K Number
K151754

Validate with FDA (Live)

Date Cleared
2016-03-22

(267 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Glide-On Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

Device Description

The Glide-On Vinyl Examination Gloves are single use only, disposable gloves intended for medical purposes to be worn on the hands or fingers of examiners. The gloves are powder-free and are made of polyvinyl chloride (PVC) with an inner coating of polyurethane. This inner polyurethane coating acts as a lubricant to aid in the donning of the gloves when hands are wet, damp or dry. The gloves are offered non-sterile and are available in small, medium, large and extra large sizes.

AI/ML Overview

The document describes the acceptance criteria and the study conducted for Medline Industries, Inc.'s "Glide-On Vinyl Examination Gloves" (K151754).

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicAcceptance Criteria (Complies with ASTM D5250-06 (reapproved 2011))Reported Device Performance (Glide-On Vinyl Examination Gloves)
Dimensions - Length≥230mmComplies with ASTM D5250-06 (reapproved 2011) ≥230mm
Dimensions - WidthSmall - 85±5mm; Medium - 95±5mm; Large - 105±5mm; Extra Large - 115±5mmSmall - 85±5mm; Medium - 95±5mm; Large - 105±5mm; Extra Large - 115±5mm
Dimensions - ThicknessPalm - 0.10±0.02mm; Finger - 0.10±0.02mmPalm - 0.10±0.02mm; Finger - 0.10±0.02mm
Physical Properties (Before/After Aging)Tensile Strength ≥11MPa; Ultimate Elongation ≥300%Tensile Strength ≥11MPa; Ultimate Elongation ≥300%
Freedom from HolesG-1, AQL 2.5 (per ASTM D5151-06 (reapproved 2011))G-1, AQL 2.5
Residual Powder<2mg per glove<2mg per glove

Biocompatibility Testing:

  • Irritation: Complies with ISO 10993-10: Irritation Intracutaneous reactivity.
  • Sensitization: Complies with ISO 10993-10: Delayed-Type Hypersensitivity (Sensitization) Buehler Test.

Frictional Properties:

  • Tested per ASTM D1894, Standard Test Method for Static and Kinetic Coefficients of Friction of Plastic Film and Sheeting under wet and dry conditions. (No specific numerical acceptance criteria provided in the document, only that a comparison was performed to a plain vinyl glove.)

User Study (Ease of Donning):

  • Evaluated the amount of time it took to don the glove.
  • Evaluated the amount of glove tears that occurred during donning. (No specific numerical acceptance criteria provided, comparison made to three other vinyl gloves coated with polyurethane.)

2. Sample Size Used for the Test Set and the Data Provenance:

  • Non-Clinical Testing (Physical Properties, Freedom from Holes, Residual Powder, Dimensions): The sample sizes are inherent to the ASTM D5250-06 and ASTM D5151-06 standards. The document doesn't explicitly state the exact number of gloves tested but indicates compliance with these standards, which define the sampling plans. The data provenance is not specified, but typically this testing is done by the manufacturer or a contract lab.
  • Biocompatibility Testing: The sample sizes are inherent to the ISO 10993 standards for biological evaluation. The document doesn't explicitly state the exact number of animals or substances used, but indicates compliance with these standards.
  • User Study: 170 subjects. The provenance (e.g., country of origin, retrospective/prospective) is not specified, but it's a prospective user study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

  • No "experts" in the context of clinical interpretation (e.g., radiologists) were used to establish ground truth for this device, as it is a patient examination glove.
  • The "ground truth" for the non-clinical tests is based on adherence to recognized international standards (ASTM, ISO). The experts involved would be the laboratory technicians and quality assurance personnel conducting and verifying these tests, but their specific qualifications are not detailed in the document.

4. Adjudication Method for the Test Set:

  • Not applicable in the typical sense of clinical image review. For non-clinical tests, compliance is determined by meeting the specified criteria of the relevant ASTM and ISO standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is an examination glove, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is an examination glove, not an algorithm.

7. The type of ground truth used:

  • Physical Properties, Dimensions, Freedom from Holes, Residual Powder: Ground truth is defined by the specifications and test methods outlined in the ASTM D5250-06 and ASTM D5151-06 standards.
  • Biocompatibility: Ground truth is established by the test methods and acceptance criteria defined in ISO 10993-10.
  • Frictional Properties: Ground truth is established by the ASTM D1894 standard. A comparison was made against a "plain vinyl glove."
  • User Study (Ease of Donning): Ground truth is based on quantitative measurements (time, tear count) observed during the study, compared to "three other vinyl gloves coated with polyurethane."

8. The Sample Size for the Training Set:

  • Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. This is not an AI/ML device that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 22, 2016

Medline Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist One Medline Place Mundelein, IL 60060

Re: K151754

Trade/Device Name: Glide-On Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 22, 2016 Received: February 23, 2016

Dear Ms. Mason:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151754

Device Name Glide-On Vinyl Examination Gloves

Indications for Use (Describe)

The Glide-On Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medline logo. The logo features the word "MEDLINE" in white, bold sans-serif font, positioned to the left of a white, stylized starburst symbol. The background is a solid dark blue color. The logo is simple, clean, and easily recognizable.

ledline Industries, Inc. One Medline Place Mundelein, IL 60060

510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]

Submitter / 510(k) Sponsor

Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060

Registration Number: 1417592

Contact Person

Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: jamason@medline.com

Summary Preparation Date March 21, 2016

Type of 510(k) Submission Traditional

Device Name / Classification

Name of Device: Glide-On Vinyl Examination Gloves Proprietary Name: Glide-On Vinyl Examination Gloves Common Name: Vinyl Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LYZ Product Class: Class I Regulation #: 21 CFR 880.6250 Classification Panel: General Hospital

Predicate Device

Powder-Free PVC Vinyl Patient Examination Glove, Clear (non-colored) K142703

Device Description

The Glide-On Vinyl Examination Gloves are single use only, disposable gloves intended for medical purposes to be worn on the hands or fingers of examiners. The gloves are powder-free and are made of poly

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, sans-serif font, and to the right of the text is a white, stylized star shape that resembles a medical cross.

edline Industries. One Medline Place Mundelein II 60060

vinyl chloride (PVC) with an inner coating of polyurethane. This inner polyurethane coating acts as a lubricant to aid in the donning of the gloves when hands are wet, damp or dry. The gloves are offered nonsterile and are available in small, medium, large and extra large sizes.

