K Number
K151014
Manufacturer
Date Cleared
2015-07-13

(88 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AURA DR 43C-AG detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general-purpose diagnostic procedures. It is not to be used for mammography.

Device Description

AURA DR 43C-AG is a digital X-ray flat panel detector which has 43cm x 43cm imaging area and communicates with a wired communication feature, Giga-bit Ethernet communication method through connecting a tether cable. The device intercepts x-ray photons and the scintillator (Gadox:Tb) and emits visible spectrum photons that illuminate an array of photo (a-SI) detector that creates electrical signals. After the electrical signals are generated, it is converted to digital values, and the images will be displayed on the monitor. This device should be integrated with an operating PC and an X-Ray generator. It can digitalize x-ray images and transfer them for radiography diagnostics. Advanced digital image processing allows considerably efficient diagnosis, all kinds of information management, and sharing of image information on network.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (AURA DR 43C-AG Digital Flat Panel X-ray Detector) and focuses on establishing substantial equivalence to a predicate device, rather than presenting a study demonstrating the device meets a specific set of acceptance criteria in the way one might for an AI/ML clinical study.

Therefore, many of the requested categories for a study proving acceptance criteria cannot be directly extracted from this document, as the submission takes a different approach to demonstrating safety and effectiveness.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document doesn't define explicit acceptance criteria in the typical "pass/fail" numerical sense for clinical performance. Instead, it demonstrates performance by showing the proposed device is equivalent or superior to a predicate device based on specific technical characteristics and non-clinical tests.

Acceptance Criteria CategoryReported Device Performance (AURA DR 43C-AG)Comparison/Remark to Predicate (LLX240AB01)
Intended UseDigital imaging solution for general radiographic system for human anatomy, intended to replace film/screen based systems. Not for mammography.Same
Technological CharacteristicsAmorphous Silicon, TFT detector, Gadox:Tb scintillator, 17x17 inch imaging area, 3072x3072 pixel matrix, 140 µm pixel pitch, 3.5 lp/mm resolution, 16 bit A/D conversion, 16384 (14bit) grayscale, RAW data output convertible to DICOM 3.0.Similar/Same (e.g., pixel pitch and A/D conversion are similar, others are same).
Operating PrincipleIntercepts x-ray photons, scintillator emits visible photons, photo detector creates electrical signals converted to digital values.Same
Design FeaturesDigital X-ray flat panel detector, 43cm x 43cm imaging area, wired communication (Giga-bit Ethernet).Similar
Performance (Non-clinical)DQE and MTF NPS values are equivalent or performed better than the predicate. Offers better resolution performance at 0-3.5 lp/mm. More efficient in utilizing input image signal at same patient exposure.Equivalent or Better
Electrical Safety & EMCComplies with IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) and IEC 60601-1-2: 2007.Complies with same standards
Software ValidationPerformed software validation and verification testing.Not explicitly compared but assumed compliance with standards
Risk ManagementAnalyzed with FMEA, specific risk controls implemented, determined all risks satisfactorily mitigated and accepted.Not explicitly compared but assumed compliance with standards

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not mention a "test set" in the context of clinical images or patient data. The non-clinical performance testing involved measurements of physical values (DQE, MTF NPS) usually conducted on a device itself or phantoms, not a patient sample. No clinical studies were conducted for this 510(k) submission.
  • Data Provenance: Not applicable as no clinical data was used for testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable as no clinical studies were performed, and thus no expert ground truth was established for a test set of images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable as no clinical studies were performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done, as this is an X-ray detector, not an AI-powered diagnostic tool. The submission focuses on device characteristics and substantial equivalence to a predicate, not on human reader performance with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is an imaging hardware device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable as no clinical images or patient data requiring ground truth were used in the testing described in this 510(k) summary. The "ground truth" for the non-clinical performance relied on standardized measurement methods for DQE, MTF, and NPS.

8. The sample size for the training set

  • Not applicable, as this is an X-ray detector and not an AI/ML device that requires a training set of data.

