K Number
K152132

Validate with FDA (Live)

Date Cleared
2015-09-25

(56 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The X-spine 30 Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the X-spine®M Pedicle Screw System is intended for posterior, non-cervical (T1-S2/ilium) pedicle spinal fixation to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Device Description

The X-spine™ Pedicle Screw System consists of pedicle screws, rods, cross bar connectors, and associated instruments. Various forms and sizes of these implants are available so that adaptations can be made to take into account the pathology and anatomy of an individual patient. The system components are manufactured from Titanium based alloy which complies with ASTM F136. Alternatively, rods are also offered manufactured from Cobalt Chromium alloy which complies with ASTM F1537. The single use only implants are provided non-sterile, and should not be reused under any circumstances.

AI/ML Overview

This document is a 510(k) summary for the X-spine™ Pedicle Screw System, which is a medical device. This type of document is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device, and thus does not require a full premarket approval application (PMA).

The information provided within this document is typical for a 510(k) submission, which focuses on comparing a new device to existing ones rather than detailing extensive clinical studies or acceptance criteria for independent device performance. Therefore, many of the requested details about acceptance criteria, specific study designs, sample sizes, and ground truth establishment, which are common in standalone clinical studies for novel AI/diagnostic devices, are not explicitly present in this type of regulatory submission.

Here's an attempt to answer your questions based on the provided text, highlighting where information is absent or inferred:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly define "acceptance criteria" in the way a clinical trial for diagnostic performance would. Instead, it focuses on demonstrating substantial equivalence to predicate devices through technological characteristics and mechanical performance. The "performance criteria" are implicitly met if the device is found substantially equivalent.

Acceptance Criterion (Implied for Substantial Equivalence to Predicate)Reported Device Performance
BiocompatibilityConforms to ASTM F136 (Titanium alloy) and ASTM F1537 (Cobalt Chrome alloy), materials with a long history of use and proven biocompatibility, corrosion-resistance, and non-toxicity.
Mechanical PerformanceEngineering evaluation shows that lengthening the screw (extension of the distal portion) has no material change to critical components or locking functionality. Performance is unaffected by modifications, raising no new safety or effectiveness concerns.
Intended Use EquivalenceProposed longer pedicle screws and inclusion of S2/ilium are considered equivalent to predicate devices which already have similar features cleared.
Material EquivalenceUses same or equivalent materials (titanium alloy and/or cobalt chrome alloy) as predicate devices.
Anatomical Region EquivalenceSame anatomical region as predicate devices.
Surgical Approach EquivalenceSame surgical approach as predicate devices.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not applicable in this context. The "test set" here refers to the new device and its components, which underwent engineering evaluations and material testing, not a clinical patient cohort.
  • Data Provenance: The document does not describe specific clinical data or patient data being used in a "test set." The evaluation is based on material specifications (ASTM standards) and engineering analysis, not patient-derived data from a specific country or study type.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Number of Experts & Qualifications: Not applicable. Ground truth, in the sense of expert consensus on diagnostic images or pathology, is not part of this submission. The "ground truth" for the material and mechanical evaluations would be the established material standards (e.g., ASTM) and engineering principles, applied by internal engineers and potentially reviewed by regulatory body experts.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. There is no "test set" in the sense of patient data requiring adjudication of clinical findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This is a submission for a spinal pedicle screw system, a physical implant, not an AI or diagnostic imaging device. Therefore, a MRMC study is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: The "ground truth" is based on:
    • Material Standards: Conformance to established ASTM standards (F136 for titanium alloy, F1537 for cobalt chrome alloy) for biocompatibility.
    • Engineering Principles: Evaluation of structural integrity and functional performance based on engineering analysis of the device's design modifications.
    • Predicate Device Performance: The underlying assumption that the predicate devices are safe and effective, and the new device is substantially equivalent based on these established technological characteristics.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. There is no "training set" in the context of an AI or machine learning model.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable as there is no training set.

In summary: The provided document is a regulatory submission for a physical medical device (pedicle screw system) and primarily relies on demonstrating substantial equivalence to already cleared predicate devices through material properties and mechanical engineering evaluations, rather than clinical efficacy studies with patient data, AI performance assessments, or complex ground truth establishment methods typically associated with diagnostic or AI-powered devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three figures intertwined to represent people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 25, 2015

X-spine Systems, Incorporated David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342

Re: K152132

Trade/Device Name: X-spinesM Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: July 30, 2015 Received: August 4, 2015

Dear Dr. Kirschman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - David Kirschman, M.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K152132

Device Name X-spinesM Pedicle Screw System

Indications for Use (Describe)

The X-spine 30 Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the X-spine®M Pedicle Screw System is intended for posterior, non-cervical (T1-S2/ilium) pedicle spinal fixation to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

X-spine™ Pedicle Screw System

Submitter/Manufacturer:X-spine Systems, Inc.452 Alexandersville Rd.Miamisburg, OH 45342Telephone (937) 847-8400FAX (937) 847-8410Email: dk@X-spine.com
Official Contact:David Kirschman, M.D.Chief Medical Officer
Date Prepared:July 30, 2015
Establishment Registration Number:3005031160
Owner/Operator Number:9063903

