K Number
K151487

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-01-14

(226 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diatron Glucose Hexokinase Method is for the in vitro quantitative determination of Glucose in serum for use with Diatron Pictus 700 Chemistry Analyzer. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes.

Diatron ISE is for the in vitro quantitative determination of Sodium (Na+), Potassium (K+), and Chloride (Cl-) concentrations in serum on the Diatron Pictus 700 Chemistry Analyzer. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements are used in monitoring electrolyte balance and in the diagnosis and treatment of diseases/conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Diatron Pictus 700 Clinical Chemistry Analyzer is a wet-chemistry analyzer for the direct determination of sodium, potassium, chloride and glucose concentrations in serum, and to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. It is for in vitro diagnostic use.

Device Description

The Pictus 700 Clinical Chemistry Analyzer is an automatic, floor-standing, wet chemistry system, designed for the qualitative and quantitative analyses of various diagnostic test systems. This premarket notification is for the direct quantitative measurements of Na (sodium), K (potassium), CI (chloride) and glucose in serum samples. Additionally, other types of chemistry assays may be performed on the analyzer, provided that suitable colorgenerating reactions or reactions with variation of color are used. The system is intended for use in clinical laboratories.

The instrument consists of an analyzer unit and an operations computer with a screen that allows the user to input commands for system operation and data display. The analyzer unit includes two temperature-controlled incubation rotors and a multi-wavelength photometer, a cooled carousel for loading barcoded sample tubes or micro cups and reagent cartridges, and two probes that deliver reagents and samples to the incubation rotors and the ISE measurement flow cell. The analyzer unit also houses containers for wash solution and waste. An ISE module, installed on the analyzer unit, is used to measure sodium, potassium and chloride ionic activity in serum.

AI/ML Overview

The provided text describes the acceptance criteria and the study performance for the Diatron ISE, Diatron Glucose Hexokinase Method, and Diatron Pictus 700 Clinical Chemistry Analyzer. This is a submission for a 510(k) premarket notification for an in vitro diagnostic device, specifically a clinical chemistry analyzer and associated assay methods. The studies presented are nonclinical performance studies to demonstrate substantial equivalence to a predicate device, not clinical effectiveness studies involving human subjects in an MRMC setting.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the claim of substantial equivalence to the predicate device and the successful demonstration of various analytical performance characteristics. The document explicitly states acceptance criteria for linearity.

