Stat Profile Prime ABG Analyzer System

K151982 · Nova Biomedical Corporation · CHL · Sep 10, 2015 · Clinical Chemistry

Device Facts

Record IDK151982
Device NameStat Profile Prime ABG Analyzer System
ApplicantNova Biomedical Corporation
Product CodeCHL · Clinical Chemistry
Decision DateSep 10, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1120
Device ClassClass 2

Intended Use

The Stat Profile Prime ABG Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for the quantitative determination of pH, PCO2, and PO2 in heparinized whole blood. PCO2, PO2, pH: Whole blood measurement of blood gases is used in the diagnosis and treatment of life threatening acid-base disturbances.

Device Story

Stat Profile Prime ABG Analyzer is a small, microprocessor-based blood gas analyzer for laboratory and point-of-care (POC) use. It accepts heparinized whole blood samples (50 uL) from syringes, tubes, or capillary devices via a peristaltic pump. Samples are processed through a replaceable microsensor card containing pH, PCO2, and PO2 sensors. The device uses traditional sensor technology (hydrogen ion-selective for pH, Severinghaus-type for PCO2, and polarographic Clark-type for PO2) to measure analytes. It provides quantitative results for blood gases and calculates parameters like bicarbonate, base excess, and oxygen saturation. Results are displayed on an integrated touch screen for clinician review to guide diagnosis and treatment of acid-base disturbances. The system includes an on-board Quality Management System (QMS) and supports internal/external liquid quality control. Waste is collected in a self-contained bag within the calibrator cartridge.

Clinical Evidence

Clinical POC study conducted at 3 sites (CVICU, MICU, PCU) with 52 personnel (respiratory therapists/nurses). Compared results from POC personnel against trained healthcare professionals using 188 syringe and 127 capillary specimens. Results showed high correlation (r=0.997-1.000) across all parameters. Total imprecision evaluated using external QC materials over 20 runs; %CV for pH was <0.1%, PCO2 1.3-4.7%, and PO2 2.8-5.0%.

Technological Characteristics

Microprocessor-based analyzer; replaceable microsensor card; peristaltic pump with TPE tubing. Sensors: Hydrogen ion-selective (pH), Severinghaus-type (PCO2), polarographic Clark-type (PO2). Connectivity: 1D/2D barcode scanner, thermal printer. User interface: 5.7" VGA color touch screen. Analog board includes amperometric/potentiometric amplifiers, air detection, and temperature control circuitry.

Indications for Use

Indicated for healthcare professionals in clinical labs and point-of-care settings for quantitative determination of pH, PCO2, and PO2 in heparinized whole blood to diagnose and treat life-threatening acid-base disturbances in critically ill patients with metabolic and pulmonary diseases.

