Dafilon Nonabsorbable Polyamide Surgical Sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neurological procedures.
Dafilon is a nonabsorbable monofilament surgical suture which is supplied sterile. Dafilon is composed of the long-chain aliphatic polymers polyamide 6 and/or 6,6. The Dafilon suture is offered undyed and dyed with FDA approved colorants Logwood extract in accordance with Title 21 CRF, §73.1410 or [Phthalocyaninato(2-)] copper in accordance with Title 21 CFR, §74.3045. The Dafilon suture will be offered in diameters ranging from USP size 11-0 through 5 and will be available in a variety of cut lengths with out needles attached.
This document is a 510(k) premarket notification for a surgical suture, not a device that involves performance criteria typically associated with AI/ML algorithms or interpretative medical devices. Therefore, many of the requested sections regarding acceptance criteria and study design are not applicable in the context of this document. This submission primarily focuses on demonstrating substantial equivalence to existing predicate devices based on material composition, indications for use, and mechanical/biocompatibility testing, rather than an AI model's diagnostic or predictive performance.
However, based on the provided text, I can extract the relevant "acceptance criteria" and "performance data" for this suture device where applicable.
Here's an interpretation of the document in the context of your request, focusing on the available information:
Device: Aesculap Dafilon Nonabsorbable Polyamide Surgical Suture
Since this is a surgical suture, the "acceptance criteria" are related to its physical properties, sterility, and biocompatibility, rather than diagnostic accuracy or sensitivity/specificity of a predictive model.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criterion (Suture Device) | Reported Device Performance |
|---|---|
| Physical Characteristics: | |
| Diameter | Meets USP Requirements. |
| Length | Meets USP Requirements. |
| Needle Attachment | Meets USP Requirements. |
| Tensile Strength | Meets USP Requirements. |
| Biocompatibility: | |
| Cytotoxicity | Testing conducted. Results demonstrate safety and effectiveness. |
| Sensitization | Testing conducted. Results demonstrate safety and effectiveness. |
| Intracutaneous Irritation | Testing conducted. Results demonstrate safety and effectiveness. |
| Systemic and Muscle Implantation (4-week) | Testing conducted. Results demonstrate safety and effectiveness. |
| Material Equivalency: | |
| Chemical Characterization (Polyamide raw material) | Results demonstrate substantial equivalence to predicate devices. |
| Sterilization & Shelf Life: | |
| Sterilization (Gamma Irradiation or EO) | Validated. |
| Shelf Life (Accelerated Aging) | Data generated to support submission. |
2. Sample size used for the test set and the data provenance
The document specifies "tests were conducted for diameter, tensile strength, and needle attachment" and "biocompatibility testing." However, it does not specify the sample sizes for these tests. The provenance is internal testing by Aesculap, Inc. against USP (United States Pharmacopeia) requirements and ISO (International Organization for Standardization) standards. The tests are likely prospective, conducted on manufactured samples of the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to a nonabsorbable surgical suture device. The "ground truth" for suture performance is established by standardized physical and biological tests (USP, ISO standards), not by human expert interpretation or consensus in the way it would be for an AI diagnostic device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable for a surgical suture. Adjudication methods are relevant for subjective interpretations of data, typically in diagnostic studies. Here, the "truth" is determined by objective measurements against established standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a surgical suture, not an AI-assisted diagnostic tool. No human reader involvement or AI assistance is relevant to its performance evaluation for this 510(k) submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical device (suture), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" is based on established industry standards and regulatory requirements:
- USP Requirements: For physical characteristics such as diameter, length, needle attachment, and tensile strength.
- ISO 10993-1: For biocompatibility testing (Cytotoxicity, Sensitization, Intracutaneous Irritation, Systemic and Muscle Implantation).
- 21 CFR §73.1410 and §74.3045: For colorant approval.
8. The sample size for the training set
Not applicable. This device does not involve a "training set" in the context of machine learning or AI.
