K Number
K151248

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-09-04

(116 days)

Product Code
Regulation Number
872.4120
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Piezosurgery White is a piezoelectric ultrasonic device, consisting of handpieces and associated tip inserts, intended for:

  • Bone cutting, osteotomy, osteoplasty and drilling in a variety of oral surgical procedures, including implantology, periodontal surgery, surgical orthodontic, and surgical endodontic procedures;
  • Scaling applications, including:
  • Scaling: All general procedures for removal of supragingival and interdental calculus & plaque deposits;
  • Periodontology: Periodontal therapy and debridement for all types of periodontal diseases, including periodontal pocket irrigation and cleaning:
  • Endodontics: All treatments for root canal reaming, irrigation, filling, gutta-percha condensation and retrograde preparation;
  • Restorative and Prosthetics: Cavity preparation, removal of prostheses, amalgam condensation, finishing of crown preparations and inlay/onlay condensation.
Device Description

The Piezosurgery White uses ultrasonic energy to generate mechanical micro-vibration of the available tip inserts to perform the dental procedures defined in its intended use. The Piezosurgery White consists of a table-top unit (console) containing the irrigation delivery system, the internal electric power supply, the ultrasonic generator, and the control keyboard. The device also includes the piezoelectric ultrasonic handpiece and foot-pedal, both connected directly to the console by means cords, and a variety of insert tips designed with different morphologies/shapes to be used for different dental procedures, according to device's intended use. Inside the console are located the ultrasonic generator, the electrical power supply module and the microprocessor electronic board that controls and supervises the functional parameters of the device. The console is connected to mains power by an electrical cord, and includes connectors for the handpiece and for the footswitch. One external peristaltic pump, intended for delivery of saline solution to the point of use, is attached to the console. The console is equipped with a keyboard that allows the surgeon to set/control the functions of the device. The keyboard also provides monitoring information, by means icons, in case of malfunction of the device. The ultrasonic power and water flow are simultaneously activated by pressing the footswitch. The ultrasonic generator of the Piezosurgery White is microprocessor-based and uses electronics to generate and control the appropriate drive signal (power and frequency) for the ultrasonic transducer, which is housed inside the handpiece. The touch panel of the console allows setting of the following treatment/operative parameters: Six different types of treatment (IMPLANT, CORTICAL, CANCELLOUS, SPECIAL, PERIO, ENDO): Seven different levels of irrigation (from 0 to 6); The "pump/flush" function. The Piezosurgery White uses piezoelectric ultrasonic technology to generate mechanical micro-vibrations of the tip insert attached to the handpiece. A piezoelectric transducer, located inside the handpiece, and driven by the ultrasonic generator electronics, induces vibrations at ultrasonic frequencies in the insert tip. The ultrasonic generator electronics and locates the resonant frequency of the transducer/insert combination, which varies according to the geometry/morphology of the tip insert in use. The functional ultrasonic frequency of the device is between approximately 24 and 36 kHz. The Piezosurgery White handpiece transducer and family of inserts are designed to resonate in this range of frequencies.

AI/ML Overview

This document is a 510(k) summary for the MECTRON S.p.A. Piezosurgery White device. It highlights the device's equivalence to a predicate device rather than providing acceptance criteria for a new AI/ML device. Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert ground truth, adjudication, MRMC studies, standalone performance, and training set details for a medical device that uses AI/ML is not available in the provided text.

The document focuses on demonstrating substantial equivalence to an existing predicate device (Piezosurgery Touch) for regulatory approval, which is a different type of evaluation than setting and meeting performance criteria for novel AI/ML functionalities.

Here's a breakdown of what is available in the document, framed against the requested items, to illustrate why the specific AI/ML performance criteria are not present:

Context of the Document:
This is an FDA 510(k) summary for a medical device called "Piezosurgery White." The purpose of a 510(k) submission is to demonstrate that a new medical device is "substantially equivalent" to a legally marketed predicate device, thereby not requiring a new premarket approval (PMA).