Study/ParameterAnalyte(s)Acceptance CriteriaReported Device Performance
Functional Sensitivity/LoD/LoQSodiumHighest dilution concentration (lowest concentration) with < 10%CV was lower than, or equal to, the low-end linearity claim.1:9 dilution resulted in 27.6 mmol/L with 3.02%CV. Low-end linearity claim: 110 mmol/L (verified).
PotassiumHighest dilution concentration (lowest concentration) with < 10%CV was lower than, or equal to, the low-end linearity claim.1:10 dilution resulted in 1.05 mmol/L with 0.00%CV. Low-end linearity claim: 1.0 mmol/L (verified).
ChlorideHighest dilution concentration (lowest concentration) with < 10%CV was lower than, or equal to, the low-end linearity claim.1:9 dilution resulted in 14.8 mmol/L with 3.02%CV. Low-end linearity claim: 50 mmol/L (verified).
GlucoseLoQ measured as 7 mg/dL.LoQ CV was 1.3%. Verified low end LOD and LOQ claim of 7 mg/dL.
LinearitySodium+/- 10% and or +/- 5 mmol/LLinear from 13 mmol/L to 196 mmol/L. Claimed range: 115 to 196 mmol/L.
Potassium+/- 10% and or +/- 1.0 mmol/LLinear from 0.9 to 12.8 mmol/L. Claimed range: 1.1 to 8.8 mmol/L.
Chloride+/- 10% and or +/- 5 mmol/LLinear from 41 mmol/L to 200 mmol/L. Claimed range: 49 mmol/L to 152 mmol/L.
Glucose+/- 10% and or +/- 5 mg/dLLinear from 13 mg/dL to 632 mg/dL. Claimed range: 13 to 500 mg/dL.
PrecisionSodium, Potassium, Chloride, Glucose%CVs within acceptable ranges (comparison to predicate's range and general analytical expectations). Specific criteria not explicitly stated but implied by successful testing according to CLSI EP5-A2 and presentation of low %CVs.Sodium: Within-run %CV: 0.5-0.7%, Total %CV: 0.9-1.4% Potassium: Within-run %CV: 0.00-0.78%, Total %CV: 1.05-1.65% Chloride: Within-run %CV: 0.5-0.6%, Total %CV: 0.8-1.2% Glucose: Within-run %CV: 0.9-1.6%, Total %CV: 3.0-3.2%
InterferencesSodium, Potassium, Chloride, Glucose≤ 10 % Change or ≤ 3 mmol/L (or mg/dL for glucose) up to specified interferent concentrations.Sodium: Hemolysis (600 mg/dL), Lipemia (2500 mg/dL), Bilirubin (24 mg/dL) Potassium: Not for hemolyzed serum, Lipemia (2500 mg/dL), Bilirubin (24 mg/dL) Chloride: Hemolysis (600 mg/dL), Lipemia (2500 mg/dL), Bilirubin (24 mg/dL) Glucose: Hemolysis (600 mg/dL), Lipemia (700 mg/dL), Bilirubin (12 mg/dL) all passed inference evaluation.
Method ComparisonSodium, Potassium, Chloride, GlucoseCorrelation coefficients and 95% CIs for slope and intercept should demonstrate agreement with the predicate device. Specific numerical acceptance criteria for correlation, slope, and intercept are not explicitly stated as hard cutoffs but implied by demonstrating substantial equivalence to the predicate.Sodium: N=69, r=0.995, Slope=1.002 (0.978-1.027), Intercept=-1.02 (-4.7 to 2.2) Potassium: N=69, r=0.998, Slope=0.999 (0.984-1.014), Intercept=0.02 (-0.05 to 0.09) Chloride: N=69, r=0.997, Slope=1.015 (0.996-1.035), Intercept=-1.90 (-3.9 to 0.1) Glucose: N=136, r=0.999, Slope=0.993 (0.987 to 1.000), Intercept=1.2 (-0.1 to 2.4)

2. Sample Sizes and Data Provenance for Test Sets

  • Glucose Functional Sensitivity/LoQ: Glucose depleted samples, tested in 5 replicates per dilution for functional sensitivity. LoQ testing was performed according to CLSI EP17-A2. The document mentions "Testing the neat sample, depleted of Glucose y standing > 24 hrs exposed to red cells in the sample draw."
  • Linearity: Eleven levels of each analyte were tested, in duplicate.
  • Precision: Control samples (low, middle, and high levels) were tested in duplicate, twice a day, for 20 days, for a total of 80 results per level.
  • Interferences: A two-level control set (low and high analyte levels) was spiked to seven or eight levels of interferent. Spiked samples plus neat samples were tested in duplicates.
  • Method Comparison:
    • Sodium: N=69 clinical specimens
    • Potassium: N=69 clinical specimens (listed under Sodium, but context implies N=69 for K as well given the table structure)
    • Chloride: N=69 clinical specimens
    • Glucose: N=136 clinical specimens
    • Data Provenance: Clinical specimens were "previously-collected serum samples that had been stored frozen and then thawed prior to analysis." The country of origin is not specified but typically for a US FDA submission, these would be collected in the US or under similar regulatory standards. The data is retrospective in that it uses "previously-collected" and "stored frozen" samples.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. This is an analytical performance study for an in vitro diagnostic device, not an AI/imaging device requiring expert interpretation for ground truth. The "ground truth" for the test samples in this context relates to their known concentrations as determined by a reference method or the predicate device.