Regulatory Classification

Identification

A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with a symbolic representation of a wing or feather extending from the head of the first profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 NOVA BIOMEDICAL CORPORATION PAUL MACDONALD CHIEF QUALITY AND REGULATORY AFFAIRS OFFICER 200 PROSPECT ST. WALTHAM MA 02454 September 10, 2015 Re: K151982 Trade/Device Name: Stat Profile® Prime ABG Analyzer System Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (PCO2, PO2) and blood pH test system Regulatory Class: II Product Code: CHL Dated: July 16, 2015 Received: July 17, 2015 Dear Paul Macdonald: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Katherine Serrano -S For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k151982 Device Name Stat Profile® Prime ABG Analyzer System Indications for Use (Describe) The Stat Profile Prime ABG Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for the quantitative determination of pH, PCO2, and PO2 in heparinized whole blood. PCO2, PO2, pH: Whole blood measurement of blood gases is used in the diagnosis and treatment of life threatening acid-base disturbances. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 | 510(K) Owner: | Nova Biomedical Corporation | |----------------------|---------------------------------------| | Registration Number: | 1219029 | | Address: | 200 Prospect St.<br>Waltham, MA 02454 | | Phone: | 781-894-0800 | | Fax Number: | 784-891-4806 | | Contact Person: | Paul W. MacDonald | | Date Prepared: | August 14, 2015 | Proprietary Name: Stat Profile® Prime ABG Analyzer Common or Usual Name: Blood gases (PCO2, PO2) and blood pH Test System Classification Name: Multiple | Classification Names: | Class No. | Reg. No. | Class | |---------------------------------|-----------|----------|-------| | Blood Gases and Blood pH system | 75CHL | 862.1120 | II | Product Codes: CHL Predicate Device: K142220 - Stat Profile® Prime ABG Analyzer System #### Device Description: The Stat Profile® Prime ABG Analyzer is a small, low cost blood gas analyzer for laboratory and point-of-care use. The sensors and flow path have been integrated into one replaceable microsensor card, which is replaced periodically according to usage. The product, consumables, installation instructions and packaging are designed for easy customer installation. Whole blood specimens are aspirated into the analyzer's microsensor card from syringes, tubes, or capillary blood collection devices using a peristaltic pump and a sampling probe. The disposable microsensor card contains the analytical flow path and the measurement sensors (pH, PCO2, and PO2). Once the analysis measurement is complete, the whole blood specimen is automatically flushed out of the microsensor card flow path and into a self-contained waste collection bag contained within the disposable calibrator cartridge. The Stat Profile Prime ABG Analyzer has an enhanced test menu and multiple quality control options. Both traditional Internal and External liquid QC shall be offered, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly. As with the predicate, the Stat Profile Prime ABG Analyzer is microprocessor-based and incorporates: - traditional sensor technology to measure blood pO2 ● - ion selective electrode technology to measure pH and pCO2 The following items are intended for use on the Stat Profile Prime ABG Analyzer and were previously cleared on the predicate Stat Profile Prime ABG Analyzer System, k142220: The Stat Profile Prime Auto QC Cartridge ABG is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat {4}------------------------------------------------ Profile Prime ABG Analyzer. The Stat Profile Prime Ampule Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime ABG Analyzer. The Stat Profile Prime Calibrator Cartridge ABG is intended for the calibration of pH, PCO2, and PO2 using the Stat Profile Prime ABG Analyzer. The Stat Profile Prime ABG Analyzer accepts Lithium heparin whole blood sample from syringes, open tubes, small cups, and capillary tubes. The minimum sample size for both syringe and capillary samples analysis is 50 uL. #### Measured Parameters: The Stat Profile Prime ABG Analyzer measures pH, PCO2, and PO2. Calculated Parameters: - pH, PCO2, PO2 (corrected to patient temperature) ● - . Bicarbonate level (HCO3-) - Total Carbon Dioxide (TCO2) ● - Base Excess of the blood (BE-b) ● - Base Excess of extracellular fluid (BE-ecf) . - Standard Bicarbonate Concentration (SBC) - Oxygen Content (O2Ct) ● - . Oxygen Capacity (O2Cap) - Alveolar Oxygen (A) - Arterial Alveolar Oxygen Tension Gradient (AaDO2) ● - Arterial Alveolar Oxygen Tension Ratio (a/A) . - Respiratory Index (RI) ● - PO2/FIO2 ratio - Oxygen Saturation (SO2%) ● #### Intended Use: The Stat Profile Prime ABG Analyzer is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for the quantitative determination of pH, PCO2, and PO2 in heparinized whole blood. PCO2, PO2, pH Whole blood measurement of blood gases is used in the diagnosis and treatment of life-threatening acid-base disturbances in critically ill patients with numerous metabolic and pulmonary diseases. #### Summary of the Technological Characteristics: The Stat Profile Prime ABG Analyzer is substantially equivalent to the previously cleared Stat Profile Prime ABG Analyzer System in intended use. It uses the same sensor technology and measurement algorithms for pH, PCO2, and PO2, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters. #### Summary of Performance Testing: Bench testing was previously completed to demonstrate that the Stat Profile Prime ABG Analyzer is substantially equivalent in performance, safety and efficacy in the predicate submission. {5}------------------------------------------------ The bench testing included: - Method Comparison Studies - . Precision/Reproducibility Studies - Run to Run Precision - Within Run Precision . The results of that testing confirmed that the performance of the Stat Profile Prime ABG Analyzer is substantially