Piezosurgery White Device Description:
The Piezosurgery White is a piezoelectric ultrasonic device used for various oral surgical procedures (bone cutting, osteotomy, drilling) and scaling applications (calculus removal, periodontal therapy, endodontics, restorative/prosthetic procedures). It uses ultrasonic energy to generate mechanical micro-vibrations in its tip inserts.

Comparison to Predicate Device (Piezosurgery Touch):
The document extensively compares the Piezosurgery White to its predicate device, the Piezosurgery Touch (K122322). The key takeaway is that both devices use identical piezoelectric ultrasonic technology, have similar components, intended uses, operational characteristics, and output parameters. The main differences noted are minor cosmetic changes to the external design and the introduction of ten new inserts, for which specific functional and design verification was performed.


Addressing the specific categories requested, based on the provided text:

  1. A table of acceptance criteria and the reported device performance:

    • N/A. The document does not describe acceptance criteria and reported performance in the context of an AI/ML device's diagnostic or predictive capabilities. Instead, it demonstrates "substantial equivalence" to a predicate device.
    • The "performance" discussed relates to compliance with recognized standards and similarity to the predicate device's operational characteristics (e.g., operational vibration frequency: 24 kHz to ~36 kHz, Max. Output: 25 Watt for both subject and predicate devices). This is about technological specifications rather than a performance metric against a disease state or outcome.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • N/A. The document does not describe a "test set" in the context of an AI/ML algorithm's performance evaluation against a dataset. The testing was primarily non-clinical (bench tests, compliance with safety and EMC standards, sterilization validation, software verification) to demonstrate the device's physical and functional equivalence to the predicate.
    • For the new inserts (PR1 and PR2), "Functional bench tests conducted on inserts PR1 and PR2 when attached to the Piezosurgery White handpiece verify that the operative resonant frequency of the inserts is within the functional ultrasonic frequency range of the Piezosurgery White." No specific sample size for these bench tests is given, nor is "data provenance" relevant here.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • N/A. Ground truth, in the context of expert consensus for AI/ML evaluation, is not applicable here as the device does not make diagnostic or predictive assessments. The evaluation is against engineering specifications and comparison to a predicate device.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • N/A. Not applicable, as there's no diagnostic "test set" and no expert adjudication process described.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A. This device does not involve "human readers" or "AI assistance" in a diagnostic or interpretive capacity. It is a surgical tool. Therefore, an MRMC study is not relevant.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • N/A. This device is a surgical instrument, not an AI/ML algorithm that operates standalone. Its "performance" is its ability to meet physical and functional specifications.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • N/A. The "ground truth" for demonstrating substantial equivalence for this type of device primarily relies on engineering specifications, compliance with international standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 17665-1, ISO 14937), and direct comparison of its technological characteristics and intended use to a legally marketed predicate device. For new inserts, Finite Element Analysis (FEA) and functional bench tests were used to verify resonant frequency.
  8. The sample size for the training set:

    • N/A. This document does not describe an AI/ML algorithm with a training set. The "software" mentioned (Page 8, Item 8) refers to the device's internal control software, which underwent "verification and validation in accordance with FDA guidance," not training on a dataset for learning tasks.
  9. How the ground truth for the training set was established:

    • N/A. Not applicable, as there is no AI/ML training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a human face in profile, with three overlapping profiles to represent the department's focus on health, human services, and the well-being of the nation. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 4, 2015

MECTRON S.p.A. c/o Mr. Roger Gray Donawa Lifescience Consulting Srl Piazza Albania, 10 00153 Rome ITALY

Re: K151248

Trade/Device Name: Piezosurgery White Regulation Number: 21 CFR 872.4120 Regulation Name: Bone cutting instrument and accessories Regulatory Class: II Product Code: DZI, ELC Dated: June 4, 2015 Received: June 8, 2015

Dear Mr. Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Roger Gray

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K151248

Device Name PIEZOSURGERY WHITE

Indications for Use (Describe)

The Piezosurgery White is a piezoelectric ultrasonic device, consisting of handpieces and associated tip inserts, intended for:

  • · Bone cutting, osteotomy, osteoplasty and drilling in a variety of oral surgical procedures, including implantology, periodontal surgery, surgical orthodontic, and surgical endodontic procedures;
  • · Scaling applications, including:
  • Scaling: All general procedures for removal of supragingival and interdental calculus & plaque deposits;
  • Periodontology: Periodontal therapy and debridement for all types of periodontal diseases, including periodontal pocket irrigation and cleaning:
  • Endodontics: All treatments for root canal reaming, irrigation, filling, gutta-percha condensation and retrograde preparation;
  • Restorative and Prosthetics: Cavity preparation, removal of prostheses, amalgam condensation, finishing of crown preparations and inlay/onlay condensation.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "mectron" in orange font. Below the word is a gray line with a series of gray circles of increasing size along the line. The circles are evenly spaced and appear to be part of a logo or branding element.

Mectron S.p.A. Via Loreto, 15/A 16042 Carasco – GE (Italy) Tel. +39 0185 35361 Fax +39 0185 351374 www.mectron.com

medical technology

P.IVA IT00177110996 N. Iscr. Reg. Imprese Genova e C.Fisc.: 01126960101 R.E.A. Genova 253624 Cap. Soc. Euro 4.286.000 i.v. mectron@mectron.com

510(k) Summary in accordance with 21 CFR 807.92

ADMINISTRATIVE INFORMATION 1.

Type of 510(k) submission:Traditional
Submission date:27 August 2015
510(k) Submitter:MECTRON S.p.aVia Loreto 1516042 Carasco - (GE) - ITALY
Phone: +39 0185 35 361Fax: +39 0185 351 374
510(k) Contact Person:Roger GrayVP, Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania, 1000153 Rome, Italy
Phone: +39 06 578 2665Fax: +39 06 574 3786email: rgray@donawa.com
Manufacturer:MECTRON S.p.aVia Loreto 1516042 Carasco - (GE) - ITALY
Establishment Registration Number: 3003933619
2.DEVICE
Trade name of the device:Piezosurgery White
Common Name/Regulation Description:Bone cutting instrument and accessories
Classification Regulation:21 CFR 872.4120
FDA Panel:Dental
Product Code:DZI ELC
Classification:Class II

3. IDENTIFICATION OF THE PREDICATE DEVICE

The substantial equivalence of the subject device is based on the following legally marketed predicate device

Trade nameManufacturerProduct Codes510(k) NumberDecision Date
Piezosurgery TouchMectron S.p.aDZI, ELCK12232212/06/2012

DEVICE DESCRIPTION 4.

The Piezosurgery White uses ultrasonic energy to generate mechanical micro-vibration of the available tip inserts to perform the dental procedures defined in its intended use.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "mectron" in orange font. Below the word is a gray line with several gray circles on it. The circles are evenly spaced along the line.

Mectron S.p.A. Via Loreto, 15/A 16042 Carasco - GE (Italy) Tel. +39 0185 35361 Fax +39 0185 351374 www.mectron.com

medical technology

P.IVA IT00177110996 N. Iscr. Reg. Imprese Genova e C.Fisc.: 01126960101 R.E.A. Genova 253624 Cap. Soc. Euro 4.286.000 i.v. mectron@mectron.com

The Piezosurgery White consists of a table-top unit (console) containing the irrigation delivery system, the internal electric power supply, the ultrasonic generator, and the control keyboard. The device also includes the piezoelectric ultrasonic handpiece and foot-pedal, both connected directly to the console by means cords, and a variety of insert tips designed with different morphologies/shapes to be used for different dental procedures, according to device's intended use.

Inside the console are located the ultrasonic generator, the electrical power supply module and the microprocessor electronic board that controls and supervises the functional parameters of the device.

The console is connected to mains power by an electrical cord, and includes connectors for the handpiece and for the footswitch. One external peristaltic pump, intended for delivery of saline solution to the point of use, is attached to the console.

The console is equipped with a keyboard that allows the surgeon to set/control the functions of the device. The keyboard also provides monitoring information, by means icons, in case of malfunction of the device. The ultrasonic power and water flow are simultaneously activated by pressing the footswitch.

