K Number
K151166

Validate with FDA (Live)

Date Cleared
2015-08-18

(109 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterlized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterilization Cycle. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.

Sterilization methods and configurations - Sterrad NX Standard Sterilization Cycle

Lumen claims for Sterrad NX Standard Sterilization Cycle

Summit Cassette ModelMinimum Inside DiameterMaximum LengthNumber of Lumens
IN-26811 mm65 mm1
IN-26813 mm200 mm1
IN-8987-S1 mm500 mm5
IN-86152.3 mm210 mm5
IN-61054 mm235 mm1

The worst case validated load by vent-to-volume calculation is the IN-2681 tray.

Device Description

Summit Medical Inc. InstruSafe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and instrument accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays do not have direct patient contact. The cassettes / trays by themselves do not maintain sterility.

The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray during the sterilization process and subsequent storage and transportation.

InstruSafe Instrument Protection System cassettes include

  • perforated base
  • perforated cover
  • silicone inserts (hold-it / hold down)
  • Handles
  • Latches
  • Feet
  • Posts (optional)
  • Divider (optional)
  • Shelf (optional)

The cassette contains components made of anodized aluminum, stainless steel, blue silicone, black silicone, polyester, ultem™ 1000

AI/ML Overview

The provided document is a 510(k) premarket notification for the InstruSafe® Instrument Protection System. It describes the device, its intended use, and a comparison to a predicate device, focusing on sterilization methods.

However, the document does not contain information typically associated with acceptance criteria and a study proving a device meets those criteria in the context of an AI/ML medical device. This document is for a non-AI medical device and primarily focuses on demonstrating substantial equivalence to a predicate device, particularly regarding a new sterilization cycle. Therefore, I cannot extract the requested information as it is not present.

Specifically, the document does NOT contain information on:

  • A table of acceptance criteria and reported device performance for an AI/ML algorithm.
  • Sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth, or training set sample size related to an AI/ML device.

The information provided pertains to the physical and functional characteristics of instrument protection cassettes and their compatibility with a specific sterilization cycle. The "performance data" mentioned refers to sterilization validation testing, not AI algorithm performance.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized depiction of human profiles facing right, arranged in a way that suggests unity and support. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the central image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 18, 2015

Summit Medical Inc. Ms. Nicole Dove QA/RA Manager 815 Northwest Parkway, Suite 100 St. Paul, MN 55121

Re: K151166

Trade/Device Name: InstruSafe® Instrument Protection System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap containers, trays, cassettes & other accessory Regulatory Class: II Product Code: KCT Dated: July 14, 2015 Received: July 20, 2015

Dear Ms. Dove:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Dove

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151166

Device Name Instru-Safe® Instrument Protection System

Indications for Use (Describe)

Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterlized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterilization Cycle. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.

Sterilization methods and configurations - Sterrad NX Standard Sterilization Cycle

Lumen claims for Sterrad NX Standard Sterilization Cycle

Summit Cassette ModelMinimum Inside DiameterMaximum LengthNumber of Lumens
IN-26811 mm65 mm1
IN-26813 mm200 mm1
IN-8987-S1 mm500 mm5
IN-86152.3 mm210 mm5
IN-61054 mm235 mm1

The worst case validated load by vent-to-volume calculation is the IN-2681 tray.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three diamond shapes, arranged horizontally. The logo is black and white.

