K Number
K151117

Validate with FDA (Live)

Date Cleared
2015-08-20

(115 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Laparoscope Lens Shield Device (LENS), a sterile, single-use and disposable laparoscopic accessory lens shield device, is intended to maintain the intra-operative view of the surgical site during minimally invasive surgery by physically shielding the laparoscope lens from debris, grease, blood, and bodily fluids.

Device Description

The Laparoscope Lens Shield Device (LENS) is a laparoscope accessory lens shielding device consisted of multi-lumen sheath that slides over the laparoscope. The sheaths assembly consists of 3 concentric sheaths: one outer and two inner sheaths. The outer sheath provides protection and cover for the inner sheath and shielding film. It is intended to maintain the intra-operative view of the surgical site during minimally invasive surgery by physically shielding the laparoscope lens from debris, grease, blood, and bodily fluids The Laparoscope Lens Shield Device (LENS) is intended to be used by clinicians through prescription use only.

AI/ML Overview

The provided text does not contain information about acceptance criteria or a study that proves a device meets such criteria in the context of AI/ML device performance. The document is a 510(k) premarket notification from the FDA regarding a "Laparoscope Lens Shield Device (LENS)."

It outlines:

  • Device Description and Intended Use: The device is a sterile, single-use, disposable accessory lens shield intended to maintain intra-operative view during minimally invasive surgery by physically shielding the laparoscope lens from debris, grease, blood, and bodily fluids.
  • Substantial Equivalence: The document asserts substantial equivalence to a predicate device (Clear-Vu System) based on similar intended use, target patient/user population, anatomical site, use location, performance, and biocompatibility.
  • Differences: Differences from the predicate device are noted in sterilization method (Ethylene Oxide vs. Radiation), energy source (None vs. CO2 Gas), and physical barrier material (PET film vs. CO2 Gas).
  • Performance Testing: General categories of performance testing are listed:
    • Biocompatibility testing (in accordance with ISO standards)
    • Sterilization validation (in accordance with ISO 11135)
    • Mechanical testing (to demonstrate design specification fulfillment and structural integrity)
    • Image quality test: "Laparoscope Lens Shield Device (LENS) had been tested for view quality during laparoscopic procedures regarding the resolution, distortion, field of view, depth of view, and color, to demonstrate that adequate view quality of laparoscopy is maintained to a safe and effective clinical use."

However, the document does not provide any specific acceptance criteria (metrics or thresholds) or detailed results from these tests. It merely states that the tests were performed and that the device meets performance specifications or demonstrates adequate view quality.

Therefore, I cannot populate the requested table and information points as there is no data on acceptance criteria, reported device performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, or ground truth details.

The device described is a physical medical device, not an AI/ML diagnostic tool, which the detailed questions about "test set," "training set," "ground truth," "MRMC," and "effect size" typically pertain to.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around a stylized eagle or bird symbol. The bird symbol is composed of three overlapping profiles, suggesting a sense of community or interconnectedness. The overall design is simple and professional, reflecting the department's role in public health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 20, 2015

Medeon Biodesign, Inc. % Greta Chang Sr. Manager of Regulatory, Quality and Clinical Affairs Lin & Associate, LLC 9223 Cambridge Manor Court Potomac, MD 20854

Re: K151117

Trade/Device Name: Laparoscope Lens Shield Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ Dated: July 13, 2015 Received: July 15, 2015

Dear Greta Chang,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151117

Device Name

Laparoscope Lens Shield Device (LENS)

Indications for Use (Describe)

Laparoscope Lens Shield Device (LENS), a sterile, single-use and disposable laparoscopic accessory lens shield device, is intended to maintain the intra-operative view of the surgical site during minimally invasive surgery by physically shielding the laparoscope lens from debris, grease, blood, and bodily fluids.

Type of Use (Select one or both, as applicable):

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92

The assigned 510(k) Number: TBD

.

1.Submitter
Mailing AddressMedeon Biodesign, Inc7F, 116, HouGang St,Taipei, Taiwan 11170
Phone: +886 2 2881 6686
Establishment Registration No.: NA (1st submission)
Contact PersonGreta ChangSr. Manager of Regulatory, Quality & Clinical Affair
Phone:+886 2 2881 6686
Fax:+886 2 2881 6685
E-mail:greta@medeonbio.com
Date PreparedApril 24, 2015
2Device Name
Common or usual nameLaparoscope Lens Shield Device (LENS)
Product CodeGCJ
DeviceEndoscope and accessories
CFR ClassificationCFR Part 876.1500
Device ClassII
Classification PanelGastroenterology/Urology
3Predicate Device Name
510(k) number:K080613
Trade or proprietary ormodel name:Clear-Vu System
Manufacturer:Minimally Invasive Devices, LLC
4Device Description:The Laparoscope Lens Shield Device (LENS) is a
laparoscope accessory lens shielding device consisted ofmulti-lumen sheath that slides over the laparoscope. Thesheaths assembly consists of 3 concentric sheaths: oneouter and two inner sheaths. The outer sheath providesprotection and cover for the inner sheath and shielding
film. It is intended to maintain the intra-operative view of

