reactivity (0.00250% to 0.71429%), with mean values near the cutoff for concentrations where a positive result was obtained.
Interference (Structurally Non-Similar Compounds, Endogenous Compounds, pH, Specific Gravity)Assay performance unaffected (correct classification) when potential interferents were spiked into drug-free urine containing target analyte at ±25% of cutoff.All tested structurally non-similar compounds and endogenous compounds showed "No" interference, meaning the -25% cutoff samples were Negative and +25% cutoff samples were Positive. No interference observed for pH values ranging from 3.0 to 11.0. No interference observed for specific gravity values ranging from 1.000 to 1.030.
Interference (Boric Acid)Assay performance unaffected.Interference identified. For a 1% w/v concentration of Boric Acid, the -25% cutoff (19 ng/mL) sample was Negative (No interference), but the +25% cutoff (31 ng/mL) sample was also Negative (Yes interference). This indicates Boric Acid interferes with the assay. Limitations have been added to the labeling.
Linearity/RecoveryAcceptable recovery (%) across a range of expected concentrations.Recovery ranged from 94.7% (at 10 ng/mL expected) to 111.9% (at 70 ng/mL expected). All reported recoveries were within acceptable ranges, typically 90-110%.
Method Comparison (Qualitative)High agreement with LC/MS confirmation.100% agreement: 40 positive and 40 negative results matched LC/MS confirmation.
Method Comparison (Semi-Quantitative)High agreement with LC/MS confirmation.100% agreement: 40 positive and 40 negative results matched LC/MS confirmation.
Calibrator Closed Vial StabilityCalibrator levels within specifications for a specified duration.All calibrator levels (1, 2, 3, and 4) for PCP were within specifications for Day 0, 8, 16, 24, 32, and 40. Supported 12 months initial expiration dating.
Calibrator Open Vial StabilityCalibrator levels within specifications for a specified duration.All calibrator levels (1, 2, 3, and 4) for PCP were within specifications for Day 0, 19, 26, 33, 41, and 60. Supported 60 days initial open vial expiration dating.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision/Cutoff Characterization Study: 80 determinations for each of the 9 concentration levels, totaling 720 determinations.
  • Specificity and Cross-Reactivity Study: Concentrations were tested to a level that would yield a positive result (equivalent to the cutoff). The specific number of determinations per compound is not explicitly stated but implied to be sufficient for result generation.
  • Interference Study (Structurally Non-Similar Compounds, Endogenous Compounds, pH, Specific Gravity, Boric Acid): Not explicitly stated, but for each compound/parameter, measurements were taken at -25% and +25% of the cutoff concentration.
  • Linearity/Recovery: Not explicitly stated, but 11 data points (expected concentrations) were used for the linearity curve.
  • Method Comparison: 80 clinical urine samples (40 positive, 40 negative) were used.
  • Calibrator Stability Studies: Data for Day 0, 8, 16, 24, 32, 40 (closed vial) and Day 0, 19, 26, 33, 41, 60 (open vial) for all 4 calibrator levels.

Data Provenance:

  • Precision/Cutoff Characterization, Specificity, Interference, Linearity/Recovery: These studies appear to be laboratory-based experiments, likely conducted in the US (Immunalysis Corporation is based in Pomona, CA). The data is prospective as it was generated to evaluate the device.
  • Method Comparison: "Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories" were used. This indicates the data is retrospective, and likely from the US, given the company's location.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention the use of human experts to establish ground truth for the test set.

4. Adjudication Method for the Test Set

Not applicable, as human experts were not used to establish ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This is an in-vitro diagnostic assay for qualitative and semi-quantitative analysis of a substance in urine, not for interpretation by human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described (Precision, Specificity, Interference, Linearity, Method Comparison, Calibrator performance) represent standalone performance of the Immunalysis PCP Urine Enzyme Immunoassay device. The device itself is an automated chemical analyzer-based test.