Part NumberMaximum # of InstrumentsEstimated Weight - Tray w/ instruments (lbs
IN-1315303.5
IN-200682
IN-2681130.64
IN-2682130.89
IN-2683131.39
IN-2688152
IN-2840368.75
IN-28413814.5
IN-2842246.5
IN-2843368.75
IN-28805612.1
IN-2900224.18
IN-2950124.1
IN-3006141.5
IN-3030349.5
IN-4000203.2
IN-4003303.25
IN-4010101.98
IN-5008108
IN-500985
IN-540162
IN-5401-0221
IN-5401-0321
IN-5401-0883.2
IN-5401-12123.25
IN-610322.15
IN-610522.15
IN-611032.15
IN-620322.75
IN-620522.75
IN-621022.75
IN-622023
IN-622323
IN-622523
IN-624022.75
IN-630323.28
IN-630523.28
IN-631023.28
IN-640323.28
IN-640523.28
IN-641023.28
IN-65003012.5
IN-701022
IN-701211.07
IN-7020308.5
IN-70303010
IN-703221.1
IN-7040255
IN-70503012
IN-7073105
IN-71204511.25
IN-71234512
IN-71304513.5
IN-71404514.5
IN-715081.9
IN-715361.7
IN-72133010
IN-72203013.5
IN-7223109.2
IN-72304514.5
IN-72344514.5
IN-72404514.5
IN-72504514.5
IN-72513010
IN-7252258
IN-72533012
IN-72604514.5
IN-7273106
IN-7274308
IN-73224514.5
IN-73234514.5
IN-73434514.5
IN-734414
IN-73604514.5
IN-74234514.5
IN-7452108
IN-7453108
IN-75404514.5
IN-75604514.5
IN-76444514.5
IN-7723157.18
IN-7724157.2
IN-7725109.5
IN-77814514.5
IN-78234514.5
IN-78304514.5
IN-78404513.5
IN-79402013.25
IN-82402013.5
IN-861026.65
IN-861226.8
IN-861326.1
IN-861525.8
IN-861625.8
IN-862037.2
IN-862147.18
IN-862247.18
IN-863036.5
IN-863236.45
IN-863336.8
IN-864045.35
IN-864245.35
IN-864355.35
IN-864545.35
IN-865045.85
IN-866045.35
IN-866245.35
IN-866335.5
IN-87004014
IN-872117.5
IN-8771110.5
IN-88102013.5
IN-8820158.75
IN-88234514
IN-8830158.75
IN-88334514
IN-88402013.75
IN-8850158.75
IN-88534514
IN-8860158.75
IN-88623010.5
IN-88634514
IN-888023.28
IN-88821612.1
IN-888323.28
IN-888445.35
IN-888512.25
IN-8886612.1
IN-8889612.1
IN-8891-S12
IN-8891-SI-12-S12
IN-8891-SI-85-S12
IN-88921212.1
IN-889397.5
IN-8894516.1
IN-889786
IN-88981010.25
IN-889976.5
IN-890112.25
IN-89022217
IN-89031513.25
IN-89042217
IN-8907712.5
IN-892018
IN-893112.4
IN-893200.5
IN-893333.75
IN-8936611.5
IN-89371614.5
IN-8938812.5
IN-89391011.6
IN-894055.18
IN-89421110
IN-894312.7
IN-894464.7
IN-894525.18
IN-894696.1
IN-8980209.5
IN-8982-01179.5
IN-8983-01169.5
IN-8984-01159.5
IN-8986-S26.5
IN-8987-S26.5
IN-8988-S26
IN-8989-S26
IN-9999-160612.1
IN-9999-162-S25.8
IN-9999-168-S25.8
IN-9999-172-S25.8
IN-9999-178-S25.8

్ స్టే

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Summit Medical. The logo has the word "SUMMIT" in large, sans-serif font. Below the word "SUMMIT" is the word "medical" in a smaller, sans-serif font. Above the word "SUMMIT" are three black diamonds.

,我们的人都是一

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA
www.summitmedicalusa.com Page 2 of 5

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with three black diamonds above it. Below "SUMMIT" is the word "medical" in a smaller, sans-serif font.

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA
www.summitmedicalusa.com

Page 3 of 5

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in large, stylized letters, with three black diamonds arranged horizontally above it. Below the word "SUMMIT" is the word "medical" in smaller, lowercase letters. The logo has a clean and professional design.

Page 4 of 5

:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three black diamonds.

2

:

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com Page 5 of 5

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Summit Medical, which includes three dark blue diamonds above the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font below. Below the logo, the text "510(k) Summary" is displayed, followed by the code "K151166".

Following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92

Submitter:Summit Medical Inc.815 Northwest Parkway, Suite 100St. Paul, MN 55121Tel: (651) 789-3939
ER Number:3008719017
ContactPerson:Nicole DoveQA/RA ManagerTel: (651) 789-3921ndove@summitmedicalusa.com
Date Prepared:July 14, 2015
SubjectDevice:Trade Name(s):InstruSafe® Instrument Protection SystemClassification Name:Sterilization wrap containers, trays, cassettes & other accessory (21 CFR 880.6850Common Name:Instrument Tray, Sterilization Tray, Sterilization Cassettes, Instrument DeliverySystemDevice Class:Class IIDevice Code:KCTPanel:General Hospital
PredicateDevice:Tradename: InstruSafe Instrument Protection System510(k) Holder: Summit Medical Inc.510(k) #: K133015
DeviceDescription:Summit Medical Inc. InstruSafe Instrument Protection System are cassettes / traysused to enclose and hold surgical instruments and instrument accessories in anorganized manner during the sterilization process and subsequent storage andtransportation. The cassettes / trays do not have direct patient contact. Thecassettes / trays by themselves do not maintain sterility.

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979

www.summitmedicalusa.com

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of three blue diamonds arranged horizontally above the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font below it. The logo is simple and modern, with a focus on the company name.