9. How the ground truth for the training set was established

  • Not applicable, as no training set was used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 13, 2015

AbyzR Co., Ltd. % Mr. David Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K151014

Trade/Device Name: AURA DR 43C-AG Digital Flat Panel X-ray Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: June 17, 2015 Received: June 18, 2015

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

For

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151014

Device Name AURA DR 43C-AG Digital Flat Panel X-ray Detector

Indications for Use (Describe)

The AURA DR 43C-AG detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general-purpose diagnostic procedures. It is not to be used for mammography.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 07

510(k) SUMMARY

{4}------------------------------------------------

1. Traditional 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.

Date 510K summary prepared : June 17, 2015

Submitter's Name, address, telephone number, a contact person:

Submitter's Name :AbyzR Co., Ltd.
Submitter's Address:#63-27, Geumgok-ro, Dongtan-myeon, Hwaseong-si,
Gyeonggi-do, 445-811, Korea
Submitter's Telephone:Tel:+ 82-70-8680-4100
Contact person:Mr. John Lim / Sr. Manager (johnlim0107@abyzr.com)
Official Correspondent:Dave Kim, MBA
Address:8310 Buffalo Speedway, Houston, TX 77025
Telephone:+713-467-2607
Fax:+713-583-8988
Email:davekim@mtech-inc.net

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary name:AURA DR 43C-AG
Regulation Name:Stationary X-ray System
Regulation Number:21 CFR 892.1680
Regulatory Class:II
Product Code:MQB
Predicate Device
ManufacturerSamsung Mobile Display Co., Ltd
DeviceLLX240AB01
Regulation NameStationary X-ray System
Regulation Number21 CFR 892.1680
Product CodeMQB
510(k) NumberK102587
Decision DateDEC 1, 2010 (Regulation Name Revised on Aug 23, 2013)

2. Device Description

AURA DR 43C-AG is a digital X-ray flat panel detector which has 43cm x 43cm imaging area and communicates with a wired communication feature, Giga-bit Ethernet communication method through connecting a tether cable. The device intercepts x-ray photons and the scintillator (Gadox:Tb) and emits visible spectrum photons that illuminate an array of photo (a-SI) detector that creates electrical signals. After the electrical signals are generated, it is converted to digital values, and the images will be displayed on the monitor. This device should be integrated with an operating PC and an X-Ray generator. It can digitalize x-ray images and transfer them for radiography diagnostics. Advanced digital image processing allows considerably efficient diagnosis, all kinds of information management, and sharing of image

{5}------------------------------------------------

information on network.

3. Indications for Use

The AURA DR 43C-AG detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all generalpurpose diagnostic procedures. It is not to be used for mammography.

4. Summary of Design Control Risk management

The AURA DR 43C-AG detector has been developed to meet the critical functional requirements and international safety standards. The risks and the hazardous impact of the device design were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the device design and production phase were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design and production were successfully mitigated and accepted.

5. Summary of the technological characteristics of the device compared to the predicate device:

The AURA DR 43C-AG detector described in this 510(k) has similar indications for use and technical characteristics as the predicate device, LLX240AB01 digital flat panel X-ray detector manufactured by Samsung Mobile Display Co., Ltd.

6. Substantial Equivalence

The AURA DR 43C-AG detector and components conform to the FDA recognized standards as like the predicate device. Based on the recognized standard conformity evidences related to electro-, mechanical-, software-,nonclinical-, and risk management, it is confirmed that AURA DR 43C-AG is substantially equivalent to the predicate device.

