K Number
K151303

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-01-21

(248 days)

Product Code
Regulation Number
868.5270
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatic population. specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 vears of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.

Device Description

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. It is provided non sterile for single patient use. The heated breathing circuit is used with Fisher and Paykel heated humidifiers.

AI/ML Overview

This document describes the regulatory submission for the AirLife Infant Heated Wire Circuit, a medical device. The information provided is primarily focused on demonstrating substantial equivalence to a predicate device rather than detailing an AI/ML study. Therefore, some of the requested information regarding AI-specific criteria (like MRMC studies, training set details, expert qualifications for ground truth in AI, etc.) is not applicable to this document.

Here's an analysis based on the provided text:

Device: AirLife Infant Heated Wire Circuit
Intended Use: To deliver and warm breathing gases before they enter the patient's airway, used with a Fisher and Paykel MR850 humidifier for neonatal and infant pediatric populations.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a single table directly comparing acceptance criteria and a single "reported device performance" value for each. Instead, it lists numerous performance tests against established standards. The general acceptance criterion for all these tests is compliance with the specified standard.

Here's a table summarizing the performance characteristics and their corresponding standards, implying that the device met these standards to be considered substantially equivalent. The "Reported Device Performance" column indicates the successful completion of the test in compliance with the standard mentioned.

Performance CharacteristicAcceptance Criteria (Referenced Standard)Reported Device Performance
Biocompatibility (Cytotoxicity, Sensitization, Irritation, Implantation, Genotoxicity, Leachables)ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010, ISO 10993-17:2002, ISO 10993-18:2005Compliant with standards
Electrical Safety (Patient Leakage Current)IEC 60601-1:2005, 3rd Edition (AAMI ES60601-1:2005)Compliant with standard
Electrical Safety (Dielectric Strength Means)IEC 60601-1:2005, 3rd Edition (AAMI ES60601-1:2005)Compliant with standard
Electrical Safety (Insulation)IEC 60601-1:2005, 3rd Edition (AAMI ES60601-1:2005)Compliant with standard
LengthISO 5367:2000, 4th Edition (within ±10% of labeled length)Compliant with standard
Resistance to Flow/Rated FlowISO 5367:2000, 4th Edition (pressure increase not to exceed 0.2 kPa at rated flow)Compliant with standard
Resistance to Flow @ 2.5 LPMISO 5367:2014, 5th Edition (flow resistance not to exceed 0.74 cmH2O/l/min)Compliant with standard
Increase in Flow Resistance w/Bending @ Rated Flow (10 LPM)ISO 5367:2000, 4th Edition (pressure not to exceed 150% of value with tube straight)Compliant with standard
Increase in Flow Resistance w/Bending @ 2.5 LPMISO 5367:2014, 5th Edition (pressure not to exceed 150% of value with tube straight)Compliant with standard
Conical ConnectorsISO 5367:2000, 4th Edition; ISO 5356-1:2004, 3rd Edition; ISO 8185:2008, 3rd Edition (Corrected Version) (correctly sized ports)Compliant with standards
Security of AttachmentISO 5367:2000, 4th Edition (adaptor not to detach at force > 45N)Compliant with standard
LeakageISO 5367:2000, 4th Edition (leakage rate not > 50 ml/min)Compliant with standard
Compliance per MeterISO 5367:2000, 4th Edition; ISO 8185:2008, 3rd Edition (Corrected Version) (compliance at 6 kPa not to exceed 10 ml*kPa-1 per meter)Compliant with standards
Humidity Output (Invasive Mode)ISO 8185:2008, 3rd Edition (Corrected Version) (delivers ≥ 33 mg/L humidity)Compliant with standard
Humidity Output (Non-Invasive Mode)ISO 8185:2008, 3rd Edition (Corrected Version) (delivers ≥ 10 mg/L humidity)Compliant with standard
Corrugate Melt (100% Duty Cycle, Zero Gas Flow)ISO 8185:2008, 3rd Edition (Corrected Version) (tubes do not collapse, occlude, or cause safety hazard)Compliant with standard
Security of engagement temperature sensorISO 8185:2008, 3rd Edition (Corrected Version) (temperature probes do not disconnect)Compliant with standard
Leakage from sensing portISO 8185:2008, 3rd Edition (Corrected Version) (sensing port not to leak > 5 ml/min)Compliant with standard
Specific enthalpyISO 8185:2008, 3rd Edition (Corrected Version) (thermal overshoot at patient connection not to exceed energy equivalent to 43°C and 100% RH, averaged over 30s)Compliant with standard (specific enthalpy not to exceed 194 kJ/kg dry gas)
Surface temperatureISO 8185:2008, 3rd Edition (Corrected Version) (accessible temperature within 25 cm of patient port not to exceed 44°C)Compliant with standard
Steady state noiseISO 8185:2008, 3rd Edition (Corrected Version) (noise at any point within 1 m of system not to exceed 50 dB)Compliant with standard
Duration of Use (30 days)N/A (Assessed circuit use over 30 day period)Assessed, results not explicitly detailed but implied compliant for safety/effectiveness
Longitudinal Temperature ProfileN/A (Assessed temperature range of wire once nominal voltage induced)Assessed, results not explicitly detailed but implied compliant for safety/effectiveness
Patient End Temperature ProfileN/A (Assessed temperature of airway gas delivered to patient with humidifier and ventilator)Assessed, results not explicitly detailed but implied compliant for safety/effectiveness

