K Number
K151280

Validate with FDA (Live)

Date Cleared
2015-09-30

(139 days)

Product Code
Regulation Number
892.1650
Age Range
12 - 120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DigiArc 100AU/ DigiArc 100AU+ is a mobile digital X-ray G-Arm diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery, it is not intended to be used in interventional procedures. The DigiArc 100AU/ DigiArc 100AU+ permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients at the age of at least eighteen.

Device Description

The DigiArc 100AU/ DigiArc 100AU+ is a mobile digital X-ray G-Arm diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery, it is not intended to be used in interventional procedures. There are two sets of X-ray tube assemblies and Image Intensifiers which are perpendicularly distributed on the G-Arm, acting as two sets of vertical X-ray source and receptor systems and providing fluoroscopy image of the patient. The two sets of X-ray tube assemblies and Image Intensifiers can operate simultaneously and separately.

AI/ML Overview

The provided FDA 510(k) summary describes the "MultiScan G-Arm System" (DigiArc 100AU & DigiArc 100AU+) by Beijing East Whale Imaging Technology Co., Ltd. The document primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical performance and safety testing against recognized standards. It does not detail specific acceptance criteria for diagnostic accuracy or device performance in the context of clinical image interpretation or AI-assisted diagnoses, nor does it present a study proving the device meets such criteria. The device is a standard X-ray fluoroscopic imaging system, not an AI-powered diagnostic tool.

Therefore, the requested information regarding acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert ground truth, MRMC studies, or standalone algorithm performance is not present in this document. The provided text exclusively covers the device's technical specifications, safety, and compliance with general medical device standards.

However, I can extract the information that is present concerning the device's compliance with safety and performance standards, which are the "acceptance criteria" discussed in this 510(k) submission.

1. A table of acceptance criteria and the reported device performance

The "acceptance criteria" in this context are compliance with various recognized safety and performance standards for medical electrical equipment and X-ray systems. The reported device performance for these criteria is consistently "Pass."

Acceptance Criteria (Standard)Reported Device Performance
AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (General Requirements for Basic Safety and Essential Performance of Medical Electrical Equipment)Pass
- Power inputPass
- Humidity preconditioning treatmentPass
- Determination of applied parts and accessible partsPass
- Legibility of markingsPass
- Durability of markingsPass
- ME equipment for connection to a power source by a plugPass
- Internal capacitive circuitsPass
- Protective earth connectionPass
- Leakage currentsPass
- Dielectric strengthPass
- Ball pressure testPass
- Measurement of creepage distance and air clearancePass
- Cord anchoragePass
- GapsPass
- Instability-overbalance in transport positionPass
- Instability-overbalance excluding transport positionPass
- Instability-overbalance from horizontal and vertical forcesPass
- Audible acoustic energyPass
- Maximum temperature during normal usePass
- SpillagePass
- Ingress of water or particulate matterPass
- CleaningPass
- Interruption of the power supply/supply mains to ME equipmentPass
- Single fault conditionsPass
- Mechanical strength testPass
- Transformer short circuitPass
- Transformer overload testPass
- ME system-leakage measurementsPass
- Half-value layerPass
- Focal spot to image receptor distancePass
- Leakage radiation in the loading statePass
- Leakage radiation when not in the loading statePass
IEC 60601-1-3 Edition 2.0 2008 (Radiation protection in diagnostic X-ray equipment)Pass
IEC 60601-2-54 Edition 1.0 2009-06 (Specific requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy)Pass
- Test for dosimetric informationPass
- Accuracy of X-ray tube voltagePass
- Accuracy of X-ray tube currentPass
- Determining the attenuation of residual radiationPass
IEC 60601-1-2 Edition 3: 2007-03 (Electromagnetic compatibility - Requirements and tests)Pass
- Harmonics on AC mainsPass
- Voltage fluctuations on AC mainsPass
- Mains terminal continuous disturbance voltagePass
- Radiated emissionPass
- Electrostatic dischargePass
- RF electromagnetic field immunity testPass
- Fast transients on AC power line, signal line and interconnecting linePass
- Injected current into AC power line, signal line and interconnecting linePass
- Surges to AC power port, signal line and interconnecting linePass
- Voltage dips and interruptions to AC power portPass
- Variations of power frequencyPass
Digital Imaging Communications in Medicine (DICOM) provisionsMeets

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The testing described is non-clinical performance and safety testing for the device itself, not a study involving patient data or clinical images for diagnostic accuracy assessment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. This type of information would be relevant for studies evaluating diagnostic performance, which is not the subject of this 510(k) submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document. Adjudication methods are used in clinical studies to establish ground truth for diagnostic decisions, which is not applicable to the non-clinical testing presented here.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not done. This device is an X-ray imaging system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study of an algorithm was not done. This device is an imaging system, not an algorithm, and the submission does not describe any AI component.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The concept of "ground truth" as it relates to diagnostic accuracy is not applicable to the non-clinical performance and safety testing described. The "ground truth" for these tests are the objective pass/fail criteria defined by the respective international and national standards (e.g., specific voltage/current ranges, radiation leakage limits, insulation resistance, etc.).