7. The Type of Ground Truth Used

  • Precision/Cutoff Characterization, Specificity, Interference, Linearity/Recovery, Calibrator Stability: Ground truth was established by preparing samples with known, spiked concentrations of PCP or other compounds in drug-free urine.
  • Method Comparison: Ground truth was established by confirmation with Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Mass Spectrometry (LC/MS), which are considered definitive analytical methods.

8. The Sample Size for the Training Set

This document describes a premarket notification for an immunoassay that operates based on established chemical reactions, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. Therefore, the concept of a training set sample size is not applicable to this device.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" for this type of immunoassay.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked one behind the other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

IMMUNALYSIS CORPORATION JOSEPH GINETE REGULATORY AFFAIRS SPECIALIST II 829 TOWNE CENTER DRIVE POMONA CA 91767

September 4, 2015

Re: K152176

Trade/Device Name: Immunalysis PCP Urine Enzyme Immunoassay and Immunalysis Multi-Drug Calibrators Regulation Number: 21 CFR 862.3100 Regulatory Class: Unclassified Product Code: LCM. DKB Dated: August 3, 2015 Received: August 4, 2015

Dear Joseph Ginete:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)K152176

Device Name

Immunalysis PCP Urine Enzyme Immunoassay and Immunalysis Multi-Drug Calibrators

Indications for Use (Describe)

The Immunalysis PCP Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 25ng/mL. The assay is intended for use in laboratories for the qualitative analysis of PCP in human urine with automated clinical chemistry analyzers. This assay is callbrated against PCP. This in-vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or permitting laboratories to establish quality control procedures.

The Immunalysis PCP Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. GC-MS or Liquid Chromatography/Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The Immunalysis Multi-Drug Calibrators are intended for in vitro diagnostic use for the calibration of assays for the analytes currently listed in the package insert: Benzoylecgonine, Morphine and PCP. The calibrators are designed for prescription use with immunoassays.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and sans-serif. The background is a solid red color, and the word is centered on the background.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).

  • A. Contact Information
1.Manufacturer:Immunalysis Corporation
2.Contact Name:Joseph Ginete
3.Contact Title:Regulatory Affairs Specialist II
4.Address:829 Towne Center Drive Pomona, CA 91767
5.Phone:(909) 482-0840
6.Fax:(909) 482-0850
7.Email:jginete@immunalysis.com
8.Summary prepared on:September 3, 2015
B. Device Information
1.Trade Name:Immunalysis PCP Urine Enzyme ImmunoassayImmunalysis Multi-Drug Calibrators
2.Common Name:Immunalysis PCP Urine Enzyme ImmunoassayImmunalysis Multi-Drug Calibrators
C. Regulatory Information
1.510(k) Number:K152176
2.Device Classification:Unclassified
3.Regulation Section:Enzyme Immunoassay, PhencyclidineCFR 862.3200 Clinical Toxicology Calibrator
4.Panel:Toxicology(91)
5.Product Code:LCMDKB

D. Legally Marketed Device to Which We are Claiming Equivalence (807.92(A)(3))

1. Predicate Device:DRI Phencyclidines AssayLZI Multiple Analyte Drugs of Abuse Calibratorsand Controls
2. Predicate Company:Diagnostic Reagents Inc.Lin-Zhi International, Inc.
3. Predicate K Number:K935320K051088

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters against a red background. The letters are bold and slightly blurred, giving them a soft, glowing effect. The red background is solid and uniform, providing a stark contrast to the white text.

