K Number
K151567

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-12-23

(196 days)

Product Code
Regulation Number
884.4160
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Endoplus Hand held Laparosopic instruments are intended for grasping, cutting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.

Device Description

The devices are commonly named hand held laparoscopic instruments. Specifically, each instrument has a different tip or jaw to define the functional use, such as scissors, forceps, graspers, biopsy, dissectors or needle holders. The devices are designed to be used through a portal, the opening maintained by an introducer or cannula which allows for insertion and removal of the device without damage to surrounding soft tissue. The devices have a linkage mechanism connecting the working tips, or jaws to a handle. This allows the surgeon to cut, dissect, grasp, clamp, cauterize, palpate, or retract at the operating site in conjunction with a laparoscope.

AI/ML Overview

The provided document is a 510(k) Summary for the Endoplus Hand-held Laparoscopic Instruments. This type of document is for demonstrating substantial equivalence to a predicate device, not typically for reporting on detailed clinical studies with specific acceptance criteria, sample sizes, and expert adjudications as might be found for a novel AI/software medical device.

Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for a diagnostic or AI-driven device is not applicable to this document.

The document focuses on comparing the Endoplus instruments to a predicate device (Pajunk GmbH Medizintechnologie's disposable inserts for handled instrument and HF Electrodes, K062072) based on indications for use, technological characteristics, and various safety and performance testing.

Here's a summary of the information that is available in the document, framed in relation to product validation rather than AI performance:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not present a table of specific acceptance criteria in the sense of quantitative thresholds for performance metrics. Instead, it lists various "Safety Testing" and "Performance Testing" studies that were conducted to demonstrate the device's safety and effectiveness and its substantial equivalence to the predicate. The "reported device performance" is implicitly that the device met the requirements of these tests, leading to the conclusion of substantial equivalence.

CategorySpecific Tests PerformedImplied "Acceptance Criteria" / Performance
Safety TestingElectrical and EMC Testing: - IEC 60601-1: 2012 3rd edition - IEC 60601-1-2 Edition 3: 2007-03 - IEC 60601-2-2: 2009Complies with IEC standards
Cleaning & Sterilization- Manual cleaning validation study and Laparoscopic Instrument Rack TOC analysis (AAMI TIR30:2011) - Manual cleaning validation study and Laparoscopic Instrument Rack Protein analysis - Steam Pre-vacuum Efficacy Validation Study - Steam Sterilization Validation Study - Shelf Life ProtocolDemonstrates effective cleaning and sterilization; maintains integrity over shelf life.
Biocompatibility(In accordance with ISO 10993-1, for externally communicating devices in contact with circulating blood for < 24 hours) - Cytotoxicity - Sensitization - Irritation - Systemic Toxicity - Intracutaneous - HemolysisComplies with ISO 10993-1; non-toxic, non-irritating, etc.
Performance Testing- PLTW Comparative Ex-Vivo Thermal Tissue Testing Validation Study - Electrical Safety Validation Report (Electrical Safety Testing) - Scissors Performance Testing - Expected Life TestingPerforms as intended for surgical use (cutting, grasping, cauterization); safe electrically; durable over expected life.

2. Sample size used for the test set and data provenance:

  • This information is not provided in the document. The studies mentioned are primarily bench testing, ex-vivo testing, and compliance with standards, rather than clinical trials with patient-specific test sets. No mention of data provenance (country of origin, retrospective/prospective) is made as it's not relevant to this type of device submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. "Ground truth" in the diagnostic/AI sense is not relevant here. The evaluation is against engineering standards and functional performance, not clinical expert consensus.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/diagnostic device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not applicable. The "ground truth" for this device's performance relies on engineering specifications, recognized standards (IEC, ISO, AAMI), and functional testing results.

8. The sample size for the training set:

  • Not applicable. This is not a machine learning device.

9. How the ground truth for the training set was established:

  • Not applicable.