4. Adjudication Method for the Test Set

Not applicable. This is an analytical performance study for an in vitro diagnostic device, where quantitative results are compared to reference methods/predicate, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size on Human Readers

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is typically done for AI-driven diagnostic imaging devices where human readers' performance (with or without AI assistance) is evaluated. This submission is for an automated clinical chemistry analyzer.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies presented are essentially "standalone" performance evaluations of the Diatron Pictus 700 Clinical Chemistry Analyzer and its associated methods (Diatron Glucose Hexokinase Method and Diatron ISE). The performance measured is that of the automated instrument itself, without direct human intervention in the analytical process being tested for. The output is a quantitative measurement of analytes.

7. The Type of Ground Truth Used

The ground truth for the analytical studies are based on:

  • Reference methods/assigned values: For linearity, samples were assigned their reference values arithmetically from labeled values. For method comparison, the predicate system served as the comparative method against which the new device's performance was evaluated.
  • Known concentrations: For precision, control samples with known low, medium, and high concentrations were used.
  • Spiked samples: For interference studies, samples were spiked with known concentrations of interferents to assess their impact.

8. The Sample Size for the Training Set

Not applicable. This is not a study involving machine learning with distinct training and test sets in the typical sense of AI algorithm development. This is a traditional analytical validation for an in vitro diagnostic device.

9. How the Ground Truth for the Training Set Was Established

Not applicable for the reason stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines beneath them.

January 14, 2016

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

DIATRON US, INC. C/O ERIKA AMMIRATI PRESIDENT 575 SHIRLYNN COURT LOS ALTOS CA 94022

Re: K151487 Trade/Device Name: Diatron ISE. Diatron Glucose Hexokinase Method Diatron Pictus 700 Clinical Chemistry Analyzer Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CFR, JGS, CEM, CGZ, JJE Dated: November 30, 2015 Received: December 1, 2015

Dear Erika Ammirati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For:

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151487

Device Name Diatron Glucose Hexokinase Method Diatron ISE Diatron Pictus 700 Clinical Chemistry Analyzer

Indications for Use (Describe)

Diatron Glucose Hexokinase Method is for the in vitro quantitative determination of Glucose in serum for use with Diatron Pictus 700 Chemistry Analyzer. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes.

Diatron ISE is for the in vitro quantitative determination of Sodium (Na+), Potassium (K+), and Chloride (Cl-) concentrations in serum on the Diatron Pictus 700 Chemistry Analyzer. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements are used in monitoring electrolyte balance and in the diagnosis and treatment of diseases/conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Diatron Pictus 700 Clinical Chemistry Analyzer is a wet-chemistry analyzer for the direct determination of sodium, potassium, chloride and glucose concentrations in serum, and to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. It is for in vitro diagnostic use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Diatron Group logo. The logo features a blue abstract shape on the left, followed by the word "Diatron" in a combination of green and blue. Below "Diatron" is the word "GROUP" in green. A green arc is above the word Diatron.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K151487.

807.92 (a)(1): Name:Diatron, US, Inc.
Address:2304 W 78th StreetHialeah, FL 33016
Phone:877-684-1139
FAX:913-344-9958
Contact:Mr. Frank Matusazak

807.92 (a)(2): Device name- trade name and common name, and classification

Diatron ISE, Diatron Glucose Hexokinase Method Trade Name: Diatron Pictus 700 Clinical Chemistry Analyzer

Common Name: Routine chemistry analyzer for glucose, and ISE technology for sodium, potassium, and chloride

Classification Name(s):

21 CFR § 862.1345- Glucose test system

21 CFR § 862.1665- Sodium test system

21 CFR § 862.1600- Potassium test system

21 CFR § 862.1170- Chloride test system

21 CFR § 862.2160- Discrete photometric chemistry analyzer for clinical use

807.92 (a)(3): Identification of the legally marketed predicate devices

Diatron Pictus 400 – K101741

807.92 (a)(4): Device Description

The Pictus 700 Clinical Chemistry Analyzer is an automatic, floor-standing, wet chemistry system, designed for the qualitative and quantitative analyses of various diagnostic test systems. This premarket notification is for the direct quantitative measurements of Na (sodium), K (potassium), CI (chloride) and glucose in serum samples. Additionally, other types of chemistry assays may be performed on the analyzer, provided that suitable colorgenerating reactions or reactions with variation of color are used. The system is intended for use in clinical laboratories.