  • E. Device Description
    • The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. 1. The antibody/ substrate reagent includes recombinant antibodies to Phencyclidine. glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes phencyclidine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide as a preservative.
      1. All of the Immunalysis Multi-Drug Calibrators are liquid and ready to use. Each contains a known concentration of a specific drug analyte as a mixture. The negative calibrator is a processed, drug-free synthetic urine matrix with sodium azide as a preservative. The Level 1, 2, 3 and 4 calibrators are prepared by spiking known concentrations of drug analyte into the negative calibrator matrix. These five calibrators (negative, Level 1, 2, 3 and 4) are sold as individual bottles. The concentration of drug analyte in the corresponding calibrators are summarized as follows:
Table 1 Immunalysis Multi-Drug Calibrators
AnalyteMulti-Drug Calibrators
Level 1Level 2Level 3Level 4
Benzoylecgonine150ng/mL300ng/mL500ng/mL1000ng/mL
Morphine100ng/mL300ng/mL500ng/mL1000ng/mL
PCP12.5ng/mL25ng/mL50ng/mL100ng/mL
  • F. Intended Use
      1. Immunalysis PCP Urine Enzyme Immunoassay

The Immunalysis PCP Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 25ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of PCP in human urine with automated clinical chemistry analyzers. This assay is calibrated against PCP. This in-vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or permitting laboratories to establish quality control procedures.

The Immunalysis PCP Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS or Liquid Chromatography/Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

    1. Immunalysis Multi-Drug Calibrators
      The Immunalysis Multi-Drug Calibrators are intended for in vitro diagnostic use for the calibration of assays for the analytes currently listed in the package insert: Benzoylecgonine, Morphine and PCP. The calibrators are designed for prescription use with immunoassays.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

  • G. Comparison of the new device with the predicate device
ItemPCP Assay K935320Immunalysis PCP Urine EIA
Intended UseFor the qualitative and semi-quantitative determination of thepresence of PCP in human urine at acutoff of 25ng/mLSame
Type of ProductAnalytical ReagentsSame
Measured AnalytesPCPSame
Test MatrixUrineSame
Cutoff Levels25ng/mL of PCPSame
Test SystemHomogeneous Enzyme ImmunoassaySame
MaterialsLiquid Ready-to-Use Two ReagentAssay (R1 and R2)Antibody/Substrate Reagents andEnzyme Labeled Conjugate
Mass SpectroscopyConfirmationRequired for preliminary positiveanalytical resultsSame
AntibodyMonoclonal antibodies to PCPRecombinant antibody to PCP
Storage2 - 8°C until expiration dateSame
ItemLZI Multiple Analyte K051088Immunalysis Multi-Drug Calibrators
Analytebenzoylecgonine, d-methamphetamine, methadone,morphine, oxazepam, secobarbital,phencyclidine, propoxyphenebenzoylecgonine, morphine,phencyclidine
MatrixUrineSame
Calibrator Levels5 Levels – See Table 2 Below4 Levels (Negative and Level 1, 2, 3 and4) - See Device Description Table 1
Storage2 – 8°C until expiration dateSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a red rectangle with the word "IMMUNALYSIS" in white letters. The letters are bolded and slightly blurred. The red background is a solid color.

  • H. The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis Phencyclidine Enzyme Immunoassay to the predicate
      1. Precision/Cutoff Characterization Study was performed for 20 days, 2 runs per day in duplicate (N=80) on concentration of ±25%, ±50%, ±75%, and ±100% of the cutoff. The study verified that the cutoff serves as a boundary between a negative and positive interpretation of a qualitative result. The instruments used for this was Beckman Coulter AU 400e.

a. The following is a summary table of the Qualitative Analysis for the 25ng/mL cutoff test data results

Table 3 - Qualitative Analysis (for 25ng/mL cutoff)
Concentration(ng/mL)% ofcutoff# ofdeterminationsResult
0-100%8080 Negative
6.25-75%8080 Negative
12.5-50%8080 Negative
19-25%8080 Negative
25Cutoff8035 Negative/45 Positive
31+25%8080 Positive
37.5+50%8080 Positive
43.75+75%8080 Positive
50+100%8080 Positive

b. The following is a summary table of the Semi-Quantitative Analysis for the 25ng/mL cutoff test data results

