K Number
K150962

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-08-28

(140 days)

Product Code
Regulation Number
866.3305
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Focus Diagnostics Simplexa™ HSV 1 & 2 Direct assay is intended for use on the 3M Integrated Cycler instrument for the qualitative detection and differentiation of herpes simples virus (HSV-1 and HSV-2) DNA present in genital lesion swabs samples from patients with signs and symptoms of HSV-1 or HSV-2 infection of the genitalia. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 genital infections.

The assay is not intended for use as a screening test for the presence of HSV-1 and HSV-2 in blood or blood products. The assay is for professional use only.

Simplexa™ HSV 1 & 2 Positive Control Pack

The Simplexa™ HSV 1 & 2 Positive Control Pack is intended to be used as a control with the Simplexa™ HSV 1 & 2 Direct kit. This control is not intended for use with other assays or systems.

Device Description

The Simplexa™ HSV 1 & 2 Direct assay system is a real-time PCR that enables the direct amplification, detection and differentiation of HSV-1 and/or HSV-2 DNA from unprocessed genital swab specimens without nucleic acid extraction. The system consists of the Simplexa™ HSV 1 & 2 Direct assay, the 3M Integrated Cycler (with 3M Integrated Cycler Studio Software), the Direct Amplification Disc and associated accessories.

In the Simplexa™ HSV 1 & 2 Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify HSV-1, HSV-2 and internal control targets. Well conserved regions of the HSV-1 and HSV-2 DNA polymerase genes are targeted to identify HSV-1 and HSV-2 DNA respectively in the specimen. An internal control is used to detect PCR failure and/or inhibition.

AI/ML Overview

The provided document describes the Simplexa™ HSV 1 & 2 Direct assay and its performance evaluation. Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct "acceptance criteria" but can be inferred from the reported performance results and the common thresholds for such diagnostic tests (e.g., typically ≥ 95% agreement/sensitivity/specificity). The document presents the performance in terms of percent agreement with expected results for reproducibility and sensitivity/specificity for clinical agreement.

Reproducibility (Inter-site, Inter-day, Inter/Intra-assay)

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
HSV-1 Result - Total % Agreement with Expected Results (all sites combined)High agreement (e.g., ≥95%)HSV-1 Low Positive: 100.0% (90/90) HSV-1 Medium Positive: 100.0% (90/90) High Negative (for HSV-1): 93.3% (84/90) Positive Control (for HSV-1): 100.0% (89/89) Total Agreement (HSV-1): 98.5% (531/539)
HSV-2 Result - Total % Agreement with Expected Results (all sites combined)High agreement (e.g., ≥95%)HSV-1 Low Positive (for HSV-2): 98.9% (89/90) HSV-1 Medium Positive (for HSV-2): 100.0% (90/90) HSV-2 Low Positive: 94.4% (85/90) HSV-2 Medium Positive: 100.0% (90/90) High Negative (for HSV-2): 94.4% (85/90) Positive Control (for HSV-2): 100.0% (89/89) Total Agreement (HSV-2): 98.0% (528/539)
DNA IC Result - Total % Agreement with Expected Results (all sites combined)High agreement (e.g., ≥95%)100.0% across all sample types (e.g., HSV-1 Low Positive: 100.0% (90/90), Positive Control: 100.0% (89/89)) Total Agreement (IC): 100.0% (539/539)

Analytical Sensitivity/Limit of Detection (LoD)

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
LoD for HSV-1 McIntyreDetect positive ≥95% of the time32/32 (100%) at 4 TCID50/mL
LoD for HSV-1 HFDetect positive ≥95% of the time32/32 (100%) at 160 TCID50/mL
LoD for HSV-2 GDetect positive ≥95% of the time32/32 (100%) at 2 TCID50/mL
LoD for HSV-2 MSDetect positive ≥95% of the time31/32 (~97%) at 10 TCID50/mL

Clinical Agreement (Prospective Study vs. Composite Comparator)

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance (HSV-1)Reported Device Performance (HSV-2)
SensitivityHigh sensitivity (e.g., ≥95%)97.4% (111/114) (95% CI: 92.5% to 99.1%)97.2% (175/180) (95% CI: 93.7% to 98.8%)
SpecificityHigh specificity (e.g., ≥95%)98.2% (560/570) (95% CI: 96.8% to 99.0%)97.8% (497/508) (95% CI: 96.2% to 98.8%)
Positive Predictive Value (PPV)High predictive value (context-dependent)91.7% (111/121) (95% CI: 85.5% to 95.4%)94.1% (175/186) (95% CI: 89.7% to 96.7%)
Negative Predictive Value (NPV)High predictive value (context-dependent)99.5% (560/563) (95% CI: 98.4% to 99.8%)99.0% (497/502) (95% CI: 97.7% to 99.6%)

Clinical Agreement (Retrospective Study vs. Validated Bi-Directional Sequencing Assay)

