K Number
K151918

Validate with FDA (Live)

Date Cleared
2016-02-11

(213 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Spherical:
The Bausch + Lomb Lomb Biotrue® ONEday (nesofilcon A) Contact Lens with NaturMoist is indicated for the daily wear correction of refractive ametropia (myopia, astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism up to 2.00 diopters or less, that does not interfere with visual acuity. The lens is to be prescribed in spherical powers ranging from +20.00D to -20.00D.

Multifocal:
The Bausch + Lomb Biotrue® ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D.

Toric:
The Bausch + Lomb Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters for daily wear.

The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.

Device Description

The Biotrue ONEday Contact Lens with NaturMoist is a soft hydrophilic contact lens. The lenses are made from the nesofilcon A material, a hydrophilic copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone, and is 78% water by weight when immersed in a sterile borate buffered saline solution with poloxamine and sodium hyaluronate. A UV-absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280nm to 315nm and less than 50% in the UVA range of 316nm to 380mm. This lens is tinted blue with Reactive Blue Dye 246. The color additive conforms to 21 CFR Part 73.3106.

AI/ML Overview

The provided document is a 510(k) premarket notification for contact lenses. It does not contain information about the acceptance criteria or a study proving that an AI-powered device meets specific acceptance criteria.

The document discusses the substantial equivalence of the "Biotrue ONEday Contact Lenses with NaturMoist" to its predicate devices, highlighting that no clinical studies were required to demonstrate the safety or effectiveness of the subject device because its technological characteristics, formulation, manufacturing, and sterilization processes are the same as the predicate device.

Therefore, I cannot provide the requested information regarding an AI device's acceptance criteria and associated studies from this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 11, 2016

Bausch & Lomb, Inc. Ms. Barbara Klube-Falso Manager, Regulatory Affairs 1400 North Goodman Street Rochester. NY 14609

Re: K151918

Trade/Device Name: Biotrue Oneday with NaturMoist; Biotrue Oneday for Presbyopia with NaturMoist; Biotrue Oneday for Astigmatism with NaturMoist Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: January 4, 2016 Received: January 5, 2016

Dear Ms. Klube-Falso:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"

(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151918

Device Name

Biotrue ONEday Contact Lenses with NaturMoist

Indications for Use (Describe)

Spherical:

The Bausch + Lomb Lomb Biotrue® ONEday (nesofilcon A) Contact Lens with NaturMoist is indicated for the daily wear correction of refractive ametropia (myopia, astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism up to 2.00 diopters or less, that does not interfere with visual acuity. The lens is to be prescribed in spherical powers ranging from +20.00D to -20.00D.

Multifocal:

The Bausch + Lomb Biotrue® ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D.

Toric:

The Bausch + Lomb Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters for daily wear.

The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date PreparedJuly 10, 2015
ApplicantBausch & Lomb Incorporated1400 North Goodman Street, Rochester, New York 14609585.338.8503
ContactBarbara Klube-Falso
Trade namesBiotrue ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lens withNaturMoistBiotrue ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) ContactLens with NaturMoistBiotrue ONEday for Astigmatism (nesofilcon A) Soft (Hydrophilic)Contact Lens with NaturMoist
Common namesoft contact lenses (hydrophilic)
ClassificationClass II (21 CFR 886.5925 (b) (1))
Predicate DevicesBiotrue ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lenses clearedunder K113703,Biotrue ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) ContactLenses cleared under K132715, andBiotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic)Contact Lenses cleared under K143632

Device Description:

The Biotrue ONEday Contact Lens with NaturMoist is a soft hydrophilic contact lens. The lenses are made from the nesofilcon A material, a hydrophilic copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone, and is 78% water by weight when immersed in a sterile borate buffered saline solution with poloxamine and sodium hyaluronate. A UV-absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280nm to 315nm and less than 50% in the UVA range of 316nm to 380mm. This lens is tinted blue with Reactive Blue Dye 246. The color additive conforms to 21 CFR Part 73.3106.