Table 4 - Semi-Quantitative Analysis (for 25ng/mL cutoff)
Concentration(ng/mL)% of cutoff# ofdeterminationsResult
0-100%8080 Negative
6.25-75%8080 Negative
12.5-50%8080 Negative
19-25%8080 Negative
25Cutoff8030 Negative/50 Positive
31+25%8080 Positive
37.5+50%8080 Positive
43.75+75%8080 Positive
50+100%8080 Positive

{7}------------------------------------------------

IMMUNALYSIS

    1. Specificity and Cross-Reactivity Structurally similar compounds were spiked into drug free urine at levels that will yield a result that is equivalent to the cutoff. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs. The instrument used for this test was a Beckman Coulter AU 400e.
a. The qualitative result summary table for the 25ng/mL cutoff is outlined
below:
Table 5 - Structurally Related Compounds (for 25 ng/mL cutoff) - Qualitative
CompoundConcentration Tested (ng/mL)ResultCross-Reactivity (%)
PCP25Positive100
Chlorpromazine140,000Positive0.01785
Clomipramine350,000Positive0.00714
Cyclobenzaprine25,000Positive0.10000
Dextromethorphan80,000Positive0.03125
Diphenhydramine220,000Positive0.01136
Doxepin90,000Positive0.02777
4 – Hydroxyphencyclidine3,500Positive0.71429
Imipramine200,000Positive0.01250
Methoxetamine36,000Positive0.06944
Thioridazine140,000Positive0.01785
Venlafaxine1,000,000Positive0.00250

b. The semi-quantitative result summary table for the 25ng/mL cutoff is outlined below:

Table 6 - Structurally Related Compounds (for 25ng/mL cutoff) – Semi-Quantitative
CompoundConcentration Tested (ng/mL)Mean ValueCross-Reactivity (%)
PCP2526.7100
Chlorpromazine140,00025.90.01785
Clomipramine350,00024.80.00714
Cyclobenzaprine25,00024.90.10000
Dextromethorphan80,00024.10.03125
Diphenhydramine220,00026.60.01136
Doxepin90,00027.10.02777
4 – Hydroxyphencyclidine3,50025.60.71429
Imipramine200,00026.40.01250
Methoxetamine36,00025.90.06944
Thioridazine140,00025.40.01785
Venlafaxine1,000,00023.90.00250
    1. Interference Structurally non-similar compounds, endogenous compounds, the effect of pH, the effect of specific gravity and boric acid was evaluated by spiking the potential interferent into drug free urine containing the target analyte at ±25% of the cutoff. All potential interferents analyzed verified that assay performance is unaffected by externally ingested compounds or an internally existing physiological condition. The instrument used for this test was a Beckman Coulter AU 400e.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The red background is in the shape of a horizontal arrow.

