K Number
K151046

Validate with FDA (Live)

Date Cleared
2015-07-17

(88 days)

Product Code
Regulation Number
866.3309
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The illumigene HSV 1&2 DNA amplification assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection and differentiation of herpes simplex virus type 1 (HSV-1) and herpes simplex virus type 2 (HSV-2) DNA in cutaneous and mucocutaneous lesion specimens from male and female patients suspected of Herpetic infections.

illumigene HSV 1&2 utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect HSV-1 and HSV-2 by targeting segments of the herpes simplex virus 1 and herpes simplex virus 2 genomes. Results from illumigene HSV 1&2 are used as an aid in the diagnosis of HSV infection in symptomatic patients.

The assay is intended for use in hospital, reference or state laboratory settings. This device is not intended for nonlaboratory point-of-care use.

WARNING: illumigene HSV 1&2 is not FDA cleared for use with cerebrospinal fluid (CSF) or to aid in the diagnosis of HSV infections of the central nervous system (CNS). The device is not intended for prenatal screening.

Device Description

The illumigene Molecular Diagnostic Test System is comprised of the illumigene® HSV 1&2 DNA Amplification Assay Test Kit, the illumigene HSV 1&2 External Controls Kit, and the illumipro-10™ Automated Isothermal Amplification and Detection System.

The illumigene HSV 1&2 molecular assay utilizes loop-mediated amplification (LAMP) technology to detect herpes simplex virus in cutaneous and mucocutaneous lesion swab specimens. Each illumigene HSV 1&2 assay is completed using illumigene Sample Preparation Apparatus III (SMP PREP III), illumigene HSV 1 Test Devices, illumigene HSV 2 Test Devices, Mineral Oil, illumigene Heat Treatment Tubes, and transfer pipettes. Using a transfer pipette, specimens are added to SMP PREP III and dispensed into Heat Treatment Tubes. Samples are heat-treated to make target and control DNA available for amplification. Each heat-treated sample is added to one illumigene HSV 1 and one illumigene HSV 2 Test Device. Mineral oil is added to each illumigene Test Device to prevent evaporation.

The illumipro-10™ heats each illumigene HSV 1 and HSV 2 Test Device containing prepared sample and control material, facilitating amplification of target DNA. When HSV-1 or HSV-2 is present in the specimen, a 208 base pair (bp) sequence of the HSV-1 glycoprotein G (US4) gene or a 189 bp sequence of the HSV-2 glycoprotein G (US4) gene is amplified and magnesium pyrophosphate is generated. Magnesium pyrophosphate forms a precipitate in the reaction mixture.

The illumipro-10™ monitors the absorbance characteristics of the reaction solutions at the assay Run Start (Signal initial, S initial) and at the assay Run End (Signal final, S final). The illumipro-10™ calculates the change in light transmission between Run End and Run Start (S final; S initial) and compares the ratio to an established cut-off value. The illumipro-10™ performs this ratio calculation to both the TEST chamber and the CONTROL chamber.

Fixed cut-off values for the TEST chamber are used to report sample results. TEST chamber S final:S initial ratios less than 82% are reported as 'POSITIVE'; TEST chamber S final:S initial ratios greater than or equal to 82% are reported as 'NEGATIVE'. Numerical values are not reported.

Fixed cut-off values for the CONTROL chamber are used to determine validity. CONTROL chamber S final:S initial ratios less than 90% are considered valid and allow for reporting of TEST chamber results (POSITIVE). CONTROL chamber S final:S initial ratios greater than or equal to 90% are considered invalid and prevent reporting of TEST chamber results. Invalid CONTROL chamber reactions are reported as "INVALID'. More stringent cut-off criteria are applied to the CONTROL chamber reaction is not inhibited, reagents are performing as intended and that sample processing was performed appropriately.

The illumigene HSV 1&2 External Controls Kit contains a combined HSV-1 and HSV-2 Positive Control and a Negative Control (Negative Control IV) for use in routine Quality Control testing. External Control reagents are provided to aid the user in detection of reagent deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct pass/fail thresholds in percentage terms for sensitivity and specificity. However, the clinical study results represent the device's performance against a reference method (ELVIS® HSV ID and D3 Typing Test System) and can be considered the demonstrated performance for substantial equivalence. The document describes combined site performance for both cutaneous and mucocutaneous lesions for HSV-1 and HSV-2. Because the data is presented separately for HSV-1 cutaneous, HSV-1 mucocutaneous, HSV-2 cutaneous, and HSV-2 mucocutaneous, I will aggregate the positive and negative counts to provide an overall performance.

