(210 days)
Intended for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine for kyphoplasty (for use with CareFusion Radiopaque Bone Cement).
The Inflatable Bone Tamp (IBT) was designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture and preventing cement leakage, while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter.
This document describes the AVAflex Vertebral Balloon System, specifically detailing its performance testing for substantial equivalence to a predicate device.
Here's an analysis of the provided text based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard/Test/FDA Guidance) | Reported Device Performance (Results Summary) |
|---|---|---|
| Inflation pressure | Constrained burst test | The balloon catheters exceeded the requirements for the minimum burst pressure in a constrained environment. |
| Inflation volume | Unconstrained burst test | The balloon catheters exceeded the requirements for the minimum burst volume in an unconstrained environment. |
| Balloon double wall thickness | Calibrated measurement | The double wall thickness of the balloons was substantially equivalent to that of the predicate device. |
| Simulated use | Cadaver | Cadaveric simulated use testing demonstrated substantial equivalence in performance. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for the "Constrained burst test," "Unconstrained burst test," or "Calibrated measurement."
For the "Cadaveric simulated use testing":
- Sample Size: Not explicitly stated (e.g., number of cadavers or balloons tested).
- Data Provenance: The cadaveric simulated use suggests human cadaver data, which is a form of ex vivo or in vitro testing that simulates a clinical environment. No country of origin is specified, but it's presumed to be associated with the manufacturer's testing facilities. The data is prospective in the sense that the tests were conducted specifically for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
It's important to note that for these types of mechanical performance tests, "ground truth" is typically established by engineering specifications, validated test methods, and industry standards, rather than expert clinical consensus in the same way it would be for diagnostic AI. The results are objective measurements against defined criteria.
4. Adjudication Method for the Test Set
Not applicable. The performance tests are objective measurements against defined criteria, not subjective assessments requiring expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The document explicitly states: "N/A - No clinical tests were conducted for this submission." This indicates no MRMC study or any clinical study was performed.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This device is a physical medical device (vertebral balloon system), not an algorithm or AI software. Therefore, the concept of "standalone algorithm performance" does not apply. The performance described is the standalone performance of the device itself.
7. Type of Ground Truth Used
The "ground truth" for the non-clinical tests is established by:
- Pre-defined engineering specifications and design requirements (for burst pressure, volume, and dimensions).
- Industry standards for medical device testing.
- The performance of the predicate device (for substantial equivalence claims).
- Observed outcomes in simulated use (cadaveric testing).
8. Sample Size for the Training Set
Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for a physical medical device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 24, 2015
CareFusion Ms. Jov Greidanus Manager, Regulatory Affairs 75 North Fairway Drive Vernon Hills, Illinois 60061
Re: K151125
Trade/Device Name: AVAflex Vertebral Balloon System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, NDN Dated: September 30, 2015 Received: October 2, 2015
Dear Ms. Greidanus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved OMB No 0910-0120
Expiration Date January 31, 2017
See PRA Statement below
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) Unknown K151125
Device Name
AVAflex Vertebral Balloon System
Indications for Use (Describe)
Intended for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine for kyphoplasty (for use with CareFusion Radiopaque Bone Cement).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda hhs gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number "
{3}------------------------------------------------
510(k) SUMMARY K151125
A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.
| SUBMITTER INFORMATION | |||
|---|---|---|---|
| Name | CareFusion | ||
| Address | 75 North Fairway Drive, Vernon Hills IL 60061 USA | ||
| Phone number | (847) 362-8103 | ||
| Fax number | (312) 949-0583 | ||
| EstablishmentRegistration Number | 1423507 | ||
| Name of contact person | Joy Greidanus | ||
| Date prepared | September 30, 2015 | ||
| NAME OF DEVICE | |||
| Trade or proprietaryname | AVAflex Vertebral Balloon System | ||
| Common or usual name | Inflatable Bone Tamp | ||
| Classification name | Arthroscope | ||
| Classification panel | Orthopedic | ||
| Regulation | Class II per 21CFR §888.1100, Procode HRXClass II per 21CFR §888.3027, Procode NDN | ||
| Legally marketeddevice(s) to whichequivalence is claimed | CareFusion Inflatable Bone Tamps, K131824Radiopaque Bone Cement, K043518 | ||
| Reason for 510(k)submission | Modifications to device | ||
| Device description | The Inflatable Bone Tamp (IBT) was designed for use in balloon kyphoplasty.The balloon serves to create a cavity in the vertebral body, thereby reducingthe fracture and preventing cement leakage, while still allowing for cementinterdigitation. The balloon catheter is the functional part of the device thatcreates a cavity and reduces the fracture. The balloon catheter provides aconduit through which the physician can inflate the balloon at the distal end ofthe catheter. | ||
| Intended use of thedevice | Intended for the reduction and fixation of fractures and/or creation of a void incancellous bone in the spine for kyphoplasty (for use with CareFusionRadiopaque Bone Cement). |
{4}------------------------------------------------
SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| Characteristic | New Device | PredicateCareFusion Vertebral Balloon K131824 |
|---|---|---|
| Compatible cannulasize | 11G | 10G |
| Balloon inflationmedium | 60% contrast recommended | 60% contrast recommended |
| Markers | Radiopaque | Radiopaque |
| Balloon shape | Cylindrical | Cylindrical |
| Maximumrecommended inflationpressure | 400 psi (27 ATM) | 400 psi (27 ATM) |
| Maximumrecommended inflationvolume | 3-6 mL | 4-8 mL |
PERFORMANCE DATA
SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
Performance Test Summary-New Device
| Characteristic | Standard/Test/FDA Guidance | Results Summary |
|---|---|---|
| Inflation pressure | Constrained burst test | The balloon catheters exceeded therequirements for the minimum burst pressure in aconstrained environment. |
| Inflation volume | Unconstrained burst test | The balloon catheters exceeded therequirements for the minimum burst volume in anunconstrained environment. |
| Balloon double wallthickness | Calibrated measurement | The double wall thickness of the balloons wassubstantially equivalent to that of the predicatedevice. |
| Simulated use | Cadaver | Cadaveric simulated use testing demonstratedsubstantial equivalence in performance. |
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
N/A - No clinical tests were conducted for this submission.
CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The results of the non-clinical tests show that the CareFusion Flexible IBT System meets or exceeds all performance requirements, and is substantially equivalent to the predicate device.
N/A