K Number
K151876

Validate with FDA (Live)

Date Cleared
2015-12-30

(174 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bio2 Fusion Implant System is indicated for fracture fixation, osteotomies, and inter-digital fusion of the fingers, toes and small bones in the presence of appropriate immobilization.

When used in the foot (including hammertoe, claw toe, and mallet toe) patients should use protected weight bearing or heel bearing until fusion or healing has occurred.

Device Description

The Bio2 Technologies Fusion Implant is a cylindrical shaped implant designed for fracture fixation repair, osteotomy and to fit into the proximal and distal sides of the proximal interphalangeal joints of the foot or hand. The implants are made of bioactive glass material that has been demonstrated to be biocompatible and osteoconductive. The implants are available in multiple diameters, lengths and in a straight and angled configuration. The implants are provided sterile and are intended for single use.

AI/ML Overview

The provided text describes the Bio2 Fusion Implant System, a medical device intended for fracture fixation, osteotomies, and inter-digital fusion of fingers, toes, and small bones. The document is a 510(k) premarket notification from the FDA, aiming to demonstrate substantial equivalence to previously marketed predicate devices.

However, the document does not contain the kind of detailed information requested about the acceptance criteria and the study that proves the device meets those criteria, specifically for an AI/CADe device. The Bio2 Fusion Implant System is a physical implant, not a software or AI-based diagnostic tool.

Therefore, for this specific input, the following requested information cannot be provided as it is not present in the document:

  1. A table of acceptance criteria and the reported device performance: Not applicable as this is a physical implant, and the performance testing described is mechanical, not related to AI metrics.
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for an AI test set. The document mentions "performance testing" of the implant but doesn't detail sample sizes or data provenance in the context of AI evaluation.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable for a physical implant; ground truth establishment as it pertains to AI/diagnostic accuracy is not mentioned.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted diagnostic device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for an AI device. The ground truth for biocompatibility and mechanical properties would be in vitro/in vivo assays and mechanical testing results.
  8. The sample size for the training set: Not applicable, as there is no AI algorithm being trained.
  9. How the ground truth for the training set was established: Not applicable.

Information that is available in the document regarding traditional device testing:

Performance Data Mentioned:
The document states: "Performance testing supports the substantial equivalence of the subject device to the predicate device."

  • Biocompatibility: Established according to blue book memorandum #G95-1 entitled "Use of International Standard ISO-10993. Biological Evaluation of Medical Devices Part-1: Evaluation and Testing."
  • In vitro bioactivity testing: (ISO 23317:2012) showed the material forms a surface apatite layer when submerged in simulated body fluid.
  • Mechanical Testing: "Static and fatigue bending tests and in vitro stability tests were conducted on the subject and predicate devices to demonstrate substantial equivalence."

Type of Ground Truth Used (for a physical device):
For biocompatibility, the ground truth would be established through standard biological evaluation methods and outcomes per ISO-10993. For bioactivity, the formation of a surface apatite layer in simulated body fluid per ISO 23317:2012 is the ground truth. For mechanical performance, established engineering standards for static, fatigue, and stability tests serve as the ground truth.

Sample Sizes for these physical tests are not specified in this summary document.
Data Provenance (for physical device testing): Not specified beyond the tests being conducted on "the subject and predicate devices."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 30, 2015

Bio2 Technologies, Incorporated Janet Krevolin, Ph.D. Chief Technical Officer 12-R Cabot Road Woburn, Massachusetts 01801

Re: K151876

Trade/Device Name: Bio2 Fusion Implant System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: November 16, 2015 Received: November 17, 2015

Dear Dr. Krevolin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K151876

Device Name

Bio2 Fusion Implant System

Indications for Use (Describe)

The Bio2 Fusion Implant System is indicated for fracture fixation, osteotomies, and inter-digital fusion of the fingers, toes and small bones in the presence of appropriate immobilization.

When used in the foot (including hammertoe, claw toe, and mallet toe) patients should use protected weight bearing or heel bearing until fusion or healing has occurred.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

510(k) Summary

Summary of 510(k) safety and effectiveness information upon which the substantial equivalence determination is based:

  • I. Submitter
Date Prepared:Device Submitter:December 21, 2015Bio2 Technologies12-R Cabot RoadWoburn, MA 01801
Phone:781-721-6309
Contact Person:Janet Krevolin, PhD
II.Device
Device Name:Bio2 Fusion Implant System
Common Name:Pin, fixation, smooth
Classification Name:Smooth or threaded metallic bone fixation fastener.
21CFR 888.3040
Regulatory Class:II
Product Code:HTY
III.Predicate Device
Predicate Device:Arthrex, Bio-Pin K050259
Bio2, Bio2 CLM•BG Bioactive Scaffold K142463
Xtremi-T, ReFIX Internal Fixation Pins K043339
Wright, Medical Phalinx K142585
  • IV. Device Description
    The Bio2 Technologies Fusion Implant is a cylindrical shaped implant designed for fracture fixation repair, osteotomy and to fit into the proximal and distal sides of the proximal interphalangeal joints of the foot or hand. The implants are made of bioactive glass material that has been demonstrated to be biocompatible and osteoconductive. The implants are available in multiple diameters, lengths and in a straight and angled configuration. The implants are provided sterile and are intended for single use.

  • V. Indications for Use
    The Bio2 Fusion Implant System is indicated for fracture fixation, osteotomies, and interdigital fusion of the fingers, toes and small bones in the presence of appropriate immobilization.

{4}------------------------------------------------

When used in the foot (including hammertoe, claw toe, and mallet toe) patients should use protected weight bearing until fusion or healing has occurred.

  • VI. Comparison of technological characteristics with the predicate device
    Bio2 Fusion Implant is a cylindrical shaped implant available in multiple sizes and in straight and angled versions. The implants are dimensionally within the range of the predicate devices. The implants are made of bioactive glass material. The resorbable material has been shown to be biocompatible and osteoconductive.

The material is radiopaque. The Bio2 Fusion Implants are intended for single use and are provided sterile to the user.

  • VII. Performance Data
    Performance testing supports the substantial equivalence of the subject device to the predicate device.

Biocompatibility of the device has been established according to blue book memorandum #G95-1 entitled "Use of International Standard ISO-10993. Biological Evaluation of Medical Devices Part-1: Evaluation and Testing." In vitro bioactivity testing (ISO 23317:2012) shows the material forms a surface apatite laver when submerged in simulated body fluid.

Static and fatigue bending tests and in vitro stability tests were conducted on the subject and predicate devices to demonstrate substantial equivalence.

VIII. Conclusions

The Bio2 Fusion Implant System when compared to the predicate has the same intended use and similar indications, technological characteristics, and principals of operation as its predicate devices. Testing demonstrates that the Bio2 Fusion Implant System is as safe and effective as the predicate device. Thus the Bio2 Fusion Implant System is substantially equivalent to the predicates.

N/A