The ultrasonic generator of the Piezosurgery White is microprocessor-based and uses electronics to generate and control the appropriate drive signal (power and frequency) for the ultrasonic transducer, which is housed inside the handpiece.

The touch panel of the console allows setting of the following treatment/operative parameters:

  • Six different types of treatment (IMPLANT, CORTICAL, CANCELLOUS, SPECIAL, PERIO, . ENDO):
  • Seven different levels of irrigation (from 0 to 6); .
  • . The "pump/flush" function.

The Piezosurgery White uses piezoelectric ultrasonic technology to generate mechanical micro-vibrations of the tip insert attached to the handpiece. A piezoelectric transducer, located inside the handpiece, and driven by the ultrasonic generator electronics, induces vibrations at ultrasonic frequencies in the insert tip.

The ultrasonic generator electronics and locates the resonant frequency of the transducer/insert combination, which varies according to the geometry/morphology of the tip insert in use. The functional ultrasonic frequency of the device is between approximately 24 and 36 kHz. The Piezosurgery White handpiece transducer and family of inserts are designed to resonate in this range of frequencies.

The Piezosurgery White complies with the electrical safety and electromagnetic compatibility requirements established by the standards IEC 60601-1 and IEC 60601-1-2.

INTENDED USE/INDICATIONS FOR USE 5.

Piezosurgery White is a piezoelectric ultrasonic device, consisting of handpieces and associated tip inserts, intended for:

  • Bone cutting, osteotomy, osteoplasty and drilling in a variety of oral surgical procedures, including . implantology, periodontal surgery, surgical orthodontic and surgical endodontic procedures;
  • · Scaling applications, including:
    • Scaling: All general procedures for removal of supragingival and interdental calculus & plaque deposits;
    • Periodontology: Periodontal therapy and debridement for all types of periodontal diseases, including periodontal pocket irrigation and cleaning:
    • Endodontics: All treatments for root canal reaming, irrigation, filling, gutta-percha condensation and retrograde preparation;
    • Restorative and Prosthetics: Cavity preparation, removal of prostheses, amalgam condensation, finishing of crown preparations and inlay/onlay condensation.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "mectron" in orange font. Below the word is a gray line with five gray circles of varying sizes along the line. The circles are evenly spaced, with the smallest circle on the left and the largest circle on the right.

Mectron S.p.A. Via Loreto, 15/A 16042 Carasco – GE (Italy) Tel. +39 0185 35361 Fax +39 0185 351374 www.mectron.com

medical technology

P.IVA IT00177110996 N. Iscr. Reg. Imprese Genova e C.Fisc.: 01126960101 R.E.A. Genova 253624 Cap. Soc. Euro 4.286.000 i.v. mectron@mectron.com

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE 6.

Piezoelectric ultrasonic technology is used both by the subject and the predicate device to generate mechanical micro-vibrations of the insert connected to the handpiece. The subject device and predicate device share many characteristics, including identical piezoelectric transducers and ultrasonic generators, as shown in the following comparison table.

