K Number
K151593

Validate with FDA (Live)

Date Cleared
2015-10-23

(133 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures.

Device Description

Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures. The CelioFlex™ UHD 5 confocal Miniprobe is a CystoFlex™ UHD R Confocal Miniprobe with an additional "Grabule" component and with an optical fiber of 3 meters vs. 2 meters. The Grabule is a metallic part glued to the distal tip and the adjacent part of the sheath of the CelioFlex™ UHD 5 Confocal Miniprobe. The Grabule has been designed as a grip for laparoscopic forceps, facilitating user control of the Miniprobe distal tip during laparoscopic procedures. Both the stainless steel 316L metal comprising the Grabule and the epoxy glue (Epotek 301) used to cement the Grabule to the Miniprobe are currently used in the CystoFlex™ UHD R Miniprobe reference device. The Grabule shields the Miniprobe tip from potential wear that repeated handling by forceps might cause.

During standard laparoscopic procedures, CelioFlex™ UHD 5 Confocal Miniprobe is inserted through a pre-installed trocar. When the CelioFlex™ UHD 5 appears on the laparoscope monitor, the surgeon grasps the distal tipped Grabule with the laparoscopic fenestrated forceps to position the Confocal Miniprobe facing the tissue to be imaged.

AI/ML Overview

The provided text is a 510(k) summary for the CelioFlex™ UHD 5 Confocal Miniprobe. It describes the device, its intended use, and its comparison to predicate and reference devices. However, it does not contain a study that demonstrates the device meets specific performance acceptance criteria related to its imaging capabilities or a comparison to a gold standard of truth for diagnostic accuracy.

The "Testing Completed" section lists engineering and manufacturing verification and validation tests, not clinical performance studies with acceptance criteria for diagnostic accuracy. These include:

  • Sterilization Efficacy Verification: "All requirements met, Test Passed"
  • Reprocessing Sterilization Validation: "All requirements met, Test Passed"
  • Biocompatibility (Cytotox) Validation: "All requirements met, Test Passed"

These tests confirm that the device is safe and functions as intended from a manufacturing and material perspective, and that additions like the "Grabule" and increased fiber length do not compromise these aspects. They are not performance metrics for diagnostic accuracy or imaging quality in a clinical context that would typically involve acceptance criteria like sensitivity, specificity, or image resolution compared to a "ground truth".

Therefore, I cannot provide the requested information about acceptance criteria and a study proving the device meets them in the context of diagnostic performance, as such a study is not present in the provided document.

To directly answer your numbered points based on the provided text's limitations:

  1. A table of acceptance criteria and the reported device performance:

    • No acceptance criteria for diagnostic accuracy/imaging performance are provided.
    • The reported device performance relates to manufacturing and safety conformity:
      • Sterilization Efficacy Verification: All requirements met, Test Passed
      • Reprocessing Sterilization Validation: All requirements met, Test Passed
      • Biocompatibility (Cytotox) Validation: All requirements met, Test Passed
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as detailed clinical performance studies for diagnostic accuracy are not described in this summary. The "tests" mentioned are likely lab-based engineering/manufacturing verification.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" in the diagnostic sense is not established in the described testing.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an endoscope component, not an AI diagnostic tool.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.

  8. The sample size for the training set: Not applicable, as this summary does not describe an AI/ML device or its training.

  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 23, 2015

Mauna Kea Technologies Mr. Michael A. Daniel Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460

Re: K151593

Trade/Device Name: CelioFlex " UHD 5 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OWN, GCJ Dated: September 24, 2015 Received: September 25, 2015

Dear Mr. Daniel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

  • Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

6. Indications for Use Statement

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(K) Number (if known)

K151593

Device Name

CelioFlex™ UHD 5

· Indications for Use (Describe)

Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (1/14)

{3}------------------------------------------------

7. Premarket Notification 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number:

Applicant Information:

Date Prepared:June 9, 2015
Name:Mauna Kea Technologies
Address:9 rue d'EnghienF-75010 Paris, France
Phone:+33 1 48 24 03 45
Fax:+33 1 48 24 12 18
Contact Person:Michael A Daniel, Consultantmadaniel@clinregconsult.com
Phone Number:(415) 407-0223
Office:(775) 392-2970
Facsimile Number:(610) 545-0799

Device Information:

Device Trade Name:CelioFlex™ UHD 5 Confocal Miniprobe to be used withCellvizio® 100 Series System
Common Name:Endoscope and Accessories
Classification Name(s):Confocal Optical Imaging
Product Code/ Regulation:OWN / GCJ 21 CFR 876.1500
Classification:Class II

Predicate Device:

  • Circon's USA Series™ Laparoscope has been cleared with K013165

Reference Devices:

  • The Cellvizio 100 Series System with Confocal Miniprobes cleared via K111047.
  • . The CystoFlex™ UHD R has been cleared with K141358

Device Description:

Cellvizio® 100 Series System with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures. The CelioFlex™ UHD 5 confocal Miniprobe is a CystoFlex™ UHD R Confocal Miniprobe with an additional "Grabule" component and with an optical fiber of 3 meters vs. 2 meters. The Grabule is a metallic part glued to the distal tip and the adjacent part of the sheath of the CelioFlex™ UHD 5 Confocal Miniprobe. The

{4}------------------------------------------------

Grabule has been designed as a grip for laparoscopic forceps, facilitating user control of the Miniprobe distal tip during laparoscopic procedures. Both the stainless steel 316L metal comprising the Grabule and the epoxy glue (Epotek 301) used to cement the Grabule to the Miniprobe are currently used in the CystoFlex™ UHD R Miniprobe reference device. The Grabule shields the Miniprobe tip from potential wear that repeated handling by forceps might cause.

During standard laparoscopic procedures, CelioFlex™ UHD 5 Confocal Miniprobe is inserted through a pre-installed trocar. When the CelioFlex™ UHD 5 appears on the laparoscope monitor, the surgeon grasps the distal tipped Grabule with the laparoscopic fenestrated forceps to position the Confocal Miniprobe facing the tissue to be imaged.

Indications for Use:

Cellvizio® 100 Series System with Confocal Miniprobes is a confocal laser system with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

CelioFlex™ UHD 5 Confocal Miniprobe is intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic surgical procedures.

Comparison to Previous Devices:

No change is being made to the fundamental technology and operating principle of the previously cleared CystoFlexTM UHD R Confocal Miniprobe (K141358). The CelioFlex™ UHD 5 is identical to this reference device, only differing by the addition of the Grabule on its distal tip and by the fiber length (2m for CystoFlex™ UHD R and 3m for CelioFlex™ UHD 5).

The Indications for Use and Intended Use of CelioFlex™ UHD R are the same as Circon's USA Series™ Laparoscopes (K013165) predicate and CystoFlex™ UHD R (K141358) reference devices, respectively.

Due to the addition of the Grabule on its distal tip, the CelioFlex™ UHD 5 is designed to fit into trocars of 4.95mm or greater.

Attached verification and validation testing demonstrates that the modifications meet the design specifications and user needs. Sterilization and biocompatibility testing provides evidence that there are no changes to the biocompatibility of the device and no changes to the efficacy and compatibility of the subject device with sterilization processes.

Testing Completed:

Sterilization Efficacy Verification:All requirements met, Test Passed
Reprocessing Sterilization Validation:All requirements met, Test Passed
Biocompatibility (Cytotox)Validation:All requirements met, Test Passed

{5}------------------------------------------------

Summary:

Based upon the Intended Use, Indications for Use, product technical information, performance, sterilization testing, and biocompatibility validation provided in this premarket notification, the CelioFlex™ UHD 5 Confocal Miniprobe, when used as part of the Cellvizio 100 Series, has been shown to be substantially equivalent to the Circon's USA Series™ Laparoscopes, and is basically identical in technological characteristics to the CystoFlexTM UHD R Confocal Miniprobe reference device. The subject device can safely and effectively be used to visualize and image the internal microstructure of tissues during endoscopic and laparoscopic surgical procedures.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.