Conclusion from the document:
The document concludes that the Endoplus Hand-held Laparoscopic instruments are "substantially equivalent to the predicate device and introduces no new issues of safety and effectiveness" based on the described safety, cleaning, sterilization, biocompatibility, and performance testing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human figures, possibly representing individuals or families, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 23, 2015

Endoplus % E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton. Marvland 21114

Re: K151567

Trade/Device Name: Endoplus Hand-held Laparoscopic Instruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: June 8, 2015 Received: June 10, 2015

Dear E.J. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

Page 2 - E.J. Smith

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Binita S. Ashar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K151567

Device Name

Endopus Hand Held Laparoscopic Instruments

Indications for Use (Describe)

Endoplus Hand held Laparosopic instruments are intended for grasping, cutting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (1/14)

Page 1 of 2

SC Publishing Services (301) 443-6740

{3}------------------------------------------------

510(k) Summary

SPONSOR
Company Name:ENDOPLUS
Company Address750 Tower Road, Suite AMundelelin, IL 60060
Telephone:800.236.5972
Contact Person:Matthew Gudeman
Summary Preparation Date:December 23, 2015
DEVICE NAME
Trade Name:ENDOPLUS Hand held Laparoscopic
Instruments Common/Usual Name:Hand-held Surgical Instruments
Classification Name:Electrosurgical, cutting & coagulation & accessories
Regulation Number:21 CFR 878.4400
Product Code:GEI
Device Class:Class II

PREDICATE DEVICE

Legally Marketed Equivalent Device

ManufacturerBrand Name510(k) Number
Pajunk GmbH MedizintechnologiePajunks disposable inserts forhandled instrument and HFElectrodesK062072

DEVICE DESCRIPTION

The devices are commonly named hand held laparoscopic instruments. Specifically, each instrument has a different tip or jaw to define the functional use, such as scissors, forceps, graspers, biopsy, dissectors or needle holders. The devices are designed to be used through a portal, the opening maintained by an introducer or cannula which allows for insertion and removal of the device without damage to surrounding soft tissue. The devices have a linkage mechanism connecting the working tips, or jaws to a handle. This allows the surgeon to cut, dissect, grasp, clamp, cauterize, palpate, or retract at the operating site in conjunction with a laparoscope.

{4}------------------------------------------------

DEVICE INDICATIONS FOR USE

Hand held Laparoscopic instruments are intended for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscope during laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation.

ParametersEndoplusPajunk GmbHComments
510(k) NumberK062072
Indications for UseEndoplus Hand heldLaparoscopicinstruments areintended forgrasping, cutting,dissecting,retracting, clamping,biopsy, and/orcauterization oftissue. The devicesare used inconjunction with alaparoscope duringlaparoscopicprocedures.Instruments should beused only by personnelcompletely familiarwith their operation.Using an instrumentimproperly for a taskwhich it was notintended may result ina damaged or brokeninstrument.Pajunks disposableinserts are instrumentsinsulated for optionalmonopolar coagulationwhich enable a surgeonto grasp, manipulate,dissect, retrieve,biopsy, cut orcoagulate internaltissue or organs whileperforminglaparoscopicprocedures.Similar
User EnvironmentFor use by physicianin the operatingroomFor use by physicianin the operatingroomIdentical
MonopolarYesYesIdentical
Length20cm to 50cm33cm to 45cmSimilar
Minimum CannulaSize5.5cm5.5cmIdentical
Handle DesignPistol GripPistol GripIdentical
Shaft RotationYesYesIdentical
Shaft FlexibilityRigidRigidIdentical
ParametersEndoplusPajunk GmbHComments
Shaft Diameter5mm5mmIdentical
ArticulatingComponentDoes not articulateDoes not articulateIdentical
ActuationMechanismHandle Operationcauses opening andclosing of endeffectorHandle Operationcauses opening andclosing of endeffectorIdentical
Flush PortYesNoDifferent
Insertion throughtrocarYesYesIdentical
Form inserted intotrocarJaws closedJaws closedIdentical
Reusable/DisposableReusable Handle,Shaft and InsertDisposable InsertReusable Handle andShaft AssemblyDifferent

