ARCHITECT 2nd Generation Testosterone

K152155 · Abbott Laboratories · CDZ · Dec 17, 2015 · Clinical Chemistry

Device Facts

Record IDK152155
Device NameARCHITECT 2nd Generation Testosterone
ApplicantAbbott Laboratories
Product CodeCDZ · Clinical Chemistry
Decision DateDec 17, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1680
Device ClassClass 1

Intended Use

The ARCHITECT 2nd Generation Testosterone assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of testosterone in human serum and plasma. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

Device Story

ARCHITECT 2nd Generation Testosterone is a chemiluminescent microparticle immunoassay (CMIA) for quantitative testosterone measurement in human serum/plasma. Operation: sample, diluent, and anti-testosterone coated paramagnetic microparticles combine; testosterone binds to particles; acridinium-labeled conjugate added; pre-trigger/trigger solutions added. Chemiluminescent reaction measured as relative light units (RLUs) via ARCHITECT i System optics. Inverse relationship between testosterone concentration and RLUs; concentration interpolated from calibration curve. Used in clinical laboratories; operated by trained personnel. Output provides testosterone levels to assist clinicians in diagnosing/treating androgen-related disorders (hypogonadism, puberty disorders, impotence, hirsutism, virilization). Device also supports calculation of Free Testosterone Index (FTI/FAI) using SHBG results to assess free testosterone status.

Clinical Evidence

Clinical study conducted in 2014 to establish FTI/FAI expected values. Study included 656 samples from four cohorts: normal males (21-49 years, n=163; ≥50 years, n=144) and normal females (premenopausal 21-49 years, n=174; postmenopausal ≥50 years, n=175). FTI/FAI calculated as [Total Testosterone] / [SHBG]. Results provided as median and 2.5th/97.5th percentiles for each demographic group. No changes to analytical performance of the underlying assay.

Technological Characteristics

Chemiluminescent microparticle immunoassay (CMIA) technology. Uses anti-testosterone (sheep, monoclonal) coated paramagnetic microparticles and acridinium-labeled conjugate. Automated system (ARCHITECT i System) measures chemiluminescence (RLUs). Quantitative, in vitro diagnostic assay. No changes to materials or sensing principle compared to predicate.

Indications for Use

Indicated for quantitative determination of testosterone in human serum and plasma to aid in diagnosis and treatment of androgen-related disorders. Patient populations include males with primary/secondary hypogonadism, delayed/precocious puberty, or impotence; and females with hirsutism or virilization due to tumors, polycystic ovaries, or adrenogenital syndromes.

