K Number
K151707

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-07-19

(25 days)

Product Code
Regulation Number
872.1800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for dagnostic use by dentists.

Device Description

EzSensor Soft, EzSensor Soft i, EzSensor Bio i are digital dental intraoral sensors which acquire digital intra oral images. Direct digital systems acquire images with a bendable sensor that is connected to a computer to produce an image almost instantaneously following exposure. The primary advantage of direct sensor systems is the image acquisition speed. For patient comfort, the ergonomic design is based on human intraoral anatomy.

AI/ML Overview

The provided text describes a Special 510(k) submission for device modifications to the EzSensor Soft, EzSensor Soft i, EzSensor Bio, and EzSensor Bio i digital dental intraoral sensors. This filing primarily focuses on demonstrating substantial equivalence to a predicate device through performance testing and does not include a comparative effectiveness study with human readers (MRMC) or a standalone (algorithm-only) performance study. The device itself is a digital dental intraoral sensor, not an AI algorithm.

Here's a breakdown of the requested information based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document doesn't explicitly state "acceptance criteria" in a pass/fail quantifiable manner for the overall device. Instead, it demonstrates performance equivalence to a predicate device against specific technical characteristics.

CharacteristicAcceptance Criteria (Implied: Equivalence to Predicate)Reported Device Performance (Proposed Device)Predicate Device Performance (K143753)
Indications for UseSubstantially equivalent to predicate.EzSensor Soft, EzSensor Soft i, EzSensor Bio and EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists.EzSensor Soft [Alternative name : EzSensor Bio] Digital Dental Intra Oral Sensor is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists.
Sensor Dimension (mm) (±10%)Slight variation within acceptable tolerance.Size "1.0": 37.8 x 26.6 Size "1.5": 40.8 x 30.6 Size "2.0": 44.0 x 32.5Size "1.0": 37.5 x 26.5 Size "2.0": 43.5 x 32.5
Sensor Thickness (mm)Equivalent to predicate.55
Active Area (mm)Slight variation within acceptable tolerance.Size "1.0": 20.01 x 30.01 Size "1.5": 23.98 x 33.00 Size "2.0": 25.99 x 35.99Size "1.0": 20 x 30 Size "2.0": 25.99 x 35.99
USB ModuleEquivalent to predicate.Integrated USB 2.0 moduleIntegrated USB 2.0 module
Pixel Pitch (µm) - Full ResolutionEquivalent to predicate.14.814.8
Pixel Pitch (µm) - Binning modeEquivalent to predicate.29.629.6
DQE 84.64 µGy 6 lp/mm - Full ResolutionEquivalent to predicate.0.0700.070
DQE 84.64 µGy 6 lp/mm - Binning modeEquivalent to predicate.0.0700.070
MTF 84.64 µGy 6 lp/mm - Full ResolutionEquivalent to predicate.0.1540.154
MTF 84.64 µGy 6 lp/mm - Binning modeEquivalent to predicate.0.1330.133
Typical dose range (µGy)N/A (Information provided for proposed device only)Incisor & Canine: 300 ~ 500 / Molar: 400 ~ 600Not specified for predicate.
Viewer SoftwareEquivalence in function and indications for use.Easydent or EzDent-i (K150747) (Note: EzDent-i 2.0 has additional features, but maintains similar indications and functionalities as EzDent-i 1.0 (K131594) from predicate)Easydent or EzDent-i (K131594)
Safety and EffectivenessNo additional safety risk identified, substantially equivalent to predicate.Performance test results indicate the subject detector performed equally to the predicate. No additional safety risk identified in bench test.N/A (Predicate performance is the benchmark)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document describes "Performance Testing" and "bench test: Non-clinical report" according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices." However, it does not specify a sample size for the test set or the data provenance (country of origin, retrospective/prospective). The testing appears to be laboratory-based ("bench test").

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The assessment relies on technical specifications (DQE, MTF, linear response to X-ray exposure) and safety testing, not on a ground truth established by medical experts for diagnostic accuracy in a clinical context. The device is a sensor, not a diagnostic algorithm that interprets images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable and not provided. As mentioned above, the evaluation is based on technical specifications and safety testing, not on clinical image interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This is a 510(k) for a digital dental intraoral sensor, not an AI-powered diagnostic tool. The document states a "Summary of Performance Testing" based on technical specifications compared to a predicate device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No standalone algorithm performance study was done. The device itself is a sensor that collects data, which is then viewed and manipulated by dentists using viewer software. It is not an algorithm making standalone diagnostic assessments.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The concept of "ground truth" as typically applied to diagnostic algorithms (expert consensus, pathology, etc.) is not directly relevant or discussed in this submission. The device's performance is assessed through technical metrics (DQE, MTF, linear response to X-ray exposure) and compliance with electrical, mechanical, and environmental safety standards (IEC 60601-1, IEC 60601-1-2), comparing these metrics against those of a predicate device to establish substantial equivalence.