SUBSTANTIALLY EQUIVALENCE COMPARISON TABLE

{5}------------------------------------------------

DIFFERENCES

  • . Endoplus includes a flush port feature for introducing cleaning fluids during the cleaning and providing a means for steam to enter during the sterilization process.
  • . Endoplus's inserts are reusable and Pajunk's inserts are disposable.

COMPARISON OF TECHNICAL CHARACTERISTICS

The Endoplus Hand held Laparoscopic instruments has a similar Indications for Use and identical technological characteristics to the predicate device. Any material differences were addressed by biocompatibility, performance and electrical testing found below:

PERFORMANCE DATA

Safety Testing

Electrical and EMC Testing

Endoplus performed the following electrical tests:

  • IEC 60601-1: 2012 3rd edition Medical electrical equipment—Part 1: . General requirements for basic safety and essential performance
  • . IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.

{6}------------------------------------------------

  • IEC 60601-2-2: 2009 Medical electrical equipment Part 2-2: Particular ● requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

Cleaning and Sterilization

The Reusable Hand-held Laparoscopic instruments are evaluated based on the following tests:

  • Manual cleaning validation study and Laparoscopic Instrument Rack TOC analysis (AAMI TIR30:2011)
  • Manual cleaning validation study and Laparoscopic Instrument Rack Protein ● analysis
  • . Steam Pre-vacuum Efficacy Validation Study
  • Steam Sterilization Validation Study
  • Shelf Life Protocol

Biocompatibility

In accordance with ISO 10993-1, Hand-held Laparoscopic instruments are categorized as externally communicating devices in contact with circulating blood for less than 24 hours. The following tests are recommended by ISO 10993-1:

  • · Cytotoxicity
  • · Sensitization
  • Irritation
  • Systemic toxicity
  • Intracutaneous
  • Hemolysis

Performance Testing:

  • PLTW Comparative Ex-Vivo Thermal Tissue Testing Validation Study
  • . Electrical Safety Validation Report (Electrical Safety Testing)
  • Scissors Performance Testing
  • Expected Life Testing

CONCLUSION

The Endoplus Hand held Laparoscopic instruments has a similar Indications for Use and identical technological characteristics to the predicate device. Biocompatibility and

{7}------------------------------------------------

performance testing has been performed to verify safe use and effective use. Endoplus concludes that Endoplus Hand held Laparoscopic instruments are substantially equivalent to the predicate device and introduces no new issues of safety and effectiveness.