ItemSubject DevicePiezosurgery WhitePredicate DevicePiezosurgery TouchEquivalence
ManufacturerMECTRON SPAVia Loreto 1516042 Carasco - ItalyMECTRON SPAVia Loreto 1516042 Carasco - ItalyN/A
510(k) NumberSubject of this submissionK122322N/A
FDA PanelDentalDentalSame
Regulation Numberand name872.4120 - Bone cutting instrument andaccessories872.4120 - Bone cutting instrument andaccessories872.4850 - Ultrasonic ScalerSame
Product CodeDZI, ELCDZI, ELCSame
Regulatory ClassClass IIClass IISame
TechnologicalCharacteristics(mechanism ofaction )Piezoelectric ultrasonic technology whichgenerates mechanical micro-vibrations ofthe insert tips. The piezoelectrictransducer uses piezoceramic disks toconvert the generator's electrical signal toultrasonic vibration of the insert tip.Piezoelectric ultrasonic technology whichgenerates mechanical micro-vibrations ofthe insert tips. The piezoelectrictransducer uses piezoceramic disks toconvert the generator's electrical signal toultrasonic vibration of the insert tip.Same
Device typeTable top deviceTable top deviceSame
Indications for Use /Intended UsePiezosurgery White is a piezoelectricultrasonic device, consisting ofhandpieces and associated tip inserts,intended for:● Bone cutting, osteotomy,osteoplasty and drilling in a varietyof oral surgical procedures, includingimplantology, periodontal surgery,surgical orthodontic, and surgicalendodontic procedures;● Scaling applications, including:- Scaling: All general procedures forremoval of supragingival andinterdental calculus & plaquedeposits;- Periodontology: Periodontaltherapy and debridement for alltypes of periodontal diseases,including periodontal pocketirrigation and cleaning;- Endodontics: All treatments forroot canal reaming, irrigation,revision, filling, gutta-perchacondensation and retrogradepreparation;- Restorative and Prosthetics:Cavity preparation, removal ofprostheses, amalgam condensation,finishing of crown preparations andinlay/onlay condensation.Piezosurgery Touch is a piezoelectricultrasonic device, consisting ofhandpieces and associated tip inserts,intended for:● Bone cutting, osteotomy,osteoplasty and drilling in a varietyof oral surgical procedures, includingimplantology, periodontal surgery,surgical orthodontic, and surgicalendodontic procedures;● Scaling applications, including:- Scaling: All general procedures forremoval of supragingival andinterdental calculus & plaquedeposits;- Periodontology: Periodontaltherapy and debridement for alltypes of periodontal diseases,including periodontal pocketirrigation and cleaning;- Endodontics: All treatments forroot canal reaming, irrigation,revision, filling, gutta-perchacondensation and retrogradepreparation;- Restorative and Prosthetics:Cavity preparation, removal ofprostheses, amalgam condensation,finishing of crown preparations andinlay/onlay condensation.Same
ItemSubject DevicePiezosurgery WhitePredicate DevicePiezosurgery TouchEquivalence
Main devicecomponentsControl unit with integrated irrigation peristaltic pump, electric power supply and ultrasonic generator. Piezoelectric ultrasonic handpiece and cord. Torque wrench to tighten the inserts on the hand piece. Footswitch with cord. Bracket/s for irrigation solution bag. Power supply cord. Range of insert tips to be used according to the dental applications defined by the intended use. Other accessories/attachments.Control unit with integrated irrigation peristaltic pump, electric power supply and ultrasonic generator. Piezoelectric ultrasonic handpiece and cord. Torque wrench to tighten the inserts on the hand piece. Footswitch with cord. Bracket/s for irrigation solution bag. Power supply cord. Range of insert tips to be used according to the dental applications defined by the intended use. Other accessories and attachments.Same
TransducertechnologicalcharacteristicsPiezoelectric:The transducer uses four hardpiezoceramic disks to convert thegenerator's electrical signal to ultrasonicvibration of the insert tip.Piezoelectric:The transducer uses four hardpiezoceramic disks to convert thegenerator's electrical signal to ultrasonicvibration of the insert tip.Same
Automaticresonance frequencytuningElectronics with automatic frequencytuning.Electronics with automatic frequencytuning.Same
Modulated amplitudeof the ultrasonicsignal modeYesYesSame
Output performancespecificationOperational Vibration frequency:24 kHz to ~ 36 kHz.Max. Output: 25 WattOperational Vibration frequency:24 kHz to ~ 36 kHz.Max. Output: 25 WattSame
Supply Voltage100 - 240 VAC100 - 240 VACSame
Output ModesSix output modes identified as Endo,Perio, Special, Cancellous, Cortical andImplant on the touch panel.Six output modes identified as Endo,Perio, Special, Cancellous, Cortical andImplant on the touch panel.Same
ActivationFootswitch connected to the controlconsole by means of a cord.Footswitch connected to the controlconsole by means of a cord.Same
Irrigation systemIrrigation bag connected to a peristalticpump. Adjustable irrigation flow rate.Irrigation bag connected to a peristalticpump. Adjustable irrigation flow rate.Same
Adjustable irrigationflow rate.Seven irrigation flow rates, from0 to 75 ml/minSeven irrigation flow rates, from0 to 75 ml/minSame
Electrical safetyComplies with IEC 60601-1Complies with IEC 60601-1Same
ElectromagneticCompatibilityComplies with IEC 60601-1-2Complies with EN 60601-1-2 which isequivalent to IEC 60601-1-2Same
Does the devicecontain software?YesYesSame
Software ValidationIn accordance with FDA Guidance:"Guidance for the Content of PremarketSubmissions for Software Contained inMedical Devices"In accordance with FDA Guidance:"Guidance for the Content of PremarketSubmissions for Software Contained inMedical Devices"Same
Parts need to besterilizedYesYesSame
Recommendedsterilization methodAutoclaving (Steam Sterilization)Autoclaving (Steam Sterilization)Same
Single use partsDiamond coated insertsDiamond coated insertsSame
ItemSubject DevicePiezosurgery WhitePredicate DevicePiezosurgery TouchEquivalence
Tip InsertsMaterial: Stainless steelThe insert tips are secured to thehandpiece by a thread, using a torquewrench.Material: Stainless steelThe insert tips are secured to thehandpiece by a thread, using a torquewrench.Same
Console DimensionsWidth: 298,5 mmHeight: 84 mmDepth: 245 mmWidth: 300 mmHeight: 95 mmDepth: 250 mmVery similar