Indications for Use

The Glide-On Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

Summary of Technological Characteristics

The Glide-On Vinyl Examination Gloves are substantially equivalent to the predicate, K142703. Both gloves have the same intended use, same material and the same device performance.

Device CharacteristicProposed DevicePredicate DeviceComparisonAnalysis
Product NameGlide-On VinylExamination GlovesPowder-free PVC VinylExam GlovesN/A
510(k) ReferenceK142703N/A
Product OwnerMedline Industries, IncHebei Grandeast PlasticProducts Co., Ltd.Different
Product CodeLYZLYZSame
Intended UseThe Glide-On VinylExamination Gloves aredisposable device intendedfor medical purposes thatare worn on the examiner'shands or fingers to preventcontamination betweenpatient and examiner.The powder-free PVC vinylexam gloves are disposabledevices intended formedical purposes that areworn on the examiner'shands or fingers to preventcontamination betweenpatient and examiner.Same
Regulation Number21 CFR 880.625021 CFR 880.6250Same
SizesSmallMediumLargeExtra-LargeSmallMediumLargeExtra-LargeSame
MaterialsPVCLubricant - Polyurethanewith a viscosity up to 18centipoisePVCLubricant - Polyurethanewith a viscosity up to 4.6centipoiseSimilar
Dimensions - LengthComplies with ASTMD5250-06 (reapproved)Complies with ASTMD5250-06 (reapproved)Same

TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image features the Medline logo, which is set against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a white, stylized starburst or cross-like symbol. The symbol is oriented diagonally, with its points extending outwards from the center. The overall design is clean and corporate, conveying a sense of professionalism and reliability.

Medline Industries, Inc. One Medline Place Mundelein, IL 60060

2011) ≥230mm2011) ≥230mm
Dimensions - WidthComplies with ASTMD5250-06 (reapproved2011)Small - 85±5mmMedium - 95±5mmLarge - 105±5mmExtra Large - 115±5mmComplies with ASTMD5250-06 (reapproved2011)Small - 85±5mmMedium - 95±5mmLarge - 105±5mmExtra Large - 115±5mmSame
Dimensions - ThicknessComplies with ASTMD5250-06 (reapproved2011)Palm - 0.10±0.02mmFinger - 0.10±0.02mmComplies with ASTMD5250-06 (reapproved2011)Palm - 0.10±0.02mmFinger - 0.10±0.02mmSame
Physical PropertiesComplies with ASTMD5250-06 (reapproved2011)Before Aging/After AgingTensile Strength ≥11MPaUltimate Elongation ≥300%Complies with ASTMD5250-06 (reapproved2011)Before Aging/After AgingTensile Strength ≥11MPaUltimate Elongation ≥300%Same
Freedom from HolesComplies with ASTMD5250-06 (reapproved2011) and ASTM D5151-06 (reapproved 2011)G-1, AQL 2.5Complies with ASTMD5250-06 (reapproved2011) and ASTM D5151-06 (reapproved 2011)G-1, AQL 2.5Same
Powder or Powder-freePowder-freePowder-freeSame
Residual PowderComplies with ASTMD5250-06 (reapproved2011)<2mg per gloveComplies with ASTMD5250-06 (reapproved2011)<2mg per gloveSame
ColorantNo colorant usedNo colorant usedSame
Sterile or Non-sterileNon-sterileNon-sterileSame

Summary of Non-Clinical Testing

The biocompatibility evaluation for the Glide-On Vinyl Examination Gloves was conducted in accordance with ANSVAAMVISO 10993-1:2009 Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. The Glide-On Vinyl Examination Gloves are classified as a surface contacting device with a limited contact duration of less than 24 hours.

The following tests were performed to evaluate the biocompatibility of the Glide-On Vinyl Examination Glove:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is written in white, with a stylized white symbol resembling a star or a stylized cross above the text. A gray bar is visible on the right side of the blue square.

ledline Industries. Inc One Medline Place Mundelein, IL 60060

  • ISO 10993-10: Irritation Intracutaneous reactivity ●
  • ISO 10993-10: Delayed-Type Hypersensitivity (Sensitization) Buehler Test ●

An evaluation was performed comparing the frictional properties of the Glide-On Vinyl Examination Gloves to a plain vinyl glove. The testing was performed per ASTM D1894, Standard Test Method for Static and Kinetic Coefficients of Friction of Plastic Film and Sheeting. Both gloves were tested under wet and dry conditions.

A user study involving one hundred seventy (170) subjects was also performed to evaluate the ease of glove donning on damp hands and dry hands compared to three other vinyl gloves coated with polyurethane. The study evaluated the amount of time it took each glove and also the amount of glove tears that occurred during the donning of each glove.

Summary of Clinical Testing

This section does not apply. No clinical testing was performed.

Conclusion

Based on the non-clinical performance testing Medline Industries, Inc. concludes that the Glide-On Vinyl Examination Gloves are substantially equivalent to the predicate, Powder-free PVC Vinyl Exam Gloves (K142703).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.