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance (HSV-1)Reported Device Performance (HSV-2)
Positive Percent Agreement (PPA)High agreement (e.g., ≥95%)100.0% (14/14) (95% CI: 78.5% to 100.0%)100.0% (14/14) (95% CI: 78.5% to 100.0%)
Negative Percent Agreement (NPA)High agreement (e.g., ≥95%)92.9% (13/14) (95% CI: 68.5% to 98.7%)100.0% (14/14) (95% CI: 78.5% to 100.0%)
Positive Predictive Value (PPV)High predictive value (context-dependent)93.3% (14/15) (95% CI: 70.2% to 98.8%)100.0% (14/14) (95% CI: 78.5% to 100.0%)
Negative Predictive Value (NPV)High predictive value (context-dependent)100.0% (13/13) (95% CI: 77.2% to 100.0%)100.0% (14/14) (95% CI: 78.5% to 100.0%)

2. Sample Size Used for the Test Set and Data Provenance

  • Reproducibility Study:

    • Test Set Size: For each of the six sample pools (HSV-1 Low Positive, HSV-1 Medium Positive, HSV-2 Low Positive, HSV-2 Medium Positive, High Negative, Positive Control), they were tested in triplicate on five different days by two operators at each of three sites. This amounts to: 6 samples * 3 replicates * 5 days * 2 operators * 3 sites = 540 data points (assuming complete datasets). The tables consolidate this, showing "Total Agreement" out of 539 samples for HSV-1, and 539 for HSV-2 and DNA IC across all categories (e.g., 531/539 for HSV-1 Total Agreement, 528/539 for HSV-2 Total Agreement, 539/539 for DNA IC Total Agreement).
    • Data Provenance: The study was conducted at "Three investigative sites," implying a multi-center study within the U.S. (typical for FDA submissions). It's a prospective design as samples were "contrived" (prepared specifically for the study) and tests were performed over different days.
  • Clinical Agreement (Prospective Study):

    • Test Set Size:
      • Initially, 718 genital swab samples were prospectively collected.
      • 9 samples were removed (not tested or invalid results on 3 assays).
      • 13 samples were removed (not tested on comparator method sufficiently).
      • 696 samples were used for the primary analysis.
      • Further exclusions for specific analyses: 6 samples excluded from discordant analysis due to insufficient volume for FDA cleared NAAT; 2 samples excluded for HSV-1/HSV-2 discrepancy due to insufficient volume for FDA cleared NAAT.
      • Final analysis for HSV-1: 684 samples.
      • Final analysis for HSV-2: 688 samples.
    • Data Provenance: "Prospectively collected from patients with signs and symptoms of genital herpes simplex virus (HSV) infection from 6 geographically diverse locations." This indicates the data is from the U.S. and is prospective.
  • Clinical Agreement (Retrospective Study):

    • Test Set Size: 28 genital swab samples (14 positive HSV-1 and 14 positive HSV-2).
    • Data Provenance: "Retrospectively collected from male patients with signs and symptoms of genital herpes simplex virus (HSV) infection." Specific country of origin is not stated but implies within the U.S. given the FDA submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • For the Reproducibility Study, the ground truth for contrived samples (low positive, medium positive, high negative, positive control) was established based on the known concentrations of HSV-1 and HSV-2 spiked into the matrix. This is an analytical study, not one requiring clinical experts for ground truth.

  • For the Clinical Agreement Studies (Prospective and Retrospective), the ground truth was established by a "composite comparator algorithm" and "validated bi-directional sequencing assay." This ground truth relies on laboratory results rather than expert interpretation of clinical data or images. Therefore, the document does not specify human experts for establishing ground truth, but rather relies on the interpretative methods of the comparator assays. Those interpreting the culture results, sequencing data, and FDA-cleared NAAT results would typically be trained laboratory personnel or clinical laboratorians.


4. Adjudication Method for the Test Set

  • Reproducibility Study: No explicit adjudication method is mentioned as the ground truth was "expected results" for contrived samples based on known concentrations. Adjudication wasn't necessary for these controlled samples.

  • Clinical Agreement (Prospective Study): An adjudication method was used for discordant results in the composite comparator algorithm:

    • "A positive result for HSV-1 and/or HSV-2 was determined by a positive test result in either the culture or the bi-directional sequencing."
    • "If both the culture and the bi-directional sequencing yielded positive results but disagreed in the differentiation of HSV-1 versus HSV-2, the results of the FDA cleared NAAT were used and a 2 out of 3 rule was followed to determine the type of the virus (e.g. if two of the methods were positive for HSV-1, the final comparator result was HSV-1 positive)." This is a form of 2-out-of-3 or majority rule adjudication.
  • Clinical Agreement (Retrospective Study): The ground truth was established by a "validated bi-directional sequencing assay." No specific adjudication process is described beyond the sequencing assay itself.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study focuses on an in vitro diagnostic (IVD) PCR assay for the detection of viral DNA, not on image interpretation or other tasks typically performed by human readers that could be augmented by AI. Therefore, there's no mention of human readers or AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, the performance presented for the Simplexa™ HSV 1 & 2 Direct assay is standalone performance. It describes the diagnostic accuracy of the assay itself (device only) against a comparator method. The device is an automated, real-time PCR system, and its output (qualitative detection and differentiation of HSV-1 and HSV-2 DNA) is directly compared. The assay operates without human "in-the-loop" interpretation for the final diagnostic call.