Overall Aggregated illumigene® HSV 1&2 DNA Amplification Assay Performance vs. ELVIS® HSV ID and D3 Typing Test System (Clinical Study)

MetricHSV-1 Performance (Combined Lesions)HSV-2 Performance (Combined Lesions)
Sensitivity200 / 211 = 94.8%179 / 181 = 98.9%
Specificity729 / 763 = 95.5%930 / 974 = 95.5%

Breakdown by Lesion Type (as provided in the document)

TargetLesion TypeCases Analyzed (ELVIS)illumigene Pos (ELVIS Pos)illumigene Pos (ELVIS Neg)illumigene Neg (ELVIS Pos)illumigene Neg (ELVIS Neg)Sensitivity (95% CI)Specificity (95% CI)
HSV-1Cutaneous264486320794.1% (84.1-98.0%)97.2% (94.0-98.7%)
HSV-1Mucocutaneous71015228852295.0% (90.5-97.5%)94.9% (92.7-96.5%)
HSV-2Cutaneous30642130251100% (91.6-100.0%)95.1% (91.8-97.1%)
HSV-2Mucocutaneous84913731267998.6% (94.9-99.6%)95.6% (93.9-96.9%)

2. Sample size used for the test set and the data provenance

  • Sample Size (Test Set):
    • Clinical Study: A total of 1158 leftover, de-identified lesion swab specimens were initially evaluated.
    • For HSV-1 performance analysis, 974 specimens were included (after exclusions).
    • For HSV-2 performance analysis, 1155 specimens were included (after exclusions).
  • Data Provenance: The document does not explicitly state the country of origin. The study was conducted between October 2014 and March 2015. The specimens were "leftover, de-identified lesion swab specimens," indicating a retrospective study using previously collected samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for the clinical performance data was established by comparison to the ELVIS® HSV ID and D3 Typing Test System. This is a laboratory-based diagnostic system, not a visual assessment by individual human experts. Therefore, the concept of "number of experts" and "qualifications of those experts" for establishing ground truth in the traditional sense (e.g., radiologists) does not apply here. The "expert" in this context is the validated reference diagnostic system itself.

4. Adjudication method for the test set

Not applicable. The ground truth was established by comparing the device's results to a predicate diagnostic system (ELVIS® HSV ID and D3 Typing Test System). There was no adjudication process involving human experts resolving discrepancies for the ground truth of the clinical samples. For discrepancies between the illumigene assay and the ELVIS system, some samples were re-tested with an "alternative, FDA-cleared molecular assay," but this is a reconciliation method, not an adjudication of primary ground truth experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs. without AI assistance

Not applicable. This device is an in vitro diagnostic (IVD) assay for molecular detection of HSV DNA, not an AI-powered image analysis tool or system designed to assist human readers. Therefore, an MRMC study and evaluation of human reader improvement with/without AI assistance are not relevant to this submission.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the clinical evaluation of the illumigene® HSV 1&2 DNA Amplification Assay against the ELVIS® HSV ID and D3 Typing Test System represents a standalone performance study of the algorithm (the assay itself) as an in vitro diagnostic device without human-in-the-loop intervention for result interpretation; the illumipro-10™ system automatically interprets signals based on fixed cut-off values.

7. The type of ground truth used

The ground truth used for the clinical performance study was the ELVIS® HSV ID and D3 Typing Test System, which is a legally marketed predicate diagnostic device for HSV detection and typing. This is a form of reference standard comparison. Additionally, for discordant results, an "alternative, FDA-cleared molecular assay" was used for confirmation, further supporting the reference standard approach.

8. The sample size for the training set

The document does not explicitly state a separate "training set" sample size for the clinical study directly tied to an external ground truth comparison. However, the "Assay cut-off" section mentions that "Development optimization includes evaluation of characterized positive and negative clinical specimens. Amplification reagent concentrations are adjusted during design as needed to ensure illumigene results are aligned with clinical specimen reported results." This implies internal development and optimization based on clinical specimens, but a specific "training set" as commonly defined in machine learning or AI context, with a stated size, is not provided.

9. How the ground truth for the training set was established

As inferred above, the ground truth for any internal optimization/development (which might be considered analogous to a training set) was established through "characterized positive and negative clinical specimens." The characterization method for these specimens is not detailed beyond "reported results," but it would logically rely on established diagnostic methods.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal are three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 9, 2015

MERIDIAN BIOSCIENCE, INC. STEFANIE JOHNS, PH.D. REGULATORY AFFAIRS AND DESIGN ASSURANCE ASSOCIATE 3471 RIVER HILLS DRIVE CINCINNATI, OH 45244

Re: K151046 Trade/Device Name: illumigene® HSV 1&2 DNA Amplification Assay, illumigene® HSV 1&2 External Control Kit Regulation Number: 21 CFR 866.3309 Regulation Name: Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel Regulatory Class: Class II Product Code: PGI Dated: April 16, 2015 Received: April 20, 2015

Dear Dr. Johns:

This letter corrects our substantially equivalent letter of July 17, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Stephen J. Lovell -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K151046

Device Name

illumigene® HSV 1&2 DNA Amplification Assay, illumigene® HSV 1&2 External Control Kit

Indications for Use (Describe)

illumigene® HSV 1&2 DNA Amplification Assay:

The illumigene HSV 1&2 DNA amplification assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection and differentiation of herpes simplex virus type 1 (HSV-1) and herpes simplex virus type 2 (HSV-2) DNA in cutaneous and mucocutaneous lesion specimens from male and female patients suspected of Herpetic infections.

illumigene HSV 1&2 utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect HSV-1 and HSV-2 by targeting segments of the herpes simplex virus 1 and herpes simplex virus 2 genomes. Results from illumigene HSV 1&2 are used as an aid in the diagnosis of HSV infection in symptomatic patients.

The assay is intended for use in hospital, reference or state laboratory settings. This device is not intended for nonlaboratory point-of-care use.

