K Number
K151943

Validate with FDA (Live)

Device Name
Vas-Q-Clip
Date Cleared
2016-03-25

(255 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Vas-Q-Clip™ ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

Device Description

Nanova Biomaterials Inc.'s Vas-Q-Clip™ causes hemostasis through vessel ligation. The technological characteristics are the same as or equivalent to the predicate device. The non-absorbable polyacetal material used in the clips is the same as the predicate device and the clips are manufactured in a similar manner.

The clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding and is sold sterile. The method of sterilization is EtO with a SAL of 10-6. The blisters are fitted into a carton in which the clips are sold.

The Vas-O-Clip™ outer structure including legs, bosses, hook and latch mechanism, and hinge are the same as the Hem-o-lok® design (Shape and dimensions are the same for the corresponding sizes). The only difference is the inner closure feature designated as "teeth". Vas-Q-Clip™ triangular protrusions (teeth) are perpendicular to the vessel whereas the Hem-O-Lok teeth are parallel.

AI/ML Overview

The provided document is a 510(k) summary for the Nanova Biomaterials Inc.'s Vas-Q-Clip. This document outlines the device's characteristics and its substantial equivalence to a predicate device, rather than describing a study to prove acceptance criteria for a new AI/imaging device. Therefore, much of the requested information regarding AI/imaging study methodology (such as expert adjudication, MRMC studies, training set details, and specific performance metrics) is not applicable or present in this document.

However, I can extract information related to the device's non-clinical performance testing and general acceptance criteria based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

The document lists "Tests performed on Vas-Q-Clip™" and their results, which can be interpreted as the device meeting predefined (though not explicitly stated numerical) acceptance criteria.

Acceptance Criteria (Test Name)Reported Device Performance
Migration ResistancePassed
Vessel OcclusionPassed
Application ForcePassed
Axial Pull-Off ForcePassed
USP 85 Bacterial Endotoxins TestPassed
USP 151 Rabbit Pyrogenicity TestPassed
Material Comparison TestPassed
ISO 10993-10 Skin SensitizationPassed
ISO 10993-6 Local Effects After ImplantationPassed
Visual Inspection of SealPassed
Bubble EmissionPassed
Dye PenetrationPassed
Peel-off Seal StrengthPassed
Accelerated and Real Time AgingPassed
SterilityPassed

2. Sample size used for the test set and the data provenance

The document states, "Non-clinical testing was completed to assess its performance." It refers to "Internal Testing" for several mechanical properties (Migration Resistance, Vessel Occlusion, Application Force, Axial Pull-Off Force). However, no specific sample sizes or data provenance (country of origin, retrospective/prospective) are mentioned for these tests. The document does not describe imaging data or a test set in the conventional sense for AI.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This document describes non-clinical performance and biocompatibility testing for a medical device (ligating clip), not an AI/imaging device that requires expert ground truth for interpretation.

4. Adjudication method for the test set

Not applicable. There is no mention of adjudication, as this is not an imaging study requiring human interpretation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device, and no MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used

For the non-clinical tests, the "ground truth" is defined by established laboratory testing procedures (e.g., USP standards for endotoxins and pyrogenicity, ISO standards for biocompatibility, and unspecified internal test methods for mechanical properties). The "Passed" results indicate adherence to the requirements of these test methods.

8. The sample size for the training set

Not applicable. This pertains to an AI/machine learning model, which is not the subject of this document.

9. How the ground truth for the training set was established

Not applicable. This pertains to an AI/machine learning model, which is not the subject of this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2016

Nanova Biomaterials, Inc. Mr. Andrew Ritts Senior Research Scientist 3806 Mojave Ct. Columbia. MO 65202

Re: K151943

Trade/Device Name: Vas-Q-Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: February 23, 2016 Received: February 26, 2016

Dear Mr. Ritts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

  • Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151943

Device Name Vas-Q-Clip^TM

Indications for Use (Describe)

Vas-Q-Clip' TM ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "NANOVA" in blue, block letters. The letters are evenly spaced and appear to be a sans-serif font. The background is white, providing a clear contrast to the blue text.