The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray during the sterilization process and subsequent storage and transportation.InstruSafe Instrument Protection System cassettes used to organize and protect other medical devices that are sterilized by a healthcare provider. InstruSafe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterrad NX Standard Sterilization Cycle. The InstruSafe Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The InstruSafe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.Sterilization methods and configurations• Sterrad NX Standard Sterilization CycleLumen claims for Sterrad NX Standard Sterilization CycleInstru-Safe Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycle. The Instru-Safe System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe System cassettes are not intended on their own to maintain sterility.Sterilization methods and configurations• Autoclave Sterilization Parameter:Cycle: Pre-vacuumTemperature: 270°F (132°C)Exposure Time: 4 minutesMinimum Dry Time: 30 minutes
Intended Use:InstruSafe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. InstruSafe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterrad NX Standard Sterilization Cycle. The InstruSafe Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The InstruSafe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.Sterilization methods and configurations• Sterrad NX Standard Sterilization CycleSummitCassette ModelMinimum Inside DiameterMaximum LengthNumber of LumensSummitCassette ModelAesculap Container Model
Lumen claims for Sterrad NX Standard Sterilization CycleIN-26811mm65mm1IN-8823-AE*JN444
SummitCassette ModelMinimum InsideDiameterMaximumLengthNumber ofLumensIN-26813mm200mm1IN-2880*JK444
IN-26811mm65mm1IN-8987-S1mm500mm5IN-6105*JN742
IN-26813mm200mm1IN-86152.3mm210mm5*Validated by Summit Medical for use in steam prevacuum sterilizers ONLY operating at 270°F (132°C) for 4 minutes exposure time. Consult container instructions to ensure that contents do not exceed the sterilization containers intended load claims.
IN-8987-S1mm500mm5IN-61054mm235mm1
IN-86152.3mm210mm5The worst case validated load by vent-to-volume calculation is the IN-2681 tray.
IN-61054mm235mm1MaterialCompositionNo changes from predicate deviceThe cassette contains components made of anodized aluminum, stainless steel, blue silicone, black silicone, polyester, ultem™ 1000
The worst case validated load by vent-to-volume calculation is the IN-2681 tray.PhysicalPropertiesInstruSafe Instrument Protection System cassettes include- perforated base- perforated coverInstru-Safe Instrument Protection System cassettes include- perforated base- perforated cover
The intended use of the subject device includes the Sterrad NX Standard Sterilization Cycle. Performance testing has been performed for the Sterrad NX Standard Sterilization Cycle. This new sterilization cycle does not affect safety and effectiveness of the InstruSafe Instrument Protection System.
Comparison ofCharacteristicsto PredicateDevice:Based on a comparison of the design, technology, materials, manufacturing, performance, specifications and methods of use, the InstruSafe Instrument Protection System is equivalent to the identified 510(k) cleared predicate device.
ElementNew DevicePredicate (K133015)

Page 2 of 10

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979

www.summitmedicalusa.com

Page 3 of 10

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Summit Medical. The logo features the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo has a clean and professional look.

- silicone inserts (hold-it / hold down)- Handles- Latches- Feet- Posts (optional)- Divider (optional)- Shelf (optional)- silicone inserts (hold-it / hold down)- Handles- Latches- Feet- Posts (optional)- Divider (optional)- Shelf (optional)
ChemicalPropertiesNot ApplicableNot Applicable
Configurations/ DimensionsVarious configurations / dimensionsSee table located in predicate devicesubmission K133015
Air permeanceNot ApplicableNot Applicable
Percent ofsurfaceperformationsNot ApplicableNot Applicable
PerformanceNew DevicePredicate (K133015)
SterilantPenetrationSterrad NX Standard Sterilization CyclePre-Vacuum SteamCycle: Pre-vacuumTemperature: 270°F (132°C)Exposure Time: 4 minutesMinimum Dry Time: 30 minutes
MicrobialBarrierProperties(PackagingIntegrity)Not ApplicableNot Applicable
MaterialCompatibilityNo changes from predicate deviceRefer to predicate device K133015
ToxicologicalProperties(Biocompatibility, includingSterilantResidueLimits)MEM Elution Cytotoxicity (ISO 10993-5)- The test samples meet the USP andISO 10993-5 requirements for this test.All controls were acceptable and thetest considered valid. The test samplesPASSED and are considered NON-CYTOTOXIC under the testconditions employed.Refer to predicate device K133015
Shelf LifeNo ChangeReusable (5 year accelerated shelflife study)
Drying TimeNot ApplicableAutoclave Sterilization Parameter:Cycle: Pre-VacuumTemperature: 270°F (132°C)

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979

www.summitmedicalusa.com

Page 4 of 10

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a large, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.