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes (e.g., number of circuits tested) for each performance test. It only states that "The following tests were performed for the proposed device to support the substantial equivalence decision." The data provenance is internal testing conducted by the manufacturer to demonstrate compliance with international standards. This is prospective testing against product specifications. There is no mention of country of origin for any data or retrospective analysis as would be relevant for an AI/ML study.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualification of Those Experts

This information is not applicable. The "ground truth" for this device is established by objective performance criteria defined in international standards (e.g., ISO, IEC). Performance is measured against these engineering and safety specifications, not against expert human interpretations or diagnoses.


4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods are typically relevant for studies involving human interpretation (e.g., medical imaging reads). For device performance testing against objective standards, the results are typically determined by measurement and adherence to predefined pass/fail criteria.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is specific to evaluating diagnostic accuracy or clinical utility where human readers are involved, often in the context of AI/ML software. This document describes the safety and performance testing of a physical medical device (heated breathing circuit).


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable as the device is a physical heated breathing circuit and does not involve an algorithm.


7. The Type of Ground Truth Used

The "ground truth" for the device's performance is objective, measurable criteria and limits set forth in recognized international standards (e.g., ISO 5367, ISO 8185, IEC 60601-1, ISO 10993 series). This is akin to engineering specifications and regulatory standards. It is not based on expert consensus, pathology, or outcomes data in the manner relevant to an AI/ML diagnostic system.


8. The Sample Size for the Training Set

This information is not applicable. There is no mention of a "training set" as this is not an AI/ML device. The device likely underwent design verification and validation testing, which is similar in concept to testing for traditional devices.


9. How the Ground Truth for the Training Set Was Established

This information is not applicable as this is not an AI/ML device and therefore has no "training set" or corresponding ground truth establishment in that context. The "ground truth" for its design and manufacturing is based on established engineering principles, material science, and compliance with the aforementioned international performance and safety standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 21, 2016

CareFusion 2200, Inc. Erika Fernandez Regulatory Affairs Manager 75 North Fairway Drive Vernon Hills, Illinois 60061

Re: K151303

Trade/Device Name: AirLife Infant Heated Wire Circuit Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE Dated: December 20, 2015 Received: December 23, 2015

Dear Ms. Fernandez,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name AirLife Infant Heated Wire Circuit

Indications for Use (Describe)

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatic population. specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 vears of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".

Summary of Safety and Effectiveness

510k summary complying with 21 CFR 807.92

1. SUBMITTER

CareFusion 2200, Inc. 75 N Fairway Drive, Vernon Hill, IL 60061

Phone: 847 362-8097 Fax: 312 949-0731

Contact Person: Erika Fernandez

Date Prepared: January 20, 2016

2. Device

Product Name:AirLife Infant Heated Wire Circuit
Device Name:Heated Breathing Circuit
Common Name:Heater, breathing system w/wo controller (not humidifier ornebulizer)
Classification Name:Breathing system heater (21 CFR 868.5270)
Regulatory Class:II
Product Code:BZE

3. Predicate Device

This submission demonstrates substantial equivalence to the Fisher and Paykel Heated Breathing Circuit K020332 cleared on July 700, 2003 and AirLife Heated Ventilator and Anesthesia Breathing Circuits, K000697 cleared on March 30, 2000. The predicate devices have not been subject to a design-related recall.

4. Device Description

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. It is provided non sterile for single patient use. The heated breathing circuit is used with Fisher and Paykel heated humidifiers.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".

5. Indication for use

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatric population, specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 years of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.

6. Comparison of technological characteristics with the predicate device

The fundamental scientific technology is the same for both proposed and predicate devices. It is based on acting as an airway conduit between a breathing machine and the patient (typically attached to an endotracheal or tracheal tube previously insert into the patient's airway). The differences in the design do not raise any different questions of safety and effectiveness.