ModularItem #'sdiameter(mm)TypeLengthDescriptions
52935insertstandardDelicate Maryland with CrossSerrations
5293L5insertbariatricDelicate Maryland with CrossSerrations
52945insertstandardJohan Forcep
5294L5insertbariatricJohan Forcep
52955insertstandardMedium Dolphin Nose DissectingForsep w/ cup
5295L5insertbariatricMedium Dolphin Nose DissectingForsep w/ cup
52965insertstandardMedium Fenestrated Bowel Grasper
5296L5insertbariatricMedium Fenestrated Bowel Grasper
52975insertstandardStrevy Grasper
5297L5insertbariatricStrevy Grasper
53005insertstandardAllis forcep
5300L5insertbariatricAllis forcep
53015insertstandardLong Allis Forcep
5301L5insertbariatricLong Allis Forcep
53025insertstandardTraumatic Grasper 2x3 Teeth
5302L5insertbariatricTraumatic Grasper 2x3 Teeth
53035insertstandardEndo Allis Forcep
5303L5insertbariatricEndo Allis Forcep
ModularItem #'sdiameter(mm)TypeLengthDescriptions
53045insertstandardGlassman Forcep
5304L5insertbariatricGlassman Forcep
53055insertstandardPennington Forcep
5305L5insertbariatricPennington Forcep
53065insertstandardStrong Fenestrated Grasping Forcep
5306L5insertbariatricStrong Fenestrated Grasping Forcep
53075insertstandardMcKernan Tri Grasping FenestratedForcep
5307L5insertbariatricMcKernan Tri Grasping FenestratedForcep
53085insertstandardAlligator Forcep
5308L5insertbariatricAlligator Forcep
53095insertstandardPointed Dissector w/ cup
5309L5insertbariatricPointed Dissector w/ cup
53105insertstandardRound Tip Allis Grasping Forcep
5310L5insertbariatricRound Tip Allis Grasping Forcep
53115insertstandardCurved Dissecting Forcep
5311L5insertbariatricCurved Dissecting Forcep
53125insertstandardShort Curved Dissecting Forcep
5312L5insertbariatricShort Curved Dissecting Forcep
53135insertstandardMixter Dissector
5313L5insertbariatricMixter Dissector
53145insertstandardDuckbill Dissector
5314L5insertbariatricDuckbill Dissector
53155insertstandardStraight Tapered Dissector
ModularItem #'sdiameter(mm)TypeLengthDescriptions
5315L5insertbariatricStraight Tapered Dissector
53165insertstandardStraight Mini Dissector
5316L5insertbariatricStraight Mini Dissector
53175insertstandardMini Right Angle Dissector
5317L5insertbariatricMini Right Angle Dissector
53185insertstandardAtraumatic Tube Forcep
5318L5insertbariatricAtraumatic Tube Forcep
53195insertstandardDelicate Cone Dissector
5319L5insertbariatricDelicate Cone Dissector
53205insertstandardMicro Dolphin Dissecting Forcep
5320L5insertbariatricMicro Dolphin Dissecting Forcep
53215insertstandardClaw Forcep 2x3 Teeth
5321L5insertbariatricClaw Forcep 2x3 Teeth
53225insertstandardBiopsy Forcep w/ Two Teeth
5322L5insertbariatricBiopsy Forcep w/ Two Teeth
53235insertstandardDorsey Fenestrated Grasper
5323L5insertbariatricDorsey Fenestrated Grasper
53245insertstandardEndo Babcock Forcep
5324L5insertbariatricEndo Babcock Forcep
53255insertstandardDebakey Tissue Forcep
5325L5insertbariatricDebakey Tissue Forcep
53265insertstandardExtracting Forcep Reverse Teeth
5326L5insertbariatricExtracting Forcep Reverse Teeth
53275insertstandardGrasping Forcep
ModularItem #'sdiameter(mm)TypeLengthDescriptions
5327L5insertbariatricGrasping Forcep
53285insertstandardFenestrated Grasping Forcep
5328L5insertbariatricFenestrated Grasping Forcep
53295insertstandardMicro Dissecting Forcep
5329L5insertbariatricMicro Dissecting Forcep
53305insertstandardBabcock Grasping Forcep
5330L5insertbariatricBabcock Grasping Forcep
