(266 days)
The IRENE Locking Compression Plate System is indicated for patients with age above 21 for fixation of fractures. Indicated facture areas include ulna, radius, humerus, femur, and tibia.
The IRENE Locking Compression Plate System is designed for internal fixation of bones that subject to fracture. The system includes bone plates and screws that can be used with the plate. Various sizes of plate are available for different fracture areas or clinical conditions. Bone screws are incorporated in the system and are intended to fix with plate with appropriate size. The plate has a limited-contact design. The plate system is provided unsterile. And the system is intended for single-use.
This document is a 510(k) Summary for the IRENE Locking Compression Plate System, which is a medical device for orthopedic fixation. It's a regulatory filing, not a research study. Therefore, most of the information requested about acceptance criteria and study details (like sample sizes, expert adjudication, MRMC studies, ground truth) is not applicable or present in this type of document because it describes a submission for regulatory clearance, not the results of a clinical effectiveness study.
Here's an analysis of the provided text based on your request, highlighting what's available and what's explicitly stated as not applicable:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table of acceptance criteria with specific numerical performance targets. Instead, it states that the device's performance was evaluated against predicate devices using recognized ASTM standards. The acceptance criterion implicitly is that the proposed device performs comparably to the predicate devices and meets the requirements of these standards.
The document states:
- "Results of testing indicates no gap between the level of performance of proposed device and the predicate device" (Page 6)
The performance data mentioned are:
- Plate: Static and dynamic testing conducted as per ASTM F382-13.
- Screw: Torsional properties, driving torque, and pull-out strength conducted as per ASTM F543-13.
No specific numerical results for these tests are provided, only the general conclusion of "no gap" in performance compared to predicate devices.
2. Sample size used for the test set and the data provenance
- Sample size: Not specified. The document refers to "mechanical testing" and "testing the mechanical properties of the proposed and predicate device" (Page 5, 6) but does not provide the number of units/samples tested for either the proposed or predicate device.
- Data provenance: The mechanical testing was conducted by Tianjin ZhengTian Medical Instrument Co., Ltd. in China, as indicated by the submitter's address (Page 3). It is preclinical (bench) testing, not clinical data from patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This was preclinical mechanical testing, not a study requiring expert clinical opinion for ground truth.
4. Adjudication method for the test set
Not applicable. This was preclinical mechanical testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a bone fixation plate system, not an AI diagnostic or assistive tool. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used
Not applicable in the conventional sense of clinical ground truth (e.g., pathology, outcomes). The "ground truth" for this submission's performance evaluation lies in the established mechanical testing standards (ASTM F382-13 and ASTM F543-13) and the performance of legally marketed predicate devices. The goal was to show the proposed device met these standards and was comparable to predicates.
8. The sample size for the training set
Not applicable. This device is a physical product, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This device is a physical product, not an AI/ML algorithm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other, representing the human element of the department's work.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Tianjin Zhengtian Medical Instrument Co., Ltd Yifan Fu Regulatory Affairs Specialist No.318, Jingyi Road, Airport Economic Zone Tianjin, 300308 CHINA
February 25, 2016
Re: K151508 Trade/Device Name: IRENE Locking Compression Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 8, 2016 Received: January 13, 2016
Dear Yifan Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
IRENE Locking Compression Plate System
Indications for Use (Describe)
The IRENE Locking Compression Plate System is indicated for patients with age above 21 for fixation of fractures. Indicated facture areas include ulna, radius, humerus, femur, and tibia.
Type of Use (Select one or both, as applicable)
즈 Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "IRENE" in white text on a blue background. The text is in a sans-serif font and is horizontally centered. The blue background appears to be a solid color, providing a clean and simple backdrop for the text.
Section 10 510(k) Summary
This document is prepared to provide sufficient information for an understanding of the basis for the determination of substantial equivalence, as required by 21 CFR section 807.92(c).