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "mectron" in orange font. Below the word is a gray line with several gray circles of varying sizes along the line. The circles are evenly spaced.

medical technology

Mectron S.p.A. Via Loreto, 15/A 16042 Carasco – GE (Italy) Tel. +39 0185 35361
Fax +39 0185 351374 www.mectron.com

P.IVA IT00177110996 N. Iscr. Reg. Imprese Genova e C.Fisc.: 01126960101 R.E.A. Genova 253624 Cap. Soc. Euro 4.286.000 i.v. mectron@mectron.com

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "mectron" in orange font. Below the word is a gray line with several gray circles of varying sizes along the line. The circles are evenly spaced.

medical technology

Mectron S.p.A Via Loreto, 15/A 16042 Carasco - GE (Italy) Tel. +39 0185 35361 Fax +39 0185 351374 www.mectron.com

P.IVA IT00177110996 N. Iscr. Reg. Imprese Genova e C.Fisc.: 01126960101 R.E.A. Genova 253624 Cap. Soc. Euro 4.286.000 i.v. mectron@mectron.com

SUBSTANTIAL EQUIVALENCE DISCUSSION 7.

The Piezosurgery White and the predicate device, Piezosurgery Touch, share identical indications for use / intended use.

Both devices combine the functions of a bone cutting instrument and of an ultrasonic dental scaler.

Piezosurgery White uses the same fundamental technology as the predicate device, utilizing the same piezoelectric ultrasonic technology to generate mechanical micro-vibration of the insert tips fitted onto the handpiece.

The ultrasonic generator of the Piezosurgery White. which generates and controls the appropriate drive signal (power and frequency) to the transducer, is identical in design, manufacture and function to the ultrasonic generator used in the predicate device, Piezosurgery Touch, including the availability of a low frequency burst mode.

Apart from the minor cosmetic differences existing between the external designs of the subject and predicate device cases, both the subject device and the predicate device consist essentially of a table-top unit with incorporated touch keyboard, peristaltic pump, footswitch, ultrasonic handpiece/s, hoses/connectors, a variety of tips for bone cutting, and a torque wrench for attaching the tips to the handpiece.

Compared to predicate device, Piezosurgery White presents the same operational characteristics and user interface, including identical, operating parameters, which can be selected/adjusted by the operator through the keyboard, specifically: Six output modes, seven pre-set flow rates of the irrigation fluid, and a 'Pump/Flush' function that allows the operator to prime the entire irrigation circuit

The touch keyboard of the predicate device Piezosurgery Touch has the additional function "Light" that allows the operator to manage the LED illumination of the handpiece, whereas on the subject device, when a handpiece incorporating an LED light is used, it is switched on by pressing the foot-pedal.