53315insertstandardDolphin Nose Dissecting Forsep w/ cup
5331L5insertbariatricDolphin Nose Dissecting Forsep w/ cup
53325insertstandardCobra Jaw Grasping Forcep
5332L5insertbariatricCobra Jaw Grasping Forcep
53335insertstandardAtraumatic Grasper
5333L5insertbariatricAtraumatic Grasper
53345insertstandardWave Jaw Grasping Forcep
5334L5insertbariatricWave Jaw Grasping Forcep
53355insertstandardPaddle Babcock
5335L5insertbariatricPaddle Babcock
53365insertstandardTri-Grasper
5336L5insertbariatricTri-Grasper
53375insertstandardLong Grasping Forcep
5337L5insertbariatricLong Grasping Forcep
53385insertstandardMaxi Fenestrated Grasping Forcep
5338L5insertbariatricMaxi Fenestrated Grasping Forcep
53395insertstandardRetrograde Grasping Forcep
ModularItem #'sdiameter(mm)TypeLengthDescriptions
5339L5insertbariatricRetrograde Grasping Forcep
53405insertstandardKocher Grasping Forcep
5340L5insertbariatricKocher Grasping Forcep
53415insertstandardCurved Maryland Dissector
5341L5insertbariatricCurved Maryland Dissector
53425insertstandardDolphin Nose Grasping Forcep
5342L5insertbariatricDolphin Nose Grasping Forcep
53435insertstandardMicro Dolphin Dissector w/ Cup
5343L5insertbariatricMicro Dolphin Dissector w/ Cup
53445insertstandardOviduct Atraumatic Forcep
5344L5insertbariatricOviduct Atraumatic Forcep
53455insertstandardTraumatic Grasper, 2x3 Teeth S/A
5345L5insertbariatricTraumatic Grasper, 2x3 Teeth S/A
53465insertstandardAtraumatic Grasper S/A
5346L5insertbariatricAtraumatic Grasper S/A
53475insertstandardGrasping Forcep w/ cup
5347L5insertbariatricGrasping Forcep w/ cup
53485insertstandardRipple Claw Forcep
5348L5insertbariatricRipple Claw Forcep
53495insertstandardMini-Micro Dissecting Forcep
5349L5insertbariatricMini-Micro Dissecting Forcep
53505insertstandardCurved Metz Scissor
5350L5insertbariatricCurved Metz Scissor
53515insertstandardStraight Micro Scissor
ModularItem #'sdiameter(mm)TypeLengthDescriptions
5351L5insertbariatricStraight Micro Scissor
53525insertstandardDelicate Curved Metz Scissor
5352L5insertbariatricDelicate Curved Metz Scissor
53535insertstandardCurved Micro Scissor
5353L5insertbariatricCurved Micro Scissor
53545insertstandardStraight Metz Scissor
5354L5insertbariatricStraight Metz Scissor
53555insertstandardFlat Dissector
5355L5insertbariatricFlat Dissector
53565insertstandardHook Scissor
5356L5insertbariatricHook Scissor
53575insertstandardDelicate S/A Scissor
5357L5insertbariatricDelicate S/A Scissor
53585insertstandardSerrated Scissor
5358L5insertbariatricSerrated Scissor
53595insertstandardBiopsy Scoop Forcep
5359L5insertbariatricBiopsy Scoop Forcep
53605insertstandardBiopsy Forcep w/ One Tooth
5360L5insertbariatricBiopsy Forcep w/ One Tooth
53615insertstandardGlassman Forcep, 20mm
5361L5insertbariatricGlassman Forcep, 20mm
53625insertstandardHeavy Curved Dissector
5362L5insertbariatricHeavy Curved Dissector
53635insertstandardDorsey Grasper, 40mm
ModularItem #'sdiameter(mm)TypeLengthDescriptions
5363L5insertbariatricDorsey Grasper, 40mm
53645insertstandardTenaculum, 5mm
5364L5insertbariatricTenaculum, 5mm
53655insertstandardMedium Curved Dissector
5365L5insertbariatricMedium Curved Dissector
53665insertstandardBump Grasper, Single Action
5366L5insertbariatricBump Grasper, Single Action
53675insertstandardSingle Tooth Grasper
5367L5insertbariatricSingle Tooth Grasper
53685insertstandardGlassman Forcep, 30mm
5368L5insertbariatricGlassman Forcep, 30mm
53695insertstandardFenestrated Bowel Grasper, 20mm