8. The sample size for the training set

Not applicable. This submission is for hardware (an X-ray sensor) and associated viewer software, not a machine learning or AI algorithm that would require a training set.

9. How the ground truth for the training set was established

Not applicable. As there is no training set for an AI algorithm, there is no ground truth established for one.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 19, 2015

Rayence Co. Ltd. % Mr. David Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K151707

Trade/Device Name: EzSensor Soft, EzSensor Soft i, EzSensor Bio, and EzSensor Bio i (Model: 1.0, 1.5, 2.0) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: June 23, 2015 Received: June 24, 2015

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151707

Device Name

EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i, Model: 1.0, 1.5, 2.0

Indications for Use (Describe)

EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for dagnostic use by dentists.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Rayence logo. The logo consists of the word "Rayence" in red, block letters. To the right of the word is a graphic of two overlapping circles in shades of orange and yellow. The logo is simple and modern.

ail. Hwaseong-si.

Date: June 23, 2015

To: Department of health & Human Services, Food and Drug Administration

Special 510(k): Device Modification

EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i Model: 1.0, 1.5, 2.0

Reference : K143753, EzSensor Soft [Alternative name- EzSensor Bio] Decision Date : March 10, 2015

The Rayence Co., Ltd. hereby submits this Special 510(k) for EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i, (model: 1.0, 1.5, 2.0) to inform the following device modifications to the predicate device, EzSensor Soft or EzSensor Bio.

  1. New device model names.

  2. New Imaging Viewer Software Version Update

  3. New sensor size 1.5

We believe these modifications are eligible for the Special 510(k) process since they have the same fundamental scientific technology and intended use as the predicate device. We consider our intent to market this device as confidential commercial information and requests that it be treated as such by FDA. We have taken precautions to protect the confidentiality of the intent to market these devices. We understand that the submission to the government of false information is prohibited by 18 U.S.C. 1001 and 21 U.S.C. 331(q).

Thank you in advance for your consideration of our application. If there are any questions, please feel free to contact me at +82-31-8015-6459.

Sincerely yours,

Kee Dock Kim, RA Team manager Rayence Co., Ltd

{4}------------------------------------------------

1. 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510K summary prepared: June 23, 2015

Submitter's Name, address, telephone number, a contact person:

Submitter's Name :Rayence Co., Ltd.
Submitter's Address:14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea
Submitter's Telephone:+82-31-8015-6459
Contact person:Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459
Official Correspondent:(U.S. Designated agent)Dave Kim (davekim@mtech-inc.net)
Address:8310 Buffalo Speedway, Houston, TX 77025
Telephone:+713-467-2607
Fax:+713-583-8988

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary Name:EzSensor Soft, EzSensor Soft i, EzSensor Bio,EzSensor Bio I,Model: 1.0, 1.5, 2.0
Regulation Number :21 CFR 872.1800
Regulation Name :Extraoral source x-ray system
Regulatory Class:II
Product Code:MUH
Predicate Device :
Manufacturer: Rayence Co., Ltd.
Device: EzSensor Soft [Alternative name : EzSensor Bio]

{5}------------------------------------------------

510(k) Number: K143753 (Decision Date – March 10, 2015
Regulation Number: 21 CFR 872.1800
Regulation Name: Extraoral source x-ray system
Regulatory Class: II
Product Code: MUH

510(k) Submission - EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i

2. Device Description

EzSensor Soft, EzSensor Soft i, EzSensor Bio i are digital dental intraoral sensors which acquire digital intra oral images. Direct digital systems acquire images with a bendable sensor that is connected to a computer to produce an image almost instantaneously following exposure. The primary advantage of direct sensor systems is the image acquisition speed. For patient comfort, the ergonomic design is based on human intraoral anatomy.

  • Excellent image quality based on advanced CMOS technology
  • · Sensor's ergonomic shape is more comfortable for the human oral structure
  • · Lower dose exposure (compared to film sensor)
  • Enhanced durability
  • · Easy-to-use USB interface

3. Indication for use

EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensors are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed and manipulated for diagnostic use by dentists.

4. Summary of Design Control Risk management

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

5. Summary of the technological characteristics of the device compared to the predicate device

EzSensor Soft, EzSensor Soft i, EzSensor Bio i Digital Dental Intra Oral Sensor described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, EzSensor Soft [Alternative name : EzSensor Bio] of Rayence Co., Ltd.

{6}------------------------------------------------

5. 1 Separate model name

The following table lists the new devise model names and their respective hardware and software combination.