8. The sample size for the training set

This information is not provided and is not applicable, as this device itself is not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established

This information is not provided and is not applicable, as this device itself is not an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design, symbolizing the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 30, 2015

Beijing East Whale Imaging Technology Co., Ltd. % Miss June Li Official Correspondent B2-2 New City Industrial Park, No. 9 KeChuang 2nd Street YiZhuang, Beijing, 100023 CHINA

Re: K151280

Trade/Device Name: MultiScan G-Arm System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO, JAA Dated: August 31, 2015 Received: September 2, 2015

Dear Miss Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ocko

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151280

Device Name MultiScan G-Arm System

Indications for Use (Describe)

The DigiArc 100AU/ DigiArc 100AU+ is a mobile digital X-ray G-Arm diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery, it is not intended to be used in interventional procedures. The DigiArc 100AU/ DigiArc 100AU+ permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients at the age of at least eighteen.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a logo that appears to be a stylized letter "E" in green, enclosed in a white square with a blue border. The logo is smaller than the word "WHALE" and is positioned slightly above the baseline of the text.

510(k) Summary of Safety and Effectiveness

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR807.92 (a) (1)]

Aug 28th 2015

2. Submitter' s Information [21 CFR807.92 (a) (1)]

Name of Sponsor: Beijing East Whale Imaging Technology Co., Ltd.

  • Address: B2-2 New City Industrial Park, No.9 KeChuang 2nd St.,
    YiZhuang, 100023, Beijing, China.

  • Contact Name: June Li

  • Telephone No.: + 86 (10) 67892355- 8968

  • Fax No.: + 86 (10) 67082218

  • Email Address: jqli@whaleimaging.com

3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)]

  • Trade Name: DigiArc 100AU & DigiArc 100AU+
  • Common Name: MultiScan G-Arm System
  • Classification: Image-intensified fluoroscopic X-ray system

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a logo that appears to be a stylized letter "E" inside of a rounded square. The letter "E" is made up of two horizontal green bars.

Product code: OXO, JAA Classification Panel: Radiology ll Device Class:

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[

Table 1- Predicate Devices

510(k) NumberK131423K141672
ApplicantBeijing East Whale Imaging Technology Co., Ltd.
Device NameMultiScan G-Arm System

5. Description of the Device [21 CFR 807.92(a)(4)]

The DigiArc 100AU/ DigiArc 100AU+ is a mobile digital X-ray G-Arm diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery, it is not intended to be used in interventional procedures.

There are two sets of X-ray tube assemblies and Image Intensifiers which are perpendicularly distributed on the G-Arm, acting as two sets of vertical X-ray source and receptor systems and providing fluoroscopy image of the patient. The two sets of X-ray tube assemblies and Image Intensifiers can operate

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a small, white square with rounded corners. Inside the square are two horizontal green bars, stacked on top of each other. The logo appears to be for a company or organization named "WHALE".

simultaneously and separately.

The DigiArc 100AU & DigiArc 100AU+ include below primary components.

Table 2- Primary components list

ComponentQuantity
Control unit1
Viewing monitor2
Control monitor1
Control panel1
G-Arm1
Image intensifier assembly2
X-ray tube assembly2
Foot switch subassembly1
Laser aimer system(optional)1
Tracking wheel system(optional)1
Printer (optional)1

6. Intended Use [21 CFR 807.92(a)(5)]

The current modifications do not change the indications for use. As previously reported and cleared: the DigiArc 100AU/ DigiArc 100AU+ is a mobile digital X-ray G-Arm diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery, it is not intended to be used in interventional procedures. The DigiArc 100AU/ DigiArc

{6}------------------------------------------------

100AU+ permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients at the age of at least eighteen.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