The instrument consists of an analyzer unit and an operations computer with a screen that allows the user to input commands for system operation and data display. The analyzer unit includes two temperature-controlled incubation rotors and a multi-wavelength photometer, a cooled carousel for loading barcoded sample tubes or micro cups and reagent cartridges, and two probes that deliver reagents and samples to the incubation rotors and the ISE measurement flow cell. The analyzer unit also houses containers for wash solution and waste. An ISE module, installed on the analyzer unit, is used to measure sodium, potassium and chloride ionic activity in serum.

2304-10 West 78th Street

Hialeah, Fl 33016

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Diatron Group. The logo features a blue abstract shape on the left, followed by the word "Diatron" in green. Below "Diatron" is the word "GROUP" in green. A green arc is above and to the right of the word "Diatron".

System operation:

After a sample tube or cup is placed into the carousel, and appropriate reagents have been installed and properly calibrated, the operator programs a test request. The analyzer pipettes reagent(s) and the sample, at appropriate volumes and times dictated by the test application stored in the analyzer memory, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator, the analyzer measures the absorbance of the test solution at wavelengths dictated by the test application. and, based on the absorbance readings, it calculates the concentration of analyte in the sample using appropriate mathematical formulas dictated by the test application. The test system can measure analytes in serum and results are available in approximately 2 minutes for the electrolyte tests and 10 minutes for glucose determinations.

Chemistry reactions:

Sodium, Potassium and Chloride.

The ISE module measures sodium, potassium, and chloride in serum using ion-selective electrode technology. The electrodes are flow-through and use selective membranes, especially formulated to be sensitive to sodium, potassium, and chloride ions. The electric potential of each electrode is measured relative to a fixed, stable voltage established by a double-junction silver/silver chloride reference electrode. The voltage developed varies with the concentration of the corresponding ion logarithmically, as expressed by the Nernst equation.

The method of measurement is comparative. First, the ISE module measures the potentials developed when the sample is positioned in front of the electrodes. Next, Calibrant A is positioned in front of the electrodes. The difference between the two potentials is related logarithmically to the concentration of the measured ions in the sample, divided by their respective concentrations in the calibrant solution. Since the difference in potentials and the concentration of the sodium, potassium, and chloride ions in the calibrant solution are known, the computer can calculate the concentration of the ions in the sample.

Glucose:

The concentration of glucose in serum is performed enzymatically, according to the hexokinase method, based on the following reactions: b.D-glucose in the sample reacts with adenosine triphosphate (ATP) under the catalytic action of hexokinase, to produce glucose 6phosphate and adenosine diphosphate (ADP). Glucose 6-phosphate consequently reacts with oxidized nicotinamide-adenine nucleotide (NAD') to produce D-gluconic acid 6-phosphate and reduced nicotinamide-adenine nucleotide (NADH), under the catalytic action of glucose-6-phosphate dehydrogenase (G-6PDH).

The reduced nucleotide (NADH) absorbs strongly at 340 nm, whereas its oxidized form (NAD+) does not. The difference in absorbance between the final reading and the blank, monitored bichromatically at 340 nm and 380 nm, is directly proportional to the concentration of b,D-glucose in the sample. The analyzer photometer reads the absorbance at 340 nm and 380 nm at time intervals dictated by the glucose application stored in the analyzer memory, and the change in absorbance is calculated automatically.

2304-10 West 78th Street

Hialeah, Fl 33016

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Diatron Group logo. The logo features a blue water droplet-like shape on the left, followed by the word "Diatron" in a combination of green and blue colors. The word "GROUP" is written in smaller blue letters below "Diatron". A green arc is above the word Diatron.