Minimum Exposure Time: 4 minutes
Minimum Dry Time: 30 minutes
Aeration TimeNot ApplicableNot Applicable
TechnologicalCharacteristics:The technological characteristics of the subject devices are equivalent to thepredicate devices. The cassettes / trays are made of standard medical gradematerials and do not incorporate any new technological characteristics.
PerformanceData:Sterilization validation testing was performed to demonstrate InstruSafe InstrumentProtection System compatibility when used in a Sterrad NX Standard SterilizationCycle with a legally marketed wrap or Aesculap rigid container.
Conclusion:Based upon intended use, performance data and technical information provided inthis pre-market notification, the InstruSafe Instrument Protection System describedherein are substantially equivalent to the predicate device [InstruSafe InstrumentProtection System (K133015)]

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, with a clean design.

Part NumberMaximum # of InstrumentsEstimated Weight - Tray w/ instruments (lbs)
IN-1315303.5
IN-200682
IN-2681130.64
IN-2682130.89
IN-2683131.39
IN-2688152
IN-2840368.75
IN-28413814.5
IN-2842246.5
IN-2843368.75
IN-28805612.1
IN-2900224.18
IN-2950124.1
IN-3006141.5
IN-3030349.5
IN-4000203.2
IN-4003303.25
IN-4010101.98
IN-5008108
IN-500985
IN-540162
IN-5401-0221
IN-5401-0321
IN-5401-0883.2
IN-5401-12123.25
IN-610322.15
IN-610522.15
IN-611032.15
IN-620322.75
IN-620522.75
IN-621022.75
IN-622023
IN-622323
IN-622523

Page 6 of 10

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Summit Medical. The logo features the word "SUMMIT" in a bold, sans-serif font, with the "U" stylized to resemble a mountain range. Above the word "SUMMIT" are three blue diamonds arranged in a horizontal line. Below the word "SUMMIT" is the word "medical" in a smaller, sans-serif font.

IN-630323.28
IN-630523.28
IN-631023.28
IN-640323.28
IN-640523.28
IN-641023.28
IN-65003012.5
IN-701022
IN-701211.07
IN-7020308.5
IN-70303010
IN-703221.1
IN-7040255
IN-70503012
IN-7073105
IN-71204511.25
IN-71234512
IN-71304513.5
IN-71404514.5
IN-715081.9
IN-715361.7
IN-72133010
IN-72203013.5
IN-7223109.2
IN-72304514.5
IN-72344514.5
IN-72404514.5
IN-72504514.5
IN-72513010
IN-7252258
IN-72533012
IN-72604514.5
IN-7273106
IN-7274308
IN-73224514.5
IN-73234514.5
IN-73434514.5

Page 7 of 10

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a bold, sans-serif font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three blue diamonds, arranged in a horizontal line. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

IN-734414
IN-73604514.5
IN-74234514.5
IN-7452108
IN-7453108
IN-75404514.5
IN-75604514.5
IN-76444514.5
IN-7723157.18
IN-7724157.2
IN-7725109.5
IN-77814514.5
IN-78234514.5
IN-78304514.5
IN-78404513.5
IN-79402013.25
IN-82402013.5
IN-861026.65
IN-861226.8
IN-861326.1
IN-861525.8
IN-861625.8
IN-862037.2
IN-862147.18
IN-862247.18
IN-863036.5
IN-863236.45
IN-863336.8
IN-864045.35
IN-864245.35
IN-864355.35
IN-864545.35
IN-865045.85
IN-866045.35
IN-866245.35
IN-866335.5
IN-87004014

Page 8 of 10

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

IN-872117.5
IN-8771110.5
IN-88102013.5
IN-8820158.75
IN-88234514
IN-8830158.75
IN-88334514
IN-88402013.75
IN-8850158.75
IN-88534514
IN-8860158.75
IN-88623010.5
IN-88634514
IN-888023.28
IN-88821612.1
IN-888323.28
IN-888445.35
IN-888512.25
IN-8886612.1
IN-8889612.1
IN-8891-S12
IN-8891-SI-12-S12
IN-8891-SI-85-S12
IN-88921212.1
IN-889397.5
IN-8894516.1
IN-889786
IN-88981010.25
IN-889976.5
IN-890112.25
IN-89022217
IN-89031513.25
IN-89042217
IN-8907712.5
IN-892018
IN-893112.4
IN-893299.5

Page 9 of 10

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

IN-893333.75
IN-8936611.5
IN-89371614.5
IN-8938812.5
IN-89391011.6
IN-894055.18
IN-89421110
IN-894312.7
IN-894464.7
IN-894525.18
IN-894696.1
IN-8980209.5
IN-8982-01179.5
IN-8983-01169.5
IN-8984-01159.5
IN-8986-S26.5
IN-8987-S26.5
IN-8988-S26
IN-8989-S26
IN-9999-160612.1
IN-9999-162-S25.8
IN-9999-168-S25.8
IN-9999-172-S25.8
IN-9999-178-S25.8

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).