WARNING: illumigene HSV 1&2 is not FDA cleared for use with cerebrospinal fluid (CSF) or to aid in the diagnosis of HSV infections of the central nervous system (CNS). The device is not intended for prenatal screening.

illumigene® HSV 1&2 External Control Kit:

The illumigene HSV 1&2 External Control Kit contains Positive Control Reagents for use with the illumigene HSV 1&2 DNA Amplification Assay. External controls are used as part of a routine quality control program to aid the user in detection of unexpected conditions that may lead to test errors.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Meridian Bioscience, Inc. The logo features a blue globe on the left side, followed by the company name in a bold, blue font. Below the company name is the tagline "Inspired Science. Trusted Solutions." in a smaller font.

illumigene® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

510(k) Summary

510(k) number:K151046Date of Preparation: 17 July 2015
Owner:Meridian Bioscience, Inc.3471 River Hills DriveCincinnati, Ohio 45244 USAPhone: (513) 271-3700Fax: (513) 272-5213
Contact:Primary Contact:Stefanie Johns, PhDRegulatory Affairs & Design Assurance Associate
Secondary Contact:Susan D. RolihExecutive Vice President, Regulatory and Quality Systems
Trade Name:illumigene® HSV 1&2 DNA Amplification Assayillumigene® HSV 1&2 External Controls
Common Name:Herpes simplex virus nucleic acid amplification assay
Classification Name:Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel(21 CFR 866.3309, Product Code PGI)
Predicate Device:Quidel Corporation Lyra™ Direct HSV 1 + 2/VZV AssayK133448

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a blue globe on the left, followed by the company name in blue, with "Meridian" on the top line and "Bioscience, Inc." on the second line. Below the company name is the slogan "Inspired Science. Trusted Solutions."

illumigene® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

Device Description:

The illumigene Molecular Diagnostic Test System is comprised of the illumigene® HSV 1&2 DNA Amplification Assay Test Kit, the illumigene HSV 1&2 External Controls Kit, and the illumipro-10™ Automated Isothermal Amplification and Detection System.

The illumigene HSV 1&2 molecular assay utilizes loop-mediated amplification (LAMP) technology to detect herpes simplex virus in cutaneous and mucocutaneous lesion swab specimens. Each illumigene HSV 1&2 assay is completed using illumigene Sample Preparation Apparatus III (SMP PREP III), illumigene HSV 1 Test Devices, illumigene HSV 2 Test Devices, Mineral Oil, illumigene Heat Treatment Tubes, and transfer pipettes. Using a transfer pipette, specimens are added to SMP PREP III and dispensed into Heat Treatment Tubes. Samples are heat-treated to make target and control DNA available for amplification. Each heat-treated sample is added to one illumigene HSV 1 and one illumigene HSV 2 Test Device. Mineral oil is added to each illumigene Test Device to prevent evaporation.

The illumipro-10™ heats each illumigene HSV 1 and HSV 2 Test Device containing prepared sample and control material, facilitating amplification of target DNA. When HSV-1 or HSV-2 is present in the specimen, a 208 base pair (bp) sequence of the HSV-1 glycoprotein G (US4) gene or a 189 bp sequence of the HSV-2 glycoprotein G (US4) gene is amplified and magnesium pyrophosphate is generated. Magnesium pyrophosphate forms a precipitate in the reaction mixture.

The illumipro-10™ monitors the absorbance characteristics of the reaction solutions at the assay Run Start (Signalmita, S) and at the assay Run End (Signalina, S). The illumipro-10™ calculates the change in light transmission between Run End and Run Start (S; S; ) and compares the ratio to an established cut-off value. The illumipro-10™ performs this ratio calculation to both the TEST chamber and the CONTROL chamber.

Fixed cut-off values for the TEST chamber are used to report sample results. TEST chamber Sy:S, ratios less than 82% are reported as 'POSITIVE'; TEST chamber S;:S; ratios greater than or equal to 82% are reported as 'NEGATIVE'. Numerical values are not reported.

Fixed cut-off values for the CONTROL chamber are used to determine validity. CONTROL chamber Sy:S, ratios less than 90% are considered valid and allow for reporting of TEST chamber results (POSITIVE). CONTROL chamber S;S; ratios greater than or equal to 90% are considered invalid and prevent reporting of TEST chamber results. Invalid CONTROL chamber reactions are reported as "INVALID'. More stringent cut-off criteria are applied to the CONTROL chamber reaction is not inhibited, reagents are performing as intended and that sample processing was performed appropriately.

The illumigene HSV 1&2 External Controls Kit contains a combined HSV-1 and HSV-2 Positive Control and a Negative Control (Negative Control IV) for use in routine Quality Control testing. External Control reagents are provided to aid the user in detection of reagent deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a blue globe on the left, followed by the company name in blue, with "Meridian" in a larger font size than "Bioscience, Inc." A thin blue line separates the two parts of the name. Below the company name is the tagline "Inspired Science. Trusted Solutions."

illumigene ® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary

Intended Use:

The illumigene HSV 1&2 DNA amplification assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection and differentiation of herpes simplex virus type 1 (HSV-1) and herpes simplex virus type 2 (HSV-2) DNA in cutaneous and mucocutaneous lesion specimens from male and female patients suspected of Herpetic infections.

illumigene HSV 1&2 utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect HSV-1 and HSV-2 by targeting segments of the herpes simplex virus 1 and herpes simplex virus 2 genomes. Results from illumigene HSV 1&2 are used as an aid in the diagnosis of HSV infection in symptomatic patients.