-
- 510(k) Submitter and Owner
Tianjin ZhengTian Medical Instrument Co., Ltd Address: No.318, Jingyi Road, Airport Economic Zone, Tianjin, P.R. China Official Correspondent: Xiang, Zhaojun Position: Regulatory Affairs Supervisor Tel: +86 10 82292929 Fax: +86 10 82290740 E-mail: xiangzhaojun@naton.cn Date of Preparation of the Summary: Feb 25th, 2016
- 510(k) Submitter and Owner
-
- Submission Contact Fu, Yifan Position: Regulatory Affairs Specialist Tel: +86 10 82292929 E-mail: fuyifan@naton.cn
-
- Proposed Device
Trade name - IRENE Locking Compression Plate System Common name - Locking compression plate and screw Class: Class 2
Regulation -21 CFR 888.3030, Product Code HRS (Bone Fixation Plate) -21 CFR 888.3040, Product Code HWC (Bone Fixation Screw)
Review Panel: Orthopedic
The IRENE Locking Compression Plate System is designed for internal fixation of bones that subject to fracture. The system includes bone plates and screws that can be used with the plate. Various sizes of plate are available for different fracture areas or clinical conditions. Bone screws are incorporated in the system and are intended to fix with plate with appropriate size. The plate has a limited-contact design. The plate system is provided unsterile. And the system is intended for
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "IRENE" in white font against a blue background. The font is sans-serif and appears to be bold. The background is a solid, light blue color, providing a clean and simple backdrop for the text.
single-use.
Indications for Use:
The IRENE Locking Compression Plate System is indicated for patients with age above 21 for fixation of fractures. Indicated facture areas include ulna, radius, humerus, femur, and tibia.
-
- Predicate Device
| Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | |
|---|---|---|---|
| 510(k) Number | K101400 | K100721 | K112819 |
| Predicate Device Name | Locking Bone Plate | Locking Bone Screw | 3.5mm Locking Screw |
| Manufacturer | Changzhou Orthmed Medical Instrument Co., Ltd. | Xiamen Double Engine Medical Material Co., Ltd. | |
| Classification | 888.3030(HRS) | 888.3040(HWC) | 888.3040(HWC) |
-
- Preclinical Data
Mechanical testing is conducted to demonstrate differences in technological characteristics of the proposed device to predicate device do not raise new question regarding safety and effectiveness, and Substantially Equivalent to the predicate device is thus concluded. The testing incorporates the following standards as test method to evaluate the performance of both proposed and predicate device:
- Preclinical Data
-
- ASTM F 382-99 (Reapproved 2008), Standard Specification and Test Method for Metallic Bone Plates.
-
- ASTM F543-13, Standard Specification and Test Methods for Metallic Medical Bone Screws.
-
- Clinical Data
No clinical data is presented in this submission.
-
- Substantial Equivalence Comparison
{5}------------------------------------------------
IRENE
| Element | Proposed DeviceIRENE Plate andScrew | Predicate DeviceOrthmed Plate andScrew | Predicate DeviceDouble Engine3.5mm LockingScrew |
|---|---|---|---|
| Class | 2 | 2 | 2 |
| Regulationnumber | Plate: 21 CFR888.3030Screw: 21 CFR888.3040 | Plate: 21 CFR888.3030Screw: 21 CFR888.3040 | Not referencedScrew: 21 CFR888.3040 |
| Intended use | Internal fixation ofbone fractures | Internal fixation ofbone fractures | Internal fixation ofbone fractures |
| Design | Plate: straight, tubularcross-section, lockingcompression screwhole, has limitedcontact area. | Plate: straight,tubular cross-section, lockingcompression screwhole, has limitedcontact area.Dimensions andcontour has slightdifference.Screw: self-tapping,fully threadedlocking screw. Thecontour has veryslight difference. | Not referencedScrew: 3.5mmdiameter, self-tapping lockingscrew. |
| Material | Plate: titaniumScrew: titanium alloy | Plate: titanium alloyScrew: titaniumalloy | Not referencedScrew: titaniumalloy |
| How Supplied | Non-sterile | Non-sterile | Non-Sterile |
| Single-UseOnly | Yes | Yes | Yes |
| Sterilization | Moist heatsterilization prior touse | Moist heatsterilization prior touse | Moist heatsterilization priorto use |
| Performancedata | Plate: static anddynamic testingconducted as perASTM F382-13Screw: torsional | Plate: static anddynamic testingconducted as perASTM F382-13Screw: torsional | Not referencedScrew: |
| properties, drivingtorque and pull-outstrength are conductedas per ASTM F543-13 | properties, drivingtorque and pull-outstrength areconducted as perASTM F543-13 | performancetesting conductedas per ASTMF543-07 |
{6}------------------------------------------------
IRENE
The proposed device is found to possess highly similar technological characteristics on design under the premises of sharing same intended use. Device performance is compared through the testing of mechanical properties of the proposed and predicate device by methods stipulated in published standards in order to evaluate whether differences existing in the design raise safety issue by compromising performance. Results of testing indicates no gap between the level of performance of proposed device and the predicate device, through which the differences are deemed not relevant to device performance and has been justified, and the conclusion of Substantial Equivalence is reached.
N/A