Except for ten new inserts, introduced with this submission, whose minimal differences compared to predicate inserts and/or to equivalent manual instruments do not raise new questions of safety or performance, all the inserts intended for use with the Piezosurgery White have been previously cleared by FDA for use with the predicate device.

Inserts PR1 and PR2, introduced in this submission, have shapes and intended use similar to existing manual periosteal elevators, having a frontal semi-circular profile with a radius of 2 mm and 5 mm respectively. The shape and dimensions of these inserts are suitable for generating ultrasonic longitudinal vibrational behaviour that mimics use of manual periosteal elevators to allow periosteum detachment by following the anatomical profile.

Inserts PR1 and PR2 have very similar shapes and designs to already-cleared inserts OP3 and EL3, having circular flat geometry of the operative part. Insert EL3 also has a very similar use to inserts PR1 and PR2, being intended for detaching/separating the sinus membrane.

Finite Element Analysis (FEA) was utilized in the design of inserts PR1 and PR2 to ensure that the resonant ultrasonic frequency of the inserts was within the functional ultrasonic frequency range of the

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "mectron" in orange font. Below the word is a gray line with several gray circles of varying sizes along the line. The circles are spaced evenly apart.

Mectron S.p.A Via Loreto, 15/A 16042 Carasco - GE (Italy) Tel. +39 0185 35361 Fax +39 0185 351374 www.mectron.com

medical technology

P.IVA IT00177110996 N. Iscr. Reg. Imprese Genova e C.Fisc.: 01126960101 R.E.A. Genova 253624 Cap. Soc. Euro 4.286.000 i.v. mectron@mectron.com

Piezosurgery White. Functional bench tests conducted on inserts PR1 and PR2 when attached to the Piezosurgery White handpiece verify that the operative resonant frequency of the inserts is within the functional ultrasonic frequency range of the Piezosurgery White.

Inserts PR1 and PR2 share the same intended use as manual periosteal elevators, have very similar operative part shapes, and functional testing has demonstrated compliance with specifications, thus no new issues of safety and effectiveness are raised.

8. SUMMARY OF NON-CLINICAL TESTING

Non-clinical testing was performed on the Piezosurgery White and associated inserts to support the claims of substantial equivalence, including:

  • Testing to confirm compliance with safety requirements of IEC 60601-1.
  • Testing to confirm compliance with EMC requirements of IEC 60601-1-2.
  • Testing to confirm compliance with the safety requirements of IEC 80601-2-60.
  • . Software documentation, verification and validation in accordance with FDA quidance for the Content of Premarket Submissions for Software Contained in Medical Devices', 11 May 2005.
  • Sterilization validation in accordance with ISO 17665-1 and ISO 14937.
  • Bench tests to demonstrate substantial equivalence to the predicate device, including:
    • Current to the handpiece. o
    • Working frequency range. O
    • Leakage current. O
    • Ultrasonic transduction power. O
    • Ultrasonic frequency range and amplitude of vibration for inserts PR1 and PR2. O

CONCLUSION 9.

The subject device, Piezosurgery White, and the predicate device, Piezosurgery Touch, share the same intended use, principle of operation, technological characteristics to generate ultrasonic mechanical vibrations of the associated tip inserts, and range of available insert tips, with no differences that raise new questions of safety and effectiveness.

The majority of tip inserts included in this submission for use with the subject device are identical to those cleared within the 510(k) submission of the predicate device. The new tip inserts introduced within this submission are manufactured from the same materials, with similar morphologies, and sterilized by the same methods as those cleared previously by FDA with the predicate device.

Based on the information contained within this submission, it is concluded that the Mectron Piezosurgery White is substantially equivalent to the identified predicate device already in interstate commerce within the USA.

§ 872.4120 Bone cutting instrument and accessories.

(a)
Identification. A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.(b)
Classification. Class II.