7. The Type of Ground Truth Used

  • Reproducibility Study: Known concentrations of spiked viral material (contrived samples) were used as ground truth.
  • Clinical Agreement (Prospective Study): A composite comparator algorithm was used as ground truth, consisting of:
    • Culture
    • Bi-directional sequencing
    • An FDA cleared NAAT (used for discordant resolution)
  • Clinical Agreement (Retrospective Study): A validated bi-directional sequencing assay was used as ground truth.

8. The Sample Size for the Training Set

  • The document describes performance evaluation studies (reproducibility and clinical agreement) for the Simplexa™ HSV 1 & 2 Direct assay. It does not explicitly mention a "training set" or "validation set" in the context of machine learning model development. This is typical for IVD assays which are developed and then verified/validated rather than 'trained'. The data presented is for validation of the a priori defined assay.

9. How the Ground Truth for the Training Set Was Established

  • As no "training set" in the machine learning sense is described, this question is not applicable based on the provided document. The ground truth for the validation and reproducibility sets is detailed in point 7.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image. The text is in all caps and is evenly spaced around the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 28, 2015

FOCUS DIAGNOSTICS, INC. SHARON YOUNG SENIOR REGULATORY AFFAIRS SPECIALIST 11331 VALLEY VIEW STREET CYPRESS, CA 90630

Re: K150962

Trade/Device Name: Simplexa HSV 1 & 2 Direct, Simplexa HSV1 & 2 Positive Control Pack

Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes simplex virus serological assays Regulatory Class: II Product Code: OQO Dated: August 25, 2015 Received: August 26, 2015

Dear Ms. Young:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Stephen J. Lovell -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150962

Device Name

Simplexa™ HSV 1 & 2 Direct and Simplexa™ HSV 1 & 2 Positive Control Pack

Indications for Use (Describe)

Simplexa™ HSV 1 & 2 Direct

The Focus Diagnostics Simplexa™ HSV 1 & 2 Direct assay is intended for use on the 3M Integrated Cycler instrument for the qualitative detection and differentiation of herpes simples virus (HSV-2) DNA present in genital lesion swabs samples from patients with signs and symptoms of HSV-2 infection of the genitalia. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 genital infections.

The assay is not intended for use as a screening test for the presence of HSV-2 in blood or blood products. The assay is for professional use only.

Simplexa™ HSV 1 & 2 Positive Control Pack

The Simplexa™ HSV 1 & 2 Positive Control Pack is intended to be used as a control with the Simplexa™ HSV 1 & 2 Direct kit. This control is not intended for use with other assays or systems.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Focus Diagnostics. The word "FOCUS" is in large, bold, sans-serif font. A curved, black shape is to the left of the word, resembling a stylized check mark or swoosh. Below "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font, underlined with a thin black line.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 1 of 12

ApplicantFocus Diagnostics, Inc.11331 Valley View StreetCypress, California 90630USA
Establishment Registration No.2023365
Contact PersonSharon Youngtel 562.240.6680fax 562.240.6529syoung@focusdx.com
Summary DateAugust 20, 2015
Proprietary NameSimplexa™ HSV 1 & 2 DirectSimplexa™ HSV 1 & 2 Positive Control Pack
Generic NameHerpes Simplex Virus serological assays
ClassificationOQO
Regulation
Product Code
Predicate DeviceOQO - artus® HSV-1/2 QS-RGQ MDx Kit (K142738)

Intended Use

Simplexa™ HSV 1 & 2 Direct

The Focus Diagnostics Simplexa™ HSV 1 & 2 Direct assay is intended for use on the 3M Integrated Cycler instrument for the qualitative detection and differentiation of herpes simples virus (HSV-1 and HSV-2) DNA present in genital lesion swabs samples from patients with signs and symptoms of HSV-1 or HSV-2 infection of the genitalia. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 genital infections.

The assay is not intended for use as a screening test for the presence of HSV-1 and HSV-2 in blood or blood products. The assay is for professional use only.

Simplexa™ HSV 1 & 2 Positive Control Pack

The Simplexa™ HSV 1 & 2 Positive Control Pack is intended to be used as a control with the Simplexa™ HSV 1 & 2 Direct kit. This control is not intended for use with other assays or systems.

Device Description

The Simplexa™ HSV 1 & 2 Direct assay system is a real-time PCR that enables the direct amplification, detection and differentiation of HSV-1 and/or HSV-2 DNA from unprocessed genital swab specimens without nucleic acid extraction. The system consists of the Simplexa™ HSV 1 & 2 Direct assay, the 3M Integrated Cycler (with 3M Integrated Cycler Studio Software), the Direct Amplification Disc and associated accessories.

