K Number
K151103

Validate with FDA (Live)

Date Cleared
2015-06-23

(60 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ASAHI Corsair Microcatheter is intended to provide support to facilitate the placement of Guide wires in the coronary and peripheral vasculatures and can be used to exchange one guide wire for another. The Corsair Microcatheter is also intended to assist in the delivery of contrast media into the coronary, peripheral, and abdominal vasculatures.

Device Description

The ASAHI Corsair Microcatheter consists of a distal tip and a shaft tube that are inserted into a vascular connector for catheter control and infusion of contrast media. No accessories are part of this device. The Corsair has a radiopaque marker coil that is imbedded into the inner layer of resin to facilitate the tip location during angiographic procedures. In addition, the device has a hydrophilic coating on the outer surface of the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair has a tapered shape and is designed to have increased flexibility towards the distal end. PTFE is applied to the inner lumen of the catheter for the purposes of a smooth transition and exchange of guidewires.

The microcatheter also contains wires to reinforce the distal tip and shaft tube to allow the physician greater control of the device during interventional procedures.

AI/ML Overview

This document describes a 510(k) premarket notification for the ASAHI Corsair Microcatheter. The notification aims to demonstrate substantial equivalence to a legally marketed predicate device (ASAHI Corsair Microcatheter, K083127). The information provided focuses on non-clinical testing and comparison to the predicate device, rather than a clinical study with detailed acceptance criteria and performance against those criteria in a typical AI/software context.

Here's an analysis of the provided text based on your request, with the understanding that this is a medical device submission focused on substantial equivalence through non-clinical testing:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state numerical acceptance criteria in a table format with corresponding performance results for each criterion in the way one might see for an AI/software device. Instead, it describes general performance categories and confirms they met "all acceptance criteria" or performed "similarly to the predicate."

  • Acceptance Criteria (Implicit): Maintain torque durability, maintain coating integrity/peel strength, biocompatibility.
  • Reported Device Performance:
    • Torque Durability: "maintained the torque durability within the acceptance criteria."
    • Coating Integrity/Peel Testing: "confirmed equivalent performance of the Subject device as compared to its predicate."
    • Biocompatibility: "supports that the modifications to the coating material continue to render the device biocompatible."
    • Overall: "The in vitro bench test demonstrated that the ASAHI Corsair Microcatheter met all acceptance criteria and performed similarly to the predicate devices."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document describes "non clinical laboratory testing" and "in vitro bench test." It does not specify sample sizes for these tests. The data provenance is implied to be from the manufacturer's internal testing ("ASAHI Intecc Co., Ltd." in Japan, with branch offices globally). This is not a clinical study; therefore, terms like retrospective/prospective clinical data or country of origin for patient data are not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the device is a physical microcatheter, and the evaluation involves bench testing for mechanical and material properties, not interpretation of medical images or data by experts. Ground truth in this context would be defined by engineering specifications and biocompatibility standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a non-clinical bench study, not a clinical study requiring adjudication of expert interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device (microcatheter), not an AI or software device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the non-clinical testing would be defined by:

  • Engineering specifications and standards: For torque durability and coating integrity/peel strength, the acceptance criteria are likely derived from internal specifications and/or industry standards for microcatheter performance.
  • Biocompatibility standards: For biocompatibility, the ground truth refers to established international standards (e.g., ISO 10993 series) for evaluating biological responses to medical devices. The specific tests mentioned (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity (acute), and hemocompatibility) are standard tests within these guidelines.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI or software device that undergoes "training."

9. How the ground truth for the training set was established

Not applicable. As there is no training set for a physical device, this question is not relevant.

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 23, 2015

ASAHI Intecc Co., Ltd. % Ms. Candace Cederman Senior Regulatory Affairs Consultant CardioMed Device Consultants, LLC 5523 Research Park Drive Suite 205 Baltimore, Maryland 21228

Re: K151103

Trade/Device Name: ASAHI Corsair Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: April 23, 2015 Received: April 24, 2015

Dear Ms. Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Hillebrand

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151103

Device Name ASAHI Corsair Microcatheter

Indications for Use (Describe)

The ASAHI Corsair Microcatheter is intended to provide support to facilitate the placement of Guide wires in the coronary and peripheral vasculatures and can be used to exchange one guide wire for another. The Corsair Microcatheter is also intended to assist in the delivery of contrast media into the coronary, peripheral, and abdominal vasculatures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

/akita-cho, Moriyama-ku, Nagoya, Aichi 463 1703 Tel. +81-52-768-1211 Fax. +81-52-768-1221 Branch offices: Tokyo, Nagoya, Osaka, Hong Kong, Amsterdam, Singapore, Beijing Research Facilities and Factories: Osaka, Seto, Thailand, Hanoi