9. How the ground truth for the training set was established
Not applicable. As there is no training set, there is no ground truth establishment for it.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, layered on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 17, 2016
Aesculap, Inc. Ms. Kathy Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K151165
Trade/Device Name: Aesculap Dafilon Nonabsorbable Polyamide Surgical Suture Regulation Number: 21 CFR 878.5020 Regulation Name: Nonabsorbable Polyamide Surgical Suture Regulatory Class: Class II Product Code: GAR Dated: January 8, 2016 Received: January 11, 2016
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151165
Device Name
Dafilon Nonabsorbable Polyamide Surgical Sutures
Indications for Use (Describe)
Dafilon Nonabsorbable Polyamide Surgical Sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neurological procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K151165- Page 1 of 3
B. 510(k) SUMMARY (as required by 21 CFR 807.92)
Aesculap Dafilon Nonabsorbable Polyamide Surgical Suture February 12, 2016
| COMPANY: | Aesculap ®, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714 |
|---|---|
| CONTACT: | Kathy A. Racosky610-984-9291 (phone)610-791-6882 (fax)kathy.racosky@aesculap.com |
| 510(k): | K151165 |
| TRADE NAME: | Dafilon Nonabsorbable Polyamide Surgical Suture |
| COMMON NAME: | Nonabsorbable Polyamide Surgical Suture |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | Suture, Nonabsorbable, Synthetic, Polyamide |
| REGULATION NUMBER: | 878.5020 |
| PRODUCT CODE: | GAR |
PURPOSE FOR PREMARKET NOTIFICATION
The Dafilon Nonabsorbable Polyamide Surgical Suture described in this submission introduces Phthalocyaninato (2)-Copper as a colorant, two packaging types and extends the available suture length range from 3 cm to 250 cm to the existing product line Dafilon Nonabsorbable Polyamide Surgical Suture K990090.
PREDICATE DEVICE
- Primary Predicate: Dafilon Nonabsorbable Polyamide Surgical Suture (K990090)
- Aesculap Optilene Nonabsorbable Suture (K133890)
DEVICE DESCRIPTION
Dafilon is a nonabsorbable monofilament surgical suture which is supplied sterile. Dafilon is composed of the long-chain aliphatic polymers polyamide 6 and/or 6,6. The Dafilon suture is offered undyed and dyed with FDA approved colorants Logwood extract in accordance with Title 21 CRF, §73.1410 or [Phthalocyaninato(2-)] copper in accordance with Title 21 CFR,
{4}------------------------------------------------
§74.3045. The Dafilon suture will be offered in diameters ranging from USP size 11-0 through 5 and will be available in a variety of cut lengths with out needles attached.
INDICATIONS FOR USE
Dafilon Nonabsorbable Polyamide Surgical Sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neurological procedures.
TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
As established in this submission, the Aesculap Dafilon suture is a synthetic nonabsorbable monofilament surgical suture offered undyed and dyed in the same range of diameters and cut lengths that are substantially equivalent to other predicate devices cleared by FDA. The subject device is shown to be substantially equivalent and has the same technological characteristics to its predicate devices through comparison in design, intended use, material composition, function and range of sizes. The device characteristics comparing the Dafilon Nonabsorbable Polyamide Surgical Suture to the predicate devices are summarized below.