The physical / optical properties of the lens are:

Specific Gravity:1.039
Refractive Index:1.374
Light Transmittance:99%
Water Content:78%
Oxygen Permeability:42 x 10-11 [cm3O2(STP) x cm]/(sec x cm2 x mmHg)@35°C(polarographic method)

{4}------------------------------------------------

The lenses will be manufactured in the following parameter ranges:

Diameter:13.5mm to 15.5mm
Center Thickness:0.05mm to 0.75mm (varies with power)
Base Curve:7.8mm to 9.5mm
Power Range:+20.00D to -20.00D
Add Power:+0.75D to +5.00D
Cylinder Powers:-0.75D to -5.00D
Cylinder Axis:0º to 180º

Indications for Use:

Bioture ONEday Contact Lenses with NaturMoist

Spherical:

The Bausch + Lomb Biotrue® ONEday (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes that exhibit refractive astigmatism up to 2.00 diopters or less, that does not interfere with visual acuity. The lens is to be prescribed in spherical powers ranging from +20.00D to -20.00D.

Multifocal:

The Bausch + Lomb Biotrue® ONEday for Presbyopia (nesofilcon A) Soft (Hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters with add power ranging from +0.75D to +5.00D.

Toric:

The Bausch + Lomb Biotrue ONEday for Astigmatism (nesofilcon A) Soft (hydrophilic) Contact Lens with NaturMoist is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of up to 5.00 diopters, that does not interfere with visual acuity. The lens provides a power range of +20.00 to -20.00 diopters for daily wear.

The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.

{5}------------------------------------------------

Technological Characteristics

PropertyBiotrue ONEdaySVS(predicate)Biotrue ONEdayMulti-Focal(predicate)Biotrue ONEdayToric(predicate)Biotrue ONEdayContact Lenseswith NaturMoist(Subject Device)
FunctionalityThe contact lens actsas a refractivemedium that focuslight rays from nearand distant objectson the retina.The contact lensacts as a refractivemedium that focuslight rays from nearand distant objectson the retina.The contact lens actsas a refractivemedium that focuslight rays from nearand distant objectson the retina.Same as predicate
ModalityDaily wear contactlensDaily wear contactlensDaily wear contactlensSame as predicate
ManufacturingMethodCast MoldCast MoldCast MoldSame as predicate
MaterialGroupGroup II(high water, no ionicpolymers)Group II(high water, no ionicpolymers)Group II(high water, no ionicpolymers)Same as predicate
USAN NameNesofilcon ANesofilcon ANesofilcon ASame as predicate
Water Content78%78%78%Same as predicate
UV BlockeryesyesyesSame as predicate
SterilizationAir over steamAir over steamAir over steamSame as predicate
Physical/optical propertiesSpecific Gravity:1.039Refractive index:1.374Surface Character:HydrophilicSpecific Gravity:1.039Refractive index:1.374Surface Character:HydrophilicSpecific Gravity:1.039Refractive index:1.374Surface Character:HydrophilicSame as predicate
OxygenPermeability$42 x 10^{-11*}$$42 x 10^{-11*}$$42 x 10^{-11*}$Same as predicate
LightTransmittance99%99%99%Same as predicate
Color AdditiveReactive Blue Dye246Reactive Blue Dye246Reactive Blue Dye246Same as predicate
Lens DesignSphericalMulti-FocalToricSame as predicates
PackagingPolypropyleneblister with plasticcoated aluminumfoil blisterPolypropyleneblister with plasticcoated aluminumfoil blisterPolypropyleneblister with plasticcoated aluminumfoil blisterSame as predicate

Bausch & Lomb Incorporated

{6}------------------------------------------------

Difference
PackagingSolutionBorate BufferedSaline withpoloxamineBorate BufferedSaline withpoloxamineBorate BufferedSaline withpoloxamineBorate BufferedSaline withpoloxamine andsodiumhyaluronate

*[cm302(STP) x cm]/(sec x cm² x mmHg)(@35°C (polarographic method)

Summary of Non-Clinical Performance Data

Non-Clinical Tests: Testing conducted followed FDA guidance, FDA Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, May 1994. The testing performed demonstrated that the devices continue to function in a safe and effective manner. Performance testing included conformance to predetermined specifications, functional test results verify that the device performs as expected and is equivalent to the predicate without creating additional risk to the user.

Clinical Tests: The technological characteristics, formulation, manufacturing and sterilization processes are the same as the predicate device, therefore, no clinical studies were required to demonstrate the safety or effectiveness of the subject device.

Substantial Equivalence:

The Biotrue ONEday Contact Lenses with NaturMoist are substantially equivalent to the predicate devices as they have the same functional and scientific technology, lens characteristics and the intended use is identical to that of the predicate device.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.