Table 7 - Structurally Non-Similar Compounds (for 25ng/mL cutoff)
Concentration-25% Cutoff (19ng/mL)+25% Cutoff (31ng/mL)
CompoundTested(ng/mL)ResultInterference?ResultInterference?
Acetaminophen500000NegativeNoPositiveNo
6-Acetylcodeine100000NegativeNoPositiveNo
6-Acetylmorphine100000NegativeNoPositiveNo
Acetylsalicylic Acid500000NegativeNoPositiveNo
Alprazolam100000NegativeNoPositiveNo
7-Aminoclonazepam100000NegativeNoPositiveNo
7-Aminoflurnitrazepam100000NegativeNoPositiveNo
7-Aminonitrazepam100000NegativeNoPositiveNo
Amitriptyline70000NegativeNoPositiveNo
Amobarbital100000NegativeNoPositiveNo
S-(+)-Amphetamine100000NegativeNoPositiveNo
Benzylpiperzine100000NegativeNoPositiveNo
Bromazepam100000NegativeNoPositiveNo
4-bromo 2-5,dimethoxyphenethylamine100000NegativeNoPositiveNo
Buprenorphine100000NegativeNoPositiveNo
Bupropion100000NegativeNoPositiveNo
Butabarbital100000NegativeNoPositiveNo
Butalbital100000NegativeNoPositiveNo
Caffeine500000NegativeNoPositiveNo
Cannabidiol100000NegativeNoPositiveNo
Cannabinol100000NegativeNoPositiveNo
Carbamazepine100000NegativeNoPositiveNo
Carisoprodol100000NegativeNoPositiveNo
Chlordiazepoxide100000NegativeNoPositiveNo
cis-Tramadol100000NegativeNoPositiveNo
Clobazam100000NegativeNoPositiveNo
Clonazepam100000NegativeNoPositiveNo
Clozapine100000NegativeNoPositiveNo
Codeine100000NegativeNoPositiveNo
Cotinine100000NegativeNoPositiveNo
Demoxepam100000NegativeNoPositiveNo
Desalkyflurazepam100000NegativeNoPositiveNo
Desipramine100000NegativeNoPositiveNo
Diazepam100000NegativeNoPositiveNo
Digoxin100000NegativeNoPositiveNo
Dihydrocodeine100000NegativeNoPositiveNo
Dehydronorketamine100000NegativeNoPositiveNo
$\Delta^9$THC100000NegativeNoPositiveNo
Table 7 - Structurally Non-Similar Compounds (for 25ng/mL cutoff)
CompoundConcentrationTested(ng/mL)-25% Cutoff (19ng/mL)Result-25% Cutoff (19ng/mL)Interference?+25% Cutoff (31ng/mL)Result+25% Cutoff (31ng/mL)Interference?
EDDP100000NegativeNoPositiveNo
EDMP100000NegativeNoPositiveNo
1R,2S Ephedrine100000NegativeNoPositiveNo
1S,2R Ephedrine100000NegativeNoPositiveNo
Ethyl-ẞ-D-Glucuronide100000NegativeNoPositiveNo
Ethylmorphine100000NegativeNoPositiveNo
Fenfluramine100000NegativeNoPositiveNo
Fentanyl100000NegativeNoPositiveNo
Flunitrazepam100000NegativeNoPositiveNo
Fluoxetine100000NegativeNoPositiveNo
Flurazepam100000NegativeNoPositiveNo
Haloperidol100000NegativeNoPositiveNo
Heroin100000NegativeNoPositiveNo
Hexobarbital100000NegativeNoPositiveNo
Hydrocodone100000NegativeNoPositiveNo
Hydromorphone100000NegativeNoPositiveNo
11-hydroxy-△' THC100000NegativeNoPositiveNo
Ibuprofen500000NegativeNoPositiveNo
Ketamine100000NegativeNoPositiveNo
Lamotrigine100000NegativeNoPositiveNo
Levorphanol Tartrate100000NegativeNoPositiveNo
Lidocaine100000NegativeNoPositiveNo
Lorazepam100000NegativeNoPositiveNo
Lorazepam Glucuronide50000NegativeNoPositiveNo
Lormetazepam100000NegativeNoPositiveNo
LSD100000NegativeNoPositiveNo