The assay is intended for use in hospital, reference or state laboratory settings. This device is not intended for nonlaboratory point-of-care use.

WARNING: illumigene HSV 1&2 is not FDA cleared for use with cerebrospinal fluid (CSF) or to aid in the diagnosis of HSV infections of the central nervous system (CNS). The device is not intended for prenatal screening.

Similarities
DEVICEillumigene® HSV 1&2 DNA AmplificationAssayPREDICATEQuidelLyra™ Direct HSV 1 + 2/VZV assayK133448
Intended UseQualitative in vitro diagnostic test for the directdetection and differentiation of herpes simplexvirus type 1 (HSV-1) and herpes simplex virustype 2 (HSV-2) DNA in cutaneous andmucocutaneous lesion specimens from male andfemale patients suspected of Herpetic infections.Qualitative detection and differentiation of herpessimplex virus type 1, herpes simplex virus type 2,and varicella-zoster virus DNA isolated andpurified from cutaneous or mucocutaneous lesionsamples obtained from symptomatic patientssuspected of active herpes simplex virus 1,herpes simplex virus 2 and/or varicella-zoster.
Assay ResultsQualitativeQualitative
Indications for UseProfessional UseProfessional Use
DNA DetectedHerpes simplex virus type 1Herpes simplex virus type 2Herpes simplex virus type 1Herpes simplex virus type 2Varicella-zoster virus
Typing of HSV-1 andHSV-2?YesYes
Specimen TypesMale and female cutaneous and mucocutaneouslesion swab specimensMale and female cutaneous and mucocutaneouslesion swab specimens
MethodDNA AmplificationDNA Amplification
DetectionInstrumentInstrument

Predicate Device Comparison:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a blue globe on the left side, followed by the company name in blue, bold font. Below the company name is the tagline "Inspired Science. Trusted Solutions."

illumigene® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015
Differences
DEVICEillumigene ® HSV 1&2 DNA Amplification AssayPREDICATEQuidelLyra™ Direct HSV 1 + 2/VZV assayK133448
AmplificationMethodologyLoop-Mediated Isothermal Amplification (LAMP)Multiplex Real-Time PCR
DetectionMethodologyLoop-Mediated Isothermal Amplification (LAMP)Target-specific fluorescent-labeled hybridizationprobes.
Reagents/ComponentsThe illumigene HSV 1&2 DNA Amplification AssayKit contains illumigene Sample PreparationApparatus III, illumigene HSV 1 Test Devices,illumigene HSV 2 Test Devices, Mineral Oil,illumigene Heat Treatment Tubes, and TransferPipettes.The Lyra™ Direct HSV 1 + 2/VZV Assay kitconsists of:- Rehydration Solution- Process Buffer Part M5050 (contains thePRC)- Lyra™ Direct HSV 1 + 2/VZV Master MixPart M5012Lyophilized Contents:◦ DNA polymerase enzyme◦ Primers and probes◦ dNTPs◦ Stabilizers
Instrumentationillumipro -10™ Automated Isothermal Amplificationand Detection SystemLife Technologies QuantStudio™ Dx, the AppliedBiosystems® 7500 Fast Dx, or the CepheidSmartCycler® II System
Reading MethodVisible Light TransmissionFluorescence detection of dual-labeled hydrolysisprobes, Ct values.
PackagingSupplied as a kit; 25 tests per kitSupplied as a kit; 96 tests per kit.
Kit Storage2-30°CReagents and Controls: 2-8°C
VZV detection?NoYes

NON-CLINICAL PERFORMANCE DATA

Analytical Performance:

Precision/Reproducibility:

Reproducibility studies were carried out by three independent laboratories. Blind-coded panels of eighteen (18) samples were supplied to participating laboratories. Samples were randomly sorted within each panel to mask sample identities. Contrived HSV-1 (strain HF) and HSV-2 (strain MS) positive and negative samples were manufactured using simulated negative matrix (cheek swab matrix in MicroTest™ M4° viral transport medium). The panel included moderate positive samples (HSV-1: 2.16 x 10 TCID-2: 4.80 x 10 TCIDs(mL), low positive samples (HSV-1: 1.08 x 104 TCID %/mL or HSV-2: 2.40 x 103 TCIDs/mL), and two HSV-contrived negative samples (HSV-1 near cut-off: 308 TCIDsy/mL, HSV-1 high negative: 29.7 TCID50/mL; HSV-2 near cut-off:

{8}------------------------------------------------

Image: Meridian Bioscience, Inc. logoillumigene® HSV 1&2 DNA Amplification Assay
Application Reference:Attachment Description:Application Date:Section 1: General InformationAttachment 007: 510(k) SummaryApril 16, 2015

24 TCIDs«/mL, HSV-2 high negative: 2.2 TCIDs;/mL). The panel also included two HSV-1 and HSV-2 negative samples and a Positive and Negative Control. Testing was performed by different operators at each site on the same day (intra-assay variability) for five days (inter-assay variability). Three lots of illumigene HSV 1&2 and five illumipro-10 instruments were used in this study. Positive and Negative Controls were tested with each panel. Results are provided in the tables below:

Reproducibility Study Summary for HSV-1
Site 1Site 2Site 3Total
Sample TypePercent AgreementPercent AgreementPercent AgreementPercent Agreement
HSV-1 Moderate Positive30/30100.0%30/30100.0%30/30100.0%90/90100.0%
HSV-1 Low Positive30/30100.0%30/30100.0%30/30100.0%90/90100.0%
HSV-1 & HSV-2 NearCut-Off20/3066.7%26/3086.7%20/3066.7%66/9073.3%
HSV-1 & HSV-2 HighNegative29/3096.7%30/30100.0%29/3096.7%88/9097.8%
HSV-1 Negative60/60100.0%60/60100.0%60/60100.0%180/180100.0%
Negative Control10/10100.0%10/10100.0%10/10100.0%30/30100.0%
Positive Control10/10100.0%10/10100.0%10/10100.0%30/30100.0%
Reproducibility Study Summary for HSV-2
Site 1Percent AgreementSite 2Percent AgreementSite 3Percent AgreementTotalPercent Agreement
Sample Type
HSV-2 Moderate Positive30/30100.0%30/30100.0%30/30100.0%90/90100.0%
HSV-2 Low Positive30/30100.0%30/30100.0%30/30100.0%90/90100.0%
HSV-1 & HSV-2 NearCut-Off25/3083.3%29/3096.7%25/3083.3%79/9087.8%
HSV-1 & HSV-2 HighNegative29/3096.7%30/30100.0%29/3096.7%88/9097.8%
HSV-2 Negative60/60100.0%60/60100.0%60/60100.0%180/180100.0%
Negative Control10/10100.0%10/10100.0%10/10100.0%30/30100.0%
Positive Control10/10100.0%10/10100.0%10/10100.0%30/30100.0%

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for Meridian Bioscience, Inc. The logo features a blue globe on the left side, followed by the company name in blue, with the word "Meridian" in a larger font size than "Bioscience, Inc." Underneath the company name is the tagline "Inspired Science. Trusted Solutions."

illumigene ® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

Detection Limit:

Analytical sensitivity studies were designed to determine the analytical limit of detection (LoD) of herpes simplex virus type 1 and type 2. The LoD is defined as the lowest concentration of analyte (TCID35/mL) that can be distinguished from negative samples with a high degree of probability (95%). Two strains of HSV-1 (HF and Maclntyre) and two strains of HSV-2 (G and MS) diluted in a simulated negative cheek swab matrix were used to establish the LoD with three kit lots of illumigene HSV 1&2. A minimum of four dilutions near the expected LoD were evaluated for each HSV strain during preliminary testing of twenty (20) individually prepared replicates. The preliminary LoD was confirmed through testing of an additional sixty (60) individually prepared replicates with HSV concentration at the LoD. The overall confirmed analytical LoD by HSV strain for illumigene HSV 1&2 is summarized below:

HSV TypeStrain DescriptionLoD Concentration(TCID50/mL)
Herpes simplex virustype 1HF Strain (VR-260)$7.20 x 10^3$
MacIntyre Strain (VR-539)$9.89 x 10^4$
Herpes simplex virustype 2G Strain (VR-734)$1.20 x 10^3$
MS Strain (VR-540)$1.60 x 10^3$

The final assigned LoD is 9.89 x 104 TCID56mL for HSV-1 and 1.60 x 103 TCID56mL for HSV-2.

Analytical sensitivity of illumigene HSV 1&2 was verified through testing of twenty (20) HSV-1 and twenty (20) HSV-2 confirmed positive clinical samples from a variety of anatomical locations (i.e. oral, genital, anorectal, etc.) at the assay LoD. Each clinical sample was quantified and diluted to 9.89 x 10° TCID=0mL or 1.60 x 10° TCIDs /mL for HSV-1 and HSV-2, respectively: samples quantified at concentrations less than the limit of detection were tested undiluted. All samples were detected by illumigene HSV 182 at LoD or lower, except one HSV-1 clinical sample, which was detected at 7.20 x 10' TCID50/mL.

Analytical Specificity:

Interference Testing:

Non-microbial contaminants potentially found in cutaneous swab specimens were evaluated for the potential to interfere with the illumigene HSV 1&2 assay. Each potentially interfering substance was added to contrived HSV-1 (strains HF and MacIntyre) and HSV-2 (strains G and MS) positive and negative samples at final concentrations of 60 µg/mL, 7% v/v, 7% w/v, or greater.

No interference was observed with the following substances:

(1) At concentrations of 7% w/v or v/v: Abreve (10% Docosano), Balneol Hygienic Cleansing Loton,
Carmex Original Lip Balm (1.7% camphor, 0.7% menthol Pharmacy Disposable), Fluoride toothpaste (Crest, 0.243% sodium fluoride), K-Y Brand Jelly, Lanacane (0.2% benzethonium chloride, 20% benzocaine), Lip Clear" (1.2% zinc oxide), Miconazole 3 (CVST's
Yeast Infection Relief: 2% miconazole nitrate), Mouthwash (Listerine " Origin menthol, 0.060% methyl salicylate, 0.064% thymol), Preparation H® Hemorrhoidal Ointment (14% mineral oil, 74.9% petrolatum, 0.25% phenylephrine HCl), Releev® (0.13% benzalkonium chloride), Tioconazole, Vagisil Regular Strength (5% benzocaine, 2% resorcinol), Yeast Guard® Gel Treatment (Candida albicans, 27X HPUS: Candida parapsilosis, 27X HPUS: Pulsatilla, 27X HPUS). Feces, Seminal fluid, Urine, Whole blood, Buffy Coat.