807.92 (a)(5): Intended Use

Diatron Glucose Hexokinase Method is for the in vitro quantitative determination of Glucose in serum for use with Diatron Pictus 700 Chemistry Analyzer. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes.

Diatron ISE is for the in vitro quantitative determination of Sodium (Na+), and Chloride (Cl-) concentrations in serum on the Diatron Pictus 700 Chemistry Analyzer. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements are used in monitoring electrolyte balance and in the diagnosis and treatment of diseases/conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Diatron Pictus 700 Clinical Chemistry Analyzer is a wet-chemistry analyzer for the direct determination of sodium, potassium, chloride and glucose concentrations in serum, and to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. It is for in vitro diagnostic use.

807.92 (a)(6): Technological Similarities and Differences to the Predicate

CharacteristicCandidate System (K151487)Diatron Pictus 700 AnalyzerPredicate SystemPictus 400, K101741
Instrument PlatformPictus 700Pictus 400
Mode of DetectionPhotometricSame
Photometric DetectorPhotodiodeSame
Optical ModesMonochromatic, Bi-chromatic,Same
Analytical MeasureEndpoint and kinetic withsample blankingSame
Wavelengths340, 380, 405, 450, 490, 505, 550,590, 620, 650, 700, 750330, 405, 450, 505, 550, 570, 600,650, 700
# of Photometer Filters129
Linear Absorbance Range-0.1 to 3.6Same
Photometric Tests/Hour600280
Reagent Volume Range180 to 700 microlitersSame
Reagents On-board Capacity7248
Sample On-board Capacity9548
Device Class, Regulation Code,Product CodeClass I, 21 CFR 862.2160 (discretephotometric chemistry analyzer), JJESame
ISE Test System, Glucose System
Device Class, Regulation CodeClass II,Sodium- 21 CFR 862.1665Potassium- 21 CFR 862.1600Chloride- 21 CFR 862.1170Same

The following chart describes similarities and differences between the two test systems.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Diatron Group. The logo features a blue abstract shape on the left, followed by the word "Diatron" in green. Below "Diatron" is the word "GROUP" in blue. A green arc is above and to the right of the word "Diatron".

Diatron Pictus 700 AnalyzerCharacteristicPredicate-Pictus 400, K101741
Sodium- JGS
Potassium- CEMSame
Classification Product CodeChloride- CGZ
Glucose- CFR
Quantitative determinations of
Intended Usesodium, potassium, and chlorideusing ion selective electrodes, andSame
glucose by photometry in serum.
Testing EnvironmentClinical labSame
ISE potentiometry (electrolytes) andSame
Test Principlephotometry (glucose)
Specimen TypeHuman serumHuman serum, plasma, or urine
Method ComparisonsSodium-0.995 (69, 120-181)Sodium-0.986 (41, 116-189)
(Correlation coefficients,Potassium-0.998 (69, 1.7-8.7)Potassium-0.980 (46, 2.1-8.4)
# of Samples, Testing Ranges)Chloride-0.997 (69, 55-134)Chloride-0.991 (46, 61-151)
[mmol/L: ISE, mg/dL: glucose]Glucose - 0.999 (136, 26 to 484)Glucose - 0.976 (62, 28-499)
Sodium- 115-196 mmol/LSodium- 115-196 mmol/L
Linearity/Reportable RangesPotassium- 1.1-8.8 mmol/LPotassium- 1.1 to 8.8 mmol/L
Chloride- 49-152 mmol/LChloride- 49-152 mmol/L
Glucose- 13-500 mg/dLGlucose 10-500 mg/dL
Sensitivity-Detection LimitsSodium- 27.6 mmol/L @ 3.02%CVN/A
(ISE Functional Sensitivity.Potassium-1.05 mmol/L @ 0%CVN/A
Glucose- standard LOB/LOD)Chloride- 14.8 mmol/L @ 3.02%CVN/A
Glucose 7 mg/dL @ 2.6%CVN/A
%CV Ranges%CVs Ranges:
Within-run PrecisionSodium- 0.4 to 0.5%CVSodium- 0.50 to 1.06 %CV
(across 3 levels- low, middle, high)Potassium-0.00 to 0.78 %CVPotassium- 0.99 to 1.15 %CV
Chloride-0.5 to 0.6 %CVChloride- 0.60 to 1.50 %CV
Glucose-0.9 to 1.6 %CVGlucose- 0.50 to 1.00%CV
%CV Ranges%CVs ranges:
Sodium-0.9-1.4 %CVSodium- 0.40 to 0.65 %CV
Between-run (Total) PrecisionPotassium-1.05-1.65 %CVPotassium- 0.40 to 2.00 %CV
Chloride- 0.8-1.2 %CVChloride- 0.73 to 1.49 %CV
Glucose- 3.0-3.2 %CVGlucose- 1.40 to 4.50 %CV