Maprotiline100000NegativeNoPositiveNo
MDA100000NegativeNoPositiveNo
MDEA100000NegativeNoPositiveNo
MDMA100000NegativeNoPositiveNo
Meperidine100000NegativeNoPositiveNo
Meprobamate100000NegativeNoPositiveNo
Methadone100000NegativeNoPositiveNo
Methylphenidate100000NegativeNoPositiveNo
R(-)-Methamphetamine100000NegativeNoPositiveNo
S(+)-Methamphetamine100000NegativeNoPositiveNo
Methaquolone100000NegativeNoPositiveNo
Methylone100000NegativeNoPositiveNo
Midazolam100000NegativeNoPositiveNo
Morphine100000NegativeNoPositiveNo
Morphine-3- ẞ-Glucuronide100000NegativeNoPositiveNo
Morphine-6ẞ-D-Glucuronide50000NegativeNoPositiveNo
Table 7 - Structurally Non-Similar Compounds (for 25ng/mL cutoff)
CompoundConcentrationTested(ng/mL)-25% Cutoff (19ng/mL)Result-25% Cutoff (19ng/mL)Interference?+25% Cutoff (31ng/mL)Result+25% Cutoff (31ng/mL)Interference?
N-Desmethyltapentadol100000NegativeNoPositiveNo
Nalorphine100000NegativeNoPositiveNo
Naloxone100000NegativeNoPositiveNo
Naltrexone100000NegativeNoPositiveNo
Naproxen100000NegativeNoPositiveNo
Nitrazepam100000NegativeNoPositiveNo
11-nor-9-carboxy-Δ9-THC100000NegativeNoPositiveNo
Norbuprenorphine50000NegativeNoPositiveNo
Norcodeine100000NegativeNoPositiveNo
Nordiazepam100000NegativeNoPositiveNo
Norketamine100000NegativeNoPositiveNo
Normorphine100000NegativeNoPositiveNo
Norproxyphene100000NegativeNoPositiveNo
Norpseudoephedrine100000NegativeNoPositiveNo
Nortriptyline100000NegativeNoPositiveNo
Olanzapine100000NegativeNoPositiveNo
Oxazepam100000NegativeNoPositiveNo
Oxycodone100000NegativeNoPositiveNo
Oxymorphone100000NegativeNoPositiveNo
Pentazocine100000NegativeNoPositiveNo
Pentobarbital100000NegativeNoPositiveNo
Phenobarbital100000NegativeNoPositiveNo
Phentermine100000NegativeNoPositiveNo
Phenylephrine100000NegativeNoPositiveNo
Phenylpropanolamine100000NegativeNoPositiveNo
Phenytoin100000NegativeNoPositiveNo
PMA100000NegativeNoPositiveNo
Prazepam100000NegativeNoPositiveNo
Propoxyphene100000NegativeNoPositiveNo
Propranolol100000NegativeNoPositiveNo
Protripyline100000NegativeNoPositiveNo
R,R Pseudoephedrine100000NegativeNoPositiveNo
S,S Pseudoephedrine100000NegativeNoPositiveNo
Ranitidine100000NegativeNoPositiveNo
Ritalinic Acid100000NegativeNoPositiveNo
Salicylic Acid100000NegativeNoPositiveNo
Secobarbital100000NegativeNoPositiveNo
Sertraline100000NegativeNoPositiveNo
Sufentanil Citrate50000NegativeNoPositiveNo
Tapentadol100000NegativeNoPositiveNo
Temazepam100000NegativeNoPositiveNo
Theophylline100000NegativeNoPositiveNo
Table 7 - Structurally Non-Similar Compounds (for 25ng/mL cutoff)
CompoundConcentrationTested(ng/mL)-25% Cutoff (19ng/mL)Result-25% Cutoff (19ng/mL)Interference?+25% Cutoff (31ng/mL)Result+25% Cutoff (31ng/mL)Interference?
Triazolam100000NegativeNoPositiveNo
Trifluoromethylphenyl-piperazine100000NegativeNoPositiveNo
Trimipramine100000NegativeNoPositiveNo
Verapamil60000NegativeNoPositiveNo
Zolpidem Tartrate100000NegativeNoPositiveNo