Regulatory Classification

Identification

A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized caduceus, which is a symbol of medicine and healthcare. The caduceus is composed of three intertwined lines that form the shape of a bird or a wing. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 17, 2015 ABBOTT LABORATORIES REGINA KUMAR DIAGNOSTICS DIVISION 100 ABBOTT PARK RD ABBOTT PARK IL 60064 Re: K152155 Trade/Device Name: ARCHITECT 2nd Generation Testosterone Regulation Number: 21 CFR 862.1680 Regulation Name: Testosterone test system Regulatory Class: I, Reserved Product Code: CDZ Dated: November 13, 2015 Received: November 16, 2015 Dear Regina Kumar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Katherine Serrano -S For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k152155 Device Name ARCHITECT 2nd Generation Testosterone #### Indications for Use (Describe) The ARCHITECT 2nd Generation Testosterone assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of testosterone in human serum and plasma. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6571 2.34286C14.0476 2.73333 14.0476 3.3619 13.6571 3.75238L3.75238 13.6571C3.3619 14.0476 2.73333 14.0476 2.34286 13.6571C1.95238 13.2667 1.95238 12.6381 2.34286 12.2476L12.2476 2.34286C12.6381 1.95238 13.2667 1.95238 13.6571 2.34286Z" fill="black" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="0.5"></path> <path d="M2.34286 2.34286C1.95238 2.73333 1.95238 3.3619 2.34286 3.75238L12.2476 13.6571C12.6381 14.0476 13.2667 14.0476 13.6571 13.6571C14.0476 13.2667 14.0476 12.6381 13.6571 12.2476L3.75238 2.34286C3.3619 1.95238 2.73333 1.95238 2.34286 2.34286Z" fill="black" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="0.5"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="1.5" stroke="black" stroke-width="0.5" width="15" x="0.5" y="0.5"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## I. 510(k) Summary The summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Date: December 14, 2015 # Applicant Name: Regina Xavier Kumar, Regulatory Affairs Project Manager Abbott Laboratories Diagnostics Division 100 Abbott Park Road Abbott Park, IL 60064 Phone: 224-668-8093 Fax: 224-667-4836 ## Device Name: ### Reagents Classification Name: Radioimmunoassay, Testosterones and Dihydrotestosterones Trade Name: ARCHITECT 2nd Generation Testosterone Governing Regulation: 21 CFR 862.1680 Device Classification: Class I, Reserved Classification Panel: Clinical Chemistry Product Code: CDZ # Legally marketed device to which equivalency is claimed: ARCHITECT 2nd Generation Testosterone (cleared under K120009 on September 11, 2012) # Intended Use/Indications for Use: The ARCHITECT 2nd Generation Testosterone assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of testosterone in human serum and plasma. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism. delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes. {4}------------------------------------------------ #### Description of Device: The ARCHITECT 2nd Generation Testosterone assay is a delayed one-step immunoassay for the quantitative determination of testosterone in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology with flexible assay protocols, referred to as Chemiflex. In the first step, sample, assay specific diluent, and anti-testosterone (sheep, monoclonal) coated paramagnetic microparticles are combined. Testosterone present in the sample binds to the anti-testosterone coated microparticles. After incubation, testosterone acridinium-labeled conjugate is added to the reaction mixture. After further incubation and washing, pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). An inverse relationship exists between the amount of testosterone in the sample and the RLUs detected by the ARCHITECT i System optics. The concentration of testosterone is interpolated from a calibration curve established with calibrators of known testosterone concentration. ### Comparison of Technological Characteristics: Both the submission device (ARCHITECT 2nd Generation Testosterone) and the predicate device (ARCHITECT 2nd Generation Testosterone [K120009]) have the same technological characteristics. Both devices use a chemiluminescent microparticle immunoassay (CMIA) technology for the quantitative determination of testosterone in human serum and plasma. There are no technological differences between the submission and predicate devices. ### Comparison of Assay Performance: The reagents, calibrators, and controls for the submission device and the predicate device are the same. The submission is to update the expected values section of the labeling to include the Free Testosterone Index (FTI) (also known as Free Androgen Index [FAI]). The risk analysis was performed and there is no change in the performance of the previously cleared assay. The performance data provided in K120009 still apply. {5}------------------------------------------------ ## Summary of Supporting Data: The ARCHITECT 2nd Generation Testosterone expected values study was conducted in 2012 and is currently provided in the existing ARCHITECT 2nd Generation Testosterone reagent insert. A second study was conducted in 2014 to calculate FTI or FAI using ARCHITECT 2nd Generation Testosterone (List Number [LN] 2P13) and ARCHITECT SHBG (LN 8K26) assay results. The %FTI or %FAI calculated as [Total Testosterone] / [SHBG] provides an index of free testosterone status. The % FTI or % FAI correlates with the value of free testosterone. # ARCHITECT 2nd Generation Testosterone FTI/FAI Performance Study Testing was performed on a minimum of 120 samples from individuals in the following categories: normal males 21-49 vears of age, normal males > 50 years of age, premenopausal normal females 21-49 years of age, and postmenopausal normal females ≥ 50 years of age not on hormone replacement therapy. The following inclusion/exclusion criteria apply to the study: Samples that were within the expected values of the ARCHITECT 2nd Generation Testosterone (LN 2P13) reagent insert and the ARCHITECT SHBG (LN 8K26) reagent insert were included in the study. {6}------------------------------------------------ | SHBG nmol/L | | | | Testosterone nmol/L [ng/dL] | | | | |---------------------------------------------------|-----|--------|---------------------|-----------------------------|-------------------|---------------------|----------------------| | Category | N | Median | 2.5th<br>Percentile | 97.5th<br>Percentile | Median | 2.5th<br>Percentile | 97.5th<br>Percentile | | Males<br>(21-49 years of age) | 163 | 31.1 | 16.2 | 68.5 | 15.33<br>[442.07] | 8.76<br>[252.73] | 27.85<br>[803.24] | | Males<br>(≥ 50 years of age) | 144 | 35.3 | 13.7 | 69.9 | 14.42<br>[415.85] | 8.58<br>[247.50] | 23.37<br>[674.13] | | Females<br>(Premenopausal,<br>21-49 years of age) | 174 | 48.6 | 14.7 | 122.5 | 1.05<br>[30.43] | 0.52<br>[14.92] | 1.72<br>[49.56] | | Females<br>(Postmenopausal,<br>≥ 50 years of age) | 175 | 49.9 | 16.7 | 124.4 | 0.76<br>[21.83] | 0.46<br>[13.34] | 1.18<br>[33.90] | The data from this study are summarized in the following tables. The % FTI or % FAI values for the different groups are summarized in the following table. | % FTI or % FAI | | | | | | | | | | |---------------------------------------------|-----|---------------|-------------------------|--------------------------|--|--|--|--|--| | Category | N | Median<br>(%) | 2.5th<br>Percentile (%) | 97.5th<br>Percentile (%) | | | | | | | Males (21-49 years of age) | 163 | 46.6 | 24.5 | 113.3 | | | | | | | Males (≥ 50 years of age) | 144 | 40.7 | 19.3 | 118.4 | | | | | | | Females (Premenopausal, 21-49 years of age) | 174 | 2.0 | 0.7 | 8.7 | | | | | | | Females (Postmenopausal, ≥ 50 years of age) | 175 | 1.5 | 0.5 | 4.7 | | | | | | {7}------------------------------------------------ # Conclusion: The submission is for a labeling change for the %FTI or %FAI Expected Values only. The risk analysis was performed and there is no change in the performance of the previously cleared assay. The performance data provided in K120009 still apply. The ARCHITECT 2nd Generation Testosterone assay (LN 2P13) is substantially equivalent to the predicate device because the proposed labeling change does not alter the intended use, indications for use, or the fundamental scientific technology of the predicate device. Furthermore, the residual risk is low because the identified and verified control measures eliminate or reduce the potential effects of failure. Both ARCHITECT 2nd Generation Testosterone and ARCHITECT SHBG (K152185) labeling will be updated to include the %FTI or %FAI expected values generated using these assays.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%