In the Simplexa™ HSV 1 & 2 Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify HSV-1, HSV-2 and internal control targets. Well conserved regions of the HSV-1 and HSV-2 DNA polymerase genes are targeted to identify HSV-1 and HSV-2 DNA respectively in the specimen. An internal control is used to detect PCR failure and/or inhibition.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of a stylized swoosh shape on the left, followed by the word "FOCUS" in a bold, sans-serif font. Below "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font, underlined with a thin line. The entire logo is in black and white.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 2 of 12

Predicate Device Information

Similarities

NameSimplexa™ HSV1 & 2 Directartus® HSV-1/2 QS-RGQ MDx KitK142738
Intended UseSimplexa™ HSV 1 & 2 DirectThe Focus Diagnostics Simplexa™ HSV 1 &2 Direct assay is intended for use on the 3MIntegrated Cycler instrument for thequalitative detection and differentiation ofherpes simplex virus (HSV-1 and HSV-2)DNA present in genital lesion swabs samplesfrom patients with signs and symptoms ofHSV-1 or HSV-2 infection of the genitalia.This test is an aid in the differential diagnosisof HSV-1 and HSV-2 genital infections.The assay is not intended for use as ascreening test for the presence of HSV-1 andHSV-2 in blood or blood products. The assayis for professional use only.Simplexa™ HSV 1 & 2 Positive Control PackThe Simplexa™ HSV 1 & 2 Positive ControlPack is intended to be used as a control withthe Simplexa™ HSV 1 & 2 Direct kit.This control is not intended for use with otherassays or systems.The artus HSV-1/2 QS-RGQ MDx Kit is an in vitroreal-time PCR DNA amplification assay performedon the QIAsymphony RGQ MDx system for thedirect qualitative detection and differentiation ofherpes simplex virus (HSV-1 and HSV-2) DNA ingenital or oral vesicular lesions from male andfemale patients suspected of HSV infection.The assay is intended for use as an aid in diagnosisof HSV infection in symptomatic patients.Warning: The artus HSV-1/2 QS-RGQ MDx Kit isnot FDA-cleared for use with cerebrospinal fluid(CSF) or for prenatal screening.
AssayTargetsHSV-1 and HSV-2 DNAHSV-1 and HSV-2 DNA
DetectionTechniquesMultiplex assay using different reporter dyesfor each target.Multiplex assay using different reporter dyes foreach target.

Differences

NameSimplexa™ HSV1 & 2 Directartus® HSV-1/2 QS-RGQ MDx KitK142738
Sample TypesGenital herpetic lesion swab samplesMale and female genital or oral herpetic lesions.
Assay MethodologyPCR-based system for detecting thepresence / absence of viral DNA in clinicalspecimensAmplification and detection are carried out using theartus HSV-1/2 QS-RGQ MDx Kit with the Rotor-Gene Q MDx (RGQ MDx) and Rotor-Gene AssayManager software.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in bold, sans-serif font, with a curved, swooping line above it. Below "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font, underlined with a thin line.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 3 of 12

REPRODUCIBILITY - Genital Swab Sample Type

Reproducibility for the Simplexa™ HSV 1 & 2 Direct assay was evaluated. Three investigative sites assessed the device's inter-site, inter-day and inter/intra-assay reproducibility. Each of the laboratories tested the positive control and a panel of five contrived sample pools including a low (approximately 1-2 times LoD) and medium positive (approximately 2-4 times LoD) for each analyte and a high negative. The high negative sample contained a small amount of HSV-1 and it was designed to be neqative approximately 95% of the time. The assays were performed in triplicate on five different days. Each site had two operators; each operator assayed the entire sample panel and positive control once per day, for a total of two sets of data per day. Combined results for all sites are presented in the tables below.

SampleSite - 1Site - 2Site - 3Total %AgreementwithExpectedResults95% CI
%AgreementwithExpectedResultsAvg.CtTotal%CV%AgreementwithExpectedResultsAvg.CtTotal%CV%AgreementwithExpectedResultsAvg.CtTotal%CV
HSV-1 LowPositive100.0%(30/30)36.02.2100.0%(30/30)36.12.6100.0%(30/30)36.32.7100.0%(90/90)95.9 to100.0%
HSV-1ResultHSV-1MediumPositive100.0%(30/30)34.41.7100.0%(30/30)34.81.2100.0%(30/30)34.61.9100.0%(90/90)95.9 to100.0%
HSV-2 LowPositive100.0%(30/30)aNANA100.0%(30/30)aNANA96.7%(29/30)aNANA98.9%(89/90)a94.0 to99.8%
HSV-2MediumPositive100.0%(30/30)aNANA96.7%(29/30)aNANA100.0%(30/30)aNANA98.9%(89/90)a94.0 to99.8%
HighNegative96.7%(29/30)a38.80.093.3%(28/30)a38.70.590.0%(27/30)a38.04.193.3%(84/90)a86.2 to96.9%
PositiveControl100.0%(30/30)29.90.8100.0%(30/30)30.41.3100.0%(29/29)29.92.8100.0%(89/89)95.9 to100.0%
TotalAgreement99.4% (179/180)98.3% (177/180)97.8% (175/179)98.5%(531/539)97.1 to99.2%
a) Expected Results of HSV-2 Low Positive, HSV-2 Medium Positive and High Negative samples are "Negative" for HSV-1.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Focus Diagnostics. The word "FOCUS" is in large, bold, black letters. Above and to the left of the word "FOCUS" is a black swoosh design. Below the word "FOCUS" is the word "Diagnostics" in smaller, black letters with a line above it.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 4 of 12