a. The following is a table of the structurally non-similar compounds for the 25ng/mL cutoff

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters against a red background. The word is written in all capital letters and is centered in the image. The red background appears to be a rectangle with rounded corners.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters against a red background. The word is written in all capital letters and appears to be bolded. The red background is a solid color, and the white letters stand out against it.

b.The following is a table of the endogenous compounds results for the 25ng/mL cutoff

Table 8 - Endogenous Compounds (for 25ng/mL cutoff)
CompoundConcentration Tested (ng/mL)-25% Cutoff (19ng/mL) ResultInterference?+25% Cutoff (31ng/mL) ResultInterference?
Acetone1.0 g/dLNegativeNoPositiveNo
Ascorbic Acid1.5 g/dLNegativeNoPositiveNo
Bilirubin0.002 g/dLNegativeNoPositiveNo
Creatinine0.5 g/dLNegativeNoPositiveNo
Ethanol1.0 g/dLNegativeNoPositiveNo
Galactose0.01 g/dLNegativeNoPositiveNo
y-Globulin0.5 g/dLNegativeNoPositiveNo
Glucose2.0 g/dLNegativeNoPositiveNo
Hemoglobin0.115 g/dLNegativeNoPositiveNo
Human Serum Albumin0.5 g/dLNegativeNoPositiveNo
Oxalic Acid0.1 g/dLNegativeNoPositiveNo
Riboflavin0.0075 g/dLNegativeNoPositiveNo
Sodium Azide1% w/vNegativeNoPositiveNo
Sodium Chloride6.0 g/dLNegativeNoPositiveNo
Sodium Fluoride1% w/vNegativeNoPositiveNo
Urea6.0 g/dLNegativeNoPositiveNo
c. The following is a table of the boric acid for the 25ng/mL cutoff results
Table 9 - Boric Acid (for 25ng/mL cutoff)
CompoundConcentration Tested (ng/mL)-25% Cutoff (19ng/mL) ResultInterference?+25% Cutoff (31ng/mL) ResultInterference?
Boric Acid1% w/vNEGNoNEGYes
d.The following is a table of the boric acid for the 25ng/mL cutoff results
Table 10 - Boric Acid (for 25ng/mL cutoff)
CompoundConcentration Tested-50% Cutoff (13ng/mL) ResultInterference?+50% Cutoff (38ng/mL) ResultInterference?

(ng/mL) 1% w/v

Yes NEG e. Boric Acid interferes with the assay and the limitations have been added to the labeling regarding this compound

No

NEG

Boric Acid

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is in the shape of a horizontal hexagon. The text is bolded and centered within the shape.

Table 11 - Effect of pH (for 25ng/mL cutoff)
Test ParameterValue-25% Cutoff (19ng/mL)+25% Cutoff (31ng/mL)
ResultInterference?ResultInterference?
pH3.0NegativeNoPositiveNo
pH4.0NegativeNoPositiveNo
pH5.0NegativeNoPositiveNo
pH6.0NegativeNoPositiveNo
pH7.0NegativeNoPositiveNo
pH8.0NegativeNoPositiveNo
pH9.0NegativeNoPositiveNo
pH10.0NegativeNoPositiveNo
pH11.0NegativeNoPositiveNo

f. The following is a table of the effect of pH results for the 25ng/mL cutoff

g.The following is a summary table of the effect of specific gravity results for the 25ng/mL cutoff

Table 12 - Effect of Specific Gravity (for 25ng/mL cutoff)
Test ParameterValue-25% Cutoff (19ng/mL)+25% Cutoff (31ng/mL)
ResultInterference?ResultInterference?
Specific Gravity1.000NegativeNoPositiveNo
Specific Gravity1.002NegativeNoPositiveNo
Specific Gravity1.005NegativeNoPositiveNo
Specific Gravity1.010NegativeNoPositiveNo
Specific Gravity1.015NegativeNoPositiveNo
Specific Gravity1.020NegativeNoPositiveNo
Specific Gravity1.025NegativeNoPositiveNo
Specific Gravity1.030NegativeNoPositiveNo
    1. Linearity/ Recovery A drug free urine pool was spiked with high concentration of the target analyte as a high value specimen. Additional pools were made by serially diluting the high value specimen. The instrument used for this test was a Beckman Coulter AU 400e.
      a.The following is a summary table of the linearity/recovery:
Table 13 - Linearity/ Recovery
Expected Concentration (ng/mL)Mean Concentration (ng/mL)Recovery (%)
109.594.7
2019.698.2
2527.0108.1
3032.4108.0
4042.2105.6
5052.4104.9
6064.6107.6
7078.3111.9
8085.4106.8
9094.6105.1

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.