{4}------------------------------------------------

MEDEON Medeon Biodesign, Inc. Traditional 510(k) Notification Laparoscope Lens Shield Device (LENS)
the surgical site during minimally invasive surgery by physically shielding the laparoscope lens from debris, grease, blood, and bodily fluids The Laparoscope Lens Shield Device (LENS) is intended to be used by clinicians through prescription use only.
5.Intended Use:Laparoscope Lens Shield Device (LENS), a sterile, single-use and disposable laparoscopic accessory lens shield device, is intended to maintain the intra-operative view of the surgical site during minimally invasive surgery by physically shielding the laparoscope lens from debris, grease, blood, and bodily fluids.
Special Conditions for Use Statement(s):For prescription use only
6.Technological Characteristics and Substantial Equivalence Comparison with Predicate:A comparison of the device features, intended use, and other information demonstrates that the Laparoscope Lens Shield Device (LENS) is substantially equivalent to the predicate device as summarized in Table 1 . The differences raise no different question of safety and effectiveness.
Similarities
Device nameSubject device: Laparoscope LensShield Device (LENS)Predicate device: Clear-Vu
Intended UseLaparoscope Lens Shield Device(LENS), a sterile, single-use anddisposable laparoscopic accessorylens shield device, is intended tomaintain the intra-operative view ofthe surgical site during minimallyinvasive surgery by physicallyshielding the laparoscope lens fromdebris, grease, blood, and bodilyfluidsClear-Vu System is a single-use,disposable laparoscopic accessorydevice intended to facilitate intra-operative defogging and cleaning ofthe lens of a laparoscope duringminimally-invasive surgery whilemaintaining visualization of thesurgical site.
Target patientPopulationPatient under laparoscopic surgerySame
Target UserPopulationClinician who is qualified to performa laparoscopic surgery.Same
AnatomicalSiteAbdominopelvic cavitySame
Where UsedHospital O.R. roomSame

Table 1

{5}------------------------------------------------

Similarities
Device nameSubject device: Laparoscope LensShield Device (LENS)Predicate device: Clear-Vu
ContraindicationsThere's no known contraindicationsfor subject deviceSame
Method ofIntroductionSubject device is introduced intoabdominopelvic cavity via a trocar.Same
CompatibilityOther DevicesLaparoscope :0°, 10mm, 30cm scopeTrocar: 12mmSame
PerformanceAble to maintain laparoscopic viewwhen it get soiled by debrisSame
Biocompatiblefor IntendedUseLimited exposure, externalcommunication device of tissuecontactPass the cytotoxicity, sensitization,irritation, and acute systemic toxicitySame
Differences
Device nameSubject device: Laparoscope LensShield Device (LENS)Predicate device: Clear-Vu
SterilizationMethodEthylene Oxide sterilization, SAL of10-6Radiation sterilization, SAL of 10-6
Energy sourceNo energy sourceCO2 Gas
PhysicalbarrierPET filmCO2 Gas

7. Performance Testing

& MEDEON

Performance testing has been carried out to demonstrate that this device meets the performance specification for its intended use. The following tests were performed on the device.

Biocompatibility testing

The biocompatibility evaluation and testing of the Laparoscope Lens Shield Device (LENS) was conducted in accordance with the following standards and guidance, as recognized by the FDA:

  • . FDA Draft Guidance - Use of International Standard ISO- 10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing", dated 04-23-2013
  • ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
  • . 10993-10, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization
  • . ISO 10993-11, Biological evaluation of medical devices- Part 11: Tests for systemic toxicity.

{6}------------------------------------------------

Sterilization validation

Sterilization validation was conducted in accordance with ISO 11135:2014 "Sterilization of health-care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices". Sterilization has been validated to achieve a Sterility Assurance Level (SAL) of 10 - .

Mechanical testing

Laparoscope Lens Shield Device (LENS)'s mechanical function and structure integrity were tested and demonstrated that the design specification from design input are fulfilled. Mechanical safety tests were also conducted to demonstrate that the insertion portion of the device remains intact during the surgery and raises no safety concern.

Image quality test

Laparoscope Lens Shield Device (LENS) had been tested for view quality during laparoscopic procedures regarding the resolution, distortion, field of view, depth of view, and color, to demonstrate that adequate view quality of laparoscopy is maintained to a safe and effective clinical use.

  1. Conclusion. Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Laparoscope Lens Shield Device (LENS) is substantially equivalent to the predicate device and raises no different question of safety or effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.