807.92 (b)(1): Brief Description of Nonclinical Data

Limits of detection/functional sensitivities:

For the glucose assay, conventional LoB/D/Q testing was performed according to CLSI EP17-A2; the LoQ was measured as 7 mg/dL.

For functional sensitivity, normal sample pools were serially diluted nine times in 10% increments (total of 10 samples), and, 5 replicates per dilution were measured for each test system. Mean values of results were calculated for each dilution, and were compared to the calculated ion activities or glucose concentrations. The highest dilution (lowest concentration) with less than 10 %CV was compared to the low linearity claim. Acceptable results were those where the highest dilution concentration was lower than, or equal to, the low-end linearity claim.

Hialeah, Fl 33016

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for Diatron Group. The word "Diatron" is written in a combination of green and blue colors, with the "Dia" in green and the "tron" in blue. Above the word "GROUP" is written in green, and there is a green arc above the word "Diatron". To the left of the word "Diatron" is a blue graphic that resembles a water droplet.

Sodium: Functional Sensitivity

The 1:9 dilution, resulting in sodium ion activity of 27.6 mmol/L, corresponded to a %CV of 3.02%; therefore, the low end claim of linearity of 110 mmol/L was verified.

Potassium: Functional Sensitivity

At all dilutions, the %CVs were substantially lower than the 10% maximum. Even at the 1:10 dilution, resulting in potassium ion activity of 1.05 mmol/L, the CV was 0.00%; therefore, the low end claim of linearity from 1.0 mmol/L was verified.

Chloride: Functional Sensitivity

The 1:9 dilution, resulting in chloride ion activity of 14.8 mmol/L, corresponded to a %CV of 3.02%; therefore, the low end claim of linearity from 50 mmol/L was verified.

Glucose: EP protocol. LOB/LOD/LOQ - Glucose depleted samples

Testing the neat sample, depleted of Glucose y standing > 24 hrs exposed to red cells in the sample draw. The LOQ CV was 1.3%. This verified the low end LOD and LOQ claim of 7 mg/dL.

Linearity:

Eleven levels of each analyte were tested. The samples were assigned their reference values arithmetically from the labeled values and were tested in duplicate by the Pictus 700. The mean Pictus (y-axis) were plotted against the assigned values (x-axis). The acceptance criteria are described below.

Acceptance Criteria: Sodium: +/- 10% and or +/- 5 mmol/L Potassium +/- 10% and or +/- 1.0 mmol/L Chloride: +/- 10% and or +/- 5 mmol/L Glucose: +/- 10% and or +/- 5 mg/dL

Sodium- The assay is linear from 13 mmol/L to 196 mmol/L. The claimed range will be 115 to 196 mmol/L.

Potassium- The assay is linear from 0.9 to 12.8 mmol/L. The claimed range will be 1.1 to 8.8 mmol/L.

Chloride- The assay is linear from 41 mmol/L to 200 mmol/L. The claimed range will be 49 mmol/L to 152 mmol/L.

Glucose- The assay is linear from 13 mg/dL to 632 mg/dL. The claimed range will be 13 to 500 mg/dL.