Section 5. 510(k) Summary

    1. Submitted By: Nanova Biomaterials, Inc 3806 Mojave Ct Columbia, MO 65202 USA 573-875-6682 Contact Person: Andrew Ritts Phone: (573) 823-3114 Secondary Contact: Richard Lebens Phone: (573) 875-6682 2) Establishment Registration No.: 3011430871 3) Date Prepared: December 14, 2015 Vas-Q-Clip™ 4) Device Trade Name: ર) Device Common Name: Ligation Clip Implantable Clip/ 878.4300/ FZP 6) Device Classification Name/ Regulation #/Product Code: Classification Panel: 7) General and Plastic Surgery Device Class: Class II 8)
    1. Predicated Devices: Vas-Q-Clip™ is believed to be substantially equivalent to Hem-O-lok® Ligating Clip (K993157).
    1. Indication for Use: Vas-Q-Clip™ ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "NANOVA" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The overall impression is clean and modern.

Section 5. 510(k) Summary - Cont.

    1. Device Description:
      Nanova Biomaterials Inc.'s Vas-Q-Clip™ causes hemostasis through vessel ligation. The technological characteristics are the same as or equivalent to the predicate device. The non-absorbable polyacetal material used in the clips is the same as the predicate device and the clips are manufactured in a similar manner.

The clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding and is sold sterile. The method of sterilization is EtO with a SAL of 10-6. The blisters are fitted into a carton in which the clips are sold.

    1. Substantial Equivalence:
      The document, "Guidance on the CDRH Premarket Notification Review Program, 6/30/86 (K86-3)" was used to determine substantial equivalence:

a) The applicant device has the same intended use as the 510(k) cleared predicate listed above.

b) The technological characteristics of this product are believed to be substantially equivalent as those for the predicate device. This device and its predicates are substantially equivalent in material and manufacturing methods. Vas-Q-Clip™ is packaged in a similar manner to its predicate and many 510k cleared products already on the market. A summary of the comparison is shown in Table 5.1 below.

The Vas-O-Clip™ outer structure including legs, bosses, hook and latch mechanism, and hinge are the same as the Hem-o-lok® design (Shape and dimensions are the same for the corresponding sizes). The only difference is the inner closure feature designated as "teeth". Vas-Q-Clip"M triangular protrusions (teeth) are perpendicular to the vessel whereas the Hem-O-Lok teeth are parallel.

Vas-O-Clip™ has the same Intended Use as the predicate devices to which it was compared, and there are no differences in technological characteristics which raise new questions of safety and / or effectiveness. Therefore, Vas-Q-Clip™ is substantially equivalent.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "NANOVA" in blue, block letters. The letters are bold and sans-serif. The word is centered and takes up most of the frame. The background is white.

NameVas-Q-ClipTMHem-O-Lok® (K993157).Comparison
Indications for useVas-Q-ClipTM ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.Hem-O-Lok® ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.Same
CompositionPolyacetalPolyacetalSame
ProcessingInjection MoldingInjection MoldingSame
SterilityEtO at 10-6 SALEtO at 10-6 SALSame
DesignTriangular protrusions are perpendicular to the vesselTriangular protrusion are parallel to the vesselDifferent
SizeML, L, XLM, ML, L, XLNo M size
Migration ResistanceInternal TestingInternal TestingHigher strength
Vessel OcclusionInternal TestingInternal TestingHigher pressure
Application ForceInternal TestingInternal TestingHigher strength
Axial Pull-Off ForceInternal TestingInternal TestingHigher strength
PackagingThe clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding.The clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding.Same
Table 5.1. Technical Comparison of Vas-Q-Clip™ and Hem-O-Lok® (K993157).
----------------------------------------------------------------------------------

13) Non-Clinical Performance Testing:

Non-clinical testing was completed to assess its performance. Table 5.2 shows the list of tests performed. The data provided in this 510(k) submission shows that the composition is safe based on the biocompatibility risk assessment conducted and a benchtop assessment.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "NANOVA" in blue, block letters. The letters are slightly rounded and have a modern, sans-serif appearance. The color of the text is a vibrant blue, which stands out against the white background.

Table 5.2. Tests performed on Vas-Q-Clip™.
NameResult
Migration ResistancePassed
Vessel OcclusionPassed
Application ForcePassed
Axial Pull-Off ForcePassed
USP 85 Bacterial Endoxins TestPassed
USP 151 Rabbit Pyrogenicity TestPassed
Material Comparison TestPassed
ISO 10993-10 Skin SensitizationPassed
ISO 10993-6 Local Effects After ImplantationPassed
Visual Inspection of SealPassed
Bubble EmissionPassed
Dye PenetrationPassed
Peel-off Seal StrengthPassed
Accelerated and Real Time AgingPassed
SterilityPassed
  1. Clinical Performance Testing: Clinical performance data was not included.

Conclusion:

Nanova Biomaterials Inc. believes that Vas-Q-Clip™ is substantially equivalent to currently legally marketed products. It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or safety risks.

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.