(2) At concentrations of: 60 ug/mL - Mucus (Mucin, bovine submaxillary gland type I-S); 1.25 mg/mL -Cornstarch: 3.3 mg/mL - Albumin: 5 mg/mL - Acetaminophen, Casein, Chlorpheniramine maleate; 7 mg/mL - Acyclovir; 10 mg/mL - Acetylsalicylic acid, Dextromethorphan hydrobromide.

{10}------------------------------------------------

Image: Meridian Bioscience, Inc. logoillumigene® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

Casein at concentrations greater than 5 mg/mL was found to interfere with the assay.

Cold-EEZE® Cold Remedy plus Sore Throat (zincum gluconicum 2X) was tested at 7% v/v and produced invalid results in all replicates.

Cross-Reactivity Study:

Crossreactivity studies employed contrived positive (HSV-1 strain HF and HSV-2 strain MS) and negative specimens inoculated with bacterial or fungal organisms to a minimum concentration of 1.0 x 10° CFU/mL (1.0 x 10° copies/mL where genomic DNA was used) or virus at a minimum of 1.0 x 10° TCID=ymL (1.0 x 10° copies/mL where genomic DNA or RNA was used). None of the following organisms or their genetic material reacted with illumigene HSV 1&2:

Acinetobacter calcoaceticus, Acinetobacter Iwoffii, Bacteroides fragilis, Bordetella bronchiseptica, Bordetella pertussis, Candida albicans, Candida quilliermondii, Candida krusei, Candida Iusitaniae, Candida parapsilosis, Candida tropicalis, Chlamydia trachomatis, Chlamydophila pneumoniae, Clostridium difficile, Clostridium perfringens, Corynebacterium diphtheriae, Enterococcus facalis, Enteroccus faecium, Escherichia coli (ESBL), Fusobacterium nucleatum, Gardnerella vaginalis, Haemophilus ducreyi, Haemophilus influenzae (Type A), Klebsiella pneumoniae, Lactobacillus acidophilus, Legionella pneumophila, Mobiluncus curtisii, Mobiluncus mulieris, Moraxella catarrhalis, Mycoplasma orale, Mycoplasma pneumoniae, Mycoplasma salivarium, Neisseria meningitidis, Prevotella melaninoqenica. Proteus mirabilis, Pseudomonas aeruginosa. Salmonella typhimurium, Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus saprophyticus, Streptoccus agalactiae, Streptococcus mitis, Streptococcus pneumoniae, Streptococus pyogenes, Streptococcus salivarius, Toxoplasma gondii, Treponema palladium, Trichomonas vaginalis, Ureaplasma urealyticum, Adenovirus, Coronavirus, Coxsackievirus, Cytomegalovirus, Enterovirus, Epstein Barr virus, Influenza A virus, Influenza B virus, Hepatitis C virus, Human herpes 6 virus, Human herpes 7 virus, Human herpes 8 virus, Human immunodeficiency virus type 1, Human metapneumovirus, Human papilloma virus, Measles virus, Parainfluenza virus, Respiratory syncytial virus, Rubella virus, Varicella zoster virus.

Human genomic DNA was nonreactive at 1.0 x 10 copies/mL. In addition, there was no competitive inhibition observed from the organisms listed above with HSV-1 or HSV-2 in the illumigene HSV 1&2 assay.

Assay cut-off:

The illumigene HSV 1&2 assay is manufactured with fixed cut-off values. The product is designed with a preselected cut-off value and amplification reagent concentrations are optimized to ensure appropriate reactions are obtained. Development optimization includes evaluation of characterized positive and negative clinical specimens. Amplification reagent concentrations are adjusted during design as needed to ensure illumigene results are aligned with clinical specimen reported results.

Cut-off values applied in the following manner:

The illumipro-10™ calculates the ratio of the Run End (Signal final or S) reads with the Run Start (Signal initial or S) reads and compares the ratio to an established cut-off value. The illumipro-10 performs this ratio calculation to both the TEST chamber and the CONTROL chamber.

Fixed cut-off values for the CONTROL chamber are used to determine validity. CONTROL chamber S;S, ratios less than 90% are considered valid and allow for reporting of TEST chamber results (POSITIVE). CONTROL chamber S;S, ratios greater than or equal to 90% are considered invalid. Results are reported as 'INVALID'; Test chamber results are not reported. More stringent cut-off criteria are applied to the Control chamber reaction to ensure amplification is not inhibited, reagents are performing as intended and that sample processing was performed appropriately.

{11}------------------------------------------------

Image: Meridian Bioscience Logoillumigene® HSV 1&2 DNA Amplification Assay
Image: Meridian Bioscience, Inc. Inspired Science. Trusted Solutions.Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

Fixed cut-off values for the TEST chamber are used to report sample results. TEST chamber S.S. ratios less than 82% are reported as 'POSITIVE'; TEST chamber S;.S; ratios greater than or equal to 82% are reported as 'NEGATIVE'. Numerical values are not reported.