CharacteristicProposedAbyzRAURA DR 43C-AGProposedSamsung Mobile DisplayLLX240AB01Remark
510(k) number-K102587
Intended UseThe AURA DR 43C-AG detector isindicated for digital imagingsolution designed for generalradiographic system for humananatomy. It is intended to replacefilm or screen based radiographicsystem in all general-purposediagnostic procedures. It is not tobe used for mammography.LLX240AB01 Digital Flat Panel X-RayDetector is indicated for digitalimaging solution designed forproviding general radiographicdiagnosis of human anatomytargeting both adult and children. Itis intended to replace film basedradiographic diagnostic systems andprovide a case diagnosis andtreatment planning for physiciansand other health care professionals.Not to be used for mammography.Same

{6}------------------------------------------------

Detector TypeAmorphous Silicon, TFTAmorphous Silicon, TFTSame
scintillatorGadox:TbGadox:TbSame
Imaging Area17 x 17 inches17 x 17 inchesSame
Pixel matrix3072 x 3072 (9.4 million)3072 x 3072 (9.4 million)Same
Pixel pitch140 μm143 μmSimilarity
Resolution3.5 lp/mm3.5 lp/mmSame
A/D conversion16 bit14 bitSimilarity
Grayscale16384 (14bit)16384 (14bit)Same
Data outputRAW*The RAW files are convertibleinto DICOM 3.0 by console S/WRAW*The RAW files are convertible intoDICOM 3.0 by console S/WSame
Dimensions460 x 460 x 15 mm500 x 496.6 x 45 mmSimilarity
ApplicationGeneral Radiology systemAvailable with upright stand,table, universal standGeneral Radiology systemAvailable with upright stand, table,universal standSame
pictureImage: detectorImage: detectorSimilarity

When compared to the predicate device (K102587), the AURA DR43C-AG presented in this submission has very similar characteristics as the predicate device in terms of:

  • Intended Use
  • Technological characteristics
  • · Operating principle
  • · Design features
  • Performance
  • · Communication Method

There are no significant difference between the AURA DR43C-AG and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.

7. Performance Testing/Data

To minimize electrical and mechanical hazards, AbyzR CO., LTD. adheres to recognized and established industry practice, and all equipment complies with the relevant FDA and international standards. For example, electrical, mechanical safety and performance testing according to standard IEC

{7}------------------------------------------------

60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) was performed and EMC testing was conducted in accordance with standard IEC 60601-1-2: 2007. EN 60601-1-2:2007

Through verification and validation activities, the safety and effectiveness of AURA DR 43C-AG is verified and validated. Engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or concerns or identify new risks.

The instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.

8. Description of non-clinical tests.

- Non-clinical study

The non-clinical performance testing constrains that the main physical values for comparison of X-ray devices like DQE, MTF NPS are basically equivalent or performed better compared to the predicate device LLX240AB01 (K102587). The results show that AURA DR 43C-AG offers better resolution performance than LLX240AB01 at 0 ~ 3.5lp/mm spatial frequencies. Moreover, the ability of AURA DR 43C-AG to utilize the input image signal is more efficient than LLX240AB01 at same patient exposure.

-Electrical safety and EMC

AURA DR43C-AG has been tested for electrical safety standard IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 *2007) + AM1 (2012) and electromagnetic compatibility IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012).

The software validation and verification testing was also performed. The results of nonclinical testing indicate that the AURA DR 43C-AG detector is as safe and effective as the predicate device.

Compliance evidences were submitted for the following standards:

  • IEC 60601-1: Test Report issued by 3rd party testing lab A
  • IEC 60601-1-2: Test Report issued by 3rd party testing lab A
  • A ISO 14971: Risk management file

9. Clinical tests

Clinical testing is not necessary to support substantial equivalence of AURA DR 43C-AG with regard to LLX240AB01, the predicate device.

10. Conclusion as to Substantial Equivalence

The AURA DR43C-AG detector is substantially equivalent to the predicate device LLX240AB01 (K102587). Both subject and predicate devices are same or very similar in the intended use, the design principle, the performance and the applicable standards. Some characteristics, for example, their appearance, pitch and weight are different. However the compliance reports, performance demonstrations in this submission STED provide demonstration that these differences do not raise any new questions of safety and effectiveness. Therefore, AbyzR CO., LTD. concludes the AURA DR43C-AG digital flat panel detector is substantially equivalent with the predicate device LLX240AB01 (K102587) of Samsung Mobile Display Co., Ltd.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.