ItemPredicate deviceDigiArc 100AUPredicate deviceDigiArc 100AU+Proposed deviceDigiArc 100AUProposed deviceDigiArc 100AU+Note
Trackingwheel systemand Laseraimer systemNoYesOptionalSE
X-ray Output:FluoroscopyRange of X-rayvoltage: 40—110kVRange of X-raycurrent: 0.1—15mARange of X-rayvoltage: 40—110kVRange of X-raycurrent: 0.1—15mARange of X-rayvoltage:40—110 kV (AU)40—120 kV (AU+)Range of X-raycurrent: 0.1—15mASE
Totalequivalentfiltration4.0 mmAl(at 80kV)4.0 mmAl(at 80kV)7.1 mmAl(at 80kV)SE
Softwareversion1.1.01.3.0DigiArc 100AU:V1.6.0.1DigiArc 100AU+:V1.6.0.2SE

Table 3- Major similarities and differences between subject and predicate devices

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a small, white square with a green design inside. The green design appears to be two horizontal bars with a white line in between them.

Others--Same
------------------

The design modifications do not alter the device's fundamental scientific technology, and the updated device has passed all the tests in according to relevant standards. The DigiArc 100AU & DigiArc 100AU+ employs the same technological characteristics as the predicate devices.

8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]

Results of performance and compliance testing conducted on DigiArc 100AU & DigiArc 100AU+ indicate conformance to all applicable standards recognized by FDA for this device.

Testing result from non-clinical demonstrates that the proposed device DigiArc 100AU & DigiArc 100AU+ are as safe and effective as the predicate devices.

Non-clinical testing:

The proposed device has been tested to compliance to the following safety and performance standards:

  • a) AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012
    Table 4- Major testing items of AAMI/ANSI ES 60601-1
Testing itemsresult
Power inputPass
Humidity precondiitioning treatmentPass
Determination of applied parts and accessible partsPass
Legibility of markingsPass
Durability of markingsPass
ME equipment for connection to a power source by a plugPass
Internal capacitive circuitsPass
Protective earth connectionPass
Leakage currentsPass
Dielectric strengthPass
Ball pressure testPass
Measurement of creepage distance and air clearancePass
Cord anchoragePass
GapsPass
Instability-overbalance in transport positionPass
Instability-overbalance excluding transport positionPass
Instability-overbalance from horizontal and vertical forcesPass
Audible acoustic energyPass
Maximum temperature during normal usePass
SpillagePass
Ingress of water or particulate matterPass
CleaningPass
Interruption of the power supply/supply mains to ME equipmentPass
Single fault conditionsPass
Mechanical strength testPass
Transformer short circuitPass
Transformer overload testPass
ME system-leakage measurementsPass
Testing itemsresult
Half-value layerPass
Focal spot to image receptor distancePass
Leakage radiation in the loading statePass
Leakage radiation when not in the loading statePass

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the "E" is a small, white square with two horizontal green bars inside. The logo is simple and clean, with a focus on the company name.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a small, square graphic with a white background and two horizontal green bars inside. The graphic is positioned next to the letter "E" in "WHALE."

b) IEC 60601-1-3 Edition 2.0 2008

Table 5- Major testing items of IEC 60601-1-3

c) IEC 60601-2-54 Edition 1.0 2009-06

Table 6- Major testing items of IEC 60601-2-54

Testing itemsresult
Test for dosimetric informationPass
Accuracy of X-ray tube voltagePass
Accuracy of X-ray tube currentPass
Determining the attenuation of residual radiationPass

d) IEC 60601-1-2 Edition 3: 2007-03

Table 7- Major testing items of IEC 60601-1-2

Testing itemsresult
Harmonics on AC mainsPass
Voltage fluctuations on AC mainsPass
Mains terminal continuous disturbance voltagePass
Radiated emissionPass
Electrostatic dischargePass
RF electromagnetic field immunity testPass

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the word "WHALE" in large, bold, blue letters. To the right of the word is a logo that appears to be a stylized letter "E" in green, enclosed in a white box with a gray border. The logo is positioned so that the "E" is partially obscured by the last letter of the word "WHALE".

Fast transients on AC power line, signal line and interconnecting linePass
Injected current into AC power line, signal line and interconnecting linePass
Surges to AC power port, signal line and interconnecting linePass
Voltage dips and interruptions to AC power portPass
Variations of power frequencyPass

Also the proposed device meets the provisions of Digital Imaging communications in Medicine (DICOM)

9. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the non-clinical result and relative information provided in this premarket notification, we concludes that DigiArc 100AU & DigiArc 100AU+ are substantially equivalent to predicate devices with regard to safety and effectiveness.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.