CLINICAL PERFORMANCE DATA

Clinical Studies:

Clinical trials for the illumigene® HSV 1&2 DNA Amplification Assay, including the illumipro-10™ Automated Isothermal Amplification and Detection System, were conducted between October 2014 and March 2015. Performance characteristics of the assay were established with cutaneous lesion swab specimens from patients suspected of herpes simplex virus type 1 (HSV-1) infection. Performance characteristics of the illumigene HSV 1&2 assay were established by comparison to the ELVIS® HSV ID and D3 Typing Test System.

A total of 1158 leftover, de-identified lesion swab specimens from symptomatic male and female patients were evaluated. One sample was HSV positive by ELVIS, however, could not be identified as HSV-1 or HSV-2 (indeterminate) and another sample was tested with contaminated ELVIS cells and could not produce a result (indeterminate). Both samples were excluded from the performance analysis. . One invalid result for HSV-1 and one invalid result for HSV-2, which could not be repeated by the illumigene® HSV 1&2 DNA Amplification Assay, were also excluded from only the HSV-1 and the HSV-2 performance analysis, respectively. In addition, the ELVIS HSV ID and D Typing Test System is unable to detect HSV-1 if HSV-2 has been identified first in coinfected specimens. Therefore, if a specimen was HSV-2 positive by ELVIS it was removed from the calculation of the HSV-1 performance analysis. One hundred and eighty one (181) specimens were identified as HSV-2 positive by ELVIS and were removed from the HSV-1 data analysis, including two specimens identified as illumigene positive for both HSV-2. The swab specimens were categorized as cutaneous (skin lesion, genital - penis) or mucocutaneous (anorectal, genital-vaginal/cervical, nasal, ocular, oral, and urethral). The tables below include performance of the remaining 974 specimens tested for HSV-1 and 1155 specimens tested for HSV-2 for all sites combined.

Reference MethodillumigenePerformance
PosNegTotalINVc95% CI
Pos486a540 (0)Sensitivity48/5194.1%84.1-98.0%
illumigeneHSV 1&2Neg3b2072100 (0)Specificity207/21397.2%94.0-98.7%
Total512132640 (0)

Combined Sites: HSV-1 Cutaneous (N=264)

6/6 samples identified as HSV-1 positive by an alternative, FDA-cleared molecular assay.

1/3 samples identified as HSV-1 negative by an alternative, FDA-cleared molecular assay.

f Initial invalid results are reported within the parentheses. The final number of invalid samples remaining after repeat testing are shown before the parenthesis.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a blue globe on the left side, followed by the company name in blue, bold font. Below the company name is the tagline "Inspired Science. Trusted Solutions."

illumigene ® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

Combined Sites: HSV-1 Mucocutaneous (N=710)

Reference Methodillumigene
PosNegTotalINVdPerformance95% CI
illumigeneHSV 1&2Pos15228b1800 (1)Sensitivity152/16095.0%90.5-97.5%
Neg8c5225300 (5)Specificity522/55094.9%92.7-96.5%
Total1605507100 (6)a

4 There were six initial INV samples by illumigene negative (ELVIS negative (ELVIS negative); one repeated as illumigene HSV 1 positive (ELVIS HSV 1 positive).

8 19/28 samples identified as HSV-1 positive by an alternative, FDA-cleared molecular assay; 3 samples could not be tested.

6 7/8 samples were identified as HSV-1 negative by an alternative, FDA-cleared molecular assay: 1 sample could not be tested.

9 Initial invalid results are reported within the parentheses. The final number of invalid samples remaining after repeat testing are shown before the parenthesis.

Combined Sites: HSV-2 Cutaneous (N=306)

Reference MethodillumigenePerformance
PosNegTotalINVc95% CI
Pos4213b550 (0)Sensitivity42/42100%91.6-100.0%
illumigeneHSV 1&2Neg02512510 (1)Specificity251/26495.1%91.8-97.1%
Total422643060 (1)a

18 There was one initial INV sample by illumigene. The sample repeated as illumigene negative (ELVIS) negative).

6 8/13 samples were identified as HSV-2 positive by an alternative, FDA-cleared molecular assay; 1 sample could not be tested.

S hitial invalid results are reported within the parentheses. The final number of invalid samples remaining after repeat testing are shown before the parenthesis.

Combined Sites: HSV-2 Mucocutaneous (N=849)

Reference MethodillumigenePerformance
PosNegTotalINVd95% CI
illumigeneHSV 1&2Pos13731b1680 (0)Sensitivity 137/139 98.6%94.9-99.6%
Neg2c6796810 (1)Specificity 679/710 95.6%93.9-96.9%
Total1397108490 (1)a

4 There was one initial INV sample by illumigene. The sample repeated illumigene negative (ELVIS negative). 6 24/31 samples were identified as HSV-2 positive by an alternative, FDA-cleared molecular assay; 4 samples could not be tested.

8 1/2 samples were identified as HSV-2 negative by an alternative, FDA-cleared molecular assay.