Element of comparisonPredicate Device 1Fisher and Paykel heatedwire breathing circuitRT131 with flow rate>4L/minPredicate Device 2AirLife Heated andAnesthesia BreathingCircuitswith flow rate >3L/minProposed DeviceAirLife Infant HeatedWire Circuitwith flow rate>4L/min
Principal of OperationResistance wires withinthe tubing generate heatto maintaintemperatures andhumidityResistance wires withinthe tubing generate heatto maintaintemperatures andhumidityResistance wireswithin the tubinggenerate heat tomaintaintemperatures andhumidity
Circuit Characteristics
Infant Inspiratory Limb withUnheated Extension5 ft5 ft5 ft
Dryline Length2 ft2 ft2 ft
Inspiratory Limb Length4 ft4 ft4 ft
Heated wirePresentPresentPresent
Expiratory Limb Length5ft 6in4 ft 3 in4 ft 5in
Water trapPresentNot PresentNot Present
Heated WireNot PresentPresentPresent
Patient populationNeonatesInfants, pediatric andAdultsInfants and Neonates
UsageDisposableDisposableDisposable

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top right. To the right of the circle is the text "CareFusion" in gray. Below the text is the phrase "has joined BD" in a smaller, lighter gray font.

Element of comparisonPredicate Device 1Fisher and Paykel heatedwire breathing circuitRT131 with flow rate>4L/minPredicate Device 2AirLife Heated andAnesthesia BreathingCircuitswith flow rate >3L/minProposed DeviceAirLife Infant HeatedWire Circuitwith flow rate>4L/min
DesignDual LimbDual LimbDual Limb
Tube Specifications
ID0.44in-0.45in0.394in0.45in-0.46in
DesignCorrugateCorrugateCorrugate
Wire Design
Minimum Resistance atambient22 Ω27 Ω22 Ω
Wire Total Length115in99in121in
Constant Power22W16W22W
Typical Wattage/foot ofwire21.92
Typical Watt/ foot ofcircuit646
ConductorCopperCopper/Ni alloy304 Steel
Anchor Position of Wire19 cm5-10 cm19 cm
Compatible HumidifierFisher and Paykel MR850Fisher and Paykel MR850Fisher and PaykelMR850

7. Performance Data

The proposed device was tested to ensure compliance to the following standards:

Biocompatibility

Tests for an externally communicating, tissue by way of gas path and direct mucosal contact with prolonged contact (greater than 24 hours but less than 30 days): Cytotoxicity, Sensitization, Irritation, Muscle Implantation, Genotoxicity and Extractables/Leachables

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller, lighter gray font.

Standards

Performance CharacteristicStandard
Biological Evaluation of Medical Devices Part 1: Evaluation andTestingFDA Guidance: Use of International Standard ISO- 10993,"Biological Evaluation of Medical Devices Part 1: Evaluationand Testing"ISO 10993-1:2009
Biological Evaluation of Medical Devices Part 3: Tests forgenotoxicity, carcinogenicity and reproductive toxicityISO 10993-3:2014
Biological Evaluation of Medical Devices Part 5: Tests for invitro cytotoxicityISO 10993-5:2009
Biological Evaluation of Medical Devices Part 6: Tests for localeffects after implantationISO 10993-6:2007
Biological Evaluation of Medical Devices Part 10: Tests forirritation and skin sensitizationISO 10993-10:2010
Biological Evaluation of Medical Devices Part 17:Establishment of allowable limits for leachable substancesISO 10993-17:2002
Biological Evaluation of Medical Devices Part 18: Chemicalcharacterization of materialsISO 10993-18:2005

Performance

The following tests were performed for the proposed device to support the substantial equivalence decision.