Table 13 - Linearity/ Recovery
Expected Concentration (ng/mL)Mean Concentration (ng/mL)Recovery (%)
100100.8100.8
110110.0100.0
    1. Method Comparison Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories were analyzed with the test device. The study verified that the product performance can be verified by Mass Spectrometry. The instrument used for this test was a Beckman Coulter AU 400e and an Agilent 6430 Liquid Chromatography Tandem Mass Spectrometry.
    • a. The following is a comparison table of qualitative assay performance for the 25ng/mL cutoff:
    • Table 14 Method Comparison for the 25ng/mL Qualitative
LC/MS Confirmation
(+)(-)
TestDevice(+)400
(-)040

b. The following is a summary table of the qualitative assay performance for the 25ng/mL cutoff

Table 15 - Assay Performance verified by LC/MS – 25ng/mL Cutoff
TypePCP ConcentrationAgreement (%)
< 12.5ng/mL12.5 ~ 24 ng/mL25 ~ 37.5 ng/mL> 37.5 ng/mL
Qualitative/ Positive00634100
Qualitative/ Negative36400100

c. The following is a comparison table of semi-quantitative assay performance for the 25ng/mL cutoff

Table 16 - Method Comparison for the 25ng/mL - Semi-Quantitative

LC/MS Confirmation
(+)(-)
Test(+)400
Device(-)040

d.The following is a summary table of semi-quantitative assay performance for the 25ng/mL cutoff

Table 16 - Assay Performance verified by LC/MS – 25ng/mL Cutoff
Type< 12.5ng/mL21.5 ~ 24 ng/mL25 ~ 37.5 ng/mL> 37.5 ng/mLAgreement (%)
Semi-Quantitative/ Positive00634100
Semi-Quantitative / Negative36400100
  1. Calibrator and Control Analytical Performance - Immunalysis Multi-Drug Calibrators

a. Immunalysis Multi-Drug Calibrators Traceability - all components of the calibrators and controls have been traced to a commercially available PCP solution.

b.Immunalysis Multi-Drug Calibrators Closed Vial Stability – A closed vial stability study was performed at 25℃ to establish the initial vial expiration

{14}------------------------------------------------

IMMUNALYSIS

dating. The stability study supported an initial expiration date of 12 months. The instrument used for this test was an Agilent 1200 Series Liquid Chromatograph coupled to Agilent 6410 Tandem Mass Spectrometer. All calibrator levels (1, 2, 3, and 4) for PCP were within specifications for Day 0, 8, 16, 24, 32, and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.

  • a. Immunalysi Multi-Drug Calibrators Open Vial Stability An open vial stability study was performed at 5℃ to establish the initial open vial expiration dating. The stability study supported an initial open vial expiration date of 60 days. The instrument used for this test was an Agilent 1200 Series Liquid Chromatograph coupled to Agilent 6410 Tandem Mass Spectrometer. All calibrator levels (1, 2, 3, and 4) for PCP were within specifications for Day 0, 19, 26, 33, 41, and 60. This stability study was performed to establish initial expiration dating.
  • b.Immunalysis Multi-Drug Calibrators Value Assignment Calibrators are manufactured and are tested by mass spectrometry. If any of the analytes are not of the acceptable range, then the calibrator is adjusted and retested. Values are assigned to the calibrators once the mass spectrometry results are within the acceptable ranges.
  • I. Conclusion

The information provided in this pre-market notification demonstrates that the Immunalysis PCP Urine Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its general intended use.

N/A