II. DEVICE NAME

Trade/Proprietary Name:X-spine Pedicle Screw System
Common Name:Pedicle screw spinal system
Device Class:Class II
Regulation Number:21 CFR §888.3070
Product Codes/Classification Names:
  • ' MNI -- Orthosis, Spinal Pedicle Fixation – Pedicle screw spinal system
  • MNH -- Orthosis, Spondylolisthesis Spinal Fixation Pedicle screw spinal ■ system

Note: Proprietary Names included in the X-spine™ Pedicle Screw System are:

  • . Fortex• Pedicle Screw System
  • . Xpress™ Minimally Invasive Pedicle Screw System

III. PREDICATE DEVICES

  • . Primary Predicate: Lanx Spinal Fixation System [Silverton; Silverton-D] (K122145) This predicate has not been subject to a design related recall. O
  • . Additional Predicate: Fortex Pedicle Screw System (K090224, K120832) [X-spine, Inc.] o This predicate has not been subject to a design related recall.

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The X-spine™ Pedicle Screw System consists of pedicle screws, rods, cross bar connectors, and associated instruments. Various forms and sizes of these implants are available so that adaptations can be made to take into account the pathology and anatomy of an individual patient. The system components are manufactured from Titanium based alloy which complies with ASTM F136. Alternatively, rods are also offered manufactured from Cobalt Chromium alloy which complies with ASTM F1537. The single use only implants are provided non-sterile, and should not be reused under any circumstances.

V. INDICATION FOR USE

The X-spine™ Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the X-spinew Pedicle Screw System is intended for posterior, non-cervical (T1-S2/ilium) pedicle spinal fixation to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The technological principle for both the subject and predicated devices is temporary posterior, non-cervical pedicle spinal fixation in order to provide immobilization of spinal segments in skeletally mature patients, including spinal fusion and including posterior, noncervical (T1-S2/ilium) pedicle spinal fixation.

Specific features of Comparison:

  • The predicate device Lanx System includes longer pedicle screws and S2/ilium in their Intended Use statement.
  • . X-spine proposes the addition of longer pedicle screws and the inclusion of S2/ilium in their Intended Use statement.

{5}------------------------------------------------

At a high level, the subject device and the predicate devices (primary and additional) are based on the following same or equivalent technological elements:

  • FDA Product Codes: MNI and MNH
  • . Implants use the same materials: titanium alloy and/or cobalt chrome alloy.
  • Multiple lengths of pedicle screws to account for variations in patient anatomy.
  • . Rods offered in multiple lengths and configurations to account for variations in patient anatomy.
  • Equivalent Intended uses.
  • . Same anatomical region.
  • Same surgical approach.
  • Mechanical Performance.
  • . Cross-connectors available in fixed and variable designs.

The addition of one more vertebra, S2/ilium, does not pose additional risk, as this same intended use has already been cleared for use in the labeling of the primary predicate device, Lanx Spinal Fixation System [Silverton; Silverton-D] (K122145).

The following minor technological difference of screw sizes exists between the subject and predicate devices. However, the basic range of sizes is equivalent, as shown in the table below.

X-spine ProposedLonger ScrewsPrimary PredicateAdditional Predicate
Lanx Pedicle Screws:,Silverton, Silverton-DCurrent X-spine FortexPedicle Screws
Diameter of screws6.5 mm to 8.25 mm4.5 mm to 8.5 mm4.75 mm to 8.25 mm
Length of screws60 mm to 100 mm25 mm to 100 mm30 mm to 55 mm

Table 5-1: Range of Screw Sizes

X-spine does not propose adding longer screws for the smaller diameters of 4.75 and 5.5.

VII. PERFORMANCE DATA

The following performance data is provided in support of the substantial equivalence determination.

Biocompatibility

Implants of the X-spine Pedicle Screw System are made of titanium alloy. The titanium alloy conforms to ASTM F136 – Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. The titanium material has a long history of use as surgical implants and has been proven to be biocompatible, corrosion-resistant, and not toxic to the biologic environment.

{6}------------------------------------------------

X-spine System rods are also available in an alternate material choice, cobalt chrome alloy. The cobalt chrome alloy conforms to ASTM F1537 – Standard Specification for Wrought Cobalt-28 Chromium-6Molybdenum Alloys for Surgical Implants. The cobalt chrome alloy material has a long history of use as surgical implants and has been proven to be biocompatible, corrosionresistant, and not toxic to the biologic environment.

The products are ultrasonically cleaned using a validated cleaning process.

All instrument pieces that contact patient anatomy are manufactured of medical-grade materials, with long history of use in the medical setting and have been proven to be biocompatible: (medical-grade stainless steel, medical-grade Radel® plastic). The tissue contact is in limitedexposure settings.

Mechanical Performance

A review of the proposed changes shows that a lengthening of the screw (extension of the distal portion of the screw shank) has no material change to the proximal aspect of the screw, no change to the screw cup, no change to the set screw, no change to the locking mechanism, and no change to the locking functionality. There is also no change to the manufacture or finishing process of the system components. Engineering evaluation of the proposed additions to the X-Spine Pedicle Screw System shows that the performance of the system is unaffected by the modifications and raises no new concerns of safety or effectiveness.

VIII. CONCLUSION

Based on a review of the information provided, X-spine finds that the X-spine Pedicle Screw System is substantially equivalent to the referenced predicate device systems.