Precision:

Following CLSI EP5-A2, control samples (low, middle, and high levels) were tested in duplicate, twice a day, for 20 days, for a total of 80 results per level. The results were tabulated and the data were analyzed for means and percent coefficient of variation (%CVs).

Hialeah, Fl 33016

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Diatron Group logo. The logo features a blue abstract shape on the left, followed by the word "Diatron" in a combination of green and blue. Above the word "GROUP" is a green arc. The logo is clean and professional, suggesting a company in the medical or scientific field.

MeanWithin-run %CVTotal %CV
Low Level126 mmol/L0.6%0.9%
Med. Level143 mmol/L0.5%1.4%
High Level159 mmol/L0.7%1.0%

Sodium values for each level assaved on the Pictus 700 Analyzer

Potassium values for each level assayed on the Pictus 700 Analyzer

MeanWithin-run %CVTotal %CV
Low Level3.4 mmol/L0.00%1.05%
Med. Level4.5 mmol/L0.00%1.65%
High Level5.8 mmol/L0.78%1.41%

Chloride values for each level assayed on the Pictus 700 Analyzer

MeanWithin-run %CVTotal %CV
Low Level87 mmol/L0.6%0.9%
Med. Level98 mmol/L0.5%1.2%
High Level109 mmol/L0.5%0.8%

Glucose values for each level assayed on the Pictus 700 Analyzer

MeanWithin-run %CVTotal %CV
Low Level (45)45 mg/dL1.6%3.2%
Med. Level (92)92 mg/dL0.9%3.0%
High Level (309)309 mg/dL1.2%3.1%

Interferences:

The study evaluated increasingly higher levels of: bilirubin, hemoglobin, and triglycerides for all parameters. A two-level control set with low and high levels of each analyte was spiked to seven or eight levels (depending on the interferent) and the spiked samples plus the neat samples were tested in duplicates by the Pictus 700 using its standard operating procedure; the results were averaged. In each case, the spiked sample result mean was compared to its neat (zero interferent) mean result, and recoveries were calculated.

≤ 10 % Change or ≤ 3 mmol/L up to:
HemolysisLipemiaBilirubin
Sodium600 mg/dL2500 mg/dL24 mg/dL
PotassiumDo not use hemolyzed serum2500 mg/dL24 mg/dL
Chloride600 mg/dL2500 mg/dL24 mg/dL
Glucose600 mg/dL700 mg/dL12 mg/dL

Method Comparisons:

Clinical specimens (n = 69, 69, 69, and 136 for sodium, chloride, and glucose, respectively), spanning the dynamic ranges, were assayed in singleton and in a blinded fashion by the Pictus 700 (y-axis) and the predicate system. The specimens were previously-collected

2304-10 West 78th Street

Hialeah, Fl 33016

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for Diatron Group. The word "Diatron" is written in a combination of green and blue colors, with the "Dia" in green and the "tron" in blue. Above the word "GROUP" is a green arc. To the left of the word "Diatron" is a blue graphic.

serum samples that had been stored frozen and then thawed prior to analysis. Deming regression analysis data appear below.

SodiumResultPotassiumResult
N69N69
Correlation Coefficient (r)0.995Correlation Coefficient (r)0.998
Slope (95% CIs)1.002(0.978 to 1.027)Slope (95% CIs)0.999(0.984-1.014)
Intercept-1.02(-4.7 to 2.2)Intercept0.02(-0.05 to 0.09)
CI = confidence intervals
ChlorideResultGlucoseResult
N69N136
Correlation Coefficient (r)0.997Correlation Coefficient (r)0.999
Slope (95% CIs)1.015(0.996 to 1.035)Slope (95% CIs)0.993(0.987 to 1.000)
Intercept-1.90(-3.9 to 0.1)Intercept1.2(-0.1 to 2.4)

807.92 (b)(2): Brief Description of Clinical Data

Not applicable

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

The Diatron Pictus 700 Clinical Chemistry Analyzer is substantially equivalent to the Pictus 400 system, (K101741), and is safe and effective for the claimed intended use.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.