The NuVasive® CoRoent Thoracolumbar Implants are devices indicated for the following:
Intervertebral Body Fusion
The NuVasive CoRoent Thoracolumbar Implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The implants are designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
Thoracic Spine:
The CoRoent Thoracolumbar Implants are intended for use for thoracic interbody fusions at one or two contiguous levels in the thoracic spine, from T1-T2 to T11-T12, or at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic disc degeneration (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain).
Lumbar Spine:
The CoRoent Thoracolumbar Implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L1-L2 to L5-S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Partial Vertebral Body Replacement
The NuVasive CoRoent Thoracolumbar Implants may also be used as a partial vertebral body replacement device indicated for use in the thoracolumbar spine (TI to LS) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The implants are intended to be used with supplemental internal spinal fixation systems (e.g., pedicle screw system) that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft material may be used at the surgeon's discretion.
The CoRoent Thoracolumbar Implants are manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026 or titanium alloy (Ti-6Al-4V) conforming to ASTM F136. The PEEK device contains radiographic markers manufactured from titanium alloy conforming to ASTM F136 or ASTM F1472. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. The hollow aperture allows for packing of autogenous bone graft. Teeth on the superior and inferior surface of the implants provide resistance to expulsion. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. This Special 510(k) is for additional implant offerings for use in the thoracic spine.
The provided document is a 510(k) summary for the NuVasive® CoRoent® Thoracolumbar Implants. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that a device meets specific acceptance criteria through a clinical study or performance data against a predefined benchmark.
Therefore, the document explicitly states:
- "No performance data was provided to demonstrate that the subject NuVasive CoRoent Thoracolumbar Implants are substantially equivalent to the predicate devices." (Page 5, "G. Performance Data")
- "Based on the technological characteristics and comparison to predicate devices, the subject NuVasive CoRoent Thoracolumbar Implants has been shown to be substantially equivalent to legally marketed predicate devices." (Page 6, "H. Conclusions")
Given this information, it is not possible to provide the details requested in "1. A table of acceptance criteria and the reported device performance" or any other points related to a study proving the device meets acceptance criteria, as such a study and performance data were not part of this 510(k) submission.
The document focuses on demonstrating substantial equivalence through:
- Device Description: Material composition (PEEK-Optima® LT-1 or titanium alloy), design features (hollow aperture for bone graft, teeth for expulsion resistance), and variations in shapes and sizes.
- Intended Use: Intervertebral body fusion for thoracic and lumbar spine, and partial vertebral body replacement in the thoracolumbar spine.
- Technological Characteristics: Comparison to predicate devices in areas including design, labeling/intended use, material composition, and function.
In summary, this document does not contain information about:
- A table of acceptance criteria and reported device performance.
- Sample sizes for test sets, data provenance, or ground truth establishment methods.
- Number of experts, their qualifications, or adjudication methods.
- MRMC comparative effectiveness studies or standalone algorithm performance.
- Ground truth types or training set sample sizes for AI/algorithmic performance, as this is a medical implant, not an AI device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2015
NuVasive, Incorporated Ms. Cynthia Adams Regulatory Affairs Specialist 7475 Lusk Boulevard San Diego, California 92121
Re: K150994
Trade/Device Name: NuVasive® CoRoent® Thoracolumbar Implants Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: PHM, MAX, MQP Dated: May 21, 2015 Received: May 22, 2015
Dear Ms. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Cynthia Adams
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150994
Device Name
NuVasive® CoRoent® Thoracolumbar Implants
Indications for Use (Describe)
The NuVasive® CoRoent Thoracolumbar Implants are devices indicated for the following:
Intervertebral Body Fusion
The NuVasive CoRoent Thoracolumbar Implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The implants are designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
Thoracic Spine:
The CoRoent Thoracolumbar Implants are intended for use for thoracic interbody fusions at one or two contiguous levels in the thoracic spine, from T1-T2 to T11-T12, or at the thoracolumbar junction (T12-L1), following discectory for the treatment of a symptomatic disc degeneration (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain).
Lumbar Spine:
The CoRoent Thoracolumbar Implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L1-L2 to L5-S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Partial Vertebral Body Replacement
The NuVasive CoRoent Thoracolumbar Implants may also be used as a partial vertebral body replacement device indicated for use in the thoracolumbar spine (TI to LS) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The implants are intended to be used with supplemental internal spinal fixation systems (e.g., pedicle screw system) that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft material may be used at the surgeon's discretion.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller, stylized font.
510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
A. Submitted by:
Cynthia Adams Regulatory Affairs Specialist NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800
Date Prepared: June 11, 2015
B. Device Name
| Trade or Proprietary Name: | NuVasive® CoRoent® Thoracolumbar Implants |
|---|---|
| Common or Usual Name: | Intervertebral Body Fusion Device; Spinal IntervertebralBody Fixation Orthosis |
| Classification Name: | Intervertebral Body Fusion Device; Spinal IntervertebralBody Fixation Orthosis |
| Device Class: | Class II |
| Classification: | 21 CFR § 888.3080 and § 888.3060 |
| Product Code: | RJ Q. 'MAX. MOP |
C. Predicate Devices
The subject NuVasive CoRoent Thoracolumbar Implants are substantially equivalent to the primary predicate device, NuVasive CoRoent Thoracolumbar Implants (K140659) and additional predicate NuVasive CoRoent System (K141665).
D. Device Description
The CoRoent Thoracolumbar Implants are manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026 or titanium alloy (Ti-6Al-4V) conforming to ASTM F136. The PEEK device contains radiographic markers manufactured from titanium alloy conforming to ASTM F136 or ASTM F1472. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. The hollow aperture allows for packing of autogenous bone graft. Teeth on the superior and inferior surface of the implants provide resistance to expulsion. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. This Special 510(k) is for additional implant offerings for use in the thoracic spine.
{5}------------------------------------------------
E. Intended Use
The NuVasive® CoRoent Thoracolumbar Implants are devices indicated for the following:
Intervertebral Body Fusion
The NuVasive CoRoent Thoracolumbar Implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The implants are designed for use with autogenous bone graft to facilitate fusion and supplemental internal spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
Thoracic Spine:
The CoRoent Thoracolumbar Implants are intended for use for thoracic interbody fusions at one or two contiguous levels in the thoracic spine, from T1-T2 to T11-T12, or at the thoracolumbar junction (T12-L1), following discectomy for the treatment of a symptomatic disc degeneration (DDD), including thoracic disc herniation (myelopathy and/or radiculopathy with or without axial pain).
Lumbar Spine:
The CoRoent Thoracolumbar Implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L1-L2 to L5-S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Partial Vertebral Body Replacement
The NuVasive CoRoent Thoracolumbar Implants may also be used as a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The implants are intended to be used with supplemental internal spinal fixation systems (e.g., pedicle screw system) that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.
F. Technological Characteristics
As was established in this submission, the subject NuVasive CoRoent Thoracolumbar Implants are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. There have been no design changes to the implants previously cleared in the predicate 510(k)s. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function.
G. Performance Data
No performance data was provided to demonstrate that the subject NuVasive CoRoent Thoracolumbar Implants are substantially equivalent to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in purple.
H. Conclusions
Based on the technological characteristics and comparison to predicate devices, the subject NuVasive CoRoent Thoracolumbar Implants has been shown to be substantially equivalent to legally marketed predicate devices.