Model NameComponent
PredicateEzSensor Soft(alternate name : EzSensor Bio)Size: 1.0, 2.0Intra Oral Sensor + Easydent (Viewer Software)Intra Oral Sensor + EzDent-i 1.0 (ViewerSoftware, K131594)
ProposedEzSensor Soft(alternate name : EzSensor Bio)Size: 1.0, 1.5, 2.0Intra Oral Sensor + Easydent (Viewer Software)
ProposedEzSensor Soft i(alternate name : EzSensor Bio i)Size: 1.0, 1.5, 2.0Intra Oral Sensor + EzDent-i 2.0 (ViewerSoftware, K150747)

5. 2 Characteristic

CharacteristicProposedRayence Co., Ltd.PredicateRayence Co., Ltd.
FeatureImage: EzSensor Soft, EzSensor Soft i, EzSensor Bio and EzSensor Bio iImage: EzSensor Soft [Alternative name : EzSensor Bio] digital dental image processing system
510(k) number-K143753
Indicationsfor useEzSensor Soft, EzSensor Soft i,EzSensor Bio and EzSensor Bio iDigital Dental Intra Oral Sensors areintended to collect dental x-rayphotons and convert them intoelectronic impulses that may bestored, viewed and manipulated fordiagnostic use by dentists.EzSensor Soft [Alternative name :EzSensor Bio] Digital Dental IntraOral Sensor is intended to collectdental x-ray photons andconvert them into electronic impulsesthat may be stored, viewed andmanipulated for diagnostic use bydentists.
Sensor Dimension(mm)(±10%)Size "1.0": 37.8 x 26.6Size "1.5": 40.8 x 30.6Size "2.0": 44.0 x 32.5Size "1.0": 37.5 x 26.5Size "2.0": 43.5 x 32.5
Sensor Thickness(mm)55
Active Area(mm)Size "1.0": 20.01 x 30.01Size "1.5": 23.98 x 33.00Size "2.0": 25.99 x 35.99Size "1.0": 20 x 30Size "2.0": 25.99 x 35.99
USB ModuleIntegrated USB 2.0 moduleIntegrated USB 2.0 module

{7}------------------------------------------------

PixelPitch (μm)Full Resolution14.814.8
Binning mode29.629.6
DQE84.64 µGy6 lp/mmFull Resolution0.0700.070
Binning mode0.0700.070
MTF84.64 µGy6 lp/mmFull Resolution0.1540.154
Binning mode0.1330.133
Typical dose range(µGy)Incisor & Canine : 300 ~ 500 / Molar: 400 ~ 600
Viewer Software*Easydent or EzDent-i(K150747)Easydent or EzDent-i(K131594)

510(k) Submission - EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i

This device is operated using viewer software. With the identical hardware configuration, the intraoral sensor model name is distinguished by the type and version of image viewing software available. The software functions include the patient information management, image capture, and viewer for captured images. Both EzDent-i 1.0 and 2.0 versions have the same indications for use and functionalities like operation software, computer platform, picture archiving and communication format, image format, image processing features, windowing, image edit, measurements and manipulation. The differences are that the EzDent-i 2.0 (K150747) has additional features such as "Hide the patient information", "Export to acquisition information", "Create New Report" and "Print Setting".

In addition, a new sensor size 1.5 is available to the existing sensor sizes 1.0 and 2.0 of the predicate device.

These differences do not raise the questions of safety or effectiveness. Based on the laboratory testing results submitted in this 510K, we conclude that the subject device is substantially equivalent to the predicate device.

6. Summary of Performance Testing

The intended use, application and detector type of EzSensor Soft i, EzSensor Soft i, EzSensor Bio, EzSensor Bio i are identical to the predicate device. Both the subject and predicate device use the same amorphous silicon alley and materials for fluorescent as the sensing means. The performance test result indicates that the EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i detector performed equally the predicated device; EzSensor Soft [Alternative name : EzSensor Bio] and no additional safety risk is identified in the bench test: Non-clinical report.

Non-clinical according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-

{8}------------------------------------------------

ray Imaging Devices" was performed.

The DQE, MTF and linear response to X-ray exposure test demonstrated that the subject sensor performed equivalently compared to the predicate device with the same dynamic range.

Electrical, mechanical, environmental safety and performance testing were performed according to standard IEC 60601-1: 2005 + CORR.1(2006) + CORR(2007) (Medical electrical equipment Part 1:General requirements for basic safety and essential performance). EMC testing was conducted in accordance with standard IEC 60601-1-2:2007.

7. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Rayence Co., Ltd. concludes that EzSensor Soft, EzSensor Soft i, EzSensor Bio, EzSensor Bio i is safe and effective and substantially equivalent to predicate device as described herein.

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.