VITEK 2 AST-Gram Negative Ertapenem (<=0.12 - >=8 ug/mL)

K152075 · bioMerieux, Inc. · LON · Apr 15, 2016 · Microbiology

Device Facts

Record IDK152075
Device NameVITEK 2 AST-Gram Negative Ertapenem (<=0.12 - >=8 ug/mL)
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateApr 15, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK® 2 Gram Negative Ertapenem is designed for antimicrobial susceptbility testing of Gram negative bacilli and is intended for use with the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ertapenem is a quantitative test. Ertapenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Escherichia coli Klebsiella pneumoniae Proteus mirabilis In vitro data available but clinical significance is unknown: Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Enterobacter cloacae Klebsiella oxytoca (excluding ESBL producing isolates) Morganella morganii Proteus vulgaris Providencia rettgeri Providencia stuartii Serratia mar Sarcens The VITEK @ 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

Device Story

VITEK® 2 AST-GN Ertapenem is a diagnostic test card used with VITEK® 2 and VITEK® 2 Compact systems for automated antimicrobial susceptibility testing (AST). The system processes isolated bacterial colonies diluted in saline, which are used to rehydrate antimicrobial media within the card. The device functions as a miniaturized version of the broth microdilution technique. The instrument automatically fills, seals, and incubates the card, monitoring bacterial growth in each well for up to 18 hours. The system generates a report providing the Minimum Inhibitory Concentration (MIC) and interpretive category results. Used in clinical laboratories, the output assists healthcare providers in determining appropriate antimicrobial therapy for patients with bacterial infections. The device benefits patients by providing rapid, standardized susceptibility data to guide clinical decision-making.

Clinical Evidence

Performance evaluated using fresh and stock clinical isolates and challenge strains compared to CLSI broth microdilution reference method (16-20 hours incubation). Testing performed on both VITEK® 2 and VITEK® 2 Compact platforms. Results demonstrated 97.9% overall Essential Agreement and 97.9% overall Category Agreement. Reproducibility and Quality Control testing met acceptance criteria.

Technological Characteristics

Miniaturized broth microdilution card containing antimicrobial concentrations. Automated incubation and optical growth monitoring system. Connectivity via VITEK® 2 or VITEK® 2 Compact instrument platforms. Software-based automated analysis of growth kinetics to determine MIC values.

Indications for Use

Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast using VITEK® 2 systems. Specifically for Ertapenem, indicated for testing Gram-negative bacilli including E. coli, K. pneumoniae, P. mirabilis, and various other Enterobacteriaceae.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 15, 2016 BIOMÉRIEUX, INC CHERECE JONES STAFF REGULATORY AFFAIRS SPECIALIST 595 ANGLUM RD. HAZELWOOD MO 63042 Re: K152075 Trade/Device Name: VITEK® 2 AST - Gram Negative Ertapenem (≤ 0.12 - ≥8 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system Regulatory Class: II Product Code: LON, LTW, LTT Dated: February 29, 2016 Received: March 1, 2016 Dear Ms. Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Ribhi Shawar -S For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K152075 Device Name VITEK® 2 AST-Gram Negative Ertapenem (≤0.12 ->>8 µg/mL) #### Indications for Use (Describe) VITEK® 2 Gram Negative Ertapenem is designed for antimicrobial susceptbility testing of Gram negative bacilli and is intended for use with the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ertapenem is a quantitative test. Ertapenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Escherichia coli Klebsiella pneumoniae Proteus mirabilis In vitro data available but clinical significance is unknown: Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Enterobacter cloacae Klebsiella oxytoca (excluding ESBL producing isolates) Morganella morganii Proteus vulgaris Providencia rettgeri Providencia stuartii Serratia marcescens The VITEK @ 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo features a stylized globe with horizontal lines on the left side and a solid dark color on the right side. A thin, curved line runs vertically through the center of the globe. Below the globe, the word "BIOMÉRIEUX" is written in a sans-serif font, with each letter slightly spaced apart. ## 510(k) SUMMARY ## VITEK® 2 AST-GN Ertapenem #### 510(k) Submission Information: | | Submitter's Name: | bioMérieux, Inc. | |-----------------|----------------------|---------------------------------------------------------------------------------------------------------------------------| | | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | | Contact Person: | Cherece L. Jones<br>Staff Regulatory Affairs Specialist | | | Phone Number: | 314-731-8684 | | | Fax Number: | 314-731-8689 | | | Date of Preparation: | July 24, 2015 | | B. Device Name: | | | | | Formal/Trade Name: | VITEK® 2 AST-Gram Negative Ertapenem<br>(≤0.12-≥8 µg/mL) | | | Classification Name: | 21 CFR 866.1645<br>Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility System<br>Product Code LON | | | Common Name: | VITEK® 2 AST-GN Ertapenem | | | C. Predicate Device: | VITEK® 2 AST-GN Doxycycline (K121546) | ## D. 510(k) Summary: VITEK® 2 Gram Negative Ertapenem is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ertapenem is a quantitative test. Ertapenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Escherichia coli {5}------------------------------------------------ Klebsiella pneumoniae Proteus mirabilis In vitro data available but clinical significance is unknown: Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Enterobacter cloacae Klebsiella oxytoca (excluding ESBL producing isolates) Morganella morganii Proteus vulgaris Providencia rettgeri Providencia stuartii Serratia marcescens The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. The antimicrobial presented in VITEK® 2 AST-GN Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card. VITEK® 2 AST-GN Ertapenem demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009). The Premarket Notification (510[k]) presents data in support of VITEK® 2 AST-GN Ertapenem. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 AST-GN Ertapenem by comparing its performance with the CLSI broth microdilution reference method incubated at 16-20 hrs. The data is representative of {6}------------------------------------------------ performance on both the VITEK®2 and VITEK® 2 Compact instrument platforms. VITEK® 2 AST-GN Ertapenem demonstrated acceptable performance of 97.9% overall Essential Agreement and 97.9% overall Category Agreement with the reference method. Reproducibility and Quality Control demonstrated acceptable results.