| Aesculap, Inc.Dafilon Nonabsorbable | Aesculap, Inc.Dafilon Nonabsorbable | Aesculap, Inc.Optilene Nonabsorbable | |
|---|---|---|---|
| Polyamide Suture | Polyamide Suture | Suture | |
| K# | K151165 | K990090 | K133890 |
| Indications | Indicated for use in all | Indicated for use in all | Indicated for use in all |
| types of general soft tissue | types of general soft tissue | types of general soft tissue | |
| approximation and/or | approximation and/or | approximation and/or | |
| ligation, including use in | ligation, including use in | ligation, including use in | |
| cardiovascular, ophthalmic, | cardiovascular, ophthalmic | cardiovascular, ophthalmic, | |
| microsurgery and | and neurological | microsurgery and neural | |
| neurological procedures. | procedures. | tissue. | |
| Material | Polyamide 6 and/or | Polyamide 6 and/or | 95% polypropylene and 5% |
| polyamide 6,6 | polyamide 6,6 | Polyethylene | |
| Dyed, Un-dyed | Un-dyed and Dyed | Un-dyed and Dyed | Dyed |
| Colorant | [Phthalocyaninato(2-)] | FD&C Blue no. 2 or | [Phthalocyaninato(2-)] |
| copper or logwood extract | logwood extract | copper | |
| Structure | Monofilament | Monofilament | Monofilament |
| Size | 11-0 through 5 (various | 11-0 through 6 (various | 10-0 through 0 (various |
| lengths) | lengths) | lengths) | |
| with or w/out needles | with or w/out needles | with or w/out needles | |
| attached | attached | attached | |
| Thread length | -3 cm to 250 cm | -5 cm to 150 cm | -8cm to 150cm |
| -ligature reels of longer | -ligature reels of longer | -ligature reels of longer | |
| length | length | length | |
| Physical: | |||
| - Diameter | All characteristics meet | All characteristics meet | All characteristics meet |
| - Length | USP Requirements. | USP Requirements | USP Requirements. |
| - Needle Attachment | |||
| - Tensile Strength |
{5}------------------------------------------------
| K151165- Page 3 of 3 | |||
|---|---|---|---|
| Aesculap, Inc.Dafilon NonabsorbablePolyamide Suture | Aesculap, Inc.Dafilon NonabsorbablePolyamide Suture | Aesculap, Inc.Optilene NonabsorbableSuture | |
| K# | K151165 | K990090 | K133890 |
| Needles | 300 or 400 series stainlesssteel | 300 or 400 series stainlesssteel | 300-series stainless steel |
| Packaging | Coated cardboard supportin double peel pouch. Innerpouch composed of PET,aluminum foil and LDPElayers. Medical gradepaper and PET/LDPElaminate outer pouch.orCoated cardboard supportin peel pouch. Medicalgrade paper and PET/LDPElaminate outer pouch.orPE suture support in aPETG tray sealed with atyvek lid. Medical gradepaper and PET/LDPElaminate outer pouch. | Oval coated cardboardsupport in double peelpouch. Inner pouchcomposed of PET,aluminum foil and LDPElayers. Medical gradepaper and PET/LDPElaminate outer pouch. | Oval HDPE support with anoval card fold card lid in atyvek peel pouch andpolyester-polyethylenerephatalatefilm. |
| Sterilization | Gamma Irradiation orEthylene Oxide (EO) | Gamma Irradiation | Ethylene Oxide (EO) |
PERFORMANCE DATA
As recommended by the FDA's Class II Special Control Guidance Document for Surgical Sutures, including mechanical testing in accordance to USP 37 and biocompatibility testing in accordance to ISO 10993-1 has been performed to demonstrate that the Dafilon Nonabsorbable Suture is substantially equivalent to other predicate devices.
Tests were conducted for diameter, tensile strength, and needle attachment. All specifications were met. The Dafilon suture is considered an implant device, tissue/bone contact device of permanent duration (>30 days). In accordance with ISO 10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing," Biocompatibility testing within this submission includes the following: Cytotoxicity, Sensitization, Intracutaneous Irritation, Systemic and Muscle Implantation (4-week).
Testing demonstrated that the device is as safe, as effective, and performs as well as the predicate devices.
Additional chemical characterization testing was conducted in order to evaluate the chemical equivalency of the polyamide raw material. The results of this testing demonstrates that the Dafilon Nonabsorbable Suture is substantially equivalent to the predicate devices.
The Dafilon Nonabsorbable Polyamide Suture is blister packed and sterilized by either gamma Irradiation or Ethylene Oxide. Accelerated aging data for the Dafilon Nonabsorbable Suture has been generated to support this submission.
§ 878.5020 Nonabsorbable polyamide surgical suture.
(a)
Identification. Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.