SampleSite - 1Site - 2Site - 3Total %AgreementwithExpectedResults95% CI
%AgreementwithExpectedResultsAvg.CtTotal%CV%AgreementwithExpectedResultsAvg.CtTotal%CV%AgreementwithExpectedResultsAvg.CtTotal%CV
HSV-2ResultHSV-1LowPositive100.0%(30/30)bNANA100.0%(30/30)bNANA96.7%(29/30)b41.10.098.9%(89/90)b94.0 to99.8%
HSV-1MediumPositive100.0%(30/30)bNANA100.0%(30/30)bNANA100.0%(30/30)bNANA100.0%(90/90)b95.9 to100.0%
HSV-2LowPositive100.0%(30/30)37.42.990.0%(27/30)37.53.593.3%(28/30)37.12.894.4%(85/90)87.6 to97.6%
HSV-2MediumPositive100.0%(30/30)35.51.9100.0%(30/30)35.62.0100.0%(30/30)35.31.6100.0%(90/90)95.9 to100.0%
HighNegative96.7%(29/30)b39.50.086.7%(26/30)b38.62.9100.0%(30/30)bNANA94.4%(85/90)b87.6 to97.6%
PositiveControl100.0%(30/30)30.21.3100.0%(30/30)30.10.6100.0%(29/29)29.91.2100.0%(89/89)95.9 to100.0%
TotalAgreement99.4% (179/180)96.1% (173/180)98.9% (176/179)98.0%(528/539)96.4 to98.9%
b) Expected Results of HSV-1 Low Positive, HSV-1 Medium Positive and High Negative samples are "Negative" for HSV-2.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Focus Diagnostics. The word "FOCUS" is in large, bold, black letters. A curved black line extends from the top left of the "F" and curves over the word. Below "FOCUS" is the word "Diagnostics" in smaller, black letters with a line above it.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 5 of 12

SampleSite - 1Site - 2Site - 3
Sample% Agreement with Expected ResultsAvg.CtTotal%CV% Agreement with Expected ResultsAvg.CtTotal%CV% Agreement with Expected ResultsAvg.CtTotal%CVTotal % Agreement with Expected Results95% CI
DNAICResultHSV-1 LowPositive100.0%(30/30)29.60.7100.0%(30/30)29.81.2100.0%(30/30)29.71.0100.0%(90/90)95.9 to100.0%
HSV-1MediumPositive100.0%(30/30)29.60.8100.0%(30/30)29.81.4100.0%(30/30)29.70.9100.0%(90/90)95.9 to100.0%
HSV-2 LowPositive100.0%(30/30)29.60.8100.0%(30/30)29.81.2100.0%(30/30)29.71.0100.0%(90/90)95.9 to100.0%
HSV-2MediumPositive100.0%(30/30)29.50.6100.0%(30/30)29.71.4100.0%(30/30)29.81.4100.0%(90/90)95.9 to100.0%
HighNegative100.0%(30/30)29.60.6100.0%(30/30)29.81.2100.0%(30/30)29.71.0100.0%(90/90)95.9 to100.0%
PositiveControl100.0%(30/30)29.50.5100.0%(30/30)29.71.4100.0%(29/29)29.70.9100.0%(89/89)95.9 to100.0%
TotalAgreement100.0% (180/180)100.0% (180/180)100.0% (179/179)100.0%(539/539)96.4 to98.9%

ANALYTICAL SENSITIVITY/LIMIT OF DETECTION - Genital Swab Sample Type

The Limit of Detection (LoD) was determined for the Simplexa™ HSV 1 & 2 Direct assay using quantified
stocks of HSV-1 and HSV-2 serially diluted into negative genital swab ma genital swabs. LoD was determined to be the lowest concentration that could be detected positive > 95% of the time.

Virus StrainLoD Concentration(TCID50/mL)QualitativeResults(#Detected/#Total)Mean Ct ± SD(from DetectedReplicates only)
HSV-1 McIntyre432/3236.4 ± 1.16
HSV-1 HF16032/3235.2 ± 1.03
HSV-2 G232/3237.5 ± 1.08
HSV-2 MS1031/3237.9 ± 1.15

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in bold, sans-serif font, with the word "Diagnostics" in a smaller font size underneath. To the left of the word "FOCUS" is a curved, black shape that resembles a swoosh or a stylized checkmark.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 6 of 12

ANALYTICAL REACTIVITY/ CROSS REACTIVITY

Analytical Reactivity -Genital Swab Sample Type

The analytical reactivity of the Simplexa™ HSV 1 & 2 Direct assay was evaluated using different strains of HSV-1 and HSV-2 that were not used in the determination of the limit of detection (LoD) for the assay. Quantified viral material was spiked into negative genital swab matrix containing male and female genital swabs using a single dilution and assayed in triplicate. The Simplexa™ HSV 1 & 2 Direct assay was able to detect other strains of HSV-1 and HSV-2 viruses.

HSVStrain/IsolateSpiked Concentration[TCID50/mL]Qualitative Result(#Detected/#Total)
HSV-1HSV-2
HSV-1 KOS163/30/3
HSV-1 F323/30/3
HSV-2 Isolate 180/33/3
HSV-2 Isolate 280/33/3
HSV-2 Isolate 380/33/3

Cross-Reactivity (Analytical Specificity) - Genital Swab Sample Type

The Simplexa™ HSV 1 & 2 Direct assay's analytical specificity was evaluated by testing the ability to exclusively identify HSV-1 and HSV-2 viruses with no cross-reactivity to organisms that are closely related, or cause similar clinical symptoms or may be present on swabs of the genital region. Thirty six (36) potential cross-reactants were spiked into negative genital swab matrix containing male and female genital swabs and assayed in triplicate. No cross-reactivity was observed.