equivalent to that of the Nova Stat Profile Prime ABG Analyzer System (predicate device). #### Summary of Point-of-Care Testing A Point-of-Care study was conducted to demonstrate that the analyzer was safe and effective for use in the POC setting. The testing compared results obtained by trained Healthcare Professionals to results obtained by POC personnel on the same specimens using the same analyzer. The Stat Profile Prime ABG Analyzer was evaluated by point-of-care (POC) personnel in 3 POC sites including a cardiovascular intensive care unit (CVICU), a medical intensive care unit (MICU) and a pulmonary care unit (PCU). A total of 35 respiratory therapy and 17 Nursing POC personnel participated from the 3 POC settings over the study. The personnel represent trained, qualified staff found in typical POC sites where blood gas analyzers are utilized. All testing was performed using quality control materials or discarded blood gas specimens. Combined method comparison data from all 3 POC settings is summarized in Tables 2-1 and 2-2. | Parameter | Total # specimens | Whole Blood Range | Slope | Intercept | r | |-----------|-------------------|-------------------|-------|-----------|-------| | pH | 188 | 6.785 - 7.767 | 0.997 | 0.018 | 0.999 | | PCO2 mmHg | 188 | 4.3 - 193.0 | 1.001 | 1.063 | 0.998 | | PO2 mmHg | 188 | 11.8 - 755.3 | 1.015 | -0.939 | 1.000 | #### Table 2-1: Prime ABG: POC v Trained Healthcare Professional (THP) - Syringe Mode #### Table 2-2: Prime ABG: POC v Trained Healthcare Professional (THP) - Capillary Mode | Parameter | Total # specimens | Whole Blood Range | Slope | Intercept | r | |-----------|-------------------|-------------------|-------|-----------|-------| | pH | 127 | 6.791 - 7.737 | 0.993 | 0.047 | 0.998 | | PCO2 mmHg | 127 | 3.2 - 192.2 | 0.984 | 1.347 | 0.997 | | PO2 mmHg | 127 | 13.1 - 672.9 | 1.041 | -2.761 | 0.999 | #### Total Imprecision Performance The total imprecision data included in the following table was obtained from different POC site personnel running 3 levels of Stat Profile Prime External Quality Control material (Levels 1-3) in duplicate each day for a total of 20 runs on 3 Stat Profile Prime ABG analyzers. The protocol was based upon methods described in CLSI "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second edition." CLSI EP5-A2. The test data is representative of the expected total imprecision between analyzer performances obtainable by POC personnel using the Stat Profile Prime ABG analyzer using external quality control materials. {6}------------------------------------------------ # Table 2-3: Prime ABG: Total Imprecision Results – Combined – External Controls | Parameter | Pooled<br>Mean | N | Within Run<br>SD (Sr) | Within Run<br>%CV | Total<br>Imprecision<br>SD (St) | Total<br>Imprecision<br>%CV | |-------------------------------------------|----------------|-----|-----------------------|-------------------|---------------------------------|-----------------------------| | Combined Total Imprecision Data - Level 1 | | | | | | | | pH | 7.152 | 120 | 0.004 | 0.056 | 0.005 | 0.070 | | PCO2 | 61.3 | 120 | 0.8 | 1.3 | 1.7 | 2.7 | | PO2 | 64.0 | 120 | 1.6 | 2.5 | 3.2 | 5.0 | | Combined Total Imprecision Data - Level 2 | | | | | | | | pH | 7.367 | 120 | 0.002 | 0.027 | 0.003 | 0.041 | | PCO2 | 41.4 | 120 | 0.4 | 1.0 | 0.6 | 1.5 | | PO2 | 101.7 | 120 | 1.4 | 1.4 | 3.0 | 2.9 | | Combined Total Imprecision Data - Level 3 | | | | | | | | pH | 7.560 | 120 | 0.006 | 0.079 | 0.007 | 0.093 | | PCO2 | 25.5 | 120 | 0.8 | 3.1 | 1.2 | 4.7 | | PO2 | 139.8 | 120 | 1.7 | 1.2 | 4.0 | 2.8 | . and the contraction of the comments of the comments of : : Point-of-Care Study - Stat Profile Prime External Quality Control Materials (N=20 runs) : 上一篇: i {7}------------------------------------------------ | Table 2-4: Comparison of Predicate and Proposed devices<br>Characteristic | K142220 - Stat Profile® Prime ABG Analyzer | Proposed: Stat Profile® Prime ABG Analyzer | |---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication For<br>Use | The Stat Profile Prime ABG Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, PCO2, and PO2 in heparinized whole blood. | The Stat Profile Prime ABG Analyzer is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for the quantitative determination of pH, PCO2, and PO2 in heparinized whole blood. | | | PCO2, PO2, pH - Whole blood measurement of blood gases is used in the diagnosis and treatment of life-threatening acid-base disturbances in critically ill patients with numerous metabolic and pulmonary diseases. | Same | | Acceptable<br>Samples | Lithium heparinized whole blood from syringes, open tubes, small cups, and capillary tubes. | Same | | Sample Volumes | 50µL (syringe and capillary) | Same | | Measurement<br>Range | | | | pH | 6.500-8.000 | Same | | PCO2 | 3.0-200 mmHg | Same | | PO2 | 5-765 mmHg | Same | | Principles of<br>Measurement | | | | pH | Hydrogen ion-selective sensor | Same | | PCO2 | Severinghaus-type sensor | Same | | PO2 | Polarographic Clark-type sensor | Same | | Touch Screen | 5.7" VGA full color display with LED backlight and integrated touch panel | Same | : ડ-ર : {8}------------------------------------------------ | Characteristic | K142220 - Stat Profile® Prime ABG Analyzer | Proposed: Stat Profile® Prime ABG Analyzer | |------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------| | Menu | Fully configurable test menu based on above sensors | Same | | Bar Code Scanner | Internal Integrated 1D/2D | Same | | Printer | 2" Roll, Thermal Transfer | Same | | Pump | Peristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT) | Same | | Analog Board | Precision low level analog front end w/ amperometrically and potentiometric amplifiers, air detector circuitry and temperature control circuitry | Same | # Conclusion: The results of software valication testing confirmed that the Stat Profile Pime ABG Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime AB .
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%