5369L5insertbariatricFenestrated Bowel Grasper, 20mm
53705insertstandardSingle Action Needle Holder
5370L5insertbariatricSingle Action Needle Holder
53715insertstandardSuture Knot Pusher
5371L5insertbariatricSuture Knot Pusher
53725insertstandardCredi Curved Dissector
5372L5insertbariatricCredi Curved Dissector
53735insertstandardSt. Joseph Dissecting Forcep
5373L5insertbariatricSt. Joseph Dissecting Forcep
53745insertstandardAtraumatic Wave Grasper
5374L5insertbariatricAtraumatic Wave Grasper
53755insertstandardCurved Debakey Tissue Forcep
ModularItem #'sdiameter(mm)TypeLengthDescriptions
5375L5insertbariatricCurved Debakey Tissue Forcep
53765insertstandardPetrovich Curved Dissector
5376L5insertbariatricPetrovich Curved Dissector
53775insertstandardAndrew Babcock Forcep
5377L5insertbariatricAndrew Babcock Forcep
53785insertstandardMaryland with Cross Serrations
5378L5insertbariatricMaryland with Cross Serrations
53795insertstandardFenestrated Bowel Grasper, 40mm
5379L5insertbariatricFenestrated Bowel Grasper, 40mm
53805insertstandardByerley Grasper
5380L5insertbariatricByerley Grasper
53815insertstandardMurray Grasper
5381L5insertbariatricMurray Grasper
53825insertstandardDouble Action Spoon Forcep, 5mm
5382L5insertbariatricDouble Action Spoon Forcep, 5mm
53835insertstandardBiopsy Forcep w/ One Tooth (onbottom)
5383L5insertbariatricBiopsy Forcep w/ One Tooth (onbottom)
53845insertstandard5mm Smoot Endo Cinch
5384L5insertbariatric5mm Smoot Endo Cinch
53855insertstandardSingle Action Spoon Forcep
5385L5insertbariatricSingle Action Spoon Forcep
53865insertstandardHarvey Grasper
5386L5insertbariatricHarvey Grasper
ModularItem #'sdiameter(mm)TypeLengthDescriptions
53875insertstandardRodriguez Grasper
5387L5insertbariatricRodriguez Grasper
53885insertstandardForsythe Bump Grasper, Single Action
5388L5insertbariatricForsythe Bump Grasper, Single Action
53895insertstandardAngled Dissector
5389L5insertbariatricAngled Dissector
53905insertstandard5mm Meah Grasper
5390L5insertbariatric5mm Meah Grasper
53915insertstandardKowalski Hunter Bowel Grasper
5391L5insertbariatricKowalski Hunter Bowel Grasper
53925insertstandardDevlin Microfenestrated Grasper
5392L5insertbariatricDevlin Microfenestrated Grasper
53935insertstandardHolmes Atraumatic Grasper
5393L5insertbariatricHolmes Atraumatic Grasper
53945insertstandardGraves Grasper
5394L5insertbariatricGraves Grasper
53955insertstandard5mm Maryland with 90 Degree Tooth
5395L5insertbariatric5mm Maryland with 90 Degree Tooth
SLRVA5handle/tubestandardTake Apart Slide Lock with Rotation
SLRVL5handle/tubebariatricTake Apart Slide Lock with Rotation
5SLR5handle/tubestandardTake Apart Slide Lock with Rotation
5SLRL5handle/tubebariatricTake Apart Slide Lock with Rotation
5MRR5handle/tubestandardMonopolar Handle with Rotation,Ratchet on Demand
5MRRL5handle/tubebariatricMonopolar Handle with Rotation
ModularItem #'sdiameter(mm)TypeLengthDescriptions
Ratchet on Demand
5MROT5handle/tubestandardMonopolar Handle with Non-LockingRotation, NO Ratchet
5MROTL5handle/tubebariatricMonopolar Handle with Non-LockingRotation, NO Ratchet

Below is a complete listing of laparoscopic instruments covered by this 510(k) submission:

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

§ 884.4160 Unipolar endoscopic coagulator-cutter and accessories.

(a)
Identification. A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.(b)
Classification. Class II (performance standards).