9 Initial invalid results are reported within the parentheses. The final number of invalid samples remaining after repeat testing are shown before the parenthesis.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows a document with the title "illumigene® HSV 1&2 DNA Amplification Assay". It includes the Meridian Bioscience, Inc. logo. The document provides information such as the application reference, which is Section 1: General Information, the attachment description, which is Attachment 007: 510(k) Summary, and the application date, which is April 16, 2015.

Prevalence of HSV-1 and HSV-2 in cutaneous and mucocutaneous lesion specimens was estimated for the 1158 specimens producing valid results with the illumigene HSV 1&2 assay. One sample generated invalid results by illumigene HSV 1 and HSV 2 and it was therefore excluded; this sample is also ELVIS indeterminate due to cell contamination. Two samples tested for HSV-1 and one sample tested for HSV-2 generated invalid results by illumigene that were not repeated due to processing errors were also excluded. One sample unable to be typed by ELVIS and two HSV-1/HSV-2 coinfected specimens were included in the prevalence calculation as they produced valid illumigene results. ELVIS HSV-2 positive samples, excluded from the HSV-1 performance calculation, were included in the prevalence calculation. Of the remaining eligible samples producing valid results. 306 were from cutaneous lesions; there were 849 specimens tested for HSV-1 and 850 specimens tested for HSV-2 from mucocutaneous lesions. The study population included specimens from pediatric, adult, and geriatric patients, with ages ranging from 1 day to 89 years. There were three samples tested for HSV-1 and two tested for HSV-2 mucocutaneous from patients with unknown age.

The prevalence of HSV-1 and HSV-2 by the illumigene HSV 1&2 assay by anatomical location and patient age is provided in the tables below.

HSV-1HSV-2
LocationTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
Genital - Penis9277.6%922830.4%
Skin Lesion2144722.0%2142712.6%

Prevalence by Anatomical Location (All Sites) - Cutaneous (N=306)

Prevalence by Anatomical Location (All Sites) - Mucocutaneous

LocationHSV-1 (N=849)HSV-2 (N=850)
Total #Total PositivePrevalenceTotal #Total PositivePrevalence
Anorectal47 (1*)714.9%46 (2*)919.6%
Genital - Vaginal/Cervical624 (2*)11217.9%62615825.2%
Nasal18950.0%1800.0%
Ocular2000.0%2000.0%
Oral Lesion1355440.0%13521.5%
Urethral5120.0%500.0%
  • Number of samples producing invalid illumigene results, which could not be resolved and therefore, were excluded from the analysis.

Prevalence by Age (All Sites) - Cutaneous (N=306)

HSV-1HSV-2
AgeTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
≤ 5 years381231.6%3812.6%
6 to 11 years14750.0%1417.1%
12 to 21 years511427.5%5147.8%
22 to 59 years1661810.8%1663621.7%
≥60 years3738.1%371335.1%
Not Provided000.0%000.0%

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a blue globe on the left side, followed by the company name in blue, bold font. Below the company name is the tagline "Inspired Science. Trusted Solutions." also in blue.

illumigene® HSV 1&2 DNA Amplification Assay
Application Reference:Section 1: General Information
Attachment Description:Attachment 007: 510(k) Summary
Application Date:April 16, 2015

Prevalence by Age (All Sites) – Mucocutaneous

AgeTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
≤ 5 years47817.0%4700.0%
6 to 11 years1200.0%1200.0%
12 to 21 years174 (1*)4626.4%1754224.0%
22 to 59 years550 (1*)11120.2%55111621.1%
≥60 years63 (1*)1828.6%63 (1*)1117.5%
Not Provided300.0%2 (1*)00.0%
  • Number of samples producing invalid illumigene results, which could not be resolved and therefore, were excluded from the analysis.

Expected Values:

The observed expected values in the clinical study are calculated using all eligible cutaneous and mucoutaneous lesion samples submitted for HSV testing. The overall incidence of HSV infection by the illumigene HSV 1&2 assay during the clinical study is 20.5% (237/1155) for HSV-1 and 19.4% (224/1156) for HSV-2. Three HSV-1 and two HSV-2 samples producing invalid illumigene results that could not be resolved were excluded from the total eligible sample population.

CONCLUSION

The illumigene® HSV 1&2 DNA amplification assay, performed on the illumipro-10™, can be used to detect HSV-1 and HSV-2 in cutaneous and mucocutaneous lesion specimens and is substantially equivalent to the predicate device.

§ 866.3309 Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel.

(a)
Identification. A herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel is a qualitative in vitro diagnostic device intended for the simultaneous detection and differentiation of different herpes viruses in cutaneous and mucocutaneous lesion samples from symptomatic patients suspected of Herpetic infections. Negative results do not preclude infection and should not be used as the sole basis for treatment or other patient management decisions. The assay is not intended for use in cerebrospinal fluid samples.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer design and selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation of a clinical study using lesion samples in which Herpes Simplex Virus 1, Herpes Simplex Virus 2, or Varicella Zoster Virus DNA detection was requested. The study must compare the device performance to an appropriate well established reference method.
(4) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(5) The device labeling must include a limitation statement that reads: “The device is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS).”
(6) Premarket notification submissions must include quality assurance protocols and a detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(7) The risk management activities performed as part of the manufacturer's 21 CFR 820.10(c) design and development activities must document an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.