510(k) Summary [as required by 21 CFR 807.92(c)] ASAHI Corsair Microcatheter 510(k) K151103

DATE PREPARED:April 29, 2015
APPLICANTASAHI Intecc Co., Ltd.1703 Wakita-cho, Moriyama-kuNagoya, Aichi 463-0024, Japan
OFFICIALCORRESPONDENTYoshi TeraiPresident, CEOASAHI Intecc USA, Inc.2500 Red Hill Avenue, Suite 210Santa Ana, CA 92705Tel: (949) 756-8252FAX: (949) 756-8165e-mail: ASAHI.ra-fda@ASAHI-intecc.com
TRADE NAME:ASAHI Corsair Microcatheter
DEVICECLASSIFICATION:Class 2 per 21 CFR §870.1250
CLASSIFICATIONNAME:Percutaneous Catheter
PRODUCT CODEDQY
PREDICATEDEVICE:ASAHI Corsair Microcatheter (K083127)

INTENDED USE/INDICATIONS FOR USE

The ASAHI Corsair Microcatheter is intended to provide support to facilitate the placement of Guide wires in the coronary and peripheral vasculatures and can be used to exchange one guide

{4}------------------------------------------------

wire for another. The Corsair Microcatheter is also intended to assist in the delivery of contrast media into the coronary, peripheral, and abdominal vasculatures.

DESCRIPTION:

The ASAHI Corsair Microcatheter consists of a distal tip and a shaft tube that are inserted into a vascular connector for catheter control and infusion of contrast media. No accessories are part of this device. The Corsair has a radiopaque marker coil that is imbedded into the inner layer of resin to facilitate the tip location during angiographic procedures. In addition, the device has a hydrophilic coating on the outer surface of the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair has a tapered shape and is designed to have increased flexibility towards the distal end. PTFE is applied to the inner lumen of the catheter for the purposes of a smooth transition and exchange of guidewires.

The microcatheter also contains wires to reinforce the distal tip and shaft tube to allow the physician greater control of the device during interventional procedures.

COMPARISON WITH PREDICATE DEVICES:

Comparisons of the ASAHI Corsair Microcatheter and predicate device show that the technological characteristics of the Subject device such as the components, design, materials, sterilization method, shelf life and operating principle are identical or similar to currently marketed predicate devices. The addition of a shorter length catheter does not raise any new questions of safety or effectiveness.

Name of DeviceASAHI Corsair MicrocatheterASAHI Corsair Microcatheter
510(k)Current ApplicationK083127
Indications for UseThe ASAHI Corsair Microcatheteris intended to provide support tofacilitate the placement of Guidewires in the coronary and peripheralvasculatures and can be used toexchange one guide wire foranother. The Corsair Microcatheteris also intended to assist in thedelivery of contrast media into thecoronary, peripheral, and abdominalvasculatures.The ASAHI Corsair Microcatheteris intended to provide support tofacilitate the placement of Guidewires in the coronary and peripheralvasculatures and can be used toexchange one guide wire foranother. The Corsair Microcatheteris also intended to assist in thedelivery of contrast media into thecoronary, peripheral, and abdominalvasculatures.
SterilizationProvided sterile via Ethylene Oxideto SAL10-6Provided sterile via Ethylene Oxideto SAL10-6
Shelf Life2 Years2 Years
Overall Length90 cm / 135 cm/ 150 cm135 cm/ 150 cm
Nominal OD0.93 mm0.93 mm
Coating Length600 mm600 mm
Coating materialHydrophilicHydrophilic

The intended use/indications between the Subject Device and its primary predicate are identical.

{5}------------------------------------------------

NON CLINICAL TESTING / PERFORMANCE DATA:

Non clinical laboratory testing was performed on the ASAHI Corsair Microcatheter to determine substantial equivalence. Torque durability testing was conducted to ensure that the shorter length of the ASAHI Corsair Microcatheter maintained the torque durability within the acceptance criteria. Coating integrity/peel testing confirmed equivalent performance of the Subject device as compared to its predicate. Confirmatory biocompatibility testing was done to ensure that the minor modifications to the coating materials had no negative impact upon biocompatibility of the device.

The in vitro bench test demonstrated that the ASAHI Corsair Microcatheter met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the Subject device functions as intended and is substantially equivalent to the predicate device.

BIOCOMPATIBILITY:

The ASAHI Corsair Microcatheter was compared to the predicate and reference devices. The ratios of the coating material were modified slightly from the ASAHI Corsair Microcatheter predicate but are the same as used on the reference ASAHI Fubuki (K141981). Confirmatory testing (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity (acute), and hemocompatibility) supports that the modifications to the coating material continue to render the device biocompatible.

CONCLUSION:

The ASAHI Corsair Microcatheter has identical intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended.

Therefore, the ASAHI Corsair Microcatheter is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).