TestDescriptionRelevant Standard
Electrical safetyPatient Leakage CurrentTo ensure, under normal systemoperating conditions, thatexcessive current does notreach the patientIEC 60601-1:2005, 3rdEdition (Equivalent toAAMI ES60601-1:2005)
Electrical safetyDielectric Strength MeansTo check the integrity of thewire insulation under excessivepower conditions.IEC 60601-1:2005, 3rdEdition (Equivalent toAAMI ES60601-1:2005)
Electrical SafetyInsulationTo ensure IEC requirements forcreepage and clearance weremet.IEC 60601-1:2005, 3rdEdition (Equivalent toAAMI ES60601-1:2005)
LengthTo ensure that the actual lengthof the device is within ±10% ofthe labeled lengthISO 5367:2000, 4th Edition
TestDescriptionRelevant Standard
Resistance to Flow/RatedFlowTo ensure, when tested per thestandard, that the pressureincrease at the rated flowthrough the breathing tube doesnot exceed 0.2 kPaISO 5367:2000, 4th Edition
Resistance to Flow @ 2.5LPMTo ensure, when tested per thestandard, that the flowresistance at 2.5 LPM throughthe breathing tube does notexceed 0.74 cmH2O/l/minISO 5367:2014, 5th Edition
Increase in Flow Resistancew/Bending @ Rated Flow(10 LPM)To ensure, when tested per thestandard, that the pressure atthe rated flow when thebreathing tube is suspendedover a metal cylinder does notexceed 150% of the valueobtained with the tube straight.ISO 5367:2000, 4th Edition
Increase in Flow Resistancew/Bending @ 2.5 LPMTo ensure, when tested per thestandard, that the pressure at2.5 LPM when the breathingtube is suspended over a metalcylinder does not exceed 150%of the value obtained with thetube straight.ISO 5367:2014, 5th Edition
Conical ConnectorsTo ensure, when tested per thestandard, that required ports arecorrectly sized.ISO 5367:2000, 4th EditionISO 5356-1:2004, 3rd EditionISO 8185:2008, 3rd Edition(Corrected Version)
Security of AttachmentTo ensure, when tested per thestandard, that the adaptor shallnot become detached from thetube at a force greater than 45N.ISO 5367:2000, 4th Edition
LeakageTo ensure, when tested per thestandard, that the systemsupplied ready to use does notleak at a rate greater than 50ml/min.ISO 5367:2000, 4th Edition
Compliance per MeterTo ensure, when tested per thestandard, that the complianceper meter at a pressure of 6 kPadoes not exceed 10 ml*kPa-1per meter length of tube.ISO 5367:2000, 4th EditionISO 8185:2008, 3rd Edition(Corrected Version)
TestDescriptionRelevant Standard
Humidity Output (InvasiveMode)To ensure, when tested per thestandard, that the systemdelivers at least the minimumhumidity of at least 33 mg/L tothe patient.ISO 8185:2008, 3rd Edition(Corrected Version)
Humidity Output (Non-Invasive Mode)To ensure, when tested per thestandard, that the systemdelivers at least the minimumhumidity of at least 10 mg/L tothe patient.ISO 8185:2008, 3rd Edition(Corrected Version)
Corrugate Melt (100% DutyCycle, Zero Gas Flow)To ensure that the breathingtubes do not collapse onbending, occlude, or otherwisecause a safety hazard at themaximum output of thehumidification system.ISO 8185:2008, 3rd Edition(Corrected Version)
Security of engagementtemperature sensorTo ensure that the temperatureprobes that accompany thecircuit do not becomedisconnected under thecondition of no flow ormaximum rated flow.ISO 8185:2008, 3rd Edition(Corrected Version)
Leakage from sensing portTo ensure, when tested per thestandard, that the sensing portdoes not leak more than 5ml/min.ISO 8185:2008, 3rd Edition(Corrected Version)
Specific enthalpyTo ensure under normal andsingle fault conditions, a thermalovershoot at the patientconnection port shall not exceedan energy equivalent to 43C and100% relative humidity (aspecific enthalpy not to exceed194 kJ/kg dry gas) whenaveraged over 30 seconds.ISO 8185:2008, 3rd Edition(Corrected Version)
Surface temperatureTo ensure that the accessibletemperature within 25 cm of thepatient port connection doesnot exceed 44°CISO 8185:2008, 3rd Edition(Corrected Version)
Steady state noiseTo ensure, when tested per thestandard, that the noise at anypoint within 1 m of the systemdoes not exceed 50 dB.ISO 8185:2008, 3rd Edition(Corrected Version)
TestDescriptionRelevant Standard
Duration of Use (30 days)To assess the circuit use over a30 day time periodN/A
Longitudinal TemperatureProfileTo assess the temperature rangeof the wire once a nominalvoltage is induced across it.N/A
Patient End TemperatureProfileTo assess the temperature of theairway gas delivered to thepatient when a given humidifiedcircuit is used with a ventilatorand the Fisher & Paykel MR850heater base.N/A

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the CareFusion logo, which includes an orange circle with a white shield-like shape inside. To the right of the logo is the text "CareFusion" in a dark gray, sans-serif font. Below the company name, in a smaller, lighter font, is the phrase "has joined BD".

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the CareFusion logo, which includes an orange circle with a white shield-like shape inside. To the right of the logo is the text "CareFusion" in a dark gray, sans-serif font. Below the company name, in a smaller, lighter font, is the phrase "has joined BD".

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, followed by the text "CareFusion" in gray. Below the text, in smaller gray font, it says "has joined BD."

8. Conclusion

The test results demonstrate that the device is as safe and effective as the predicate and therefore substantially equivalent to the predicate device.

§ 868.5270 Breathing system heater.

(a)
Identification. A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.(b)
Classification. Class II (performance standards).