No.Potential Cross-ReactantsTested ConcentrationQualitative Result(#Detected/#Total)
HSV-1HSV-2
1None (Baseline)Not Applicable0/150/15
2Bacteroides fragilis1.00 X 106 cfu/mL0/30/3
3Bacteroides ureolyticus*Not ApplicableNot ApplicableNot Applicable
4Candida albicans1.00 X 106 cfu/mL0/30/3
5Chlamydia trachomatis1.00 X 106 IFU/mL0/30/3
6Clostridium sordellii1.00 X 106 cfu/mL0/30/3
7Corynebacterium genitalium1.00 X 106 cfu/mL0/30/3
8Cytomegalovirus AD169 strain1.00 X 105 TCID50/mL0/30/3
9Enterococcus faecalis vanB1.00 X 106 cfu/mL0/30/3
10Enterovirus 711.00 X 105 TCID50/mL0/81/8
11Epstein Barr Virus (B95-8)1.00 X 105 copies/mL0/30/3
12Escherichia coli O157H71.00 X 106 cfu/mL0/30/3
13Gardnerella vaginalis1.00 X 106 cfu/mL0/30/3
14Hepatitis B1.00 X 105 IU/mL0/30/3
15Hepatitis C1.00 X 105 IU/mL0/30/3
16HHV-6 (Z29 Strain)1.00 X 105 TCID50/mL0/30/3
17HHV-7 SB1.00 X 105 TCID50/mL0/30/3
18HIV-1 IIIB1.00 X 105 copies/mL0/30/3

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in large, bold, black letters. To the left of the word is a curved, black shape that resembles a stylized check mark or swoosh. Below the word "FOCUS" is the word "Diagnostics" in a smaller font, underlined with a thin black line.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 7 of 12

No.Potential Cross-ReactantsTested ConcentrationQualitative Result(#Detected/#Total)
HSV-1HSV-2
19HIV-2 NIHZ**Not Available0/30/3
20HPV18 Recombinant1.00 X 105 pfu/mL0/30/3
21Lactobacillus acidophilus1.00 X 106 cfu/mL0/30/3
22Mobiluncus mulieris1.00 X 106 cfu/mL0/30/3
23Mycoplasma genitalium**Not ApplicableNot ApplicableNot Applicable
24Mycoplasma hominis1.00 X 106 CCU/mL0/30/3
25Neisseria gonorrhoeae1.00 X 106 cfu/mL0/30/3
26Proteus vulgaris1.00 X 106 cfu/mL0/30/3
27Rubella1.00 X 105 TCID50/mL0/30/3
28Staphylococcus aureus (MRSA),ATCC 7006991.00 X 106 cfu/mL0/30/3
29Staphylococcus epidermidis (MRSE),ATCC 298871.00 X 106 cfu/mL0/30/3
30Staphylococcus saprophyticus1.00 X 106 cfu/mL0/30/3
31Streptococcus mitis1.00 X 106 cfu/mL0/30/3
32Streptococcus pyogenes, M11.00 X 106 cfu/mL0/30/3
33Toxoplasma gondii1.00 X 106 tachyzooites/mL0/30/3
34Treponema pallidum**Not ApplicableNot ApplicableNot Applicable
35Trichomonas vaginalis1.00 X 106 trophozoites/ml0/30/3
36Ureaplasma urealyticum1.00 X 106 CCU/mL0/30/3
37VZV1.00 X 105 copies/mL0/30/3

" Microorganism was not available for testing therefore in-silico NCBI BLAST analysis was performed and found no predicted cross reactivity. ** Quantified material was not available to test; instead the vendor provided a culture fluid with a known Ct value. The site was directed to dilute the stock to a relevant Ct value; 1:50 dilution factor.

INTERFERENCE - Genital Swab Sample Type

The performance of the Simplexa™ HSV 1 & 2 Direct assay was evaluated with potentially interfering substances that may be present on swabs of the genital region at the concentrations indicated in the table below. A total of five (26) potentially interfering substances were tested in a low positive HSV-1 and HSV-2 sample (4 times LoD) in negative genital swab matrix containing male and female genital swabs and assayed in triplicate. No interference was observed.

Potential InterferentInterferent Concentration#Detected/#Total
HSV-1HSV-2IC
Abreva cold sore treatment7% w/v3/33/33/3
Acyclovir2.5 mg/mL3/33/33/3
Acyclovir Cream*7% w/v8/85/88/8
Albumin10 mg/mL3/33/33/3
Balneol Hygienic Cleansing lotion7% w/v3/33/33/3
Casein10 mg/mL3/33/33/3
Cidofovir2.5 mg/mL3/33/33/3
Clotrimazole vaginal cream7% w/v3/33/33/3
Denavir2.5 mg/mL3/33/33/3

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, uppercase letters. A curved line extends from the left side of the "F", arching over the word. Below "FOCUS" is the word "Diagnostics" in a smaller, non-bold font, underlined with a thin line.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 8 of 12

#Detected/#Total
Potential InterferentInterferent ConcentrationHSV-1HSV-2IC
Douche7% w/v3/33/33/3
Famciclovir2.5 mg/mL3/33/33/3
Feces2.5 mg/mL3/33/33/3
Gynol II (Contraceptive jelly)7% w/v3/33/33/3
KY Jelly5% v/v3/33/33/3
Monistat 17% w/v3/33/33/3
Monistat 37% w/v3/33/33/3
Mucin7% w/v3/33/33/3
Preparation H Hemorrhoid cream7% w/v3/33/33/3
Releev cold sore treatment7% w/v3/33/33/3
Urine10% v/v3/33/33/3
Vagicaine Anti-Itch Cream7% w/v3/33/33/3
Vagisil creme7% w/v3/33/33/3
VagiStat 17% w/v3/33/33/3
Valacyclovir2.5 mg/mL3/33/33/3
Whole Blood10% v/v3/33/33/3
YeastGard Suppositories7% w/v3/33/33/3
* One out of three (1/3) initial replicates and two out of five (2/5) confirmation replicates were negative for HSV-2

CLINICAL AGREEMENT - Genital Swab Sample Type

Prospective Study

A total of 718 genital swab samples were prospectively collected from patients with signs and symptoms of genital herpes simplex virus (HSV) infection from 6 geographically diverse locations. Of the 718 samples collected. 9 samples were removed from the analysis because they were either not tested or had invalid results on the 3 assays (Simplexa™ HSV 1 & 2 Direct, Culture or bi-directional sequencing). Of the 709 remaining samples, 13 samples were removed from the analysis because they were not tested on the tests included in the comparator method sufficient to generate a final comparator result. A total of 696 samples were used for the analysis. The clinical performance of the Simplexa HSV 1 & HSV 2 assay was evaluated by comparing the positive and negative percent agreement to a composite comparator algorithm consisting of; culture, bi-directional sequencing and an FDA cleared NAAT. A positive result for HSV-1 and/or HSV-2 was determined by a positive test result in either the culture or the bi-directional sequencing. If both the culture and the bi-directional sequencing yielded positive results but disagreed in the differentiation of HSV-1 versus HSV-2, the results of the FDA cleared NAAT were used and a 2 out of 3 rule was followed to determine the type of the virus (e.g. if two of the methods were positive for HSV-1, the final comparator result was HSV-1 positive). All sites collected the genital swab samples on the Simplexa™ HSV-1 and HSV-2 Direct and sent samples to a central lab for culture testing. For culture, each sample was tested for HSV- 2 first and if positive for HSV-2 no further testing was performed. Samples that were HSV-2 culture negative were further tested for HSV-1 culture positivity. Dual positives could not be identified in the culture assay.

The available retained samples were sent to Focus Diagnostics and tested in a validated bi-directional sequencing assay. Of the 24 discordant samples that were positive for HSV-2 by the culture method but positive for HSV-1 by the bi-directional sequencing assay, 18 samples had valid results and 1 sample had an invalid result when tested on an FDA cleared NAAT. There were 5 samples that were not tested on the FDA cleared NAAT due to insufficient volume and therefore 6 out of 24 samples were excluded from the

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Focus Diagnostics. The word "FOCUS" is in large, bold, sans-serif font. A curved, swooping shape is to the left of the word. Below the word "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font, underlined with a thin line.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 9 of 12

analysis. There were 2 samples that were positive for HSV-1 by the culture method but positive for HSV-2 by the bi-directional sequencing assay that were not tested on the FDA cleared NAAT for insufficient volume and therefore were excluded from analysis. Results for Simplexa™ HSV 1 & 2 Direct compared to the composite comparator algorithm are presented in the following tables.

Simplexa™ HSV 1 & 2 Direct Compared to Composite Comparator Result (HSV-1)

Simplexa™ HSV 1 & 2 DirectResults HSV-1Composite Comparator Result (HSV-1)
DetectedNot DetectedTotal
Detected11110121
Not Detected3560563
Total114570684
Sensitivity97.4%(111/114)95% CI: 92.5% to 99.1%Specificity98.2%(560/570)95% CI: 96.8% to 99.0%
PPV91.7%(111/121)95% CI: 85.5% to 95.4%NPV99.5%(560/563)95% CI: 98.4% to 99.8%

Simplexa™ HSV 1 & 2 Direct Compared to Composite Comparator Result (HSV-2)

Simplexa™ HSV 1 & 2 DirectComposite Comparator Result (HSV-2)
Results HSV-2DetectedNot DetectedTotal
Detected17511186
Not Detected5497502
Total180508688
Sensitivity97.2%(175/180)95% Cl: 93.7% to 98.8%Specificity97.8%(497/508)95% CI: 96.2% to 98.8%
PPV94.1%(175/186)95% Cl: 89.7% to 96.7%NPV99.0%(497/502)95% CI: 97.7% to 99.6%

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Focus Diagnostics. The word "FOCUS" is in large, bold, black letters. Above and to the left of the word is a curved, black shape. Below the word "FOCUS" is a horizontal line, and below that line is the word "Diagnostics" in a smaller font.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 10 of 12

Retrospective Study

A total of 28 genital swab samples (14 positive HSV-1 and 14 positive HSV-2) were retrospectively collected from male patients with signs and symptoms of genital herpes simplex virus (HSV) infection and contained preselected positive and negative samples were tested at Focus Diagnostics using the SimplexaTM HSV 1 & 2 Direct and a validated bi-directional sequencing assay.

HSV-1 Validated Bi-Directional Sequencing Assay Results

Simplexa™ HSV 1 & 2 DirectResultsComposite Comparator Result (HSV-1)
DetectedNot DetectedTotal
Detected14115
Not Detected01313
Total141428
PPA100.0%(14/14)NPA92.9%(13/14)
95% CI: 78.5% to 100.0%95% CI: 68.5% to 98.7%
PPV93.3%(14/15)NPV100.0%(13/13)
95% CI: 70.2% to 98.8%95% CI: 77.2% to 100.0%

HSV-2 Validated Bi-Directional Sequencing Assay Results

Simplexa™ HSV 1 & 2 DirectResultsComposite Comparator Result (HSV-2)
ResultsDetectedNot DetectedTotal
Detected14014
Not Detected01414
Total141428
PPA100.0%(14/14)95% CI: 78.5% to 100.0%NPA100.0%(14/14)95% CI: 78.5% to 100.0%
PPV100.0%(14/14)95% CI: 78.5% to 100.0%NPV100.0%(14/14)95% CI: 78.5% to 100.0%

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in a bold, sans-serif font, with the word "Diagnostics" in a smaller font below it. A curved, black shape is to the left of the word "FOCUS", which gives the logo a modern and professional look. The logo is simple, clean, and easy to read.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 11 of 12

EXPECTED VALUES - Genital Swab Sample Type

The observed expected values using the Simplexa™ HSV 1 & 2 Direct assay are presented below for patients 17 years of age or older from the prospective study population. The data is stratified by age, and gender and collection sites.

Sample Demographics by Simplexa™ HSV 1 & 2 Direct Assay
GenderAge GroupSimplexa™ HSV 1 & 2 Direct Result
AllHSV-1 &HSV-2 NotDetectedHSV-1 Detected& HSV-2 NotDetectedHSV-1 NotDetected & HSV-2DetectedHSV-1 & HSV-2Dual PositiveNot-Evaluable
Female17 Years ofage to 21Years of age9841.8% (41/98)25.5% (25/98)30.6% (30/98)0.0% (0/98)2.0% (2/98)
More than 21Years of age50354.3% (273/503)16.9% (85/503)25.8% (130/503)0.8% (4/503)2.2% (11/503)
All60152.2% (314/601)18.3% (110/601)26.6% (160/601)0.7% (4/601)2.2% (13/601)
Male17 Years ofage to 21Years of age1442.9% (6/14)21.4% (3/14)35.7% (5/14)0.0% (0/14)0.0% (0/14)
More than 21Years of age6668.2% (45/66)10.6% (7/66)18.2% (12/66)1.5% (1/66)1.5% (1/66)
All8063.8% (51/80)12.5% (10/80)21.3% (17/80)1.3% (1/80)1.3% (1/80)
All68153.6% (365/681)17.6% (120/681)26.0% (177/681)0.7% (5/681)0.7% (5/681)**

volume, daily PC/NTC failure,

** Excluded from this table are 35 samples from 17 years of age and 2 samples where the age of the patient was not t available.

Hypothetical PrevalenceHSV-1HSV-2
PPV*NPV**PPV*NPV**
1.0%35.3%100.0%30.9%100.0%
2.0%52.5%99.9%47.4%99.9%
3.0%62.6%99.9%57.7%99.9%
5.0%74.0%99.9%69.9%99.8%
10.0%85.7%99.7%83.1%99.7%
15.0%90.5%99.5%88.6%99.5%
20.0%93.1%99.3%91.7%99.3%
30.0%95.9%98.9%95.0%98.8%
50.0%98.2%97.4%97.8%97.2%

*Positive Predictive Value (PPV) was calculated using:

(Sensitivity x Prevalence)/(Sensitivity x Prevalence + [1 - Specificity] x [1 - Prevalence]).

**Negative Predictive Value (NPV ) was calculated using:

(Specificity x [1 - Prevalence])/([1 - Sensitivity] x Prevalence + Specificity x [1 - Prevalence]).

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Focus Diagnostics. The word "FOCUS" is in large, bold, black letters. A curved black line extends from the top left of the "F" and curves around the word. Below "FOCUS" is the word "Diagnostics" in smaller, black letters, underlined with a thin black line.

510(k) Summary

Simplexa™ HSV 1 & 2 Direct Catalog No. MOL2150 Simplexa™ HSV 1 & 2 Positive Control Pack Catalog No. MOL2160 August 20, 2015 Page 12 of 12

FRESH VS FROZEN STUDY

Storage conditions were validated using the following transport media types BD VTM, M4, M4RT, M5, M6, and UTM by spiking media with organism at concentrations ranging from 3 times LoD to 50 times LoD and at different storage temperatures and durations.

Samples should be transported on ice and stored at 2 to 8°C for up to 7 days post collection. If there is a greater than 7 day delay before processing of the sample, store the sample at -70° C.

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate the device is as safe and effective as the legally marketed device identified above.

§ 866.3305 Herpes simplex virus serological assays.

(a)
Identification. Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.(b)
Classification. Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).